31 :WGEN Commands
:WGEN:MODulation:TYPE
(see page 790)
Command Syntax
:WGEN:MODulation:TYPE <type>
<type> ::= {AM | FM | FSK}
The :WGEN:MODulation:TYPE command selects the modulation type:
• AM (amplitude modulation) — the amplitude of the original carrier
signal is modified according to the amplitude of the modulating signal.
Use the :WGEN:MODulation:AM:FREQuency command to set the
modulating signal frequency.
Use the :WGEN:MODulation:AM:DEPTh command to specify the amount
of amplitude modulation.
• FM (frequency modulation) — the frequency of the original carrier
signal is modified according to the amplitude of the modulating signal.
Use the :WGEN:MODulation:FM:FREQuency command to set the
modulating signal frequency.
Use the :WGEN:MODulation:FM:DEViation command to specify the
frequency deviation from the original carrier signal frequency.
• FSK (frequency- shift keying modulation) — the output frequency "shifts"
between the original carrier frequency and a "hop frequency" at the
specified FSK rate.
The FSK rate specifies a digital square wave modulating signal.
Use the :WGEN:MODulation:FSKey:FREQuency command to specify the
"hop frequency".
Use the :WGEN:MODulation:FSKey:RATE command to specify the rate at
which the output frequency "shifts".
Query Syntax
:WGEN:MODulation:TYPE?
The :WGEN:MODulation:TYPE? query returns the selected modulation type.
Return Format
<type><NL>
<type> ::= {AM | FM | FSK}
See Also
• ":WGEN:MODulation:AM:DEPTh" on page 672
• ":WGEN:MODulation:AM:FREQuency" on page 673
• ":WGEN:MODulation:FM:DEViation" on page 674
• ":WGEN:MODulation:FM:FREQuency" on page 675
• ":WGEN:MODulation:FSKey:FREQuency" on page 676
• ":WGEN:MODulation:FSKey:RATE" on page 677
682
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
Related documents
VDO MS 2000X - Technical data
Agilent Technologies InfiniiVision 3000 DSO-X 3054A Technical data
Agilent Technologies InfiniiVision 5000 Series Technical data
Agilent Technologies Ininiium 90000 Technical data
Agilent Technologies Ininiium 9000 Technical data
Agilent Technologies Sprinkler 54753a User's Manual
Agilent Technologies 6000 Series Sprinkler User Manual
Bedienung und Grundeinstellungen
Agilent VisualBasic VisualC#による オシロスコープ 制御入門
KEYENCE Vision Sensor MODEL: CV-X100