Download Directed Electronics 1500X User`s guide
Transcript
AngioDynamics, Inc. 8.3 Model 1500X User’s Guide and Service Manual Maintenance The RF Generator is designed for indoor use in a dry operating room/procedure room environment. The RF Generator requires no maintenance or calibration by the user. The user should not remove the cover. Removal of unit cover voids the warranty. All maintenance and calibration must be referred to Customer Service at AngioDynamics, Inc. Refer to Section 1 for applicable phone numbers. 8.4 Cleaning and Disinfecting the RF Generator The RF Generator should be given reasonable care and be kept clean and sanitary. The RF Generator may be cleaned with a damp disposable wipe moistened with 70/30 isopropyl alcohol and water. WARNING: Electric shock hazard. Do not saturate the RF Generator with liquids. Do not allow liquids to run inside the unit. Do not immerse the RF Generator in water. Shut off the RF Generator, and disconnect power before cleaning. Do not sterilize the RF Generator. CAUTION: Do not use abrasives, caustics, or mineral spirits. Use of these agents to clean the RF Generator, or any of its accessories may cause damage and voids the warranty. All electrical connections must be air-dried before use. 8.5 Fuse Replacement 1) Turn the Generator off with the switch in the back and unplug the power cord. Get a small flat blade screwdriver and pry open the fuse access panel on the Line Input module. 160-103990 Rev 03 2) Below is Line Input module with fuse access panel open. Note, there are two fuse modules. They are marked with arrows. Pull on each of the arrows to slide the fuse module out. Page 17