4339B
To Retrieve Data Eciently
To Perform a Measurement Sequence
The following commands perform the sequence measurement:
:ARM:SOUR
:ARM:DEL
:TRIG:SOUR
:TRIG:TIM
:TRIG:COUN
Warning
When the 4339B is in the sequence measurement mode, triggering a
measurement may output a dangerous voltage levels up to 1000 Vdc. Do
not touch the UNKNOWN terminals or the electrodes of the accessory,
when the V output indicator is ON.
To perform the single measurement sequence:
:
OUTPUT 717;":SOUR:VOLT 10"
Test Voltage: 10 V
OUTPUT 717;":ARM:SOUR BUS"
Measurement Sequence: ON, Trigger Source: Bus
OUTPUT 717;":ARM:DEL 60"
Charge time: 60 s
OUTPUT 717;":TRIG:SOUR INT" Single mode
OUTPUT 717;":INIT:CONT ON"
OUTPUT 717;"*TRG"
ENTER 717;S,D
PRINT S,D
:
To perform the continuous measurement sequence:
OPTION BASE 1
DIM D(20)
:
OUTPUT 717;":SOUR:VOLT 10"
Test Voltage: 10 V
OUTPUT 717;":ARM:SOUR BUS"
Measurement sequence: ON, Trigger Source: Bus
OUTPUT 717;":ARM:DEL 60"
Charge time: 60 s
OUTPUT 717;":TRIG:SOUR TIM" Continuous mode
OUTPUT 717;":TRIG:TIM 30"
Interval time: 30 s
OUTPUT 717;":TRIG:COUN 10"
Memory size: 10
OUTPUT 717;":INIT:CONT ON"
OUTPUT 717;"*TRG"
ENTER 717;D(*)
PRINT D(*)
:
Remote Operation 4-25