Chapter 2 Command System
STATus:OPERation:ENABle
Syntax
Description
Parameters
Explanation
STATus:OPERation:ENABle <enable_value>
STATus:OPERation:ENABle?
inset or query the enable register for the Operation Status Register set.
Name
Type
Range
Default
<enable_value>
Integer
0 to 65535
None

When the events correspond to the bits that are set to 1 in the register occurs, the bits
are then reported to the status byte register.

Bit 14 (Busy) will be set to 1 while the instrument is executing a time-consuming
command, such as the MMEMory:IMPort:CONFig? command.

Sending the STATus:PRESet command will set this register to 0.

The bit definitions
below.
Bit
Bit Name
0
Calibrating
1
Self Test
2
Not Used
3
Not Used
4
Scanning
5
WFT
6
Not Used
7
USB MSD
detected
8
Config
Changed
9
Mem
Threshold
10
11
12
13
14
15
Return
Format
Example
Instrument
Locked
Settings
Changed
Not Used
Global
Error
Busy
Not Used
for the operation status enable register are as shown in the table
Weight
1
2
4
8
16
32
64
128
256
512
Definition
The instrument is calibrating.
The instrument is performing a self-test.
Always be 0.
Always be 0.
The instrument is scanning.
The instrument is waiting for a trigger.
Always be 0.
A USB storage device (external memory) has been
detected.
The scan lsit configuration has changed.
1024
The number of readings in the memory has
exceeded the memory threshold setting (refer to
the DATA:POINts:EVENt:THReshold comamnd).
The instrument is locked.
2048
The instrument's settings have changed.
4096
8192
Always be 0.
An error occurs (the error queue is not empty).
16384
32768
The instrument is busy.
Always be 0.
The query returns an integer that corresponds to the binary-weighted sum of all the bits in
the register. For example, if bit 1 (2 in decimal) and bit 4 (16 in decimal) are enabled, the
query will return 18 (#b100010010).
STAT:OPER:ENAB 18
STAT:OPER:ENAB?
The query returns 18.
Related
commands
STATus:OPERation:CONDition?
STATus:OPERation[:EVENt]?
STATus:PRESet
M300 Programming Guide
2-193