Download Agilent Technologies HP 3325B Specifications
Transcript
Testing Performance Perform an operational accuracy calibration Acquire the data 1 Press the RUN key. The clicking of attenuator relays should be audible. Verify 2 3 4 5 6 7 8 See Also resistance readings on the digital multimeter of 50 Ω ± 0.5 Ω (49.5 to 50.5 Ω). Record the reading in the performance test record. In the Channel menu change the Coupling field to 1MΩ / DC. The clicking of attenuator relays should be audible. Press the RUN key. Verify resistance readings on the digital multimeter of 1 MΩ ± 10 kΩ (0.990 to 1.010 MΩ). Record the reading in the performance test record. In the Channel menu change the Coupling field to 50 Ω/DC and V/Div to 200 mV/Div. Repeat steps 1 through 3. In the Channel menu change the Coupling field to 50 Ω /DC and V/Div to 1 V/Div. Repeat steps 1 through 3. In the Channel menu change the Coupling field to 50 Ω /DC and V/Div to 4 V/Div. Repeat steps 1 through 3. Connect the male end of the BNC tee to the channel 2 input of the oscilloscope module. Repeat from "Set up the logic analyzer" for channel 2, replacing channel 1 with channel 2 where applicable. If a reading is not within limits, then the attenuator for the out-of-bounds channel should be replaced (see chapter 6). Perform an operational accuracy calibration At this point, an operational accuracy calibration should be performed. Follow the procedure "To calibrate the oscilloscope" in chapter 4, "To calibrate the oscilloscope." 3–72