Download Agilent Technologies 6051A Operating instructions
Transcript
Query Syntax *SRE? Returned Parameters <NR1> Suffix None Related Commands *PSC *STB? Value: 0 to 255 Read Status Byte Query Type Description Device Status This query reads the Status Byte register. Note that the MSS (Master Summary Status) bit and not the RQS bit is returned in Bit 6. This bit indicates whether or not the electronic load has at least one reason for requesting service. *STB? does not clear the Status Byte register, which is cleared only when subsequent action has cleared all its set bits. Refer to Chapter 5 - Status Reporting for more information about this register. Bit Position Condition Bit Weight Parameters Returned Parameters <NR1> None Related Commands None Status Byte Register 5 4 3 2 ESB MAV QUES CSUM 32 16 8 4 1 Always zero. 1 0 2 1 0 0 1 1 Value: 0 to 255 Immediate Trigger Command Type Description Command Syntax Parameters Related Commands 38 6 MSS 64 None Suffix *TRG 7 OPER 128 Device Trigger This command which is essentially the same as the Group Execute Trigger (<GET>), generates a trigger to the electronic load only if TRIG:SOUR is set to BUS (refer to the TRIGger Subsystem Root commands). *TRG None <GET> Language Dictionary TRIG:SOUR