Dionex ICS-1100 Ion Chromatography System
5.10 Replacing the Conductivity Cell
1. Turn off the Dionex ICS-1100 power.
2. Open the front door and disconnect the tubing from the CELL IN and CELL
OUT fittings (see Figure 5-12, View A). Loosen the two screws on the cell
front cover. The screws remain attached (see Figure 5-12, View B).
Unscrew
Disconnect
Unscrew
View B
View A
Figure 5-12. Removing the Cell Front Cover
3. Grip the two loosened screws
and pull the cell cover toward
you to remove it from the
component panel. Do not pull
from the CELL IN and CELL
OUT fittings.
4. Continue pulling on the cover
until the attached cell assembly
slides out the front of the
Dionex ICS-1100 (see
Figure 5-13).
Figure 5-13. Removing the Cell Assembly
104
Doc. 065289-03 10/12