ContaminationSensor CS 1000 Setting the measuring mode Mode M4: Filter to continuously monitor cleanliness class Application: Control of stationary offline filtration unit Data output: Display & RS485 & analog output & switching output Purpose: Establish continuous monitoring of cleanliness class between min./max. limit values. Function: Control of a filter unit, continuous measurement of solid contamination. If min./max. limit values are pre-programmed, the CS switches the filtration unit on/off to maintain the cleanliness within the limit value range. Once the target cleanliness has been reached (5x undershooting of the TARGET), the set test cycle time in minutes appears on the display. The test cycle time expires. T 1 After the test cycle time has elapsed, the switching output is closed and a measurement is started. If the result is still below the TARGET cleanliness, the test cycle time (CYCLE) begins again. Mode "SINGLE" measurement Application: Stand-alone sensor Data output: Display & RS485 & analog output Purpose: Perform a single measurement and "stop" the result. Function: Single measurement of solid contamination without switching functions When Single mode is selected in the PowerUp menu, the display jumps directly to the following message after switching to the Measuring menu or after switching the CS on: The CS begins with individual measurement after the message has been confirmed by pressing HYDAC FILTER SYSTEMS GMBH BeWa CS1000 3764916 300 en-us 2012-08-29.doc o.k. . en(us) START? Page 29/112 2012-08-29