Download Emerson Oxymitter 4000 Specifications
Transcript
k. Fault 11, Bad Cell. The O2 CELL flashes three times, pauses for three seconds, and repeats (Figure 5-11). HEATER T/C HEATER O2 CELL CALIBRATION 1. The bad cell alarm activates when the cell exceeds the maximum resistance value. SW2 ON DIAGNOSTIC ALARMS 2. The cell should be replaced. See paragraph 6-8, Cell Replacement, for cell replacement instructions. CALIBRATION RECOMMENDED INC INC HIGH GAS LOW GAS DEC DEC TP1 J1 TP2 TP3 TP4 RED YEL GRN ORG TEST POINTS O2 CELL mV + O2 CELL mv HEATER T/C + HEATER T/C - CAL TEST GAS + PROCESS % O2 TP5 TP6 22220042 Figure 5-11. Fault 11, Bad Cell IB-106-340 5-13