Download Agilent Technologies E8364C User`s guide
Transcript
E8362/3/4C Option H85 Example: Making High Power Measurements with Option H85 External Reference Mixer Switch Setup 1. In the Trace/Chan drop-down menu select Channel > Hardware Setup > Test Set. Figure 12 Trace/Chan 2. In the application window, select External: flow through R1 Loop. Figure 13 R1 Input Path 3. Measure the output power at Test Port 1, using a power meter. Verify the power measurement. If you are measuring a highly reflective device, a high power isolators should be inserted between the 20 dB coupler and CPLR THRU front panel ports to protect Port 1 Source. 26 User’s Guide