E4438C Option HEC Repeatable Triggering for Multiple Baseband Sources 3. Select ARB Setup > More (1 of 3) > More (2 of 3) > VCO CLOCK Ext > [Return] ARB “On.” Refer to Figure 11. 4. Adjust the trigger level control on the oscilloscope for a stable display. 5. Align the DACs: [Mode] Dual ARB > ARB Setup > More (1 of 3) > More (2 of 3) > Align DAC NOTE Changing or disrupting the external clock frequency can cause the DACs to become misaligned. Press the align DACs softkey after changing the clock frequency. Figure 11 I & Q Output for the “Master” Instrument 16 User’s and Service Guide Supplement