Post-Run Procedures Introduction Post-run procedures include the removal and weighing of reagents, and an instrument wash. A water wash is required after each run, with the option after a high output run to perform a maintenance wash instead. Illumina recommends a maintenance wash. Regular instrument washes ensure continued performance in the following ways: } Flush any remaining reagents and sample from the fluidics tubing and sippers. } Prevent salt accumulation and crystallization in the fluidics tubing and sippers. } Prevent cross-contamination from the previous run, including cross-contamination after switching modes. NOTE A water wash performed at the end of a run washes the system and checks fluidics. When setting up a run, the software confirms that a water wash or maintenance wash has been performed within the last 24 hours. A maintenance wash is required every 10 days. When 10 days elapse after the last maintenance wash, the software prompts you to perform a maintenance wash. 98 Part # 15035786 Rev. D