VSPP Vertical Sync Pulse Polarity Class Description Command syntax Format parameter setting - Synchronization Establishes the logic sense of the digital vertical sync outputs. Setting polarity to 1 causes the leading edge of vertical sync to be a low-to-high transition. Setting polarity to 0 causes the leading edge of vertical sync to be a high-to-low transition. The VSPP? query returns the current polarity of VSPP. VSPP polarity polarity 0 = high-to-low transition 1 = low-to-high transition Query syntax Query response Other required commands Example 566 VSPP? 0 or 1 In order to use digital vertical sync, it must be gated on with the VSPG command and digital separate H and V sync must be selected with the SSST command. The FMTU command instructs the generator to use the new setting. The ALLU command updates hardware to the new setting and redraws the test image. VSPP 1 VSPG 1 SSST 1 FMTU // // // // Set active hi V sync in buffer Enable V sync output in buffer Select H&V sync type in buffer Update hardware to current buffer contents Appendix A Command Reference