n. Fault 14, Invalid Constant. The CALIBRATION LED flashes twice, pauses for three seconds, and repeats (Figure 5-14). HEATER T/C HEATER O2 CELL CALIBRATION 1. After a calibration has been performed, the electronics calculates a cell constant value. SW2 ON DIAGNOSTIC ALARMS 2. If the cell constant value is outside of the range, -4 mV to 10 mV, the alarm will activate. See paragraph 6-2, Calibration, and verify the last calibration was performed correctly. CALIBRATION RECOMMENDED INC INC HIGH GAS LOW GAS DEC DEC TP1 J1 TP2 TP3 TP4 RED YEL GRN ORG TEST POINTS O2 CELL mV + O2 CELL mv HEATER T/C + HEATER T/C - CAL TEST GAS + PROCESS % O2 3. Power down the Oxymitter 4000 and remove it from the stack. TP5 TP6 4. Replace the cell per paragraph 6-8, Cell Replacement. 22220045 Figure 5-14. Fault 14, Invalid Constant IB-106-340 5-16