Measurement Units and Functions Compliance Figure 7-10 Power Compliance Output Area If you specify power compliance, SMUs can be swept at their maximum output limits because the 4155C/4156C changes the V (or I) output range and I (or V) compliance range during a V (or I) sweep. Figure 7-11 shows an example of the difference in SMU output when power compliance is set and when power compliance is not set. Figure 7-11 Allowable I Output when the MPSMU Sweeps Voltage (0 V to 100 V) If you specify power compliance, the measurement time increases slightly because of the range changing for every step. When the ranges are changed during a sweep to accommodate power compliance, the SMU output is momentarily set to 0 V. 7-28 Agilent 4155C/4156C User’s Guide Vol.2, Edition 6