Download manuales
Transcript
ScaleResource.com 11.15 CAL 12. CAL d cal done 13. F1.3 0 expanded display, 0 is normal, 1 is expanded. 14.Push Print until SAVE is displayed and push Print again. If Abort or dFAULT is displayed push Function until Save is displayed. TOLEDO PANTHER 1. 2. 3. 4. 5. 6. 7. Open up the indicator and close SW1-1 Press Enter and Zero simultaneously. F1 will be displayed, Press ENTER. Press SELECT until F1.2 is displayed This is Cal units, 1 is lb, 2 is kg. Press ENTER Press SELECT to go to CAL, display should read CAL 1 then press ENTER, use SELECT to toggle to 1. 8. E SCL empty scale press enter. 9. 15 CAL delay 10.Add Ld Place test weight on scale. ENTER. 11.000000 Enter test weight value 12.15 CAL 13.CAL d cal done 14.F1.6 x zero cal adjust 0 is skip, 1 to store, press clear to abort. 15.F1.7 x Span cal adj., 16.Press clear to exit. TOLEDO 8140 1. 2. 3. 4. Jump W2 Push TARE and CLEAR at the same time. PRINT will enter and advance. Advance to CAL C1. 27