Perform a TruSeq v3 Run Load Reagents for Read 2 After completion of Read 1 and any index reads, load paired-end reagents for Read 2 resynthesis and freshly prepared ICB for Read 2. Load Paired-End Reagents 1 Record the weight of each reagent on the lab tracking form. 2 Make sure that the paired-end rack is not in use on the opposite flow cell for Read 2 resynthesis, Index 1 (i7) Read preparation, or Index 2 (i5) Read preparation. 3 Raise the sippers for the paired-end reagent rack using the following motion: a Pull the handle towards you and raise the handle. b Release the handle into the slot on the top end of the groove. Make sure that the handle rests securely in the slot. 4 Slide the reagent rack out of the reagent compartment using the rack handle. 5 Remove the caps from each reagent tube. 6 Place each of the reagent tubes onto the rack in the associated numbered positions. Table 9 Paired-End Reagent Positions Position Reagent 10 11 12 13 14 15 16 18 19 RMR LMX2 BMX AMX2 APM2 AT2 HP7 or HP11 HP3 HT2 Description Resynthesis Mix Linearization Mix 2 Blocking Mix Amplification Mix 2 AMX2 Premix 100% Formamide Read 2 Sequencing Primer Denaturation Solution Wash Buffer NOTE For dual-indexed paired-end runs, RMR is loaded with indexing reagents before starting the run. For single-indexed or non-indexed runs, RMR is loaded with paired-end reagents. 66 Part # 15035786 Rev. D