Testing Performance To test the time interval accuracy (logic analyzer) 4 Set up the Waveform menu. a Press the Waveform key. b Move the cursor to the sec/Div field, then use the RPG knob to dial in 100 ns. c Select the Markers Off field, then select Pattern. d Select the Specify Patterns field. Select X entering 1 and O entering 1. e f g h Select Done to exit the Specify Patterns menu. Move the cursor to the X-pat field. Type 1, then press Done. Move the cursor to the O-pat field. Type 1901, then press Done. Select the Markers Patterns field, then select Statistics. Select Reset Statistics to initialize the statistics fields. 3–63