VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Medical electrical equipment
Part 1: General requirements for basic safety and essential performance
Report Reference No. ................... :
1307033001
Date of issue ................................. :
22 August, 2013
..................................................
Total number of pages.................. :
203
..................................................
CB Testing Laboratory ................. :
Victronic Technology Corporation.
Address ......................................... :
4th Fl 130 Ln 235 Baoqiao Rd Xindian Dist New Taipei, 231
Taiwan.
Applicant’s name .......................... :
Onyx Healthcare Inc.
Address ......................................... :
2F., No.135, Ln. 235, Baoqiao Rd., Xindian Dist., New Taipei City
231, Taiwan
Test specification:
Standard ........................................ :
IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007)
EN 60601-1 : 2006
Test procedure.............................. :
CB Scheme
Non-standard test
method…………..:
None
Test Report Form No. ................... :
60601-1TRF_A
Test Report Form Originator ........ :
Victronic Technology Corporation
Master TRF .................................... :
2011/11/29
Copyright © 2010 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE),
Geneva, Switzerland. All rights reserved.
This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as
copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from
the reader's interpretation of the reproduced material due to its placement and context.
If this Test Report Form is used by non-IECEE members, the IECEE/IEC logo shall be removed
This report is not valid as a CB Test Report unless signed by an approved CB Testing Laboratory and
appended to a CB Test Certificate issued by an NCB in accordance with IECEE 02.
Test item description .................... :
Slim Medical Panel PC
Trade Mark .................................... :
Manufacturer ................................. :
Onyx Healthcare Inc.
2F., No.135, Ln. 235, Baoqiao Rd., Xindian Dist., New Taipei City
231, Taiwan
Model/Type reference ................... :
xxxxx-ONYX-2121DTy-xxxxxxx (where y can be T or blank; x
can be 0-9, A-Z, “-” or blank for marketing purpose)
Input Rating :
Slim Medical Panel PC : 24Vdc, 2.91A
Power Adapter : 100-240Vac, 50/60Hz, 1.8-0.8A
Ratings .......................................... :
Test Report No.: 1307033001
60601-1TRF_A
Page 1 of 203
Issued: 22 August, 2013
VICTRONIC TECHNOLOGY CORPORATION
List of Attachments (including a total number of pages in each attachment):
- Photo Documentation (6 pages)
- Technical Documentation (6 pages)
Summary of testing
Tests performed (name of test and test clause):
Testing location:
Power Input Test (4.11)
Victronic Technology
Corporation.
Humidity Preconditioning Treatment (5.7)
Legibility of Markings (7.1.2)
Durability of Marking Test (7.1.3)
Dielectric Voltage Withstand (8.8.3)
Ball Pressure (8.8.4.1)
Temperature Test (11)
Overflow, Spillage, Leakage, Cleaning, Sterilization And Disinfection, Harmful
Ingress of Liquids (11.6)
Interruption of Power Supply (11.8)
Abnormal Operation and Single Fault Conditions (13)
Enclosure Mechanical Strength:(15.3)
Drop Impact Test: (15.3.4)
Mold Stress Relief Test: (15.3.6)
Reversed Battery Connection/Overcharging: (15.4.3)
Leakage Current Test (8.7)
Suspension System Without Safety Device Loading Test (28.4 US Deviation)
Summary of compliance with National Differences
List of countries addressed:
N/A
The product fulfils the requirements of EN 60601-1:2006, CAN/CSAC22.2 No. 60601-1:08, ANSI/AAMI
ES60601-1:2005
Test Report No.: 1307033001
60601-1TRF_A
Page 3 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
Copy of marking plate
The artwork below may be only a draft. The use of certification marks on a product must be
authorized by the respective NCBs that own these marks.
Test Report No.: 1307033001
60601-1TRF_A
Page 4 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
GENERAL INFORMATION
Test item particulars (see also Clause 6):
Classification of installation and use................................ :
transportable / portable / stationary /
mobile / fixed / permanently installed /
hand-held
Device type (component/sub-assembly/ equipment/
system) ................................................................................ :
equipment
Intended use (Including type of patient, application
location) .............................................................................. :
as part of medical patient monitoring
system for information display
Mode of operation .............................................................. :
Continuous / non-continuous
Supply connection ............................................................. :
internally powered /permanently installed /
appliance coupler / non-detachable cord
Accessories and detachable parts included ..................... :
Power adaptor type GMPU70A-6 (Eljintek
Inc.)
Other options include ......................................................... :
N/A
Testing
Date of receipt of test item(s) .............................................. : 18 July, 2013
Dates tests performed ......................................................... : 19 July, 2013 to 05 August, 2013
Possible test case verdicts:
- test case does not apply to the test object ...................:
N/A
- test object does meet the requirement ..........................:
Pass (P)
- test object was not evaluated for the requirement ......... :
N/E
- test object does not meet the requirement....................:
Fail (F)
Abbreviations used in the report:
- normal condition................................................. : N.C.
- means of Operator protection ........................... : MOOP
- single fault condition ...............: S.F.C.
- means of Patient protection ...: MOPP
General remarks:
"(see Attachment #)" refers to additional information appended to the report.
"(see appended table)" refers to a table appended to the report.
The tests results presented in this report relate only to the object tested.
This report shall not be reproduced except in full without the written approval of the testing laboratory.
List of test equipment must be kept on file and available for review.
Additional test data and/or information provided in the attachments to this report.
Throughout this report a
comma /
point is used as the decimal separator.
Manufacturer’s Declaration per sub-clause 6.2.5 of IECEE 02:
Test Report No.: 1307033001
60601-1TRF_A
Page 5 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
The application for obtaining a CB Test Certificate
includes more than one factory location and a
declaration from the Manufacturer stating that the
sample(s) submitted for evaluation is (are)
representative of the products from each factory
has been provided .................................................. :
Yes
Not applicable
When differences exist; they shall be identified in the General product information section.
Name and address of factory (ies) ........................ :
- Onyx Healthcare Inc.
2F., No.135, Ln. 235, Baoqiao Rd., Xindian Dist., New Taipei City 231, Taiwan
General product information:
The 22” Slim Medical Panel PC consists of an external certified switching power adapter, TFT display with
1366 x 768 resolution and with 22” LCD respectively, support 2.5” Hard Disk Drive storage function and one
mini card slot for WLAN expansion, feature flexible I/O ports, such as six USB2.0 each USB load 0.5A (6
USB ports total load 15W), two isolated COM (RS-232) port and two Gigabit LANs.
Equipment provided additional isolated protection at SIP/SOP connector on RS-232 x 2 and Gigabit LAN x 2,
which provide equivalent protection of Patient Connection.
The power adaptor source Eljintek Inc. type GMPU70A-6 which intended to supply the equipment have been
evaluated and certified by TÜ V Rheinland according to IEC 60601-1:2005 and EN 60601-1:2006 standards.
(Report no. 11027380 001 and Certificate no. DE 2-015508) and (Report no. 11035109 001 and Certificate
no. TA 50221530)
Two MOPP (means of patient protection) are provide between primary circuit and secondary SIP/SOP, one
MOPP (means of patient protection) are provide between secondary circuit and isolated protection at
SIP/SOP.
The maximum specified operational ambient temperature is +40C.
Model Differences Series model xxxxx-ONYX-2121DTy-xxxxxxx (where y can be T or blank; x can be 0-9, A-Z, “-” or blank)
different only for panel size and marking purpose.
Test Report No.: 1307033001
60601-1TRF_A
Page 6 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
INSULATION DIAGRAM
Test Report No.: 1307033001
60601-1TRF_A
Page 7 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
TABLE: To insulation diagram
Pass
Pollution degree .................................... : 2
—
Overvoltage category ............................ : II
—
Altitude .................................................. : 0 – 3000 m
—
Additional details on parts considered
None
Areas ________________
as applied parts ..................................... : (See Clause 4.6 for details)
—
Number and
CTI
type of Means
(IIIb,
Area
of Protection: unless is
MOOP, MOPP known)
Working voltage
Vrms
Vpk
Required
creepage
(mm)
Required Measured Measured
clearance creepage clearance
(mm)
(mm)
(mm)
Remarks
1
1 BOP
IIIb
250
--
3.0
1.6
--
--
Certified
Power Supply.
2
2 MOPP
IIIb
364
--
11.2
7.0
--
--
Certified
Power Supply
3
2 MOPP
IIIb
250
--
8.0
5.0
--
--
Certified
Power Supply
4
1 MOPP
IIIb
250
--
4.0
2.5
--
--
Certified
Power Supply
5
2 MOPP
IIIb
250
--
8.0
5.0
--
--
Certified
Power Supply
6
2 MOPP
IIIb
250
--
8.0
5.0
The
equipme
nt is
supplied
via
external
certified
power
supply
The
equipme
nt is
supplied
via
external
certified
power
supply
7
1 MOPP
IIIb
250
--
4.0
2.5
5.0
5.0
8
1 MOPP
IIIb
250
--
4.0
2.5
The
equipme
nt is
supplied
via
external
certified
power
supply
Test Report No.: 1307033001
60601-1TRF_A
Page 8 of 203
Adapter
Primary to
equipment
SIP/SOP
SIP/SOP to
Touch Panel
and Plastic
enclosure
The
Adapter Earth
equipme
to Touch
nt is
Panel and
supplied
Plastic
via
enclosure
external
certified
power
supply
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
INSULATION DIAGRAM CONVENTIONS and GUIDANCE:
A measured value must be provided in the value columns for the device under evaluation. The symbol >
(greater than sign) must not be used. Switch-mode power supplies must be re-evaluated in the device under
evaluation therefore N/A must not be used with a generic statement that the component is certified.
Insulation diagram is a graphical representation of equipment insulation barriers, protective impedance and
protective earthing. If feasible, use the following conventions to generate the diagram:
- All isolation barriers are identified by letters between separate parts of diagram, for example separate
transformer windings, optocouplers, wire insulation, creepage and clearance distances.
- Parts connected to earth with large dots are protectively earthed. Other connections to earth are functional
- Applied parts are extended beyond the equipment enclosure and terminated with an arrow.
- Parts accessible to the operator only are extended outside of the enclosure, but are not terminated with an
arrow.
Test Report No.: 1307033001
60601-1TRF_A
Page 9 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
4
GENERAL REQUIREMENTS
Pass
4.1
Requirements of this standard applied in
NORMAL USE and reasonably foreseeable misuse
Pass
4.2
A RISK MANAGEMENT PROCESS complying with
ISO 14971 was performed ............................... :
See Appended RM Results
Table 4.2
4.3
ESSENTIAL PERFORMANCE functions identified
according to MANUFACTURER’S policy for RISK
acceptability in RISK MANAGEMENT FILE .............. :
The equipment offers the
highly scalable bedside
infotainment solution to
enhance patient experiences
within the hospital.
N/A
ESSENTIAL PERFORMANCE functions maintained
following particular tests as applicable
--
N/A
4.4
EXPECTED SERVICE LIFE stated in RISK
MANAGEMENT FILE .............................................. :
The Slim Medical Panel PC
expected live was 50,000
hours Completed by
inspection
4.5
Alternative means of addressing particular
No alternative means applied
RISKS considered acceptable based on
in the evaluation of RM
MANUFACTURER’S justification that RESIDUAL RISKS report
resulting from application of alternative means
equal to or less than RESIDUAL RISKS resulting
from requirements of this standard................. :
N/A
4.6
RISK MANAGEMENT PROCESS identifies parts that
can come into contact with PATIENT but not
defined as APPLIED PARTS, subjected to the
requirements for APPLIED PARTS, except for
Clause 7.2.10.................................................... :
N/A
4.7
ME EQUIPMENT remained SINGLE FAULT SAFE, or
The requirements of Clause
the RISK remained acceptable as determined by 8.1, item a), b) and Clause 13
Clause 4.2 ........................................................ : are checked by inspection
and/or by test and
compliance with IEC 60601-1
requirements.
No applied part
Pass
Pass
Pass
See Appended RM Results
Table 4.7
Failure of any one component at a time that
See Appended RM Results
could result in a HAZARDOUS SITUATION, including Table 4.7
those in 13.1, simulated physically or
theoretically .................................................... :
Pass
RISK associated with failure of component
during EXPECTED SERVICE LIFE of ME EQUIPMENT
taken into account to evaluate if a component
should be subjected to failure simulation
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 10 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
4.8
All components and wiring whose failure could All components are used
result in a HAZARDOUS SITUATION used according within their specified ratings
to their applicable ratings, except as specified,
See Appended RM Results
or by RISK MANAGEMENT PROCESS....................... :
Table 4.8
Pass
Reliability of components used as MEANS OF
PROTECTION assessed for conditions of use in
ME EQUIPMENT , and they complied with one of
the following:
Pass
a) Applicable safety requirements of a relevant
IEC or ISO standard
Verdict
See table 8.10 list of critical
components
b) Requirements of this standard applied in the Power adapter complied with
absence of a relevant IEC or ISO standard
the requirements of EN
60601-1:2006, IEC 60601-1:
2005, ANSI/AAMI ES60601-1
4.9
A COMPONENT WITH HIGH-INTEGRITY
CHARACTERISTICS provided because a fault in a
particular component can generate an
unacceptable RISK ............................................ :
No high integrity component
used
COMPONENTS WITH HIGH-INTEGRITY
CHARACTERISTICS selected and evaluated
consistent with their conditions of use and
reasonable foreseeable misuse during
EXPECTED SERVICE LIFE of ME EQUIPMENT by
reviewing RISK MANAGEMENT FILE ..................... :
Pass
Pass
N/A
N/A
4.10
Power supply
Pass
4.10.1
ME EQUIPMENT is suitable for connection to a
SUPPLY MAINS, specified to be connected to a
separate power supply, can be powered by an
INTERNAL ELECTRICAL POWER SOURCE, or a
combination of the three ................................. :
4.10.2
Maximum rated voltage for ME EQUIPMENT
Not Hand-Held device
intended to be connected to SUPPLY MAINS is 250
V for HAND-HELD ME EQUIPMENT (V) ..................... :
The product is specified to
be connected to a separate
power supply
– 250 V d.c. or single-phase a.c., or 500 V
For adapter: single phase:
polyphase a.c. for ME EQUIPMENT and ME SYSTEMS 100-240Vac, current: 1.8-0.8
with a RATED input ≤ 4 kVA (V) ......................... : A; for Panel PC unit: 24Vdc,
2.91A
– 500 V for all other ME EQUIPMENT and ME
Pass
N/A
Pass
N/A
SYSTEMS
4.11
Power input
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 11 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Steady-state measured input of ME EQUIPMENT or See appended Table 4.11
ME SYSTEM at RATED voltage and at operating
settings indicated in instructions for use did
not exceed marked rating by more than 10% .. :
Pass
– Measurements on ME EQUIPMENT or a ME
See appended Table 4.11
SYSTEM marked with one or more RATED voltage
ranges made at both upper and lower limits of
the range .......................................................... :
Pass
Measurements made at a voltage equal to the
No such marking
mean value of the range when each marking of
RATED input was related to the mean value of
relevant voltage range
Power input, expressed in volt-amperes,
measured with a volt-ampere meter or
calculated as the product of steady state
current (measured as described above) and
supply voltage ................................................. :
N/A
See appended Table 4.11
Pass
5
GENERAL REQUIREMENTS FOR TESTING ME EQUIPMENT
Pass
5.1
TYPE TESTS determined in consideration of
Clause 4, in particular 4.2
Pass
Test not performed when analysis indicated
condition being tested was adequately
evaluated by other tests or methods............... :
No such condition
Results of RISK ANALYSIS used to determine
combination(s) of simultaneous faults to be
tested
See Appended RM Results
Table 5.1
Pass
See Appended RM Results
Table 5.1
Pass
5.2
TYPE TESTS conducted on one representative
sample under investigation; multiple samples
used simultaneously when validity of results
was not significantly affected .......................... :
Pass
5.3
a) Tests conducted within the environmental
conditions specified in technical description
Pass
Temperature (ºC), Relative Humidity (%) ........ :
—
10-40ºC;
30-75%RH;
Atmospheric Pressure (kPa)............................ :
b) ME EQUIPMENT shielded from other
No such situation
influences that might affect the validity of tests
N/A
c) Test conditions modified and results
adjusted accordingly when ambient
temperature could not be maintained ............ :
N/A
Test Report No.: 1307033001
60601-1TRF_A
—
70-106kPa;
Page 12 of 203
No such situation
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
5.4
a) ME EQUIPMENT tested under least favourable See Appended RM Results
working conditions specified in instructions for Table 5.4a
use and identified during RISK ANALYSIS, except
as noted ........................................................... :
b) ME EQUIPMENT with adjustable or controlled
operating values by anyone other than SERVICE
PERSONNEL adjusted to values least favourable
for the relevant test per instructions for use
This equipment was tested
with the maximum normal
load, see test results table for
detail.
N/A
c) When test results influenced by inlet
pressure and flow or chemical composition of
a cooling liquid, tests performed within the
limits in technical description ......................... :
No cooling system provided
inside the equipment
N/A
d) Potable water used for cooling
5.5
5.6
5.7
Supply voltage during tests was the least
favourable of the voltages specified in 4.10 or
voltages marked on ME EQUIPMENT (V) ............. :
N/A
Rated voltage: 100-240 V ac
Pass
Tested at 90/100/240/264 V ac
ME EQUIPMENT marked with a RATED frequency
Rated frequency: 50/60 Hz
range tested at the least favourable frequency
within the range (Hz) ........................................ :
Pass
ME EQUIPMENT with more than one RATED
Tested at 90/100/240/264V ac,
voltage, or both a.c./ d.c. tested in conditions
Single phase supply, AC
(see 5.4) related to the least favourable voltage, current.
nature of supply, and type of current .............. :
Pass
ME EQUIPMENT tested with alternative
No such construction
ACCESSORIES and components specified in
ACCOMPANYING DOCUMENTS to result in the least
favourable conditions ...................................... :
N/A
ME EQUIPMENT connected to a separate power
supply as specified in instructions for use
N/A
No such construction
When failure occurred or probability of future
failure detected during sequence of tests, per
agreement with manufacturer, all tests
affecting results conducted on a new sample
Pass
Alternatively, upon repair and modification of
the sample, only the relevant tests conducted
Pass
ME EQUIPMENT or parts thereof affected by
See Appended RM Results
climatic conditions were set up completely, or Table 5.7
partially, with covers detached and subjected
to a humidity preconditioning prior to tests of
Clauses 8.7.4 and 8.8.3 ...................................... :
Pass
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 13 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Manually detachable parts removed and
treated concurrently with major parts and
manually removable ACCESS COVERS were
opened and detached
No such construction
ME EQUIPMENT heated to a temperature between 21.9 ºC
T and T + 4 °C for at least 4 h and placed in a
93 %RH
humidity chamber with a relative humidity of 93 48 h
% ± 3 % and an ambient within 2 °C of T in the
range of + 20 °C to + 32 °C for 48 h
N/A
Pass
When RISK MANAGEMENT PROCESS indicated ME
EQUIPMENT can be exposed to high humidity for
extended periods (i.e., out-door use), test time
extended proportionally (h) ............................. :
N/A
5.8
Unless stated otherwise, tests in this standard
sequenced as in Annex B to prevent results of
one test on a subsequent test
Pass
5.9
Determination of APPLIED PARTS and ACCESSIBLE PARTS
Pass
5.9.1
APPLIED PARTS identified by inspection and
See clause 4.6
reference to ACCOMPANYING DOCUMENTS............ :
Bedside Terminal was
defined for the equipment.
Pass
5.9.2
ACCESSIBLE PARTS
Pass
5.9.2.1
Accessibility, when necessary, determined
using standard test finger of Fig 6 applied in a
bent or straight position
See Appended Table 5.9.2
Openings preventing entry of test finger of Fig.
6 mechanically tested with a straight un-jointed
test finger of the same dimensions with a force
of 30 N
Pass
When the straight un-jointed test finger
entered, test with the standard test finger (Fig
6) was repeated, if necessary, by pushing the
finger through the opening
5.9.2.2
5.9.2.3
N/A
Test hook of Fig. 7 inserted in all openings of
Openings designed to
ME EQUIPMENT and pulled with a force of 20 N for prevent the possibility for
10 s
test hook to insert.
Pass
All additional parts that became accessible
checked using standard test finger and by
inspection
N/A
Conductive parts of actuating mechanisms of No accessible conductive
electrical controls accessible after removal of
part
handles, knobs, levers and the like regarded as
ACCESSIBLE PARTS............................................. :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 14 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Conductive parts of actuating mechanisms
not considered ACCESSIBLE PARTS when
removal of handles, knobs, etc. required use
of a TOOL, and inspection of RISK MANAGEMENT
FILE indicated the relevant part is unlikely to
detach unintentionally during EXPECTED
SERVICE LIFE of ME EQUIPMENT ...........................:
N/A
6
CLASSIFICATION OF ME EQUIPMENT AND ME SYSTEMS
Pass
6.2
CLASS I ME EQUIPMENT, externally powered
Pass
Class I
CLASS II ME EQUIPMENT, externally powered
N/A
INTERNALLY POWERED ME EQUIPMENT
N/A
EQUIPMENT with
means of connection to a
Device is a Slim Medical
with CLASS I or CLASS II ME Panel PC which connected to
EQUIPMENT requirements when so connected,
a switching power adapter.
and when not connected to SUPPLY MAINS with
INTERNALLY POWERED ME EQUIPMENT requirements
N/A
TYPE B APPLIED PART
N/A
SUPPLY MAINS complied
No applied part
TYPE BF APPLIED PART
N/A
TYPE CF APPLIED PART
N/A
DEFIBRILLATION-PROOF APPLIED PARTS
N/A
6.3
ENCLOSURES classified according to degree of
IPX0
protection against ingress of water and
See Appended RM Result
particulate matter (IPN1N2) as per IEC 60529..... :
Table 11.6.5
N/A
6.4
ME EQUIPMENT or its parts intended to be
No such function for the
sterilized classified according to method(s) of product.
sterilization in instructions for use.................. :
The application of this clause
shall be evaluated in end
application
N/A
6.5
ME EQUIPMENT and ME SYSTEMS intended for use
in an OXYGEN RICH ENVIRONMENT classified for
such use and complied with 11.2.2
N/A
Equipment not suitable for
use in the presence of
OXYGEN RICH ENVIRONMENT
6.6
CONTINUOUS or Non-CONTINUOUS OPERATION ...... : CONTINUOUS
Pass
7
ME EQUIPMENT IDENTIFICATION, MARKING, AND DOCUMENTS
Pass
7.1.1
RISK of poor USABILITY associated with the
design of ME EQUIPMENT'S identification and
marking addressed in a USABILITY ENGINEERING
PROCESS .............................................................. :
Test Report No.: 1307033001
60601-1TRF_A
Page 15 of 203
The equipment which used
for information display and
its intended performance
will not be affect by
usability.
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
7.1.2
Legibility of Markings Test for Markings
See appended Tables 7.1.2
specified in Clause 7.2-7.6.................................. :
Pass
7.1.3
Required markings can be removed only with a
TOOL or by appreciable force, are durable and
remain CLEARLY LEGIBLE during EXPECTED SERVICE
LIFE of ME EQUIPMENT in NORMAL USE
Pass
a) After tests, adhesive labels didn’t loosen up See appended Tables 7.1.3
or curl up at edges and markings complied with
requirements in Clause 7.1.2 ............................. :
Pass
b) Markings required by 7.2-7.6 remained
See appended Tables 7.1.3
CLEARLY LEGIBLE after marking durability test ... :
Pass
7.2
Marking on the outside of ME EQUIPMENT or ME EQUIPMENT parts
Pass
7.2.1
At least markings in 7.2.2, 7.2.5, 7.2.6 (not for
See marking plate for detail.
PERMANENTLY INSTALLED ME EQUIPMENT ), 7.2.10,
and 7.2.13 were applied when size of
EQUIPMENT , its part, an ACCESSORY , or
ENCLOSURE did not permit application of all
required markings ............................................. :
Pass
Remaining markings fully recorded in
ACCOMPANYING DOCUMENTS .................................. :
Pass
7.2.2
Result - Remark
Verdict
Markings applied to individual packaging when No such markings
impractical to apply to ME EQUIPMENT
N/A
A material, component, ACCESSORY, or ME
No single use construction
EQUIPMENT intended for a single use, or its
packaging marked “Do Not Reuse” or with
symbol 28 of Table D.1 (ISO 7000-1051,
DB:2004-01) .................................................................................... :
N/A
MANUFACTURER’s name or trademark marked on
ME EQUIPMENT and detachable components ....... :
Pass
provided on the label
7.2.3
Misidentification does not present an
unacceptable risk
See Appended RM Results
Table 7.2.2
Pass
MODEL OR TYPE REFERENCE also marked, except
when misidentification would not present an
unacceptable RISK ............................................. :
Model name xxxxx-ONYX2121DTy-xxxxxxx (where y
can be T or blank; x can be
0-9, A-Z, “-” or blank for
marketing purpose)
Pass
Software forming part of a PEMS identified with
a unique identifier, such as revision level or
date of release/issue, and identification are
available to designated persons ...................... :
No such part
Symbol 11 on Table D.1 (ISO 7000-1641, DB:
2004-01) used, optionally, advice to OPERATOR
to consult ACCOMPANYING DOCUMENTS
Provided on installation
Instruction, see Installation
Instruction for detail
Test Report No.: 1307033001
60601-1TRF_A
Page 16 of 203
N/A
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
7.2.4
7.2.5
7.2.6
Requirement + Test
Result - Remark
Verdict
Safety sign 10 on Table D.2 (safety sign IEC
60878 Safety 01) used, advising OPERATOR that
ACCOMPANYING DOCUMENTS must be consulted
N/A
ACCESSORIES marked with name or trademark of No accessories
MANUFACTURER or supplier, and with a MODEL or
TYPE REFERENCE ........................................................................... :
N/A
Markings applied to individual packaging when
not practical to apply to ACCESSORIES
N/A
MODEL or TYPE REFERENCE of equipment to be
connected to ME EQUIPMENT to provide power, is
marked adjacent to the relevant connection
point when this connection could result in an
unacceptable RISK .............................................. :
No such construction
See Appended RM Results
Table 7.2.5
Connection to the Supply Mains
Pass
Except for PERMANENTLY INSTALLED ME
EQUIPMENT , marking appearing on the outside
of part containing SUPPLY MAINS connection
and, adjacent to connection point
Pass
For PERMANENTLY INSTALLED ME EQUIPMENT,
NOMINAL supply voltage or range marked inside
or outside of ME EQUIPMENT, preferably, adjacent
to supply connection terminals
Not permanently installed
me equipment
N/A
– RATED supply voltage(s) or RATED voltage
range(s) with a hyphen (-) between minimum
and maximum voltages (V, V-V) ........................ :
For Adapter: 100-240Vac
Pass
For Panel PC: 24Vdc
Multiple RATED supply voltages or multiple RATED No Multiple RATED supply
supply voltage ranges are separated by (V/V) .. : voltages or multiple RATED
supply voltage ranges
– Nature of supply (e.g., No. of phases, except
single-phase) and type of current ..................... :
7.2.7
Single phase
N/A
Pass
Symbols 1-5, Table D.1 (symbols of IEC 60417- Symbol IEC 60417-5032 was
5032, 5032-1, 5032-2, 5031, and 5033, all DB:
used for external adapter
2002-10) used, optionally, for same parameters:
Pass
– RATED supply frequency or RATED frequency
range in hertz..................................................... :
50/60Hz
Pass
– Symbol 9 of Table D.1 (symbol IEC 604175172, DB: 2003-02) used for CLASS II ME
EQUIPMENT ........................................................... :
Class I equipment
RATED input in amps or volt-amps, or in watts
when power factor exceeds 0.9 (A, VA, W) ...........:
Rated input in amps marked
Test Report No.: 1307033001
60601-1TRF_A
N/A
Page 17 of 203
N/A
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
RATED input for one or more RATED voltage
ranges provided for upper and lower limits of
the range or ranges when the range(s) is/are
greater than ± 10 % of the mean value of
specified range (A, VA,W) ................................. :
For Adapter: 1.8-0.8A
Input at mean value of range marked when
range limits do not differ by more than 10 %
from mean value (A, VA, W) .............................. :
No such marking
N/A
Marking includes long-time and most relevant
No such marking
momentary volt-ampere ratings when provided,
each plainly identified and indicated in
ACCOMPANYING DOCUMENTS (VA) .......................... :
N/A
Marked input of ME EQUIPMENT provided with
means for connection of supply conductors of
other electrical equipment includes RATED and
marked output of such means (A, VA, W) ......... :
Verdict
Pass
For Panel PC: 2.91A
Panel PC Unit : 24Vdc, 2.91A
Pass
Power Adapter: 100-240Vac,
50/60 Hz, 1.8-0.8A
7.2.8
Output connectors
Pass
7.2.8.1
See 16.9.2.1 b) for MULTIPLE SOCKET-OUTLETS
integral with ME EQUIPMENT
7.2.8.2
Output connectors are marked, except for
Single output for external
MULTIPLE SOCKET-OUTLETS or connectors intended adapter and the rating
for specified ACCESSORIES or equipment
marked on the label.
No MULTIPLE SOCKET-OUTLETS
N/A
Pass
Rated Voltage (V), Rated Current (A) ............... :
—
Rated Power (W), Output Frequency (Hz) ........ :
—
7.2.9
ME EQUIPMENT or its parts marked with the IP
The equipment classified
environmental Code per IEC 60529 according to IPX0 need not be marked.
classification in 6.3 (Table D.3, Code 2) ............ :
Pass
7.2.10
Degrees of protection against electric shock as
classified in 6.2 for all APPLIED PARTS marked
with relevant symbols as follows (not applied to
parts identified according to 4.6):
Pass
TYPE B APPLIED PARTS with symbol 19 of Table D.1
(IEC 60417-5840, DB: 2002-10), not applied in
such a way as to give the impression of being
inscribed within a square in order to distinguish
it from symbol IEC 60417-5333.......................... :
N/A
TYPE BF APPLIED PARTS with symbol 20 of Table
D.1 (IEC 60417-5333, DB: 2002-10) .................... :
N/A
TYPE CF APPLIED PARTS with symbol 21 of Table
D.1 (IEC 60417-5335, DB: 2002-10) .................... :
N/A
DEFIBRILLATION-PROOF APPLIED PARTS marked with
symbols 25-27 of Table D.1 (IEC 60417-5841, IEC
60417-5334, or IEC 60417-5336, all DB: 2002-10):
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 18 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
7.2.11
Requirement + Test
Result - Remark
Proper symbol marked adjacent to or on
connector for APPLIED PART, except marked on
APPLIED PART when there is no connector, or
connector used for more than one APPLIED PART
and different APPLIED PARTS with different
classifications ................................................... :
N/A
Safety sign 2 of Table D.2 (ISO 7010-W001)
placed near relevant outlet when protection
against effect of discharge of a cardiac
defibrillator is partly in the PATIENT cable .......... :
N/A
An explanation indicating protection of ME
EQUIPMENT against effects of discharge of a
cardiac defibrillator depends on use of proper
cables included in instructions for use ............. :
N/A
ME EQUIPMENT not marked to the contrary
assumed to be suitable for CONTINUOUS
OPERATION
7.2.12
7.2.13
Verdict
The equipment work
continuously without
specified marking
Pass
DUTY CYCLE for ME EQUIPMENT intended for nonCONTINUOUS OPERATION appropriately marked to
provide maximum “on” and “off” time.............. :
N/A
Type and full rating of a fuse marked adjacent to Evaluated in part of power
ACCESSIBLE fuse-holder
supply
N/A
Fuse type ........................................................... :
—
Voltage (V) and Current (A) rating .................... :
—
Operating speed (s) and Breaking capacity .... :
—
A safety sign CLEARLY LEGIBLE and visible after
No physiological effects
INSTALLATION in NORMAL USE applied to a
prominent location of EQUIPMENT that produce
physiological effects capable of causing HARM to
PATIENT or OPERATOR not obvious to OPERATOR .. :
N/A
Nature of HAZARD and precautions for avoiding
or minimizing the associated RISK described in
instructions for use ........................................... :
N/A
See Appended RM Results
Table 7.2.13
7.2.14
HIGH VOLTAGE TERMINAL DEVICES on the outside of No such construction
ME EQUIPMENT accessible without the use of a
TOOL marked with symbol 24 of Table D.1
(symbol IEC 60417-5036, DB: 2002-10)
N/A
7.2.15
Requirements for cooling provisions marked
(e.g., supply of water or air)............................... :
No special air shall be
provided for cooling but
nature air, no marking is
required
N/A
7.2.16
ME EQUIPMENT with
No special limitation in
mechanical stability
N/A
Test Report No.: 1307033001
60601-1TRF_A
limited mechanical stability
Page 19 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.2.17
Packaging marked with special handling
instructions for transport and/or storage.......... :
Packaging with special
transportation and storage
condition – see carton label
for detail
Pass
Permissible environmental conditions for
transport and storage marked on outside of
packaging ......................................................... :
See carton label for detail
Pass
Packaging marked with a suitable safety sign
No such packaging
indicating premature unpacking of ME EQUIPMENT
could result in an unacceptable RISK................. :
N/A
Packaging of sterile ME EQUIPMENT or
ACCESSORIES marked sterile
No sterilized packaging
N/A
7.2.18
RATED maximum supply pressure from an
external source marked on ME EQUIPMENT
adjacent to each input connector..................... :
No such construction
N/A
7.2.19
Symbol 7 of Table D.1 (IEC 60417-5017,
DB:2002-10) marked on FUNCTIONAL EARTH
TERMINAL ............................................................. :
No functional earth terminal
N/A
7.2.20
Protective means, required to be removed to
use a particular function of ME EQUIPMENT with
alternate applications, marked to indicate the
necessity for replacement when the function is
no longer needed .............................................. :
No such protective means is
required.
N/A
No marking applied when an interlock provided
N/A
7.3
Marking on the inside of ME EQUIPMENT or ME EQUIPMENT parts
N/A
7.3.1
Maximum power loading of heating elements or No heating elements or
lamp-holders designed for use with heating
lamp-holders provided
lamps marked near or in the heater (W) ............ :
N/A
A marking referring to ACCOMPANYING DOCUMENTS
provided for heating elements or lamp-holders
designed for heating lamps that can be changed
only by SERVICE PERSONNEL using a TOOL
N/A
7.3.2
Symbol 24 of Table D.1 (symbol IEC 60417-5036,
DB: 2002-10), or safety sign 3 of Table D.2 used
to mark presence of HIGH VOLTAGE parts ............ :
N/A
7.3.3
Type of battery and mode of insertion when
applicable is marked ......................................... :
An identifying marking provided referring to
instructions in ACCOMPANYING DOCUMENTS for
batteries intended to be changed only by
SERVICE PERSONNEL using a TOOL ........................ :
Test Report No.: 1307033001
60601-1TRF_A
Page 20 of 203
RTC battery provided.
Pass
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
A warning provided indicating replacement of
lithium batteries or fuel cells when incorrect
replacement by inadequately trained personnel
would result in an unacceptable RISK (e.g.,
excessive temperatures, fire or explosion) ...... :
An identifying marking also provided referring
to instructions in ACCOMPANYING DOCUMENTS ..... :
7.3.4
7.3.5
N/A
See appended RM Results
Table 7.3.3
Pass
Fuses, replaceable THERMAL CUT-OUTS and OVER- Evaluated in part of power
CURRENT RELEASES, accessible by use of a TOOL,
supply
marked by type and full rating at the component
or by reference to ACCOMPANYING DOCUMENTS
N/A
Type .................................................................. :
—
Voltage (V) and Current (A) rating .................... :
—
Operating speed (s) and Breaking capacity .... :
—
PROTECTIVE EARTH TERMINAL marked with symbol
6 of Table D.1 (IEC 60417-5019, DB: 2002-10),
except for the PROTECTIVE EARTH TERMINAL in an
APPLIANCE INLET according to IEC 60320-1
Evaluated in part of power
supply
N/A
Markings on or adjacent to PROTECTIVE EARTH
TERMINALS not applied to parts requiring removal
to make the connection, and remained visible
after connection made
N/A
7.3.6
Symbol 7 of Table D.1 (IEC 60417-5017, DB:
No FUNCTIONAL EARTH
2002 -10) marked on FUNCTIONAL EARTH TERMINALS TERMINALS
N/A
7.3.7
Terminals for supply conductors marked
adjacent to terminals, except when no HAZARD
would result when interchanging connections
N/A
Evaluated in part of power
supply
Terminal markings included in ACCOMPANYING
DOCUMENTS when ME EQUIPMENT too small to
accommodate markings
N/A
Terminals exclusively for neutral supply
conductor in PERMANENTLY INSTALLED ME
EQUIPMENT marked with Code 1 of Table D.3
(Code in IEC 60445)
Not PERMANENTLY INSTALLED
Marking for connection to a 3-phase supply, if
necessary, complies with IEC 60445
Single phase
N/A
Markings on or adjacent to electrical connection No such construction
points not applied to parts requiring removal to
make connection, and remained visible after
connection made
Test Report No.: 1307033001
60601-1TRF_A
Page 21 of 203
N/A
ME EQUIPMENT
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
7.3.8
“For supply connections, use wiring materials
Evaluated in part of power
suitable for at least X °C” (where X > than max
supply
temperature measured in terminal box or wiring
compartment under NORMAL USE), or equivalent,
marked at the point of supply connections
N/A
Statement not applied to parts requiring
removal to make the connection, and CLEARLY
LEGIBLE after connections made
N/A
7.4
Marking of controls and instruments
N/A
7.4.1
The “on” & “off” positions of switch to control
power to ME EQUIPMENT or its parts, including
mains switch, marked with symbols 12 and 13
of Table D.1 (IEC 60417-5007, DB: 2002-10, and
IEC 60417-5008, DB: 2002-10), or
– indicated by other unambiguous means
N/A
The “on/off” positions of push button switch
No ON/OFF switch provided
with bi-stable positions marked with symbol 14
of Table D.1 (IEC 60417-5010 DB: 2002-10), and
N/A
– status indicated by adjacent indicator light
N/A
– status indicated by other unambiguous
means
N/A
No ON/OFF switch provided
N/A
– status indicated by adjacent indicator light
N/A
– status indicated by other unambiguous
means
N/A
Different positions of control devices/switches The symbol
is provided
indicated by figures, letters, or other visual
on stand-by switch of the
means
equipment.
No setting / control will
result risk to patient when
normal use.
Pass
N/A
– an indication of direction in which magnitude
of the function changes
N/A
Numeric indications of parameters on ME
No such function
EQUIPMENT expressed in SI units according to
ISO 31 except the base quantities listed in Table
1 expressed in the indicated units
N/A
Test Report No.: 1307033001
60601-1TRF_A
N/A
N/A
Controls provided with an associated
indicating device when change of setting of a
control could result in an unacceptable RISK to
PATIENT in NORMAL USE, or
7.4.3
No ON/OFF switch provided
– indicated by an adjacent indicator light, or
The “on/off” positions of push button switch
with momentary on position marked with
symbol 15 of Table D.1 (symbol 60417-5011
DB: 2002-10), or
7.4.2
Verdict
Page 22 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
7.5
Requirement + Test
Result - Remark
Verdict
ISO 1000 applied for application of SI units,
their multiples, and certain other units
N/A
All Markings in Sub-clause 7.4 complied with
tests and criteria of 7.1.2 and 7.1.3
N/A
Safety signs
N/A
Markings used to convey a warning,
prohibition or mandatory action mitigating a
RISK not obvious to OPERATOR are safety signs
from ISO 7010
No such situation
N/A
Affirmative statement together with safety sign
placed in instructions for use if insufficient
space on ME EQUIPMENT
N/A
Specified colours in ISO 3864-1 used for safety
signs .................................................................. :
N/A
Safety notices include appropriate precautions
or instructions on how to reduce RISK(S)
N/A
Safety signs including any supplementary text
or symbols described in instructions for use
N/A
7.6
Symbols
Pass
7.6.1
Meanings of symbols used for marking
described in instructions for use ..................... :
Explanation provided for
each related marking – see
“Safety Symbol
Description”
Pass
7.6.2
Symbols required by this standard conform to
IEC or ISO publication referenced
Complied with table D.1
Pass
7.6.3
Symbols used for controls and performance
conform to the IEC or ISO publication where
symbols are defined, as applicable
No such function
7.7
Colours of the insulation of conductors
7.7.1
PROTECTIVE EARTH CONDUCTOR identified by
green and yellow insulation
7.7.2
Insulation on conductors inside ME EQUIPMENT
forming PROTECTIVE EARTH CONNECTIONS
identified by green and yellow at least at
terminations
Pass
7.7.3
Green and yellow insulation identify only
following conductors:
Pass
– PROTECTIVE EARTH CONDUCTORS
Pass
Evaluated in part of power
supply
Evaluated in part of power
supply
– conductors specified in 7.7.2
60601-1TRF_A
Pass
Pass
Pass
– POTENTIAL EQUALIZATION CONDUCTORS
Test Report No.: 1307033001
N/A
Page 23 of 203
No such construction
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– FUNCTIONAL EARTH CONDUCTORS
No such construction
N/A
7.7.4
Neutral conductors of POWER SUPPLY CORDS are
“light blue” specified in IEC 60227-1 or IEC
60245-1
No power supply cord
provides.
N/A
7.7.5
Colours of conductors in POWER SUPPLY CORDS
in accordance with IEC 60227-1 or IEC 60245-1
No power supply cord
provides.
N/A
7.8
Indicator lights and controls
7.8.1
Red indicator lights mean: Warning (i.e.,
immediate response by OPERATOR required)
Pass
No red colour indicator light
Yellow indicator lights mean: Caution (i.e.,
prompt response by OPERATOR required)
Green indicator lights mean: Ready for use
N/A
Green Light for power on
Other colours, if used: Meaning other than red,
yellow, or green (colour, meaning) .................. :
7.8.2
Red used only for emergency control
7.9
ACCOMPANYING DOCUMENTS
7.9.1
ME EQUIPMENT accompanied by documents
containing at least instructions for use, and a
technical description
No such function
N/A
Pass
User Manual provided
Pass
Pass
– Name or trade-name of MANUFACTURER and an
address the RESPONSIBLE ORGANIZATION can be
referred to ......................................................... :
ONYX HEALTHCARE INC or
Pass
– MODEL or TYPE REFERENCE ............................... :
xxxxx-ONYX-2121DTyxxxxxxx (where y can be T
or blank; x can be 0-9, A-Z,
“-” or blank for marketing
purpose)
Pass
When ACCOMPANYING DOCUMENTS provided
See Appended RM Results
electronically (e.g., on CDROM ), RISK
Table 7.9.1
MANAGEMENT PROCESS includes instructions as
to what is required in hard copy or as markings
on ME EQUIPMENT (for emergency operation)
Pass
ACCOMPANYING DOCUMENTS specify special skills, See User Manual: Chapter 1
training, and knowledge required of OPERATOR or - 1.3 Specification
RESPONSIBLE ORGANIZATION and environmental
restrictions on locations of use
Pass
ACCOMPANYING DOCUMENTS written at a level
consistent with education, training, and other
needs of individuals for whom they are intended
Instructions for use include the required information
Test Report No.: 1307033001
60601-1TRF_A
Pass
N/A
ACCOMPANYING DOCUMENTS identify ME EQUIPMENT
by the following, as applicable:
7.9.2
N/A
Page 24 of 203
N/A
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.9.2.1
– intended use of ME EQUIPMENT,
See User Manual: Chapter 1
- 1.1 Introduction
– frequently used functions, and
No specified function be
defined as frequently use
– known contraindication(s) to use of ME
Pass
N/A
N/A
EQUIPMENT
Classifications as in Clause 6, all markings per
Clause 7.2, and explanation of safety signs and
symbols marked on ME EQUIPMENT
Instructions for use are in a language
acceptable to the intended operator
7.9.2.2
Pass
Written in English which is
international common-used
language
Pass
Instructions for use include all warning and
safety notices
Pass
Warning statement for CLASS I ME EQUIPMENT
See User Manual: “Safety &
indicating: “WARNING: To avoid risk of electric
Warranty”
shock, this equipment must only be connected
to a supply mains with protective earth”
Pass
Warnings regarding significant RISKS of
reciprocal interference posed by ME EQUIPMENT
during specific investigations or treatments
N/A
Information on potential electromagnetic or
other interference and advice on how to avoid
or minimize such interference
N/A
Warning statement for ME EQUIPMENT supplied
No multiple socket-outlet
with an integral MULTIPLE SOCKET-OUTLET
indicating, “connecting electrical equipment to
MSO effectively leads to creating an ME SYSTEM,
and can result in a reduced level of safety”
N/A
The RESPONSIBLE ORGANIZATION is referred to this No system been defined
standard for the requirements applicable to ME
N/A
SYSTEMS
7.9.2.3
Statement on ME EQUIPMENT for connection to a Specified power supply was
separate power supply indicating “power
indicated on the label
supply is specified as a part of ME EQUIPMENT or
combination is specified as a ME SYSTEM ”
7.9.2.4
Warning statement for mains- operated ME
EQUIPMENT with additional power source not
automatically maintained in a fully usable
condition indicating the necessity for periodic
checking or replacement of power source
The power supply is full
usable condition, no
warning statement is
required
N/A
Warning to remove primary batteries when ME
No primary battery used.
EQUIPMENT is not likely to be used for some time
when leakage from battery would result in an
unacceptable RISK ............................................ :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 25 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Specifications of replaceable INTERNAL
ELECTRICAL POWER SOURCE when provided ....... :
Warning indicating ME EQUIPMENT must be
connected to an appropriate power source
when loss of power source would result in an
unacceptable RISK ............................................ :
7.9.2.5
N/A
See Appended RM Results
Table 7.9.2.4
Instructions for use include a description of ME See User Manual: Chapter 1
EQUIPMENT , its functions, significant physical
and performance characteristics together with
the expected positions of OPERATOR, PATIENT, or
other persons near ME EQUIPMENT in NORMAL USE
N/A
Pass
Information provided on materials and
ingredients PATIENT or OPERATOR is exposed to
when such exposure can constitute an
unacceptable RISK
N/A
Restrictions specified on other equipment or
NETWORK/ DATA COUPLINGS, other than those
forming part of an ME SYSTEM , to which a SIGNAL
INPUT/OUTPUT PART may be connected
N/A
APPLIED PARTS specified
No APPLIED PARTS
7.9.2.6
Information provided indicating where the
installation instructions may be found or
information on qualified personnel who can
perform the installation
See User Manual: Chapter 2
Pass
7.9.2.7
Instructions provided indicating not to position
ME EQUIPMENT to make it difficult to operate the
disconnection device when an APPLIANCE
COUPLER or separable plug is used as isolation
means to meet 8.11.1 a)
The disconnection device of
the Bedside Terminal is
power supply cord, which
will not be difficult to
disconnect.
Pass
7.9.2.8
Necessary information provided for OPERATOR
to bring ME EQUIPMENT into operation including
initial control settings, and connection to or
positioning of PATIENT prior to use of ME
EQUIPMENT , its parts, or ACCESSORIES
See User Manual: Chapter 2
Pass
7.9.2.9
Information provided to operate ME EQUIPMENT
See User Manual: Chapter 1
including explanation of controls, displays and
signals, sequence of operation, connection of
detachable parts or ACCESSORIES, replacement
of material consumed during operation
Pass
Meanings of figures, symbols, warning
statements, abbreviations and indicator lights
described in instructions for use
Test Report No.: 1307033001
60601-1TRF_A
Page 26 of 203
See User Manual: “Safety
Symbol Description”
N/A
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.9.2.10
A list of all system messages, error messages,
and fault messages provided with an
explanation of messages including important
causes and possible action(s) to be taken to
resolve the problem indicated by the message
N/A
7.9.2.11
Information provided for the OPERATOR to safely See User Manual: Chapter 2
terminate operation of ME EQUIPMENT
Pass
7.9.2.12
Information provided on cleaning, disinfection, See User Manual: Appendix
and sterilization methods, and applicable
A
parameters that can be tolerated by ME
EQUIPMENT parts or ACCESSORIES specified
Pass
Components, ACCESSORIES or ME EQUIPMENT
marked for single use, except when required
by MANUFACTURER to be cleaned, disinfected, or
sterilized prior to use
7.9.2.13
Instructions provided on preventive
inspection, calibration, maintenance and its
frequency
No calibration/ maintenance
are required for this Slim
Medical Panel PC because
normal lifetime is expected
to be 50,000 hours.
N/A
Information provided for safe performance of
routine maintenance necessary to ensure
continued safe use of ME EQUIPMENT
N/A
Parts requiring preventive inspection and
maintenance to be performed by SERVICE
PERSONNEL identified including periods of
application
N/A
Instructions provided to ensure adequate
maintenance of ME EQUIPMENT containing
rechargeable batteries to be maintained by
anyone other than SERVICE PERSONNEL
7.9.2.14
N/A
See User Manual: “Safety &
Warranty”
Pass
A list of ACCESSORIES, detachable parts, and
materials for use with ME EQUIPMENT provided
N/A
Other equipment providing power to ME SYSTEM
sufficiently described (e.g. part number, RATED
VOLTAGE, max or min power, protection class,
intermittent or continuous service)
N/A
7.9.2.15
RISKS associated with disposal of waste
See User Manual: Appendix
products, residues, etc., and of ME EQUIPMENT
A
and ACCESSORIES at the end of their EXPECTED
SERVICE LIFE are identified, and instructions
provided on minimizing these RISKS ................. :
Pass
7.9.2.16
Instructions for use include information
specified in 7.9.3 or identify where it can be
found (e.g. in a service manual)
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 27 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.9.3
Technical description
Pass
7.9.3.1
All essential data provided for safe operation, See below
transport, storage, and measures or conditions
necessary for installing ME EQUIPMENT, and
preparing it for use including the following:
Pass
– information as in clause 7.2
Pass
– permissible environmental conditions of use
including conditions for transport and storage
See User Manual: Chapter 1
- 1.3 Specification
Pass
– all characteristics of ME EQUIPMENT including
range(s), accuracy, and precision of displayed
values or where they can be found
See User Manual: Chapter 1
- 1.3 Specification
Pass
– special installation requirements such as
max. permissible apparent impedance of
N/A
SUPPLY MAINS
– permissible range of values of inlet pressure
and flow, and chemical composition of cooling
liquid used for cooling
– a description of means of isolating ME
EQUIPMENT from SUPPLY MAINS, when such
means not in ME EQUIPMENT
N/A
Medical grade power supply
provided as means of
insulation
– a description of means for checking oil level
in partially sealed oil filled ME EQUIPMENT or its
parts when applicable
N/A
– a warning statement addressing HAZARDS that See below
can result from unauthorized modification of
ME EQUIPMENT according to following examples
Pass
“warning: No modification of this equipment is See User Manual - Safety &
allowed”
Warranty
Pass
“warning: Do not modify this equipment
without authorization of the manufacturer”
N/A
“warning: If this equipment is modified,
appropriate inspection and testing must be
conducted to ensure continued safe use of
equipment”
N/A
Technical description separable from instructions for use contains required
information, as follows
N/A
– information as in clause 7.2
N/A
– all applicable classifications in Clause 6,
warning and safety notices, and explanation of
safety signs marked on ME EQUIPMENT
N/A
– a brief description of ME EQUIPMENT, how it
functions, and its significant physical and
performance characteristics
N/A
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 28 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
7.9.3.2
Requirement + Test
Result - Remark
Verdict
MANUFACTURER’S optional requirements for
minimum qualifications of SERVICE PERSONNEL
documented in technical description
N/A
The technical description contains the following required information
N/A
–type and full rating of fuses used in SUPPLY
Not permanently installed
MAINS external to PERMANENTLY INSTALLED ME
me equipment
EQUIPMENT , when type and rating of fuses are
not apparent from information on RATED current
and mode of operation of ME EQUIPMENT .......... :
N/A
– a statement for ME EQUIPMENT with a nonDetachable power supply
DETACHABLE POWER SUPPLY CORD if POWER SUPPLY cord used.
CORD is replaceable by SERVICE PERSONNEL, and
if so, instructions for correct connection and
anchoring to ensure compliance with 8.11.3
N/A
– instructions for correct replacement of
interchangeable or detachable parts specified
by MANUFACTURER as replaceable by SERVICE
PERSONNEL, and
N/A
No such instruction is
required
– warnings identifying nature of HAZARD when
replacement of a component could result in an
unacceptable RISK, and when replaceable by
SERVICE PERSONNEL all information necessary to
safely replace the component
N/A
7.9.3.3
Technical description indicates, MANUFACTURER See User Manual: “Safety &
will provide circuit diagrams, component part
Warranty”
lists, descriptions, calibration instructions to
assist to SERVICE PERSONNEL in parts repair
Pass
7.9.3.4
Means used to comply with requirements of
Isolation power supply
8.11.1 clearly identified in technical description provided as source of
electricity
Pass
8
PROTECTION AGAINST ELECTRICAL HAZARDS FROM ME EQUIPMENT
Pass
8.1
Limits specified in Clause 8.4 not exceeded for See appended RM Results
ACCESSIBLE PARTS and APPLIED PARTS in NORMAL
Table 8.1 b(1), 8.1 b(2) and
or SINGLE FAULT CONDITIONS
8.1 b(3)
Pass
NORMAL CONDITION considered as simultaneous
occurrence of situations identified in 8.1a)
Pass
SINGLE FAULT CONDITION considered to include
See appended RM Results
the occurrences as specified in Clause 8.1b) ... : Table 8.1 b(1), 8.1 b(2) and
8.1 b(3)
Pass
ACCESSIBLE PARTS determined according to 5.9
Pass
LEAKAGE CURRENTS measured according to 8.7
Pass
Requirements related to power sources
Pass
8.2
Test Report No.: 1307033001
60601-1TRF_A
Page 29 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
8.2.1
When ME EQUIPMENT specified for connection to Separated power supply
a separate power source other than SUPPLY
was considered as part of
MAINS, separate power source considered as
ME Equipment
part of ME EQUIPMENT or combination
considered as an ME SYSTEM
Pass
Tests performed with ME EQUIPMENT connected
to separate power supply when one specified
Pass
8.2.2
8.3
Result - Remark
Verdict
When a generic separate power supply
specified, specification in ACCOMPANYING
DOCUMENTS examined
Specification of specified
power supply examined
Pass
No HAZARDOUS SITUATION other than absence of
ESSENTIAL PERFORMANCE developed when a
connection with wrong polarity made for ME
EQUIPMENT from an external d.c. source
Design of plug will prevent
wrong polarity connection
Pass
ME EQUIPMENT connected with correct polarity
did not present an unacceptable RISK
See appended RM Results
Table 8.2.2
Pass
Protective devices that can be reset by anyone No re-settable protective
without a TOOL restore correct operation on
device
reset
N/A
Classification of APPLIED PARTS
N/A
a) APPLIED PART specified in ACCOMPANYING
DOCUMENTS as suitable for DIRECT CARDIAC
APPLICATION is TYPE CF
Not suitable for DIRECT
CARDIAC APPLICATION is TYPE
CF
b) An APPLIED PART provided with a PATIENT
CONNECTION intended to deliver electrical
energy or an electrophysiological signal to or
from PATIENT is TYPE BF or CF APPLIED PART
c) An APPLIED PART not covered by a) or b) is a
TYPE B, BF, or CF
N/A
No applied part
N/A
d) Requirements of a TYPE B APPLIED PART
applied to a part in 4.6 to be subjected to
requirements for an APPLIED PART (except
marking)
Requirements for a TYPE BF or CF APPLIED PART
applied as in RISK MANAGEMENT PROCESS
N/A
N/A
See Appended RM Results
Table 8.3d
N/A
8.4
Limitation of voltage, current or energy
N/A
8.4.1
PATIENT CONNECTIONS intended
N/A
to deliver Current
Limits in 8.4.2 not applied to currents intended
to flow through body of PATIENT to produce a
physiological effect during NORMAL USE
8.4.2
ACCESSIBLE PARTS
Test Report No.: 1307033001
60601-1TRF_A
including APPLIED PARTS
Page 30 of 203
N/A
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
a) Currents from, to, or between PATIENT
See Appended Table 8.7
CONNECTIONS did not exceed limits for PATIENT
LEAKAGE CURRENT and PATIENT AUXILIARY CURRENT
per Tables 3 and 4 when measured according
to Clause 8.7.4.................................................. :
Pass
b) LEAKAGE CURRENTS from, to, or between
ACCESSIBLE PARTS did not exceed limits for
TOUCH CURRENT in Cl. 8.7.3 c) when measured
per Clause 8.7.4 (mA)....................................... :
See Appended Table 8.7
Pass
c) Limits specified in b) not applied to parts
when probability of a connection to a PATIENT,
directly or through body of OPERATOR, is
negligible in NORMAL USE, and the OPERATOR is
appropriately instructed
See appended Table 8.7 and
RM Results Table 8.4.2 c)
Pass
– accessible contacts of connectors
N/A
– contacts of fuseholders accessible during
replacement of fuse
N/A
– contacts of lampholders accessible after
removal of lamp
N/A
– parts inside an ACCESS COVER that can be
opened without a TOOL, or where a TOOL is
needed but the instructions for use instruct an
OPERATOR other than SERVICE PERSONNEL to open
the relevant ACCESS COVER
N/A
Voltage to earth or to other ACCESSIBLE PARTS
Evaluated in part of power
did not exceed 42.4 V peak a.c. or 60 V d.c. for supply
above parts in NORMAL or single fault condition
(V a.c. or d.c.) ................................................... :
N/A
Limit of 60 V d.c applied with no more than
10% peak-to-peak ripple, and when ripple
larger than specified value, 42.4 V peak limit
applied (V d.c.) ................................................. :
N/A
Energy did not exceed 240 VA for longer than
60 s or stored energy available did not exceed
20 J at a potential up to 2 V (VA or J) .............. :
N/A
LEAKAGE CURRENT limits referred to in 8.4.2 b)
applied when voltages higher than limits in
8.4.2 c) were present (mA) ............................... :
N/A
d) Voltage and energy limits specified in c)
above also applied to the following:
– internal parts, other than contacts of plugs,
connectors and socket-outlets, touchable by
test pin in Fig 8 inserted through an opening in
an ENCLOSURE; and
Test Report No.: 1307033001
60601-1TRF_A
Page 31 of 203
Pass
Product was designed with
construction that internal
live part can’t be touch by
test pin , rod and hook
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.4.3
8.4.4
Requirement + Test
Verdict
– internal parts touchable by a metal test rod
with a diameter of 4 mm and a length of 100
mm, inserted through any opening on top of
ENCLOSURE or through any opening provided
for adjustment of pre-set controls using a TOOL
Pass
Test pin or the test rod inserted through
relevant openings with minimal force of no
more than 1 N
Pass
Test rod inserted in every possible position
through openings provided for adjustment of
pre-set controls that can be adjusted in NORMAL
USE, with a force of 10 N
Pass
Test repeated with a TOOL specified in
instructions for use
N/A
Test rod freely and vertically suspended
through openings on top of ENCLOSURE
N/A
e) Devices used to de-energize parts when an Need tool to open access
ACCESS COVER opened without a TOOL gives
cover.
access to parts at voltages above levels
permitted by this Clause comply with 8.11.1 for
mains isolating switches and remain effective
in SINGLE FAULT CONDITION
N/A
A TOOL is required when it is possible to
prevent the devices from operating
N/A
Worst case voltage between pins of plug and
between either supply pin and ENCLOSURE did
not exceed 60 V one s after disconnecting the
plug of ME EQUIPMENT or its parts (V)................ :
Evaluated in part of power
supply
N/A
A triggering circuit used to ensure
disconnection occurred at peak of supply
voltage waveform
N/A
When voltage exceeded 60 V, calculated or
Evaluated in part of power
measured stored charge didn’t exceed 45 μC .. : supply
N/A
Residual voltage of conductive parts of
Evaluated in part of power
capacitive circuits, having become accessible supply
after ME EQUIPMENT was de-energized after
removal of ACCESS COVERS, didn’t exceed 60V or
calculated stored charge didn’t exceed 45μC .. :
N/A
A device manually discharging capacitors
used when automatic discharging was not
possible and ACCESS COVERS could be removed
only with aid of a TOOL
N/A
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 32 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Capacitor(s) and connected circuitry marked
with symbol 24 of Table D.1 (IEC 60417-5036,
DB: 2002-10), and manual discharging device
specified in technical description ................... :
N/A
8.5
Separation of parts
Pass
8.5.1
MEANS OF PROTECTION (MOP)
Pass
8.5.1.1
Two MEANS of PROTECTION provided for ME
EQUIPMENT to prevent APPLIED and other
ACCESSIBLE PARTS from exceeding limits in 8.4
Pass
Each MEANS OF PROTECTION categorized as a
MEANS OF PATIENT PROTECTION or a MEANS OF
OPERATOR PROTECTION, taking into account
Clause 4.6, and flow chart in Fig A.12
Evaluated in part of power
supply - 2 MOPP provided
between primary to
secondary circuit
Varnishing, enameling, oxidation, and similar
protective finishes and coatings with sealing
compounds replasticizing at temperatures
expected during operation and sterilization
disregarded as MEANS OF PROTECTION
No such construction
Coatings and other insulation intended as a
MOPP considered and
MEANS OF PROTECTION complying with IEC 60950- provided.
1:2001 considered acceptable as a MEANS OF
OPERATOR PROTECTION but not automatically as a
Pass
N/A
N/A
MEANS OF PATIENT PROTECTION
8.5.1.2
RISK MANAGEMENT PROCESS taken into
consideration for MEANS OF PATIENT PROTECTION
N/A
Components and wiring forming a MEANS OF
PROTECTION comply with 8.10
N/A
Insulation, CREEPAGE, CLEARANCES, components
or earth connections not complying with
8.5.1.2 and 8.5.1.3 not considered as MEANS OF
PROTECTION, and failure of these parts regarded
as NORMAL CONDITION
N/A
MEANS OF PATIENT PROTECTION
(MOPP)
Pass
Solid insulation forming a MEANS OF PATIENT
PROTECTION complied with dielectric strength
test of Clause 8.8 at test voltage of Table 6
CREEPAGE and CLEARANCES forming a MEANS OF
PATIENT PROTECTION complied with Table 12
Evaluated in part of power
supply - 2 MOPP provided
between primary to
secondary circuit
Pass
PROTECTIVE EARTH CONNECTIONS forming a MEANS Evaluated in part of power
supply - 2 MOPP provided
OF PATIENT PROTECTION complied with Cl. 8.6
between primary to
secondary circuit
Pass
Test Report No.: 1307033001
60601-1TRF_A
N/A
Page 33 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
A Y1 capacitor complying with IEC 60384-14
and having passed dielectric strength test for
two MEANS OF PATIENT PROTECTION considered
equivalent to one MEANS OF PATIENT PROTECTION
Evaluated in part of power
supply
N/A
Two capacitors used in series, each RATED for Evaluated in part of power
total WORKING VOLTAGE across the pair and have supply - 2 MOPP provided
between primary to
the same NOMINAL capacitance
secondary circuit
Pass
—
Voltage Total Working (V) and C Nominal (F) ............ :
8.5.1.3
MEANS OF OPERATOR PROTECTION
(MOOP)
N/A
Solid insulation forming a MEANS OF OPERATOR
PROTECTION complied with:
N/A
– dielectric strength test of 8.8 at test voltage
of Table 6; or
N/A
– requirements of IEC 60950-1 for INSULATION
CO- ORDINATION
N/A
CREEPAGE and CLEARANCES forming a MEANS OF
OPERATOR PROTECTION complied with:
N/A
– limits of Tables 13 to 16 (inclusive); or
Evaluated in part of power
supply - 2 MOPP provided
between primary to
secondary circuit
N/A
– requirements of IEC 60950-1 for INSULATION
CO- ORDINATION
N/A
PROTECTIVE EARTH CONNECTIONS forming a MEANS Evaluated in part of power
OF OPERATOR PROTECTION complied with Cl. 8.6,
supply
or
N/A
– requirements and tests of IEC 60950-1 for
Evaluated in part of power
protective earthing............................................ : supply
N/A
A Y2 capacitor complying with IEC 60384-14
and passing dielectric strength test for one
MEANS OF OPERATOR PROTECTION considered
equivalent to one MEANS OF OPERATOR
PROTECTION ......................................................... :
N/A
A Y1 capacitor complying with IEC 60384-14
Evaluated in part of power
and having passed dielectric strength test for supply
two MEANS OF OPERATOR PROTECTION considered
equivalent to two MEANS OF OPERATOR
PROTECTION ......................................................... :
N/A
Two capacitors used in series each RATED for
Evaluated in part of power
total WORKING VOLTAGE across the pair and have supply
the same NOMINAL capacitance
N/A
Voltage Total Working (V) and C Nominal (F) ............ :
Test Report No.: 1307033001
60601-1TRF_A
MOPP provided
Page 34 of 203
—
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Points at which impedances of components,
CREEPAGE, CLEARANCES, PROTECTIVE EARTH
CONNECTIONS or insulation, prevent ACCESSIBLE
PARTS from exceeding limits in 8.4 examined
whether a failure at any of these points is to be
regarded as a NORMAL or SINGLE FAULT CONDITION
N/A
A MEANS OF PROTECTION protecting APPLIED
PARTS, or parts identified by 4.6 as parts
subject to the same requirements, considered
MEANS OF PATIENT PROTECTION ............................ :
N/A
A MEANS OF PROTECTION protecting other parts
considered MEANS OF OPERATOR PROTECTION .... :
N/A
8.5.2
Separation of PATIENT CONNECTIONS
N/A
8.5.2.1
PATIENT CONNECTIONS of F-TYPE APPLIED PART
separated from all other parts by equivalent to
one MEANS OF PATIENT PROTECTION for a WORKING
VOLTAGE equal to maximum MAINS VOLTAGE and
complied with limit for PATIENT LEAKAGE
CURRENT at 110 % of max. MAINS VOLTAGE ........ :
Separation requirement not applied between
multiple functions of a single F-TYPE APPLIED
No PATIENT CONNECTIONS
provided
N/A
N/A
PART
8.5.2.2
PATIENT CONNECTIONS treated as one APPLIED
PART in the absence of electrical separation
between PATIENT CONNECTIONS of same or
another function
N/A
MANUFACTURER has defined if multiple
functions are to be considered as all within
one APPLIED PART or as multiple APPLIED PARTS . :
N/A
Classification as TYPE BF, CF, or DEFIBRILLATIONPROOF applied to one entire APPLIED PART
N/A
LEAKAGE CURRENT tests conducted per 8.7.4 .... :
N/A
Dielectric strength test conducted per 8.8.3 .... :
N/A
CREEPAGE and CLEARANCES measured per 8.9
and Tables 11 to 16 as applicable
N/A
A protective device connected between PATIENT
CONNECTIONS of an F-TYPE APPLIED PART and
ENCLOSURE to protect against excessive
voltages did not operate below 500 V r.m.s
N/A
PATIENT CONNECTIONS of a TYPE B APPLIED PART
not PROTECTIVELY EARTHED are separated by one
MEANS OF PATIENT PROTECTION from metal
ACCESSIBLE PARTS not PROTECTIVELY EARTHED .... :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 35 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.5.2.3
Requirement + Test
Result - Remark
Verdict
– except when metal ACCESSIBLE PART is
physically close to APPLIED PART and can be
regarded as a part of APPLIED PART; and
N/A
– RISK that metal ACCESSIBLE PART will make
contact with a source of voltage or LEAKAGE
CURRENT above permitted limits is acceptably
low
N/A
LEAKAGE CURRENT tests conducted per 8.7.4 .... :
N/A
Dielectric strength test conducted per 8.8.3 .... :
N/A
Relevant CREEPAGE and CLEARANCES measured
per 8.9 and Tables 11 to 16 as applicable
N/A
The RISK MANAGEMENT FILE reviewed
N/A
A connector on a PATIENT lead located at the end of the lead remote from
PATIENT, with conductive part not separated from all PATIENT CONNECTIONS by
one MEANS OF PATIENT PROTECTION for a WORKING VOLTAGE equal to MAXIMUM
N/A
MAINS VOLTAGE
8.5.3
- cannot be connected to earth or hazardous
voltage while the PATIENT CONNECTIONS are in
contact with PATIENT .......................................... :
N/A
– conductive part of connector not separated
from all PATIENT CONNECTIONS did not come into
contact with a flat conductive plate of not less
than 100 mm diameter
N/A
– CLEARANCE between connector pins and a flat
surface is at least 0.5 mm
N/A
– conductive part pluggable into a mains
socket protected from making contact with
parts at MAINS VOLTAGE by insulation with a
CREEPAGE DISTANCE of at least 1.0 mm, a 1500 V
dielectric strength and complying with 8.8.4.1
N/A
– required test finger did not make electrical
contact with conductive part when applied
against access openings with a force of 10 N,
except when RISK MANAGEMENT PROCESS
indicated no unacceptable RISK existed from
contact with objects other than a mains socket
or a flat surface .................................................. :
N/A
MAXIMUM MAINS VOLTAGE
Pass
– MAXIMUM MAINS VOLTAGE determined to be the 240 V ac
highest RATED supply voltage for single-phase
or d.c. SUPPLY MAINS powered ME EQUIPMENT, as
well as INTERNALLY POWERED ME EQUIPMENT with a
means of connection to a SUPPLY MAINS (V) ..... :
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 36 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.5.4
Requirement + Test
Result - Remark
Verdict
When less than 100 V, MAXIMUM MAINS VOLTAGE
was 250 V
N/A
– MAXIMUM MAINS VOLTAGE was the highest RATED Not poly-phase ME
phase to neutral supply voltage for poly-phase EQUIPMENT
ME EQUIPMENT (V) ............................................... :
N/A
– for other INTERNALLY POWERED ME EQUIPMENT,
maximum mains voltage was 250 V
N/A
Not INTERNALLY POWERED ME
EQUIPMENT
WORKING VOLTAGE
Pass
– Input supply voltage to ME EQUIPMENT was
RATED voltage or voltage within RATED range
resulting in highest measured value (V) ......... :
240 V ac
Pass
– WORKING VOLTAGE for d.c. voltages with
superimposed ripple was average value when
peak-to-peak ripple less than 10% of average
value or peak voltage when peak-to-peak ripple
exceeding 10% of average value (V) ............... :
N/A
– WORKING VOLTAGE for each MEANS OF
PROTECTION forming DOUBLE INSULATION was
voltage DOUBLE INSULATION, as a whole,
subjected to (V) ................................................ :
See Insulation Diagram and
Insulation Table - Evaluated
in part of power supply
– Intentional or accidental earthing of PATIENT
regarded as a NORMAL CONDITION for WORKING
VOLTAGE involving a PATIENT CONNECTION not
connected to earth
No patient connections
N/A
– WORKING VOLTAGE between PATIENT
CONNECTIONS of an F-TYPE APPLIED PART and
ENCLOSURE was highest voltage appearing
across insulation in NORMAL USE including
earthing of any part of APPLIED PART (V) .......... :
No patient connections
N/A
– WORKING VOLTAGE for DEFIBRILLATION-PROOF
APPLIED PARTS determined disregarding
possible presence of defibrillation voltages
No Defibrillation-proof
applied parts
N/A
– WORKING VOLTAGE was equal to resonance
voltage in case of motors provided with
capacitors between the point where a winding
and a capacitor are connected together and a
terminal for external conductors (V) ............... :
No motor provided
N/A
8.5.5
DEFIBRILLATION-PROOF APPLIED PARTS
No Applied Part was defined
for the product
N/A
8.5.5.1
Classification “DEFIBRILLATION-PROOF APPLIED
PART” applied to one APPLIED PART in its
entirety, but not separate functions of same
Pass
N/A
APPLIED PART
Test Report No.: 1307033001
60601-1TRF_A
Page 37 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Possibility of an OPERATOR receiving a shock
from such parts taken into consideration in
N/A
RISK MANAGEMENT PROCESS
Isolation of PATIENT CONNECTIONS of a
DEFIBRILLATION- PROOF APPLIED PART from other
parts of ME EQUIPMENT accomplished as follows:
N/A
a) No hazardous electrical energies appear
during a discharge of cardiac defibrillator ..... :
N/A
b) ME EQUIPMENT complied with relevant
requirements of this standard, providing BASIC
SAFETY and ESSENTIAL PERFORMANCE following
exposure to defibrillation voltage, and recovery
time stated in ACCOMPANYING DOCUMENTS .......... :
N/A
8.5.5.2
Means provided to limit energy delivered to a
See Appended Table 8.5.5.2
100 Ω load to at least 90% of energy delivered
to this load with ME EQUIPMENT disconnected .. :
N/A
8.6
Protective and functional earthing and potential equalization of ME EQUIPMENT
N/A
8.6.1
Requirements of 8.6.2 to 8.6.8 applied
N/A
8.6.2
Parts complying with IEC 60950-1 for
protective earthing and serving as MEANS OF
OPERATOR PROTECTION but not PATIENT
PROTECTION exempted from requirements of
8.6.2 to 8.6.8
No protective earth
N/A
PROTECTIVE EARTH TERMINAL is suitable for
connection to an external protective earthing
system by a PROTECTIVE EARTH CONDUCTOR in a
POWER SUPPLY CORD and a suitable plug or by a
FIXED PROTECTIVE EARTH CONDUCTOR .................. :
No such construction
N/A
Clamping means of PROTECTIVE EARTH TERMINAL
of ME EQUIPMENT for FIXED supply conductors or
POWER SUPPLY CORDS comply with 8.11.4.3, and
cannot be loosened without TOOL
No such construction
N/A
Screws for internal PROTECTIVE EARTH
CONNECTIONS completely covered or protected
against accidental loosening from outside .... :
No Screw provided
N/A
Earth pin of APPLIANCE INLET forming supply
connection to ME EQUIPMENT regarded as
Evaluated in part of power
supply
N/A
PROTECTIVE EARTH TERMINAL not used for
Evaluated in part of power
mechanical connection between different parts supply
of ME EQUIPMENT or securing components not
related to protective or functional earthing
N/A
PROTECTIVE EARTH TERMINAL
Test Report No.: 1307033001
60601-1TRF_A
Page 38 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
8.6.3
PROTECTIVE EARTH CONNECTION not used for a
Evaluated in part of power
moving part, except when MANUFACTURER
supply
demonstrated in RISK MANAGEMENT FILE
connection will remain reliable during EXPECTED
SERVICE LIFE ....................................................... :
N/A
8.6.4
a) PROTECTIVE EARTH CONNECTIONS carried fault
Evaluated in part of power
currents reliably and without excessive voltage supply
drop
N/A
b) Allowable TOUCH CURRENT and PATIENT
Evaluated in part of power
LEAKAGE CURRENT in SINGLE FAULT CONDITION were supply
not exceeded, when impedance of PROTECTIVE
EARTH CONNECTIONS exceeded values in 8.6.4 a)
and Table 8.6.4, due to limited current
capability of relevant circuits ........................... :
N/A
Surface coatings
N/A
8.6.5
Result - Remark
Poorly conducting surface coatings on
conductive elements removed at the point of
contact
8.6.6
Verdict
No such construction
N/A
Coating not removed when requirements for
impedance and current-carrying capacity met
N/A
Plugs and sockets
N/A
PROTECTIVE EARTH CONNECTION where
connection between SUPPLY MAINS and ME
EQUIPMENT or between separate parts of ME
EQUIPMENT made via a plug and socket was
made before and interrupted after supply
connections
N/A
- applied also where interchangeable parts are
N/A
PROTECTIVELY EARTHED
8.6.7
Terminal for connection of a POTENTIAL EQUALIZATION CONDUCTOR
N/A
– Terminal is accessible to OPERATOR with ME
EQUIPMENT in any position of NORMAL USE
N/A
– RISK of accidental disconnection minimized
in NORMAL USE
N/A
– Terminal allows conductor to be detached
without a TOOL
N/A
– Terminal not used for a PROTECTIVE EARTH
N/A
CONNECTION
– Terminal marked with symbol 8 of Table D.1
(i.e., symbol IEC 60417-5021)
Test Report No.: 1307033001
60601-1TRF_A
Page 39 of 203
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– Instructions for use contain information on
function and use of POTENTIAL EQUALIZATION
CONDUCTOR together with a reference to
requirements of this standard
N/A
POWER SUPPLY CORD does not incorporate a
N/A
POTENTIAL EQUALIZATION CONDUCTOR
8.6.8
FUNCTIONAL EARTH TERMINAL not used to provide No FUNCTIONAL EARTH
a PROTECTIVE EARTH CONNECTION
N/A
8.6.9
Class II ME EQUIPMENT
N/A
Third conductor of POWER SUPPLY CORD
connected to protective earth contact of MAINS
PLUG provided with CLASS II ME EQUIPMENT with
isolated internal screens used as functional
earth connection to the screen’s FUNCTIONAL
EARTH TERMINAL, coloured green and yellow
Class I equipment
Two MEANS OF PROTECTION provided by
insulation of internal screens and all internal
wiring connected to them with a related
explanation in technical description ............... :
N/A
N/A
8.7
LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS
Pass
8.7.1
a) Electrical isolation providing protection
See appended Tables 8.7
against electric shock limits currents to values
in 8.7.3 ............................................................... :
Pass
b) Specified values of EARTH LEAKAGE, TOUCH,
See appended Tables 8.7
PATIENT LEAKAGE, and PATIENT AUXILIARY
CURRENTS applied in combination of conditions
in appended Table 8.7....................................... :
Pass
Allowable values specified in 8.7.3 applied
under SINGLE FAULT CONDITIONS of 8.1 b), except
Pass
– where insulation used in conjunction with a
PROTECTIVE EARTH CONNECTION, insulation short
circuited only under conditions in 8.6.4 b)
N/A
8.7.2
8.7.3
– the only SINGLE FAULT CONDITION for EARTH
LEAKAGE CURRENT was interruption of one
supply conductor at a time
Pass
– LEAKAGE CURRENTS and PATIENT AUXILIARY
Patient leakage for
CURRENT not measured in SINGLE FAULT
reference only
CONDITION of short circuiting of one constituent
part of DOUBLE INSULATION
Pass
SINGLE FAULT CONDITIONS not applied at same
time as special test conditions of MAXIMUM
MAINS VOLTAGE on APPLIED PARTS and nonPROTECTIVELY EARTHED parts of ENCLOSURE
Pass
Allowable Values
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 40 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.7.4
Requirement + Test
Result - Remark
Verdict
a) Allowable values in 8.7.3 b), c), and d)
See appended Tables 8.7
measured based on, and are relative to
currents in Fig 12 a), or by a device measuring
frequency contents of currents as in Fig 12 b . :
Pass
b) Allowable values of PATIENT LEAKAGE and
See appended Tables 8.7
AUXILIARY CURRENTS are according to Tables 3 & and Patient leakage for
4, and values of a.c. are relative to currents
reference only
having a frequency not less than 0.1Hz ........... :
Pass
c) T OUCH CURRENT did not exceed 100 μA in
See appended Tables 8.7
NORMAL CONDITION and 500 μA in SINGLE FAULT
CONDITION (ITNC, ITSFC) .......................................... :
Pass
d) EARTH LEAKAGE CURRENT did not exceed 5 mA See appended Tables 8.7
in NORMAL CONDITION and 10 mA in SINGLE FAULT
CONDITION (IENC, IESFC) .......................................... :
Pass
Higher values of EARTH LEAKAGE CURRENT
Not permanently installed
permitted for PERMANENTLY INSTALLED ME
ME equipment
EQUIPMENT connected to a supply circuit
supplying only this ME EQUIPMENT according to
local regulations or IEC 60364-7-710 ............... :
N/A
e) LEAKAGE CURRENTS, regardless of waveform See appended Table 8.7
and frequency, did not exceed 10 mA r.m.s. in
NORMAL or in SINGLE FAULT CONDITION (measured
with a non-frequency-weighted device ............ :
Pass
LEAKAGE and PATIENT AUXILIARY CURRENTS
Pass
See appended Table 8.7
measurements .................................................. :
8.8
Insulation
Pass
8.8.1
Insulation relied on as MEANS OF PROTECTION,
including REINFORCED INSULATION and insulation
between parts of opposite polarity of MAINS
PART on SUPPLY MAINS side of mains fuse or
OVER-CURRENT RELEASE
Pass
Insulation exempted from test (complies with
clause 4.8)
IEC 60601-1: 2005 and ANSI/
AAMI ES60601 of Certified
Power Supply.
Insulation forming MEANS OF OPERATOR
PROTECTION and complying with IEC 60950-1 for
INSULATION CO- ORDINATION not tested as in 8.8
8.8.2
Pass
Distance through solid insulation or use of thin sheet material
N/A
Solid insulation forming SUPPLEMENTARY or
REINFORCED INSULATION for a PEAK WORKING
VOLTAGE greater than 71 V provided with:
N/A
a) 0.4 mm, min, distance through insulation, or
N/A
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 41 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
b) does not form part of an ENCLOSURE and not
subject to handling or abrasion during NORMAL
USE, and comprised of:
N/A
– at least two layers of material, each passed
the appropriate dielectric strength test, or
N/A
– three layers of material, for which all
combinations of two layers together passed
the appropriate dielectric strength test
N/A
Dielectric strength test for one or two layers
was same as for one MEANS OF PROTECTION for
N/A
SUPPLEMENTARY INSULATION
Dielectric strength test for one or two layers
was same as for two MEANS OF PROTECTION for
N/A
REINFORCED INSULATION
BASIC, SUPPLEMENTARY , and REINFORCED
INSULATION required between windings of
Evaluated in part of power
supply
N/A
wound components separated by interleaved
insulation complying with a) or b), or both,
except when
c) Wire with solid insulation, other than solvent No such construction
based enamel, complying with a)
N/A
d) Wire with multi-layer extruded or spirally
wrapped insulation complying with b) and
complying with Annex L
Evaluated in part of power
supply
N/A
e) Finished wire with spirally wrapped or multi- Evaluated in part of power
layer extruded insulation, complying with
supply
Annex L
N/A
– BASIC INSULATION: minimum two wrapped
layers or one extruded layer
N/A
– SUPPLEMENTARY INSULATION: minimum two
layers, wrapped or extruded
N/A
– REINFORCED INSULATION: minimum three layers,
wrapped or extruded
N/A
In d) and e), for spirally wrapped insulation
Evaluated in part of power
with CREEPAGE DISTANCES between layers less
supply
than in Table 12 or 16 (Pollution Degree 1)
depending on type of insulation, path between
layers sealed as a cemented joint in 8.9.3.3 and
test voltages of TYPE TESTS in L.3 equal 1.6
times of normal values
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 42 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.8.3
Requirement + Test
Result - Remark
Verdict
Protection against mechanical stress provided Evaluated in part of power
where two insulated wires or one bare and one supply
insulated wire are in contact inside wound
component, crossing at an angle between 45°
and 90° and subject to winding tension.......... :
N/A
Finished component complied with routine
dielectric strength tests of 8.8.3 ...................... :
Evaluated in part of power
supply
N/A
Tests of Annex L not repeated since material
data sheets confirm compliance ..................... :
Evaluated in part of power
supply
N/A
Dielectric Strength
Pass
Solid insulating materials with a safety
See appended Table 8.8.3
function withstood dielectric strength test
voltages ............................................................ :
Pass
8.8.4
Insulation other than wire insulation
Pass
8.8.4.1
Resistance to heat retained by all insulation
and insulating partition walls during EXPECTED
SERVICE LIFE of ME EQUIPMENT
Pass
ME EQUIPMENT and RISK MANAGEMENT FILE
See Appended RM Results
examined in conjunction with resistance to
Table 8.8.4.1
moisture, dielectric strength, and mechanical
strength tests .................................................... :
Pass
Satisfactory evidence of compliance provided See appended RM Results
by manufacturer for resistance to heat............ : Table 8.8.4.1
Pass
Tests conducted in absence of satisfactory
evidence for resistance to heat ........................ :
Pass
a) ENCLOSURE and other external parts of
See appended Table 8.8.4.1
insulating material, except insulation of flexible
cords and parts of ceramic material, subjected
to ball-pressure test using apparatus of Fig 21
Pass
b) Parts of insulating material supporting
Evaluated in part of power
uninsulated parts of MAINS PART subjected to
supply
ball-pressure test in a), except at 125 °C ± 2 ° C
or ambient indicated in technical description
±2°C plus temperature rise determined during
test of 11.1 of relevant part, if higher (°C) ........ :
N/A
Test not performed on parts of ceramic
material, insulating parts of commutators,
brush-caps, and similar, and on coil formers
not used as REINFORCED INSULATION
N/A
8.8.4.2
Resistance to environmental stress
Test Report No.: 1307033001
60601-1TRF_A
Page 43 of 203
No such construction
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Insulating characteristics and mechanical
No such construction
strength of all MEANS OF PROTECTION not likely to
be impaired by environmental stresses
including deposition of dirt resulting from wear
of parts within EQUIPMENT, potentially reducing
CREEPAGE and CLEARANCES below 8.9
N/A
Ceramic and similar materials not tightly
sintered, and beads alone not used as
SUPPLEMENTARY or REINFORCED INSULATION
No such construction
N/A
Insulating material with embedded heating
conductors considered as one MEANS OF
PROTECTION but not two MEANS OF PROTECTION
No such construction
N/A
Parts of natural latex rubber aged by
No natural latex rubber
suspending samples freely in an oxygen
provided
cylinder containing commercial oxygen to a
pressure of 2.1 MPa ± 70 kPa, with an effective
capacity of at least 10 times volume of samples
N/A
There were no cracks visible to naked eyes
after samples kept in cylinder at 70 °C ± 2 °C
for 96h, and afterwards, left at room
temperature for at least 16h
N/A
8.9
CREEPAGE DISTANCES and AIR CLEARANCES
Pass
8.9.1.1
CREEPAGE DISTANCES and AIR CLEARANCES are 
to values in Tables 11 to 16 (inclusive), except
as specified in Clauses 8.9.1.2 to 8.9.1.15
Pass
8.9.1.2
Tables 11 to 16 (inclusive) not applied to
CREEPAGE and CLEARANCES forming MEANS OF
OPERATOR PROTECTION per IEC 60950-1 for
INSULATION CO- ORDINATION and used under
conditions compliance was tested
8.9.1.3
Specified min CLEARANCE applied as min
No such material
CREEPAGE for CREEPAGE DISTANCES across glass,
mica, ceramic and other inorganic insulating
materials with similar tracking characteristics
8.9.1.4
When min CREEPAGE derived from Tables 11 to
16 (inclusive) was less than min applicable
CLEARANCE, value of min CLEARANCE applied as
min CREEPAGE DISTANCE
Pass
8.9.1.5
ME EQUIPMENT RATED to operate at an altitude of
2000 m
N/A
ME EQUIPMENT RATED to operate at an altitude
specified by MANUFACTURER (m) ....................... :
Test Report No.: 1307033001
60601-1TRF_A
Page 44 of 203
No such construction
3000 m
N/A
N/A
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.9.1.6
Requirement + Test
Result - Remark
Operating altitude corresponding to actual air
pressure for ME EQUIPMENT intended for
pressurized environments (e.g., aircraft) used
to determine multiplication factor from Table 8,
and AIR CLEARANCE was multiplied by this factor
Pass
CREEPAGE DISTANCES not subjected to
multiplication factors, but were at least as
large as the resulting value for AIR CLEARANCE
Pass
When WORKING VOLTAGE was between those in
Tables 11 to 16 (inclusive), CREEPAGE and
CLEARANCES calculated as follows:
Pass
– CREEPAGE DISTANCES determined by linear
interpolation between the nearest two values,
and the calculated spacing rounded off to the
next higher 0.1 mm increment (mm) ............... :
8.9.1.7
Verdict
See Insulation Diagram and
To insulation diagram Table.
Pass
– CLEARANCES for PEAK WORKING VOLTAGES above Evaluated in part of power
2800 V peak or d.c. determined by linear
supply
interpolation between the nearest two values,
and the calculated spacing rounded off to the
next higher 0.1 mm increment (mm) ............... :
N/A
– for AIR CLEARANCES corresponding to PEAK
WORKING VOLTAGE up to 2800 V peak or d.c., the
higher of the two values applied
N/A
Material groups classified in accordance with
Table 9 (Material Group) .................................. :
N/A
Material group evaluated using 50 drops of
solution A based on test data for material
according to IEC 60112.................................... :
N/A
Material of unknown group considered IIIb
Client didn’t mention the
material group, and treated
as most worse condition:
Pass
IIIb
8.9.1.8
– Pollution degree 1: Micro-environment sealed
to exclude dust and moisture
– Pollution degree 2: Micro-environment with
non-conductive pollution, except occasional
conductivity caused by condensation
Pass
– Pollution degree 3: Micro-environment
subject to conductive pollution, or dry nonconductive pollution that could become
conductive due to expected condensation
N/A
– Pollution degree 4: Micro-environment where
continuous conductivity occurs due to
conductive dust, rain, or other wet conditions
N/A
Test Report No.: 1307033001
60601-1TRF_A
N/A
Page 45 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.9.1.9
Requirement + Test
Result - Remark
Verdict
Pollution degree 4 not used for insulation
providing a MEANS OF PROTECTION
N/A
Where insulation between MAINS PART and earth
might be compromised, measures such as
maintenance ensure that micro-environment is
mitigated to a lower pollution degree
N/A
Overvoltage category classification; value of
MAINS TRANSIENT VOLTAGE determined from
overvoltage category per IEC60664-1 and
NOMINAL a.c. MAINS VOLTAGE using Table 10
Evaluated in part of power
supply
V MT Peak (V) .................................................... :
2500
—
V MN r.m.s (V) .................................................... :
300
—
8.9.1.10
AIR CLEARANCE for MAINS PARTS (operating on
RATED MAINS VOLTAGES up to 300 V) were values
for r.m.s. or d.c. RATED MAINS VOLTAGE in Table
13 plus additional CLEARANCE in Table 14 for
PEAK WORKING VOLTAGE
MOPP provided
Pass
8.9.1.11
SUPPLY MAINS
overvoltage category II applied
according to IEC 60664-1
Overvoltage category II
Pass
8.9.1.12
For ME EQUIPMENT intended for overvoltage
category III, Tables 13 to 15 (inclusive) not
used for clearance, instead values in the next
MAINS TRANSIENT VOLTAGE column upwards used
N/A
When PATIENT protection (Table 12) is required
for use of ME EQUIPMENT on overvoltage
category III SUPPLY MAINS, guidance provided on
values required in the rationale for Cl. 8.9 used
N/A
A SECONDARY CIRCUIT derived from a SUPPLY
For Slim Medical Panel PC:
MAINS, normally, considered to be overvoltage
24Vdc
category I according to IEC 60664-1 when the
MAINS PART is overvoltage category II (Table 15)
Table 15 applied to earthed SECONDARY CIRCUIT
or INTERNALLY POWERED ME EQUIPMENT
Evaluated in part of power
supply
Pass
N/A
Requirements for primary circuits in Tables 13
and 14 used for an unearthed SECONDARY
CIRCUIT derived from a SUPPLY MAINS
N/A
Table 15 applied when SECONDARY CIRCUIT was
separated from MAINS PART by a functionally
earthed or PROTECTIVELY EARTHED metal screen
or transients in SECONDARY CIRCUIT were below
the levels expected for overvoltage category I
N/A
Table 15 column for circuits not subject to
transient overvoltages applied to:
N/A
Test Report No.: 1307033001
60601-1TRF_A
N/A
Page 46 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.9.1.13
8.9.1.14
Requirement + Test
Result - Remark
Verdict
– d.c. SECONDARY CIRCUITS reliably connected to Evaluated in part of power
earth and have capacitive filtering limiting
supply
peak-to-peak ripple to 10 % of d.c. voltage, and
N/A
– circuits in INTERNALLY POWERED ME EQUIPMENT
No circuits in internally
powered ME equipment
N/A
For PEAK WORKING VOLTAGES above 1400 V peak
or d.c. Table 15 not applied since all the
following conditions were met:
Evaluated in part of power
supply
N/A
– CLEARANCE was at least 5 mm
N/A
– insulation complied with dielectric strength
test of 8.8.3 using an a.c. test voltage with an
r.m.s. value equal to 1.06 times PEAK WORKING
VOLTAGE, or
N/A
– a d.c. test voltage equal to peak value of a.c.
test voltage with an r.m.s. value equal to 1.06
times PEAK WORKING VOLTAGE, and
N/A
– CLEARANCE path was partly or entirely through
air or along the surface of an insulating
material of material group I
N/A
Dielectric strength test conducted only across
part(s) of the path that are through air when
CLEARANCE path was also partly along surface
of a non- group I material
N/A
Minimum CREEPAGE DISTANCES for two MEANS OF
OPERATOR PROTECTION obtained by doubling
values in Table 16 for one MEANS OF OPERATOR
MOPP provided
N/A
PROTECTION
8.9.1.15
CREEPAGE DISTANCES and AIR CLEARANCES for
No defibrillation-proof
DEFIBRILLATION- PROOF APPLIED PARTS are 4 mm or applied parts
more to meet 8.5.5.1
N/A
8.9.2
a) Short circuiting of each single one of
Evaluated in part of power
CREEPAGE DISTANCES and CLEARANCES in turn did supply
not result in a HAZARDOUS SITUATION for
insulation in MAINS PART between parts of
opposite polarity, therefore, min CREEPAGE and
CLEARANCES not applied .................................... :
N/A
b) Contribution to CREEPAGE DISTANCES of
grooves or air gaps less than 1 mm wide
limited to widths
N/A
c) Relative positioning of CLEARANCE providing Evaluated in part of power
a MEANS OF PROTECTION is such that the relevant supply
parts are rigid and located by molding, or there
is no reduction of a distance below specified
value by deformation or movement of parts
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 47 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Normal or likely limited movements of relevant
parts taken into consideration when
calculating minimum AIR CLEARANCE
N/A
8.9.3
Spaces filled by insulating compound
N/A
8.9.3.1
Only solid insulation requirements applied
No insulating compound
where distances between conductive parts
provided
filled with insulating compound were such that
CLEARANCES and CREEPAGE DISTANCES don’t exist
N/A
Thermal cycling, humidity preconditioning, and
dielectric strength tests in 8.9.3.2 and 8.9.3.4 or
8.9.3.3 and 8.9.3.4 conducted
N/A
For insulating compound forming solid
insulation between conductive parts, a single
sample subjected to thermal cycling
PROCEDURE of 8.9.3.4 followed by humidity
preconditioning per 5.7 (for 48 hours), followed
by dielectric strength test (clause 8.8.3), test
voltage multiplied by 1.6................................... :
N/A
Cracks or voids in insulating compound
affecting homogeneity of material didn’t occur
N/A
Where insulating compound forms a cemented
joint with other insulating parts, three samples
tested for reliability of joint
N/A
A winding of solvent-based enameled wire
replaced for the test by a metal foil or by a few
turns of bare wire placed close to cemented
joint, and three samples tested as follows:
N/A
– One sample subjected to thermal cycling
PROCEDURE of 8.9.3.4, and immediately after the
last period at highest temperature during
thermal cycling, it was subjected to dielectric
strength test of 8.8.3 except at 1.6 times the
test voltage ....................................................... :
N/A
– The other two samples subjected to humidity
preconditioning of 5.7, except for 48 hours
only followed by a dielectric strength test of
8.8.3 at 1.6 times the test voltage
N/A
8.9.3.4
One sample containing the cemented joint
subjected to a sequence of temperature
cycling tests for 10 times ................................. :
N/A
8.10
Components and wiring
Pass
8.10.1
Components of ME EQUIPMENT likely to result in See Appended RM Results
an unacceptable RISK by their movements
Table 8.10.1
mounted securely as indicated in RISK
MANAGEMENT FILE ................................................ :
Pass
8.9.3.2
8.9.3.3
Test Report No.: 1307033001
60601-1TRF_A
Page 48 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.10.2
Conductors and connectors of ME EQUIPMENT
See Appended RM Results
adequately secured or insulated to prevent
Table 8.10.2
accidental detachment in a HAZARDOUS
SITUATION ........................................................... :
Pass
Conductors and connectors of ME EQUIPMENT
when breaking free at their joint are not
capable of touching circuit points resulting in a
HAZARDOUS SITUATION as indicated in RISK
Pass
MANAGEMENT FILE
Breaking free of one means of mechanical
restraint considered a SINGLE FAULT CONDITION
Stranded conductors are not solder-coated
when secured by clamping means to prevent
HAZARDOUS SITUATIONS due to poor contact
Pass
No such construction
N/A
8.10.3
Flexible cords detachable without a TOOL used No such construction
to interconnect different parts of ME EQUIPMENT
provided with means for connection to comply
with requirements for metal ACCESSIBLE PARTS
of 8.4 when a connection is loosened or
broken as shown by measurement or using
test finger
N/A
8.10.4
Cord-connected HAND-HELD parts and cord-connected foot-operated control
devices
N/A
8.10.4.1
Control devices of ME EQUIPMENT and their
No Cord-connected HANDconnection cords contain only conductors and HELD parts and cordcomponents operating at 42.4 V peak a.c., max, connected foot-operated
or 60 V d.c. in circuits isolated from MAINS PART control devices
by two MEANS OF PROTECTION
N/A
d.c. limit of 60 V applied to d.c. with no more
than 10 % peak-to-peak ripple
N/A
42.4 V peak limit applied when ripple exceeded
10 % peak-to-peak limit
N/A
Connection and anchorage of a flexible cord to
a HAND-HELD or foot-operated control device of
ME EQUIPMENT at both ends of cable to control
device complied with 8.11.3 when breaking free
or shorting between conductors could result in
a HAZARDOUS SITUATION
N/A
This requirement applied to other HAND-HELD
parts when disturbance or breaking of one or
more of connections could result in a
N/A
8.10.4.2
HAZARDOUS SITUATION
8.10.5
Mechanical protection of wiring
Test Report No.: 1307033001
60601-1TRF_A
Page 49 of 203
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
a) Internal cables and wiring adequately
See Appended RM Results
protected against contact with a moving part or Table 8.10.5
from friction at sharp corners and edges where
damage to insulation could result in a
HAZARDOUS SITUATION ......................................... :
Pass
b) Wiring, cord forms, or components are not
likely to be damaged during assembly or
during opening or closing of ACCESS COVERS
where such damage could result in a
HAZARDOUS SITUATION as shown by manual tests
and RISK MANAGEMENT FILE
Pass
8.10.6
Guiding rollers of insulated conductors
No such construction
prevent bending of movable insulated
conductors around a radius of less than five
times the outer diameter of the lead concerned
in NORMAL USE
8.10.7
a) Insulating sleeve that can only be removed
by breaking or cutting, or secured at both
ends, is used on internal wiring of when
needed .............................................................. :
See appended Table 8.10
b) Sheath of a flexible cord not used as a
MEANS OF PROTECTION inside ME EQUIPMENT when
it is subject to mechanical or thermal stresses
beyond its RATED characteristics
No such construction
c) Insulated conductors subject to
No such situation
temperatures > 70 °C in NORMAL USE provided
with insulation of heat-resistant material when
compliance is likely to be impaired due to
deterioration of insulation ................................ :
N/A
Pass
N/A
N/A
8.11
MAINS PARTS, components and layout
Pass
8.11.1
a) ME EQUIPMENT provided with means of
See appended Table 8.10
electrically isolating its circuits from SUPPLY
MAINS simultaneously on all poles .................... :
Pass
PERMANENTLY INSTALLED ME EQUIPMENT connected Not Permanently installed
to a poly-phase SUPPLY MAINS equipped with a
ME equipment
device not interrupting neutral conductor,
provided local installation conditions prevent
voltage on neutral conductor from exceeding
limits in 8.4.2 c)
N/A
b) Means of isolation incorporated in ME
See appended Table 8.10
EQUIPMENT , and external means described in
technical description ....................................... :
Pass
c) A SUPPLY MAINS switch used to comply with
Not Permanently installed
8.11.1 a) complies with CREEPAGE and
ME equipment
CLEARANCES in IEC 61058-1 for a MAINS
TRANSIENT VOLTAGE of 4 kV ................................. :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 50 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
d) A SUPPLY MAINS switch not incorporated in a
POWER SUPPLY CORD or external flexible lead
Not Permanently installed
ME equipment
e) Direction of movement of actuator of a
No supply mains switch
SUPPLY MAINS switch used to comply with 8.11.1
a) complies with IEC 60447
N/A
f) A suitable plug device such as an APPLIANCE
COUPLER or a flexible cord with a MAINS PLUG
used in non-PERMANENTLY INSTALLED ME
EQUIPMENT to isolate it from SUPPLY MAINS
considered to comply with 8.11.1 a) ............... :
N/A
g) A fuse or a semiconductor device not used
as an isolating means
Provided IEC 60601-1: 2005
and ANSI/ AAMI ES60601
Certified Power Supply.
Pass
h) ME EQUIPMENT not provided with a device
causing disconnection of ME EQUIPMENT from
SUPPLY MAINS by producing a short circuit
resulting in operation of an overcurrent
protection device
Such device is not provided
Pass
i) Parts within ENCLOSURE of ME EQUIPMENT with a Evaluated in part of power
circuit > 42.4 V peak a.c. or 60 V d.c. that
supply
cannot be disconnected from its supply by an
external switch or a plug device accessible at
all times is protected against touch even after
opening ENCLOSURE by an additional covering
N/A
A clear warning notice is marked on outside of Not exceed allowable
ME EQUIPMENT to indicate it exceeds allowable
voltage, no warning is
touch voltage (symbol 10 of Table D.1 is
required
insufficient)
N/A
For a part that could not be disconnected from
supply by an external switch or a plug device
accessible at all times, the required cover or
warning notice complied with this clause
N/A
Standard test finger of Fig 6 applied
N/A
8.11.2
MULTIPLE SOCKET-OUTLETS integral with ME
EQUIPMENT complied with 16.2 d), second dash;
and 16.9.2
8.11.3
POWER SUPPLY CORDS
8.11.3.1
MAINS PLUG not fitted with more than one POWER Evaluated in part of power
SUPPLY CORD
supply
8.11.3.2
POWER SUPPLY CORDS are no less robust than
ordinary tough rubber sheathed flexible cord
(IEC 60245-1:2003, Annex A, designation 53) or
ordinary polyvinyl chloride sheathed flexible
cord (IEC 60227-1:1993, Annex A, design. 53) .. :
Test Report No.: 1307033001
60601-1TRF_A
N/A
No Multiple socket-outlets
N/A
Pass
Page 51 of 203
N/A
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Only polyvinyl chloride insulated POWER SUPPLY PVC power supply cord
CORD with appropriate temperature rating used used and no metal parts of
for ME EQUIPMENT having external metal parts
the outside equipment.
with a temperature > 75 °C touchable by the
cord in NORMAL USE ............................................ :
Pass
8.11.3.3
NOMINAL cross-sectional area of conductors of
POWER SUPPLY CORDS of ME EQUIPMENT is not less
than in Table 17 (mm2 Cu) ............................... :
Pass
8.11.3.4
APPLIANCE COUPLERS complying with IEC 603201 are considered to comply with 8.11.3.5 and
8.11.3.6 ............................................................... :
N/A
8.11.3.5
Cord anchorage (for APPLIANCE COUPLERS not complying with IEC 60320-1)
N/A
a) Conductors of POWER SUPPLY CORD provided
with strain relieve and insulation protected
from abrasion at point of entry to ME EQUIPMENT
or a MAINS CONNECTOR by a cord anchorage
N/A
b) Cord anchorage of POWER SUPPLY CORD is
made of and arranged as follows when a total
insulation failure of POWER SUPPLY CORD caused
conductive non-PROTECTIVELY EARTHED
ACCESSIBLE PARTS to exceed limits of 8.4:
N/A
– insulating material, or
N/A
– metal, insulated from conductive ACCESSIBLE
PARTS non-PROTECTIVELY EARTHED by a MEANS OF
PROTECTION, or
N/A
– metal provided with an insulating lining
affixed to cord anchorage, except when it is a
flexible bushing forming part of the cord guard
in 8.11.3.6, and complying with the
requirements for one MEANS OF PROTECTION
N/A
c) Cord anchorage prevents cord from being
clamped by a screw bearing directly on cord
insulation
N/A
d) Screws to be operated when replacing
POWER SUPPLY CORD do not serve to secure any
components other than parts of cord
anchorage
N/A
e) Conductors of POWER SUPPLY CORD arranged
to prevent PROTECTIVE EARTH CONDUCTOR against
strain as long as phase conductors are in
contact with their terminals when cord
anchorage fails
N/A
f) Cord anchorage prevents POWER SUPPLY CORD
from being pushed into ME EQUIPMENT or MAINS
N/A
CONNECTOR
Test Report No.: 1307033001
60601-1TRF_A
Page 52 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.11.3.6
Requirement + Test
Result - Remark
Verdict
Conductors of POWER SUPPLY CORD supplied by
MANUFACTURER disconnected from terminals or
from MAINS CONNECTOR and cord subjected 25
times to a pull applied with no jerks, each time
for 1 s, on sheath of the value in Table 18 ....... :
N/A
Cord subjected to a torque in Table 18 for 1
min immediately after pull tests
N/A
Cord anchorage did not allow cord sheath to
be longitudinally displaced by more than 2 mm
or conductor ends to move over a distance of
more than 1 mm from their connected position
N/A
CREEPAGE and CLEARANCES not reduced below
limits in 8.9
N/A
It was not possible to push the cord into ME
EQUIPMENT or MAINS CONNECTOR to an extent the
cord or internal parts would be damaged
N/A
POWER SUPPLY CORDS other than for STATIONARY
ME EQUIPMENT protected against excessive
bending at inlet opening of equipment or of
MAINS CONNECTOR by means of an insulating
cord guard or by means of an appropriately
shaped opening
Not stationary ME
equipment
N/A
Cord guard complied with test of IEC 603351:2001, Clause 25.14, or
N/A
ME EQUIPMENT placed such that axis of cord
guard projected at an angle of 45° with cord
free from stress, and a mass equal 10 x D2
gram attached to the free end of cord (g) ...... :
N/A
Cord guard of temperature-sensitive material
tested at 23 °C ± 2 °C, and flat cords bent in the
plane of least resistance
N/A
Curvature of the cord radius, immediately after
mass attached, was not less than 1.5 x D...... :
N/A
8.11.4
MAINS TERMINAL DEVICES
N/A
8.11.4.1
PERMANENTLY INSTALLED and ME EQUIPMENT with
non-DETACHABLE POWER SUPPLY CORD
replaceable by SERVICE PERSONNEL provided
with MAINS TERMINAL DEVICES ensuring reliable
connection
Terminals alone are not used to keep
conductors in position, except when barriers
are provided such that CREEPAGE and
CLEARANCES cannot be reduced below 8.9 if any
conductor breaks away
Test Report No.: 1307033001
60601-1TRF_A
Page 53 of 203
Appliance inlet provided
N/A
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Terminals of components other than terminal
blocks complying with requirements of this
Clause and marked according to 7.3.7 used as
terminals intended for external conductors
N/A
Screws and nuts clamping external conductors
do not serve to secure any other component,
except they also clamp internal conductors
when unlikely to be displaced when fitting the
supply conductors
N/A
Arrangement of MAINS TERMINAL DEVICES
N/A
a) Terminals provided for connection of
AC inlet is used on power
external cords or POWER SUPPLY CORDS together adapter
with PROTECTIVE EARTH TERMINAL grouped to
provide convenient means of connection
N/A
b) PROTECTIVE EARTH CONDUCTOR connections
complied with 8.6
N/A
c) Marking of MAINS TERMINAL DEVICES complied
with 7.3
N/A
d) MAINS TERMINAL DEVICES not accessible
without use of a TOOL
N/A
e) A MEANS OF PROTECTION are not short
circuited when one end of a flexible conductor
with NOMINAL cross-sectional area is stripped 8
mm and a single free wire is bent in each
possible direction
N/A
8.11.4.3
Internal wiring not subjected to stress and
CREEPAGE and CLEARANCES not reduced below
8.9 after fastening and loosening a conductor
of largest cross-sectional area 10 times
N/A
8.11.4.4
Terminals with clamping means for a rewirable
flexible cord did not require special
preparation of conductors and conductors
were not damaged and did not slip out when
clamping means tightened as verified by test
of 8.11.3.4
N/A
8.11.4.5
Adequate space provided inside ME EQUIPMENT
designed for FIXED wiring or a re-wirable POWER
SUPPLY CORD to allow for connection of
conductors, and covers fitted without damage
to conductors or their insulation
N/A
Correct connection and positioning of
conductors before ACCESS COVER was fitted
verified by an installation test
N/A
Mains fuses and OVER-CURRENT RELEASES
N/A
8.11.4.2
8.11.5
Test Report No.: 1307033001
60601-1TRF_A
Page 54 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
A fuse or OVER-CURRENT RELEASE provided in
Evaluated in part of power
each supply lead for CLASS I and CLASS II ME
supply
EQUIPMENT with a functional earth connection
per clause 8.6.9, and in at least one supply lead
for other single-phase CLASS II ME EQUIPMENT.... :
N/A
– neutral conductor not fused for PERMANENTLY
N/A
INSTALLED ME EQUIPMENT
8.11.6
Not permanently installed
ME equipment
– fuses or OVER-CURRENT RELEASES omitted due No such situation
to provision of two MEANS OF PROTECTION
between all parts of opposite polarity within
MAINS PART , and between all parts of MAINS PART
and earth, and such provisions continued
within all components
N/A
Effect of short-circuit fault conditions in other
circuits taken into consideration before
eliminating fuses or OVER-CURRENT RELEASES
N/A
Protective devices have adequate breaking
See appended Table 8.10
capacity to interrupt the maximum fault current
including the available short-circuit ............... :
N/A
A fuse or OVER-CURRENT RELEASE not provided
in a PROTECTIVE EARTH CONDUCTOR
N/A
Fuses complying with IEC 60127 have high
breaking capacity (1 500 A) and prospective
short-circuit current > 35 A or 10 times current
rating of the fuse, whichever is greater
No such situation
N/A
Justification for omission of fuses or OVERCURRENT RELEASES is in RISK MANAGEMENT FILE
Evaluated in part of power
supply
N/A
Internal wiring of the MAINS PART
a) Cross-sectional area of internal wiring in a
MAINS PART between MAINS TERMINAL DEVICE and
protective devices is not less than minimum
required for POWER SUPPLY CORD as in clause
8.11.3.3 (mm2 Cu) ............................................. :
N/A
Evaluated in part of power
supply
N/A
b) Cross-sectional area of other wiring in MAINS Evaluated in part of power
PART and sizes of tracks on printed wiring
supply
circuits sufficient to prevent fire in case of fault
currents ............................................................. :
N/A
When necessary, ME EQUIPMENT connected to a
SUPPLY MAINS with max available short-circuit
fault, and subsequent simulation of a fault in a
single insulation in MAINS PART did not result in
any of the HAZARDOUS SITUATIONS in 13.1.2
N/A
Test Report No.: 1307033001
60601-1TRF_A
Verdict
Page 55 of 203
Evaluated in part of power
supply
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9
PROTECTION AGAINST MECHANICAL HAZARDS OF ME EQUIPMENT AND
ME SYSTEMS
Pass
9.1
ME EQUIPMENT complies with Clause 4 for
design and manufacture, and mechanical
strength (15.3)
Pass
9.2
HAZARDS associated with moving parts
N/A
9.2.1
When ME EQUIPMENT with moving parts
See Appended RM Results
PROPERLY INSTALLED, used per ACCOMPANYING
Table 9.2.1
DOCUMENTS or under foreseeable misuse, RISKS
associated with moving parts reduced to an
acceptable level .................................................:
N/A
RISK from contact with moving parts reduced
to an acceptable level using protective
measures, (access, function, shape of parts,
energy, speed of motion, and benefits to
PATIENT considered)
N/A
RESIDUAL RISK associated with moving parts
considered acceptable when exposure was
needed for ME EQUIPMENT to perform its
function
N/A
Warnings marked on ME EQUIPMENT or included
in instructions for use when HAZARDS persisted
after implementing all reasonable protective
measures ...........................................................:
N/A
9.2.2
T RAPPING ZONE
N/A
9.2.2.1
ME EQUIPMENT with a TRAPPING ZONE complied
with one or more of the following as feasible:
No Trapping zone
Verdict
N/A
– Gaps in Clause 9.2.2.2, or
N/A
– Safe distances in Clause 9.2.2.3, or
N/A
– GUARDS and protective measures in 9.2.2.4,
or
N/A
– Continuous activation in Clause 9.2.2.5
N/A
Control of relevant motion complied with
9.2.2.6 when implementation of above
protective measures were inconsistent with
INTENDED USE of ME EQUIPMENT or ME SYSTEM
N/A
9.2.2.2
A TRAPPING ZONE considered not to present a
No Trapping zone
MECHANICAL HAZARD when gaps of TRAPPING
ZONE complied with dimensions per Table 20 ..:
N/A
9.2.2.3
A TRAPPING ZONE considered not to present a
No Trapping zone
MECHANICAL HAZARD when distances separating
OPERATOR, PATIENT , and others from TRAPPING
ZONES exceeded values in ISO 13852 .............. :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 56 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Distances measured from expected positions
of OPERATOR, PATIENT, and others near
EQUIPMENT in NORMAL USE or under foreseeable
misuse
N/A
9.2.2.4
GUARDS and protective measures
N/A
9.2.2.4.1
A TRAPPING ZONE considered not to present a
No Trapping zone
MECHANICAL HAZARD when GUARDS and
protective measures were of robust
construction, not easy to bypass or render
non-operational, and did not introduce
additional unacceptable RISK based on results
of applicable tests in 15.3 for ENCLOSURES .......:
N/A
9.2.2.4.2
FIXED GUARDS held in place by systems that
cannot be dismantled without a TOOL
N/A
9.2.2.4.3
Movable GUARDS that can be opened without a
TOOL remained attached when GUARD was open
N/A
– they are associated with an interlock
preventing relevant moving parts from starting
to move while TRAPPING ZONE is accessible, and
stops movement when the GUARD is opened,
N/A
– absence or failure of one of their
components prevents starting, and stops
moving parts
N/A
Movable GUARDS complied with all applicable
tests as confirmed by review of RISK
N/A
MANAGEMENT FILE
9.2.2.4.4
9.2.2.5
Protective measures provided in control
system prevented moving parts from starting
to move while in reach of persons ...................:
N/A
– protective measures prevented TRAPPING
ZONE from reach, or, when it was reached,
system movement stopped once ME EQUIPMENT
started to move, and in the latter case, no
HAZARD or damage resulted
N/A
– when protective measure was in a SINGLE
FAULT CONDITION, and an unacceptable RISK
could arise, one or more emergency stopping
device(s) provided
N/A
RISK MANAGEMENT FILE reviewed and all
conditions confirmed
N/A
Continuous activation
N/A
T RAPPING ZONE not considered to present a
MECHANICAL HAZARD where impractical to make
TRAPPING ZONE inaccessible when:
Test Report No.: 1307033001
60601-1TRF_A
Page 57 of 203
No Trapping zone
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
a) movement was in OPERATOR'S field of view
N/A
b) movement of ME EQUIPMENT or its parts was
possible only by continuous activation of
control by OPERATOR as long as OPERATOR
response to deactivate device relied upon to
prevent HARM
N/A
Manually operated movements complied with
this clause since mass and velocity allowed
adequate control of positioning without
causing an unacceptable RISK
N/A
c) when in a SINGLE FAULT CONDITION of
continuous activation system an unacceptable
RISK could arise, one or more emergency
stopping device(s) provided in ME EQUIPMENT ..:
N/A
Speed of movement(s) positioning parts of ME No Trapping zone
EQUIPMENT or PATIENT , when contact with ME
EQUIPMENT could result in a HAZARDOUS
SITUATION, limited to allow OPERATOR control of
positioning without resulting in an
unacceptable RISK..............................................:
N/A
Over travel (stopping distance) of such
movement occurring after operation of a
control to stop movement, did not result in an
unacceptable RISK
N/A
9.2.3
Other HAZARDS associated with moving parts
N/A
9.2.3.1
Controls positioned, recessed, or protected by No moving parts
other means and could not be accidentally
actuated to result in unacceptable RISK, except
when ergonomic considerations for a PATIENT
with special needs require otherwise
N/A
9.2.3.2
RISK due to over travel (past range limits) of
ME EQUIPMENT parts reduced to an acceptable
level, and stops or other means with
mechanical strength to withstand intended
loading in NORMAL USE and foreseeable misuse
provided limiting measure in NORMAL and
SINGLE FAULT CONDITION .......................................:
N/A
9.2.4
Emergency stopping devices
N/A
Where necessary to have one or more
No Emergency stopping
emergency stopping device(s), emergency
devices
stopping device complied with all the
following, except for actuating switch capable
of interrupting all power....................................:
N/A
a) Emergency stopping device reduced RISK to
an acceptable level
N/A
9.2.2.6
Test Report No.: 1307033001
60601-1TRF_A
Page 58 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
9.2.5
Requirement + Test
Verdict
b) Proximity and response of OPERATOR to
actuate emergency stopping device could be
relied upon to prevent HARM
N/A
c) Emergency stopping device actuator was
readily accessible to OPERATOR
N/A
d) Emergency stopping device(s) are not part
of normal operation of ME EQUIPMENT
N/A
e) Emergency switching operation or stopping
means neither introduced further HAZARD nor
interfered with operation necessary to remove
original HAZARD
N/A
f) Emergency stopping device was able to
break full load of relevant circuit, including
possible stalled motor currents and the like
g) Means for stopping of movements operate
as a result of one single action
h) Emergency stopping device provided with
an actuator in red and easily distinguishable
and identifiable from other controls
i) An actuator interrupting/opening mechanical
movements marked on or immediately
adjacent to face of actuator with symbol 18 of
Table D.1 (symbol IEC 60417-5638, DB:200210) or “STOP”
N/A
j) Emergency stopping device, once actuated,
maintained ME EQUIPMENT in disabled condition
until a deliberate action, different from that
used to actuate it, was performed
N/A
k) Emergency stopping device is suitable for
its application
N/A
Means provided to permit quick and safe
No such construction
release of PATIENT in event of breakdown of ME
EQUIPMENT or failure of power supply,
activation of a protective measure, or
emergency stopping, and
N/A
– Uncontrolled or unintended movement of ME
EQUIPMENT that could result in an unacceptable
RISK prevented
N/A
– Situations where PATIENT is subjected to
unacceptable RISKS due to proximity of moving
parts, removal of normal exit routes, or other
HAZARDS prevented
N/A
– Measures provided to reduce RISK to an
acceptable level when after removal of
counterbalanced parts, other parts of ME
EQUIPMENT can move in a hazardous way
N/A
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 59 of 203
N/A
N/A
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
9.3
Rough surfaces, sharp corners and edges of
See Appended RM Results
ME EQUIPMENT that could result in an
Table 9.3
unacceptable RISK avoided or covered ............:
Pass
9.4
Instability HAZARDS
N/A
9.4.1
ME EQUIPMENT, other than FIXED and hand-held,
for placement on a surface did not
overbalance (tip over) or move unexpectedly,
to the degree that it could present an
unacceptable RISK to PATIENT, or OPERATOR as
tested in 9.4.2 to 9.4.4
N/A
9.4.2
Instability – overbalance
N/A
9.4.2.1
ME EQUIPMENT or its parts did not overbalance
when prepared per ACCOMPANYING DOCUMENTS,
or when not specified, as in 9.4.2.2, and placed
on a 10° inclined plane from horizontal
consisting of a hard and flat surface (e.g.,
concrete floor covered with 2 to 4 mm thick
vinyl material) ....................................................:
N/A
9.4.2.2
Instability excluding transport
N/A
ME EQUIPMENT or its parts prepared based on
a) to g), inclusive, did not overbalance when
placed in different positions of NORMAL USE,
except transport positions, on a 5° inclined
plane from horizontal (hard and flat surface)
N/A
A warning provided, stating “Transport only
under conditions described in instructions for
use or marked on ME EQUIPMENT with an
indication of RESIDUAL RISK if ME EQUIPMENT or
its parts overbalances” when overbalance
occurred during 10° inclined plane test
N/A
Instability from horizontal and vertical forces
N/A
a) ME EQUIPMENT with a mass of 25 kg or more,
other than FIXED ME EQUIPMENT for use on floor,
did not overbalance due to pushing or resting
N/A
Surfaces of ME EQUIPMENT where a RISK of
overbalancing exists from pushing, leaning,
resting etc., permanently marked with a
CLEARLY LEGIBLE warning of the RISK (e.g.,
safety sign 5 of Table D.2, safety sign ISO
7010-P017)
N/A
ME EQUIPMENT did not overbalance when
placed on a horizontal plane, and a force of
25% of its weight, but not more than 220 N,
applied in different directions, except a
direction with an upward component
N/A
9.4.2.3
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 60 of 203
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
b) ME EQUIPMENT, other than FIXED ME
EQUIPMENT , for use on the floor or on a table,
did not overbalance due to sitting or stepping,
except when a legible warning of this RISK
provided on ME EQUIPMENT (e.g., safety signs 6
and 7 of Table D.2, safety signs ISO 7010P018, or ISO 7010-P019 as appropriate)
N/A
ME EQUIPMENT did not overbalance when
placed on a horizontal plane, and a constant
force of 800 N applied at the point of maximum
moment to working surfaces, offering an
foothold or sitting surface of a min 20 x 20 cm
area, and at a height  1 m from the floor .......:
N/A
9.4.2.4
Castors and wheels
N/A
9.4.2.4.1
Means used for transportation of MOBILE ME
EQUIPMENT (e.g., castors or wheels) did not
result in an unacceptable RISK when MOBILE ME
EQUIPMENT moved or parked in NORMAL USE
9.4.2.4.2
Force required to move MOBILE ME EQUIPMENT
along a hard and flat horizontal surface did
not exceed 200 N applied at a height of 1 m
above floor or highest point on ME EQUIPMENT
when < 1 m high, except when instructions
indicated more than one person needed (N) ....:
N/A
9.4.2.4.3
MOBILE ME EQUIPMENT exceeding 45 kg
configured with a SAFE WORKING LOAD, moved
10 times in forward direction over a solid
vertical plane obstruction with wheels
impacting the obstruction at a speed of 0.4
m/s ± 0.1 m/s for manual or with max speed for
motor driven MOBILE ME EQUIPMENT ....................:
N/A
ME EQUIPMENT went up the obstruction without
overbalancing or any other unacceptable RISK
as determined by examination of RISK
MANAGEMENT FILE, ME EQUIPMENT and its parts....:
N/A
There was no reduction of CREEPAGE and
CLEARANCES below 8.9, no access to parts
exceeding limits in 8.4, and no access to
moving parts capable of causing HARM , and
N/A
– Assessment criteria in Clause 9 and 11.6
used
N/A
– Dielectric strength test of 8.8.3 conducted to
evaluate integrity of solid SUPPLEMENTARY or
N/A
No such construction
N/A
REINFORCED INSULATION
Test Report No.: 1307033001
60601-1TRF_A
Page 61 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– CREEPAGE DISTANCES and AIR CLEARANCES
measured compared favourably with min
distances in clause 8.9
N/A
Small chips not adversely affecting protection
against electric shock or moisture,
disregarded
N/A
9.4.3
Instability from unwanted lateral movement (including sliding)
N/A
9.4.3.1
a) Brakes of power-driven MOBILE ME EQUIPMENT No Such construction
normally activated and could only be released
by continuous actuation of a control
N/A
b) MOBILE ME EQUIPMENT provided with locking
means to prevent unwanted movements of ME
EQUIPMENT or its parts in transport position
N/A
c) No unacceptable RISK due to unwanted
lateral movement resulted when MOBILE ME
EQUIPMENT placed in its transport position or
worst case NORMAL USE position with SAFE
WORKING LOAD, and locking device activated,
on a 10° inclined hard flat surface with castors
in the worst-case position
N/A
Following initial elastic movement, creepage,
and pivoting of castors, no further movement
of MOBILE ME EQUIPMENT > 50 mm (in relation to
inclined plane) occurred (mm) ......................... :
N/A
RISK due to any initial movement assessed
taking into consideration NORMAL USE of ME
N/A
EQUIPMENT
9.4.3.2
Instability excluding transport
N/A
a) Further movement of ME EQUIPMENT (after
No Such construction
initial elastic movement) was less than 50 mm
when MOBILE ME EQUIPMENT with a SAFE WORKING
LOAD positioned on a 5 inclined hard flat
surface with wheel locked or braking system
activated (mm) .................................................. :
N/A
RISK due to initial movements assessed taking
into consideration NORMAL USE of ME EQUIPMENT
N/A
b) TRANSPORTABLE or STATIONARY ME EQUIPMENT
for use on the floor and with a SAFE WORKING
LOAD prepared as in 9.4.2.2 and placed on a
horizontal plane with locking device activated
and castors, when supplied, in their worst –
case position
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 62 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
9.4.4
Requirement + Test
Result - Remark
Verdict
Further movement of ME EQUIPMENT (after initial
elastic movement), was no more than 50 mm
when a force of 25 % of weight of unit, but less
than 220 N, applied in different directions,
except a direction with an upwards
component, at highest point of ME EQUIPMENT
but  1.5 m from floor ....................................... :
N/A
RISK due to initial movements assessed taking
into consideration NORMAL USE of ME EQUIPMENT
N/A
Grips and other handling devices
N/A
a) ME EQUIPMENT other than PORTABLE
No such construction
EQUIPMENT or its part with a mass of over 20 kg
requiring lifting in NORMAL USE or transport
provided with suitable handling means, or
ACCOMPANYING DOCUMENTS specify safe lifting
method, except when handling is obvious and
causing HAZARDS
N/A
Handles, when supplied, suitably placed to
enable ME EQUIPMENT or its part to be carried by
two or more persons and by examination of
EQUIPMENT , its part, or ACCOMPANYING
N/A
DOCUMENTS
b) PORTABLE ME EQUIPMENT with a mass > 20 kg No such construction
provided with one or more carrying-handles
suitably placed to enable carrying by two or
more persons as confirmed by actual carrying
N/A
c) Carrying handles and grips and their means No such construction
of attachment withstood loading test .............. :
N/A
9.5
Expelled parts HAZARD
N/A
9.5.1
Suitability of means of protecting against
No expelled part
unacceptable RISK of expelled parts
determined by assessment and examination of
RISK MANAGEMENT FILE ........................................:
N/A
9.5.2
Cathode ray tube(s) complied with IEC
60065:2001, Clause 18, or IEC 61965 ...............:
N/A
9.6
Acoustic energy (including infra- and ultrasound) and vibration
N/A
9.6.1
Human exposure to acoustic energy and
No Acoustic energy produce
vibration from ME EQUIPMENT doesn’t result in
by the device
unacceptable RISK as confirmed in RISK
MANAGEMENT FILE including audibility of
auditory alarm signals, PATIENT sensitivity, and
tests of 9.6.2 and 9.6.3......................................:
N/A
9.6.2
Acoustic energy
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 63 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
9.6.2.1
PATIENT, OPERATOR, and other persons are not No Acoustic energy produce
exposed to acoustic energy from ME EQUIPMENT by the device
in NORMAL USE, except for auditory alarm
signals
9.6.2.2
Result - Remark
Verdict
N/A
– 80 dBA for a cumulative exposure of 24 h
over a 24 h period (dBA) .................................. :
—
- 83 dBA (when halving the cumulative
exposure time) (dBA) ....................................... :
—
– 140 dB un-weighted sound pressure level for
impulsive or impact acoustic energy (dB) .......... :
—
RISK MANAGEMENT FILE examined for RISKS
associated with infrasound or ultrasound,
when present, addressed in RISK MANAGEMENT
No Acoustic energy produce
by the device
N/A
PROCESS
9.6.3
Hand-transmitted vibration
N/A
Means provided, except for INTENDED USE
vibrations, to protect PATIENT and OPERATOR
when hand-transmitted frequency-weighted
r.m.s. acceleration generated in NORMAL USE
exceeds specified values measured at points
of hand contact with PATIENT or OPERATOR
No vibration produce by the
device
N/A
– 2.5 m/s2 for a cumulative time of 8 h during a
24 h period (m/s2) ............................................. :
N/A
– Accelerations for different times, inversely
proportional to square root of time (m/s2) ....... :
N/A
9.7
Pressure vessels and parts subject to pneumatic and hydraulic pressure
N/A
9.7.1
Requirements of this clause applied to vessels No Pressure vessels and
and parts of ME EQUIPMENT subject to pressure parts subject to pneumatic
resulting in rupture and unacceptable RISK
and hydraulic pressure
N/A
Parts of a pneumatic or hydraulic system used
as a support system, comply with 9.8
N/A
Pneumatic and hydraulic parts of ME
No Pressure vessels and
EQUIPMENT or ACCESSORIES met following
parts subject to pneumatic
requirements based on examination of RISK
and hydraulic pressure
MANAGEMENT FILE ...............................................:
N/A
– No unacceptable RISK resulted from loss of
pressure or loss of vacuum
N/A
– No unacceptable RISK resulted from a fluid
jet caused by leakage or a component failure
N/A
– Elements of ME EQUIPMENT or an ACCESSORY,
especially pipes and hoses leading to an
unacceptable RISK protected against harmful
external effects
N/A
9.7.2
Test Report No.: 1307033001
60601-1TRF_A
Page 64 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
9.7.3
Requirement + Test
Result - Remark
Verdict
– Reservoirs and similar vessels leading to an
unacceptable RISK are automatically
depressurized when ME EQUIPMENT is isolated
from its power supply
N/A
Means provided for isolation, or local
depressurizing reservoirs and similar vessels,
and pressure indication when above not
possible
N/A
– All elements remaining under pressure after
isolation of ME EQUIPMENT or an ACCESSORY
from its power supply resulting in an
unacceptable RISK provided with clearly
identified exhaust devices, and a warning to
depressurize these elements before setting or
maintenance activity
N/A
Maximum pressure a part of ME EQUIPMENT can
be subjected to in NORMAL and SINGLE FAULT
CONDITIONS considered to be highest of
following:
No Pressure vessels and
parts subject to pneumatic
and hydraulic pressure
N/A
a) RATED maximum supply pressure from an
external source
N/A
b) Pressure setting of a pressure-relief device
provided as part of assembly
N/A
c) Max pressure that can develop by a source
of pressure that is part of assembly, unless
pressure limited by a pressure-relief device
N/A
9.7.4
Max pressure in NORMAL and SINGLE FAULT
See Appended RM Results
CONDITIONS did not exceed MAXIMUM PERMISSIBLE Table 9.7.4
WORKING PRESSURE for EQUIPMENT part, except
as allowed in 9.7.7, confirmed by examination
of ME EQUIPMENT and RISK MANAGEMENT FILE, and
by functional tests............................................:
N/A
9.7.5
A pressure vessel withstood a HYDRAULIC TEST See Appended Table 9.7.5
PRESSURE when pressure was > 50 kPa, and
product of pressure and volume was more
than 200 kPaI ....................................................:
N/A
9.7.6
Pressure-control device regulating pressure in See Appended RM Results
ME EQUIPMENT with pressure-relief device
Table 9.7.6
completed 100,000 cycles of operation under
RATED load and prevented pressure from
exceeding 90 % of setting of pressure-relief
device in different conditions of NORMAL USE .:
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 65 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9.7.7
Pressure-relief device(s) used where MAXIMUM See Appended RM Results
PERMISSIBLE WORKING PRESSURE could otherwise Table 9.7.7
be exceeded met the following, as confirmed
by MANUFACTURER’S data, ME EQUIPMENT, RISK
MANAGEMENT FILE, and functional tests .............:
N/A
a) Connected as close as possible to pressure
vessel or parts of system it is to protect
N/A
b) Installed to be readily accessible for
inspection, maintenance, and repair
N/A
c) Could be adjusted or rendered inoperative
without a TOOL
N/A
d) With discharge opening located and
directed as to not to release material towards
any person
N/A
e) With discharge opening located and
directed as to not to deposit material on parts
that could result in an unacceptable RISK
N/A
f) Adequate discharge capacity provided to
ensure that pressure will not exceed MAXIMUM
PERMISSIBLE WORKING PRESSURE of system it is
connected to by more than 10 % when failure
occurs in control of supply pressure
N/A
g) No shut-off valve provided between a
pressure-relief device and parts it is to protect
N/A
h) Min number of cycles of operation 100 000,
except for one-time use devices (bursting
disks)
N/A
9.8
HAZARDS associated with support systems
Pass
9.8.1
ME EQUIPMENT parts designed to support loads See appended RM Results
or provide actuating forces when a
Table 9.8.1
mechanical fault could constitute an
unacceptable RISK ............................................:
Pass
– Construction of support, suspension, or
actuation system complied with Table 21 and
Pass
TOTAL LOAD
Equipment weight: 6.4kg
Safety Factor: 6
Total load: 38.4kg
– Means of attachment of ACCESSORIES prevent No accessories
possibility of incorrect attachment that could
result in an unacceptable RISK
N/A
– RISK ANALYSIS of support systems included
HAZARDS from static, dynamic, vibration,
impact and pressure loading, foundation and
other movements, temperature,
environmental, manufacture and service
conditions
N/A
Test Report No.: 1307033001
60601-1TRF_A
Verdict
Page 66 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– RISK ANALYSIS included effects of failures
See appended RM Results
such as excessive deflection, plastic
Table 9.8.1
deformation, ductile/brittle fracture, fatigue
fracture, instability (buckling), stress-assisted
corrosion cracking, wear, material creep and
deterioration, and residual stresses from
manufacturing PROCESSES
Pass
– Instructions on attachment of structures to a
floor, wall, ceiling, included in ACCOMPANYING
DOCUMENTS making adequate allowances for
quality of materials used to make the
connection and list the required materials
N/A
Additional instructions provided on checking
adequacy of surface of structure parts will be
attached to
N/A
Support systems maintain structural integrity
during EXPECTED SERVICE LIFE, and TENSILE
SAFETY FACTORS are not less than in Table 21,
except when an alternative method used to
demonstrate structural integrity throughout
EXPECTED SERVICE LIFE, or for a foot rest
Pass
Compliance with 9.8.1 and 9.8.2 confirmed by See Appended RM Results
examination of ME EQUIPMENT, RISK MANAGEMENT Table 9.8.2
FILE, specifications and material processing...:
Pass
When test results were part of information,
testing consisted of application of a test load
to support assembly equal to TOTAL LOAD times
required TENSILE SAFETY FACTOR while support
assembly under test was in equilibrium after 1
min, or not resulted in an unacceptable RISK ..:
N/A
9.8.3
Strength of PATIENT or OPERATOR support or suspension systems
N/A
9.8.3.1
ME EQUIPMENT parts supporting or immobilizing No patient or operator
PATIENTS minimize RISK of physical injuries and support systems
accidental loosening of secured joints
N/A
SAFE WORKING LOAD of ME EQUIPMENT or its parts
supporting or suspending PATIENTS or
OPERATORS is sum of mass of PATIENTS or mass
of OPERATORS plus mass of ACCESSORIES
supported by ME EQUIPMENT or its parts
N/A
Supporting and suspending parts for adult
human PATIENTS or OPERATORS designed for a
PATIENT or OPERATOR with a min mass of 135 kg
and ACCESSORIES with a min mass of 15 kg,
unless stated by MANUFACTURER
N/A
9.8.2
Test Report No.: 1307033001
60601-1TRF_A
Page 67 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Maximum mass of PATIENT included in SAFE
WORKING LOAD of ME EQUIPMENT or its parts
supporting or suspending PATIENTS adapted
when MANUFACTURER specified applications
N/A
Max allowable PATIENT mass < 135 kg marked
on ME EQUIPMENT and stated in ACCOMPANYING
N/A
DOCUMENTS
Max allowable PATIENT mass > 135 kg stated in
N/A
ACCOMPANYING DOCUMENTS
Examination of markings, ACCOMPANYING
DOCUMENTS, and RISK MANAGEMENT FILE
confirmed compliance .....................................:
9.8.3.2
Part of SAFE WORKING LOAD representing mass
of PATIENTS or OPERATORS is distributed on
support/suspension surface representing
human body as in Fig A.19
No patient or operator
support systems
N/A
Part of SAFE WORKING LOAD representing mass
of ACCESSORIES deployed as in NORMAL USE and,
when not defined, at worst case position
permitted by configuration or ACCESSORIES
attachment on support/suspension parts
N/A
a) Entire mass of PATIENT or OPERATOR
2
distributed over an area of 0.1 m on a foot
rest temporarily supporting a standing PATIENT
or OPERATOR ......................................................:
N/A
Compliance confirmed by examination of ME
EQUIPMENT , RISK MANAGEMENT FILE,
specifications of materials and their
processing, and tests.......................................:
N/A
PATIENT support/suspension system
positioned horizontally in most
disadvantageous position in NORMAL USE, and
a mass 2 x 135 kg or twice intended person’s
load (the greater used), applied to foot rest
over an area of 0.1 m2 for 1 min (Kg) ...............:
N/A
Damage or deflection resulting in an
unacceptable RISK did not occur on foot rest
and its secured joints
N/A
b) Deflection of a support surface from PATIENT
or OPERATOR loading on an area of support/
suspension where a PATIENT or OPERATOR can
sit did not result in an unacceptable RISK........:
N/A
Compliance confirmed by examination of ME
EQUIPMENT , RISK MANAGEMENT FILE,
specifications of materials and their
processing, and by a test .................................:
N/A
Test Report No.: 1307033001
60601-1TRF_A
N/A
Page 68 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
PATIENT support/suspension system set in
most unfavourable NORMAL USE position, and a
mass of 60 % of part of SAFE WORKING LOAD
simulating PATIENT or OPERATOR, or a min 80
kg, placed on support or suspension system
with centre of load 60 mm from outer edge of
support or suspension system for at least one
minute (Kg) .......................................................:
N/A
Deflection of support/suspension system
resulting in an unacceptable RISK not occur
N/A
Dynamic forces that can be exerted on
No patient or operator
equipment parts supporting or suspending a
support systems
PATIENT or OPERATOR in NORMAL USE did not
result in an unacceptable RISK as confirmed by
following test:
N/A
PATIENT support/suspension system set in
most unfavourable NORMAL USE position, and a
mass equal to SAFE WORKING LOAD simulating
PATIENT or OPERATOR dropped from 150 mm
above seat area on an area of support/
suspension a PATIENT or OPERATOR can sit....... :
N/A
9.8.4
Systems with MECHANICAL PROTECTIVE DEVICES
N/A
9.8.4.1
a) A MECHANICAL PROTECTIVE DEVICE provided
No mechanical protective
when a support system or its parts impaired
device provided
by wear have a TENSILE SAFETY FACTOR  to
values in Table 21, rows 5 and 6, but less than
3 and 4 .............................................................:
N/A
b) MECHANICAL PROTECTIVE complies with the
requirements as follows:
N/A
– Designed based on TOTAL LOAD, and
includes effects of SAFE WORKING LOAD when
applicable
N/A
– Has TENSILE SAFETY FACTORS for all parts not
less than Table 21, row 7
N/A
– Activated before travel (movement)
produced an unacceptable RISK
N/A
– Takes into account Clauses 9.2.5 and 9.8.4.3
N/A
Compliance confirmed by examination of ME
EQUIPMENT , RISK MANAGEMENT FILE,
specifications of materials and their
processing .......................................................:
N/A
9.8.3.3
Test Report No.: 1307033001
60601-1TRF_A
Page 69 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
9.8.4.2
Activation of MECHANICAL PROTECTIVE DEVICE is No mechanical protective
made obvious to OPERATOR when ME EQUIPMENT device provided
can still be used after failure of suspension or
actuation means and activation of a
MECHANICAL PROTECTIVE DEVICE (e.g., a
secondary cable)
N/A
MECHANICAL PROTECTIVE DEVICE requires use of
a TOOL to be reset or replaced
N/A
MECHANICAL PROTECTIVE DEVICE intended to function once
N/A
– Further use of ME EQUIPMENT not possible
No mechanical protective
until replacement of MECHANICAL PROTECTIVE
device provided
DEVICE ...............................................................:
N/A
– ACCOMPANYING DOCUMENTS instruct once
MECHANICAL PROTECTIVE DEVICE is activated,
SERVICE PERSONNEL shall be called, and
MECHANICAL PROTECTIVE DEVICE must be
replaced before ME EQUIPMENT can be used
N/A
– ME EQUIPMENT permanently marked with
safety sign 2 of Table D.2 (i.e., safety sign
7010-W001)
N/A
– Marking is adjacent to MECHANICAL
PROTECTIVE DEVICE or its location relative to
MECHANICAL PROTECTIVE DEVICE is obvious to
service personnel
N/A
– Compliance confirmed by examination of ME
EQUIPMENT , ACCOMPANYING DOCUMENTS, RISK
MANAGEMENT FILE, specifications and
processing of materials, and following test ....:
N/A
A chain, cable, band, spring, belt, jack screw
nut, pneumatic or hydraulic hose, structural
part or the like, employed to support a load,
defeated by a convenient means causing
maximum normal load to fall from most
adverse position permitted by construction of
N/A
9.8.4.3
Result - Remark
Verdict
ME EQUIPMENT
9.8.5
Load included SAFE WORKING LOAD in 9.8.3.1
when system was capable of supporting a
PATIENT or OPERATOR
N/A
No evidence of damage to MECHANICAL
PROTECTIVE DEVICE affecting its ability to
perform its intended function
N/A
Systems without MECHANICAL PROTECTIVE DEVICES
Pass
Support system parts have TENSILE SAFETY
See appended RM Results
Table 9.8.5
FACTORS  to values in Table 21, rows 1 and 2,
and are not impaired by wear ..........................:
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 70 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Support system parts impaired by wear,
however, they have TENSILE SAFETY FACTORS 
to values in Table 21, rows 3 and 4
N/A
Examination of ME EQUIPMENT and RISK
MANAGEMENT FILE confirmed compliance
N/A
10
PROTECTION AGAINST UNWANTED AND EXCESSIVE RADIATION HAZARDS
Pass
10.1
X-Radiation
N/A
10.1.1
X-radiation dose-rate was  36 pA/kg (5 μSv/h) No X-Radiation
(0.5 mR/h) 5 cm from surface of ME EQUIPMENT
including background radiation for ME
EQUIPMENT not producing therapeutic/diagnostic
X-radiation but producing ionizing radiation . :
N/A
Amount of radiation measured by means of an
ionizing chamber radiation monitor with an
2
effective area of 10 cm or by other instruments
producing equal results
N/A
ME EQUIPMENT operated as in NORMAL USE at
most unfavourable RATED MAINS VOLTAGE and
controls adjusted to emit maximum radiation
N/A
Internal pre-set controls not intended for
adjustment during EXPECTED SERVICE LIFE of ME
EQUIPMENT not taken into consideration
N/A
10.1.2
RISK from unintended X-radiation from ME
EQUIPMENT producing X-radiation for diagnostic
and therapeutic purposes addressed in RISK
MANAGEMENT PROCESS as indicated in RISK
MANAGEMENT FILE (see IEC 60601-1-3 & 1.3) .... :
N/A
10.2
RISK associated with alpha, beta, gamma,
neutron, and other particle radiation, when
applicable, addressed in RISK MANAGEMENT
PROCESS as shown in RISK MANAGEMENT FILE .... :
No alpha, beta, gamma,
neutron, and other particle
radiation
N/A
10.3
RISK associated with microwave radiation,
when applicable, addressed in RISK
MANAGEMENT PROCESS as indicated in RISK
MANAGEMENT FILE ............................................... :
No microwave radiation
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 71 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
10.4
Relevant requirements of IEC 60825-1:1993
applied to lasers, light emitting diodes (LEDs),
and laser light barriers or similar products
No lasers and laser light
barriers or similar products
RISK associated with visible electromagnetic
radiation other than emitted by lasers and LEDS,
when applicable, addressed in RISK
MANAGEMENT PROCESS as indicated in RISK
MANAGEMENT FILE ............................................... :
RISK associated with infrared radiation other
than emitted by lasers and LEDS, as applicable,
addressed in RISK MANAGEMENT PROCESS as
indicated in RISK MANAGEMENT FILE ................... :
RISK associated with ultraviolet radiation other
than emitted by lasers and LEDS, as applicable,
addressed in RISK MANAGEMENT PROCESS as
indicated in RISK MANAGEMENT FILE ................... :
See Appended RM Results
Table 10.5
N/A
No such hazard
N/A
No UV
N/A
10.5
10.6
10.7
Verdict
Pass
This product contains only
LEDs (Class 1) operating in
the range of 400 - 700 nm
wavelength. No IEC60825-1
evaluation was deemed
necessary. Additional review
may be required at the
discretion of the accepting
NCB.
11
PROTECTION AGAINST EXCESSIVE TEMPERATURES AND OTHER
HAZARDS
Pass
11.1
Excessive temperatures in ME EQUIPMENT
Pass
11.1.1
Temperatures on ME EQUIPMENT parts did not
See appended Table 11.1.1
exceed values in Tables 22 and 23 operating in and appended RM Results
worst-case NORMAL USE at maximum rated
Table 11.1.1
ambient operating temperature T .................... :
Pass
Surfaces of test corner did not exceed 90 °C
Pass
THERMAL CUT -OUTS did
CONDITION
not operate in NORMAL
11.1.2
Temperature of APPLIED PARTS
11.1.2.1
Temperatures, hot or cold surfaces, and when
appropriate, clinical effects of APPLIED PARTS
supplying heat to a PATIENT determined and
documented in RISK MANAGEMENT FILE and
instructions for use ......................................... :
Test Report No.: 1307033001
60601-1TRF_A
No THERMAL CUT-OUTS
N/A
N/A
Page 72 of 203
No applied part
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
11.1.2.2
APPLIED PARTS not supplying heat to a PATIENT
met Table 24 with max surface temperatures >
41 °C disclosed in instructions for use, and
clinical effects regarding maturity of PATIENTS,
body surface, surface pressure, medications
taken, as shown in RISK MANAGEMENT FILE ........ :
See appended Table 11.1.2.2
and appended RM Results
Table 11.1.2.2
Surfaces of APPLIED PARTS cooled below
ambient temperatures that can also result in
HAZARD evaluated as part of RISK MANAGEMENT
N/A
N/A
PROCESS
11.1.3
Measurements not made when engineering
judgment and rationale by MANUFACTURER
indicated temperature limits could not exceed,
as documented in RISK MANAGEMENT FILE ......... :
Temperature test was
conducted. No such
situation
N/A
Test corner not used where engineering
judgment and rationale by MANUFACTURER
indicated test corner will not impact
measurements, as documented in RISK
The test corner is used and
its surfaces shall not exceed
90 °C.
N/A
No Applied part or parts
likely to be touched by
patient
N/A
No such construction
N/A
MANAGEMENT FILE
Probability of occurrence and duration of
contact for parts likely to be touched and for
APPLIED PARTS documented in RISK MANAGEMENT
FILE
11.1.4
GUARDS preventing contact with hot or cold
accessible surfaces removable only with a
TOOL
11.2
Fire prevention
Pass
11.2.1
ENCLOSURE has strength and rigidity necessary
to prevent a fire caused by reasonably
foreseeable misuse and met mechanical
strength tests for ENCLOSURES in 15.3
Pass
11.2.2
Me equipment and me systems used in conjunction with OXYGEN RICH
N/A
ENVIRONMENTS
11.2.2.1
RISK of fire in an OXYGEN RICH ENVIRONMENT
Equipment not suitable for
reduced by means limiting spread of fire under use in the presence of
NORMAL or SINGLE FAULT CONDITIONS when source OXYGEN RICH ENVIRONMENT
of ignition in contact with ignitable material .. :
N/A
Requirements of 13.1.1 applied to oxygen
concentrations up to 25 % at one atmosphere
or partial pressures up to 27.5 kPa for higher
atmospheric pressures
N/A
a) No sources of ignition discovered in an
OXYGEN RICH ENVIRONMENT in NORMAL and SINGLE
FAULT CONDITIONS under any of the following
conditions .......................................................... :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 73 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Verdict
1) when temperature of material raised to its
ignition temperature
N/A
2) when temperatures affected solder or solder
joints causing loosening, short circuiting, or
other failures causing sparking or increasing
material temperature to its ignition temperature
N/A
3) when parts affecting safety cracked or
changed outer shape exposing temperatures
higher than 300°C or sparks due to overheating
N/A
4) when temperatures of parts or components
exceeded 300°C, atmosphere was 100 %
oxygen, contact material solder, and fuel
cotton
N/A
5) when sparks provided adequate energy for
ignition by exceeding limits of Figs 35 to 37
(inclusive), atmosphere was 100 % oxygen,
contact material solder, and fuel cotton
N/A
Deviations from worst case limits in 4) and 5)
above based on lower oxygen concentrations
or less flammable fuels justified and
documented in RISK MANAGEMENT FILE
N/A
Alternative test in this clause did not identify
existence of ignition sources at highest
voltage or current, respectively ....................... :
N/A
A safe upper limit determined by dividing
upper limit of voltage or current, respectively,
with safety margin factor of three ................... :
N/A
b) RESIDUAL RISK of fire in an OXYGEN RICH
ENVIRONMENT as determined by application of
RISK MANAGEMENT PROCESS is based on following
configurations, or in combination ................... :
N/A
1) Electrical components in an OXYGEN RICH
ENVIRONMENT provided with power supplies
having limited energy levels lower than those
considered sufficient for ignition in 11.2.2.1 a)
as determined by examination, measurement
or calculation of power, energy, and
temperatures in NORMAL and SINGLE FAULT
CONDITIONS identified in 11.2.3........................... :
N/A
2) Max oxygen concentration measured until it
did not exceed 25 % in ventilated
compartments with parts that can be a source
of ignition only in SINGLE FAULT CONDITION and
can be penetrated by oxygen due to an
undetected leak (%) ......................................... :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 74 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
3) A compartment with parts or components
that can be a source of ignition only under
SINGLE FAULT CONDITION separated from another
compartment containing an OXYGEN RICH
ENVIRONMENT by sealing all joints and holes for
cables, shafts, or other purposes
N/A
Effect of possible leaks and failures under
SINGLE FAULT CONDITION that could cause ignition
evaluated using a RISK ASSESSMENT to determine
maintenance intervals by examination of
documentation and RISK MANAGEMENT FILE ....... :
N/A
4) Fire initiated in ENCLOSURE of electrical
components in a compartment with OXYGEN
RICH ENVIRONMENT that can become a source of
ignition only under SINGLE FAULT CONDITIONS
self-extinguished rapidly and no hazardous
amount of toxic gases reached PATIENT as
determined by analysis of gases .................... :
N/A
11.2.2.2
RISK of
ignition under least favourable
conditions did not occur and oxygen
concentration did not exceed 25% in
immediate surroundings due to location of
external exhaust outlets of an OXYGEN RICH
ENVIRONMENT when electrical components
mounted outside of ME EQUIPMENT or ME SYSTEM
N/A
11.2.2.3
Electrical connections within a compartment
containing an OXYGEN RICH ENVIRONMENT under
NORMAL USE did not produce sparks due to
loosening or breaking, except when limited in
power and energy to values in 11.2.2.1 a) 5)
N/A
– Screw-attachments protected against
loosening during use by varnishing, use of
spring washers, or adequate torques
N/A
– Soldered, crimped, and pin-and-socket
connections of cables exiting ENCLOSURE
include additional mechanical securing means
N/A
SINGLE FAULT CONDITIONS related to OXYGEN RICH ENVIRONMENTS ME EQUIPMENT and
ME SYSTEMS considered
N/A
– Failure of a ventilation system constructed in Equipment not suitable for
accordance with 11.2.2.1 b) 2) .......................... : use in the presence of
N/A
11.2.3
OXYGEN RICH ENVIRONMENT
– Failure of a barrier constructed in
accordance with 11.2.2.1 b) 3) .......................... :
N/A
– Failure of a component creating a source of
ignition (as defined in 11.2.2.1 a) ..................... :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 75 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
11.3
Requirement + Test
Verdict
– Failure of solid insulation or creepage and
clearances providing equivalent of at least one
MEANS OF PATIENT PROTECTION but less than two
MEANS OF PATIENT PROTECTION that could create a
source of ignition defined in 11.2.2.1 a) ........... :
N/A
– Failure of a pneumatic component resulting
in leakage of oxygen-enriched gas .................. :
N/A
Constructional requirements for fire ENCLOSURES of ME EQUIPMENT
Pass
ME EQUIPMENT met this clause for alternate
See Appended RM Results
means of compliance with selected HAZARDOUS Table 11.3
SITUATIONS and fault conditions in 13.1.2.......... :
Pass
Constructional requirements were met, or
Pass
- constructional requirements specifically
See Appended RM Results
analysed in RISK MANAGEMENT FILE .................... : Table 11.3
Pass
Justification, when requirement not met ......... : The constructional
requirements were met
Pass
a) Flammability classification of insulated wire
within fire ENCLOSURE is FV-1, or better, based
on IEC 60695 series as determined by
examination of data on materials
See appended Table 8.10
Pass
Flammability classification of connectors,
See appended Table 8.10
printed circuit boards, and insulating material
on which components are mounted is FV-2, or
better, based on IEC 60695-11-10 as decided by
examination of materials data .......................... :
Pass
If no FV Certification, FV tests based on IEC
60695-11-10 conducted on 3 samples of
complete parts (or sections of it), including
area with min. thickness, ventilation openings
N/A
b) Fire ENCLOSURE met following:
Pass
1) No openings at bottom or, as specified in
Fig 39, constructed with baffles as in Fig 38, or
made of perforated metal as in Table 25, or a
metal screen with a mesh  2 × 2 mm centre to
centre and wire diameter of at least 0.45 mm
Numerous openings
provided. No hazardous part
within projection area at a 5°
angle from edge of the
opening.
Pass
2) No openings on the sides within the area
included within the inclined line C in Fig 39
Same above
Pass
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 76 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
11.4
11.5
11.6
Requirement + Test
Result - Remark
Verdict
3) ENCLOSURE, baffles, and flame barriers have Enclosure rated V-0
adequate rigidity and made of appropriate
See appended Table 8.10
metal or of non-metallic materials, except
constructions based on Table 25 and a mesh;
FV-2 or better for TRANSPORTABLE ME EQUIPMENT,
FV-1 or better for fixed EQUIPMENT, or
STATIONARY EQUIPMENT per IEC 60695-11-10,
determined by ENCLOSURE examination or
flammability classification based on 11.3a)..... :
Pass
ME EQUIPMENT and ME SYSTEMS intended for use with flammable anaesthetics
N/A
ME EQUIPMENT, ME SYSTEMS and parts described
in ACCOMPANYING DOCUMENTS for use with
flammable anaesthetics (CATEGORY AP) or
anaesthetics with oxidants (CATEGORY APG)
comply with Annex G
N/A
Not suitable to use with
AP/APG
ME EQUIPMENT and ME SYSTEMS intended for use in conjunction with flammable
agents
N/A
MANUFACTURER’S RISK MANAGEMENT PROCESS
Not intended to use in
addresses possibility of fire and associated
conjunction with flammable
mitigations as confirmed by examination of
agents
RISK MANAGEMENT FILE ........................................ :
N/A
Overflow, spillage, leakage, ingress of water or particulate matter, cleaning,
disinfection, sterilization and compatibility with substances used with the ME
Pass
EQUIPMENT
11.6.1
Sufficient degree of protection provided
See table 11.6.1 for detail
against overflow, spillage, leakage, ingress of
water or particulate matter, cleaning,
disinfection and sterilization, and compatibility
with substances used with ME EQUIPMENT ....... :
Pass
11.6.2
Overflow in ME EQUIPMENT
N/A
Liquid reservoir liable to overflow in NORMAL
USE completely filled and 15 % of its capacity
poured in for over 1 min, and except when
restricted, TRANSPORTABLE ME EQUIPMENT tilted
through an angle of 15° in least favourable
direction(s), and when necessary refilled
starting from position of NORMAL USE ............... :
N/A
ME EQUIPMENT met dielectric strength and
LEAKAGE CURRENT tests and uninsulated
electrical parts or electrical insulation of parts
that could result in a HAZARDOUS SITUATION were
not wet ............................................................... :
N/A
Spillage on ME EQUIPMENT and ME SYSTEM
N/A
11.6.3
Test Report No.: 1307033001
60601-1TRF_A
Page 77 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
ME EQUIPMENT and ME SYSTEMS handling liquids No liquid handling when
in NORMAL USE positioned as in 5.4 a) and liquid Normal use
with composition, volume, duration of spill,
point of contact, and test conditions based on
RISK MANAGEMENT PROCESS poured steadily on a
point on top of ME EQUIPMENT ............................ :
N/A
ME EQUIPMENT met dielectric strength and
LEAKAGE CURRENT tests and uninsulated
electrical parts or electrical insulation of parts
that could result in a HAZARDOUS SITUATION were
not wet .............................................................. :
N/A
11.6.4
Leakage
N/A
11.6.5
Ingress of water or particulate matter into ME EQUIPMENT and ME SYSTEMS
N/A
ME EQUIPMENT with IP Code placed in least
favourable position of NORMAL USE and
subjected to tests of IEC 60529 (IP Code)....... :
N/A
11.6.6
11.6.7
ME EQUIPMENT met dielectric strength and
LEAKAGE CURRENT tests and there were no
bridging of insulation or electrical components
that could result in a HAZARDOUS SITUATION in
NORMAL CONDITION or in a SINGLE FAULT CONDITION
N/A
Cleaning and disinfection of ME EQUIPMENT and ME SYSTEMS
Pass
ME EQUIPMENT/ME SYSTEM and their parts and
See appended Table 11.6.1
ACCESSORIES cleaned or disinfected once using
methods specified in instructions for use
including any cooling or drying period ........... :
Pass
ME EQUIPMENT met dielectric strength and
See appended RM Results
LEAKAGE CURRENT tests, with no deterioration
Table 11.6.6.
resulting in an unacceptable RISK present ....... :
See appended Tables 8.7
and 8.8.3
Pass
Effects of multiple cleanings/disinfections
during EXPECTED SERVICE LIFE of EQUIPMENT
evaluated by MANUFACTURER and assurance that
no unacceptable RISK will occur verified by RISK
MANAGEMENT FILE review .................................... :
N/A
Expected service life will not
be impacted since cleaning
test shows solution will
cause no harm / degradation
to device insulation
Sterilization of ME EQUIPMENT and ME SYSTEMS
N/A
ME EQUIPMENT, ME SYSTEMS and their parts or
Not intended to be sterilized
ACCESSORIES intended to be sterilized assessed
and documented according to ISO 11134, ISO
11135, or ISO 11137 as appropriate ................. :
N/A
After the test, ME EQUIPMENT complied with the
appropriate dielectric strength and LEAKAGE
CURRENT tests and there was no deterioration
resulting in an unacceptable RISK .................... :
N/A
Test Report No.: 1307033001
60601-1TRF_A
IPX0
Page 78 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
11.6.8
RISKS associated
with compatibility of
substances used with ME EQUIPMENT addressed
in RISK MANAGEMENT PROCESS as confirmed by
examination of RISK MANAGEMENT FILE ............... :
N/A
11.7
ME EQUIPMENT, ME SYSTEM , and ACCESSORIES
coming into direct or indirect contact with
biological tissues, cells, or body fluids
assessed and documented per ISO 10993
N/A
11.8
Interruption and restoration of power supply
did not result in a HAZARDOUS SITUATION, except
interruption of its intended function
Pass
12
ACCURACY OF CONTROLS AND INSTRUMENTS AND PROTECTION
AGAINST HAZARDOUS OUTPUTS
N/A
12.1
RISKS associated
with accuracy of controls and No function for accuracy of
instruments stated in RISK MANAGEMENT PROCESS controls and instruments and
confirmed by RISK MANAGEMENT FILE review ..... : protection against hazardous
outputs
N/A
12.2
RISK of poor USABILITY, including identification,
marking, and documents addressed in a
USABILITY ENGINEERING PROCESS as confirmed by
review of provided records.............................. :
N/A
12.3
The need for alarm systems as a means of RISK
CONTROL and RISKS associated with operation or
failure of alarm system addressed in RISK
MANAGEMENT PROCESS ....................................... :
N/A
12.4
Protection against hazardous output
N/A
12.4.1
RISKS associated
12.4.2
When applicable, need for indication of
parameters associated with hazardous output
addressed in RISK MANAGEMENT PROCESS as
confirmed in RISK MANAGEMENT FILE .................... :
N/A
12.4.3
RISKS associated
N/A
12.4.4
with hazardous output arising
from intentional exceeding of safety limits
addressed in RISK MANAGEMENT PROCESS as
confirmed in RISK MANAGEMENT FILE .................... :
with accidental selection of
excessive output values for ME EQUIPMENT with
a multi-purpose unit designed to provide low
and high-intensity outputs for different
treatments addressed in RISK MANAGEMENT
PROCESS, confirmed in RISK MANAGEMENT FILE .. :
When applicable, RISKS associated with
incorrect output addressed in RISK MANAGEMENT
PROCESS as confirmed by review of RISK
MANAGEMENT FILE ............................................... :
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 79 of 203
Verdict
No function for accuracy of
controls and instruments and
protection against hazardous
outputs
N/A
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
12.4.5
Diagnostic or therapeutic radiation
12.4.5.1
Adequate provisions to protect OPERATORS,
PATIENTS, other persons and sensitive devices
in vicinity of unwanted or excessive radiation
emitted by ME EQUIPMENT designed to produce
radiation for diagnostic/therapeutic purposes
12.4.5.2
12.4.5.3
12.4.5.4
12.4.6
Result - Remark
Radiation safety ensured by compliance with
requirements of appropriate standards
RISKS associated with diagnostic X-rays
addressed in RISK MANAGEMENT PROCESS as
confirmed in RISK MANAGEMENT FILE .................. :
RISKS associated with radiotherapy addressed
in RISK MANAGEMENT PROCESS as confirmed by
review of RISK MANAGEMENT FILE ....................... :
RISKS associated with ME EQUIPMENT producing
diagnostic or therapeutic radiation other than
diagnostic X-rays and radiotherapy addressed
in RISK MANAGEMENT PROCESS as confirmed by
examination of RISK MANAGEMENT FILE .............. :
When applicable, RISKS associated with
diagnostic or therapeutic acoustic pressure
addressed in RISK MANAGEMENT PROCESS as
confirmed in RISK MANAGEMENT FILE .................. :
Verdict
N/A
No function for accuracy of
controls and instruments and
protection against hazardous
outputs
N/A
N/A
N/A
N/A
N/A
No function for accuracy of
controls and instruments and
protection against hazardous
outputs
N/A
13
HAZARDOUS SITUATIONS AND FAULT CONDITIONS
Pass
13.1
Specific HAZARDOUS SITUATIONS
Pass
13.1.1
None of HAZARDOUS SITUATIONS in 13.1.2-13.1.4,
inclusive, occurred when SINGLE FAULT
CONDITIONS applied, one at a time, as in 4.7 and
13.2
Pass
13.1.2
Emissions, deformation of ENCLOSURE or exceeding maximum temperature
Pass
– Emission of flames, molten metal, poisonous
or ignitable substance in hazardous quantities
did not occur
Pass
– Deformation of ENCLOSURE impairing
compliance with 15.3.1 did not occur
Pass
– Temperatures of APPLIED PARTS did not
No applied part
exceed allowable values in Table 24 when
measured as in 11.1.3 ....................................... :
N/A
– Temperatures of ME EQUIPMENT parts that are See appended table 13.2 for
not APPLIED PARTS likely to be touched did not
detail
exceed values in Table 23 when measured and
adjusted as in 11.1.3 ......................................... :
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 80 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
–Allowable values for “other components and
materials” in Table 22 times 1.5 minus 12.5 °C
were not exceeded
Pass
Limits for windings in Tables 26, 27, and 31 not Evaluated in part of power
exceeded
supply
N/A
Table 22 not exceeded in all other cases
N/A
Evaluated in part of power
supply
Temperatures measured according to 11.1.3
Pass
SINGLE FAULT CONDITIONS in 4.7, 8.1 b), 8.7.2, and
13.2.2 relative to emission of flames, molten
metal, or ignitable substances, not applied to
parts and components where:
Pass
– Supply circuit was unable to supply 15 W
one minute after 15 W drawn from supply
circuit, or
N/A
– Parts and components completely contained
within a fire ENCLOSURE complying with 11.3 as
verified by review of design documentation
Pass
After tests of this Clause, settings of THERMAL
CUT-OUTS and OVER-CURRENT RELEASES did not
change sufficiently to affect their safety
function
No THERMAL CUT-OUTS and
OVER-CURRENT RELEASES
N/A
– limits for LEAKAGE CURRENT in SINGLE FAULT
See appended Table 8.7.3
CONDITION based on 8.7.3 did not exceed.......... :
Pass
– voltage limits for ACCESSIBLE PARTS including See appended Tables 8.7.3
APPLIED PARTS in 8.4.2 did not exceed ............... :
Pass
13.1.4
ME EQUIPMENT complied with the requirements
of 9.1 to 9.8 for specific MECHANICAL HAZARDS
Pass
13. 2
SINGLE FAULT CONDITIONS
Pass
13.2.1
During application of SINGLE FAULT CONDITIONS in See appended Table 13.2
13.2.2 -13.2.13, inclusive, NORMAL CONDITIONS in
8.1 a) applied in least favourable combination :
Pass
13.2.2 –
13.2.12
ME EQUIPMENT complied with 13.2.2 -13.2.12..... : See appended Table 13.2
and RM Results Table 13.2.6
Pass
13.2.13
ME EQUIPMENT remained safe after tests of
13.2.13.2 to 13.2.13.4 (inclusive), and cooling
down to room temperature
Pass
ME EQUIPMENT examined for compliance or
appropriate tests such as dielectric strength of
motor insulation according to 8.8.3 conducted
Pass
13.1.3
Test Report No.: 1307033001
60601-1TRF_A
Page 81 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
13.2.13.2
Requirement + Test
Verdict
For insulation of thermoplastic materials relied
upon as a MEANS OF PROTECTION (see 8.8), the
ball-pressure test specified in 8.8.4.1 a)
performed at a temperature 25 °C higher than
temperature of insulation measured during tests
of 13.2.13.2 to 13.2.13.4 (inclusive).
Pass
ME EQUIPMENT with heating elements
N/A
a 1) thermostatically controlled ME EQUIPMENT
No heating elements
with heating elements for building-in, or for
unattended operation, or with a capacitor not
protected by a fuse connected in parallel with
THERMOSTAT contacts met tests of 13.2.13.2 b) &
13.2.13.2 c)
N/A
a 2) ME EQUIPMENT with heating elements RATED
for non-CONTINUOUS OPERATION met tests of
13.2.13.2 b) and 13.2.13.2 c)
N/A
a 3) other ME EQUIPMENT with heating elements
met test of 13.2.13.2 b)
N/A
When more than one test was applicable to
same ME EQUIPMENT, tests performed
consecutively
N/A
Heating period stopped when a heating
element or an intentionally weak part of a nonSELF-RESETTING THERMAL CUT- OUT ruptured, or
current interrupted before THERMAL STABILITY
without possibility of automatic restoration
N/A
Test repeated on a second sample when
interruption was due to rupture of a heating
element or an intentionally weak part
N/A
Both samples met 13.1.2, and open circuiting
of a heating element or an intentionally weak
part in second sample not considered a failure
by itself
N/A
b) ME EQUIPMENT with heating elements tested
per 11.1without adequate heat discharge, and
supply voltage set at 90 or 110 % of RATED
supply voltage, least favourable of the two (V)
.......................................................................... :
N/A
Operating period stopped when a non-SELFRESETTING THERMAL CUT-OUT operated, or current
interrupted without possibility of automatic
restoration before THERMAL STABILITY
N/A
ME EQUIPMENT switched off as soon as THERMAL
STABILITY established and allowed to cool to
room temperature when current not interrupted
N/A
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 82 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
13.2.13.3
Requirement + Test
Verdict
Test duration was equal to RATED operating
time for non-CONTINUOUS OPERATION
N/A
c) Heating parts of ME EQUIPMENT tested with ME
EQUIPMENT operated in NORMAL CONDITION at 110
% of RATED supply voltage and as in 11.1, and
N/A
1) Controls limiting temperature in NORMAL
CONDITION disabled, except THERMAL CUT -OUTS
N/A
2) When more than one control provided, they
were disabled in turn
N/A
3) ME EQUIPMENT operated at RATED DUTY CYCLE
until THERMAL STABILITY achieved, regardless of
RATED operating time
N/A
ME EQUIPMENT with motors
N/A
a 1) For the motor part of the ME EQUIPMENT,
No motor
compliance checked by tests of 13.2.8- 13.2.10,
13.2.13.3 b), 13.2.13.3 c), and 13.2.13.4, as
applicable
N/A
To determine compliance with 13.2.9 and
13.2.10 motors in circuits running at 42.4 V
peak a.c./ 60 V d.c. or less are covered with a
single layer of cheesecloth which did not ignite
during the test
N/A
a 2) Tests on ME EQUIPMENT containing heating
parts conducted at prescribed voltage with
motor & heating parts operated simultaneously
to produce the least favourable condition
N/A
a 3) Tests performed consecutively when more
tests were applicable to the same ME EQUIPMENT
N/A
b) Motor met running overload protection test
of this clause when:
N/A
1) it is intended to be remotely or automatically
controlled by a single control device with no
redundant protection, or
N/A
2) it is likely to be subjected to
OPERATION while unattended
N/A
CONTINUOUS
Motor winding temperature determined during
each steady period and maximum value did not
exceed Table 27 (Insulation Class, Maximum
temperature measured C) .............................. :
N/A
Motor removed from ME EQUIPMENT and tested
separately when load could not be changed in
appropriate steps
N/A
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 83 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
13.2.13.4
Requirement + Test
Result - Remark
Verdict
Running overload test for motors operating at
42.4 V peak a.c./60 V d.c. or less performed
only when examination and review of design
indicated possibility of an overload
N/A
Test not conducted where electronic drive
circuits maintained a substantially constant
drive current
N/A
Test not conducted based on other
justifications (justification) .............................. :
N/A
c) ME EQUIPMENT with 3-phase motors operated
with normal load, connected to a 3-phase
SUPPLY MAINS with one phase disconnected, and
periods of operation per 13.2.10
N/A
ME EQUIPMENT RATED for NON-CONTINUOUS OPERATION
N/A
ME EQUIPMENT (other than HAND-HELD) operated
under normal load and at RATED voltage or at
upper limit of RATED voltage range until
increase in temperature was  5 °C in one hour,
or a protective device operated
N/A
CONTINUOUS OPERATION
When a load-reducing device operated in
NORMAL USE, test continued with ME EQUIPMENT
running idle
N/A
Motor winding temperatures did not exceed
values in 13.2.10 .............................................. :
N/A
Insulation Class ............................................... :
—
Maximum temperature measured (C) ............ :
—
14
PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS)
N/A
14.1
Requirements of this clause not applied to
PESS when it provided no BASIC SAFETY or
ESSENTIAL PERFORMANCE, or
N/A
- when application of ISO 14971 showed that
failure of PESS does not lead to unacceptable
RISK .................................................................... :
N/A
Every PROCESS has been followed throughout
the PEMS DEVELOPMENT LIFE-CYCLE and a RECORD
of PROCESS has been made available as
confirmed by RISK MANAGEMENT FILE REVIEW and
assessment of PROCESSES cited in this Clause
N/A
Test Report No.: 1307033001
60601-1TRF_A
Since failure of PESS does
not lead to unacceptable risk
for this device, requirement
of this clause will not be
applied.
Page 84 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
MANUFACTURER considered the need for
additional RISK CONTROL measures when unable
to follow all PROCESSES identified in Clause 14
for each constituent component of PEMS as
confirmed by RISK MANAGEMENT FILE review and
assessment of PROCESSES cited in this Clause
N/A
Assessment of PROCESSES cited in this Clause
made by internal audits
N/A
14.2
Documents produced from application of
Clause 14 are maintained and form a part of
RISK MANAGEMENT FILE in addition to RECORDS
and documents required by ISO 14971 ............ :
N/A
14.3
RISK MANAGEMENT plan required by 3.5 of ISO
14971 includes reference to PEMS VALIDATION
plan
N/A
14.4
A PEMS DEVELOPMENT LIFE-CYCLE including a set
of defined milestones has been documented
N/A
At each milestone, activities to be completed,
and VERIFICATION methods to be applied to
activities have been defined
N/A
Each activity including its inputs and outputs
defined, and each milestone identifies RISK
MANAGEMENT activities that must be completed
before that milestone
N/A
PEMS DEVELOPMENT LIFE-CYCLE tailored for a
specific development by making plans
detailing activities, milestones, and schedules
N/A
PEMS DEVELOPMENT LIFE-CYCLE includes
documentation requirements
N/A
A documented system for problem resolution
within and between all phases and activities of
PEMS DEVELOPMENT LIFE-CYCLE has been
developed and maintained where appropriate
N/A
Problem resolution system meets the
prescribed criteria depending on type of
product:
N/A
– it is documented as a part of PEMS
DEVELOPMENT LIFE-CYCLE
N/A
– it allows reporting of potential or existing
problems affecting BASIC SAFETY or ESSENTIAL
N/A
14.5
PERFORMANCE
– it includes an assessment of each problem
for associated RISKS
Test Report No.: 1307033001
60601-1TRF_A
Page 85 of 203
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– it identifies criteria that must be met for the
issue to be closed
N/A
– it identifies the action to be taken to resolve
each problem
N/A
14.6
RISK MANAGEMENT PROCESS
N/A
14.6.1
MANUFACTURER considered HAZARDS associated
with software and hardware aspects of PEMS
including NETWORK/ DATA COUPLING, components
of third-party origin, legacy subsystems when
compiling list of known or foreseeable HAZARDS
N/A
In addition to the material in ISO 14971, Annex
D, list of possible sources for HAZARDS
associated with PEMS includes specified causes
N/A
– failure of NETWORK/ DATA COUPLING to provide
characteristics necessary for PEMS to achieve
its BASIC SAFETY or ESSENTIAL PERFORMANCE
N/A
– undesired feedback [physical and data] (such
as unsolicited/ out of range/ inconsistent input
or input from electromagnetic interference)
N/A
– unavailable data
N/A
– lack of integrity of data
N/A
– incorrect data
N/A
– incorrect timing of data
N/A
– unintended interactions within & among PESS
N/A
– unknown aspects or quality of third-party
software
N/A
– unknown aspects or quality of third-party
N/A
PESS
– lack of data security, particularly
vulnerability to tampering, unintended
interaction with other programs and viruses
N/A
14.6.2
Suitably validated tools and PROCEDURES
assuring each RISK CONTROL measure reduces
identified RISK(S) satisfactorily provided in
addition to PEMS requirements in Clause 6.1 of
ISO 14971 ........................................................... :
N/A
14.7
A documented requirement specification for
PEMS and each of its subsystems (e.g. for a
PESS) which includes ESSENTIAL PERFORMANCE
and RISK CONTROL measures implemented by
that system or subsystem ................................. :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 86 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
14.8
An architecture satisfying the requirement is
specified for PEMS and each of subsystems ..... :
N/A
The architecture specification makes use of
considers the specified items to reduce RISK to
an acceptable level, where appropriate:
N/A
a) COMPONENTS WITH HIGH-INTEGRITY
N/A
CHARACTERISTICS
b) fail-safe functions
N/A
c) redundancy
N/A
d) diversity;
N/A
e) partitioning of functionality
N/A
f) defensive design potentially limiting
hazardous effects by restricting available
output power or by introducing means to limit
travel of actuators
N/A
g) allocation of RISK CONTROL measures to
subsystems and components of PEMS
N/A
h) failure modes of components and their
effects;
N/A
i) common cause failures
N/A
j) systematic failures
N/A
k) test interval duration and diagnostic
coverage
N/A
l) maintainability
N/A
m) protection from reasonably foreseeable
misuse
N/A
n) NETWORK/ DATA COUPLING specification, when
applicable
N/A
14.9
Design is broken up into subsystems, each
with a design and test specification where
appropriate, and descriptive data on design
environment included in RISK MANAGEMENT FILE :
N/A
14.10
A VERIFICATION plan containing the specified
information used to verify and document
functions implementing BASIC SAFETY , ESSENTIAL
PERFORMANCE, or RISK CONTROL measures ......... :
N/A
– milestone(s) when VERIFICATION is to be
performed for each function
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 87 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– selection and documentation of VERIFICATION
strategies, activities, techniques, and
appropriate level of independence of the
personnel performing the VERIFICATION
N/A
– selection and utilization of VERIFICATION tools
N/A
– coverage criteria for VERIFICATION
N/A
A PEMS VALIDATION plan containing validation of
BASIC SAFETY & ESSENTIAL PERFORMANCE and
requiring checks for unintended functioning of
PEMS to perform and document PEMS VALIDATION
N/A
The person with overall responsibility for PEMS
VALIDATION is independent of design team, and
no member of a design team is responsible for
PEMS VALIDATION of their own design
N/A
All professional relationships of members of
PEMS VALIDATION team with members of design
team documented in RISK MANAGEMENT FILE
providing methods & results of PEMS VALIDATION
N/A
14.12
Continued validity of previous design
documentation assessed under a documented
modification/change PROCEDURE
N/A
14.13
Technical description includes the following
information when PEMS is to be connected to
other equipment outside control of PEMS
MANUFACTURER by NETWORK/DATA COUPLING ....... :
N/A
a) characteristics of NETWORK/ DATA COUPLING
necessary for PEMS to achieve its INTENDED USE
N/A
b) list of HAZARDOUS SITUATIONS resulting from a
failure of NETWORK/DATA COUPLING to provide the
specified characteristics
N/A
c) instructions to RESPONSIBLE ORGANIZATION
containing required information and warnings
N/A
– connection of PEMS to a NETWORK/ DATA
COUPLING that includes other equipment could
result in previously unidentified RISKS and
RESPONSIBLE ORGANIZATION shall identify,
analyze, and control such RISKS
N/A
– subsequent changes to NETWORK/DATA
COUPLING introducing new RISKS and requiring
new analysis; and changes to NETWORK/DATA
COUPLING include:
N/A
– NETWORK/ DATA COUPLING configuration change
N/A
– connection of additional items to
NETWORK/ DATA COUPLING
N/A
14.11
Test Report No.: 1307033001
60601-1TRF_A
Page 88 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– disconnecting items from NETWORK/ DATA
N/A
COUPLING
– update of equipment connected to
NETWORK/ DATA COUPLING
N/A
– upgrade of equipment connected to
NETWORK/ DATA COUPLING
N/A
15
CONSTRUCTION OF ME EQUIPMENT
Pass
15.1
RISKS associated with arrangement of controls
and indicators of ME EQUIPMENT addressed in
RISK MANAGEMENT PROCESS, as confirmed by
examination of RISK MANAGEMENT FILE .............. :
Arrangement of controls and
indicators of this device only
display the status of device
and will not impact the basic
safety
N/A
15.2
Parts of ME EQUIPMENT subject to mechanical
wear, electrical, environmental degradation or
ageing resulting in unacceptable RISK when
unchecked for a long period, are accessible for
inspection, replacement, and maintenance
No unacceptable risk after
examining, see appended
test table 15.3 for related test
results.
Pass
Inspection, servicing, replacement, and
adjustment of parts of ME EQUIPMENT can easily
be done without damage to or interference with
adjacent parts or wiring
N/A
15.3
Mechanical strength
Pass
15.3.1
Mold stress relief, push, impact, drop, and
rough handling tests did not result in
unacceptable RISK and ME EQUIPMENT displayed
adequate mechanical strength
Pass
15.3.2
Push test conducted by subjecting external
See Appended Table 15.3
parts of ENCLOSURE to a steady force of 250 N ± and RM Results Table 15.3.2
10 N for 5 s applied to a circular (30mm) plane
surface, except bottom of ENCLOSURE of an ME
EQUIPMENT >18 kg, using a suitable test tool ... :
Pass
No damage resulting in an unacceptable RISK
sustained as determined by examination of
Pass
RISK MANAGEMENT FILE
15.3.3
Impact test conducted by subjecting a
See Appended Table 15.3
complete ENCLOSURE or its largest nonand RM Results Table 15.3.3
reinforced area, except for HAND-HELD ME
EQUIPMENT and parts, to a free falling 500 g ± 25
g solid smooth steel ball, approx. 50 mm in
diameter from a height of 1.3 m ....................... :
Pass
Test not applied to flat panel displays, platen
glass of ME EQUIPMENT, or cathode ray tubes
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 89 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
No damage resulting in an unacceptable RISK
sustained as shown in RISK MANAGEMENT FILE
Pass
15.3.4
Drop test
Pass
15.3.4.1
Sample of HAND-HELD ME EQUIPMENT and HANDSee Appended Table 15.3
HELD part with SAFE WORKING LOAD allowed to fall
freely once from each of 3 different positions
as in NORMAL USE from height specified in
ACCOMPANYING DOCUMENTS, or from 1 m onto a
50 mm ± 5 mm thick hardwood board lying flat
on a concrete or rigid base............................... :
Pass
No unacceptable RISK resulted
Pass
Sample of PORTABLE ME EQUIPMENT and PORTABLE Test conducted with 5cm.
part with SAFE WORKING LOAD lifted to a height
See Appended Table 15.3
as in Table 29 above a 50 ± 5 mm thick
and RM Table 15.3.4.2
hardwood board lying flat on a concrete floor
or rigid base, dropped 3 times from each
orientation in NORMAL USE (cm) .........................:
Pass
No damage resulting in an unacceptable RISK
sustained as determined by examination of
sample and RISK MANAGEMENT FILE
N/A
Each sample of MOBILE ME EQUIPMENT and MOBILE No mobile part.
part with SAFE WORKING LOAD and in most
adverse condition in NORMAL USE passed Rough
Handling tests ................................................... :
N/A
a) Ascending step shock test conducted on the
sample by pushing it 3 times in its normal
direction of travel at 0.4 m/s ± 0.1 m/s against
an ascending hardwood step obstruction
without the sample going over the obstruction
N/A
b) Descending step shock test conducted on
the sample by pushing it 3 times in its normal
direction of travel at 0.4 m/s ± 0.1 m/s in order
to fall over a vertical step affixed flat on a rigid
base with direction of movement perpendicular
to face of the step until full descent achieved
N/A
c) Door frame shock test conducted on the
sample by moving it 3 times in its normal
direction of travel at 0.4 m/s ± 0.1 m/s, or for
motor driven EQUIPMENT, at maximum possible
speed against a hardwood vertical obstacle
higher than EQUIPMENT contact point(s)
N/A
No damage resulting in an unacceptable RISK
sustained as determined by examination of
sample and RISK MANAGEMENT FILE
N/A
15.3.4.2
15.3.5
Test Report No.: 1307033001
60601-1TRF_A
Page 90 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
15.3.6
Examination of ENCLOSURE made from molded
or formed thermoplastic material indicated that
material distortion due to release of internal
stresses by molding or forming operations will
not result in an unacceptable RISK
Pass
Mold-stress relief test conducted by placing
See table 15.3.6 for detail
one sample of complete ME EQUIPMENT,
ENCLOSURE or a portion of larger ENCLOSURE, for
7 hours in a circulating air oven at 10°C over
the max temperature measured on ENCLOSURE
in 11.1.3, but no less than 70 °C ........................ :
Pass
No damage resulting in an unacceptable RISK
Pass
15.3.7
Result - Remark
INTENDED USE, EXPECTED SERVICE LIFE, and
conditions for transport and storage were
taken into consideration for selection and
treatment of materials used in construction of
Verdict
See RMF xxxxx-ONYX2121DTy-xxxxxxx
_RMF_2013_07_19, Section Version data
Pass
ME EQUIPMENT
Based on review of EQUIPMENT, ACCOMPANYING
DOCUMENTS, specifications and processing of
materials, and MANUFACTURER’S relevant tests or
calculations, corrosion, ageing, mechanical
wear, degradation of biological materials due
to bacteria, plants, animals and the like, will
not result in an unacceptable RISK
Pass
15.4
ME EQUIPMENT components and general assembly
N/A
15.4.1
Incorrect connection of accessible connectors, No any outlet is provided on
removable without a TOOL, prevented where an the power adapter except for
unacceptable RISK exists, in particular............. : dc output connector itself.
N/A
a) Plugs for connection of PATIENT leads cannot No patient connections
be connected to other outlets on same ME
EQUIPMENT intended for other functions, except
when RISK MANAGEMENT FILE provides proof that
no unacceptable RISK could result ................... :
N/A
b) Medical gas connections on ME EQUIPMENT
No Medical gas connections
for different gases to be operated in NORMAL
USE are not interchangeable as verified by
review of RISK MANAGEMENT FILE ........................ :
N/A
15.4.2
Temperature and overload control devices
N/A
15.4.2.1
a) T HERMAL CUT-OUTS and OVER-CURRENT
No such component is
RELEASES with automatic resetting not used in
incorporated.
ME EQUIPMENT when their use could result in a
HAZARDOUS SITUATION by resetting action as
verified by review of RISK MANAGEMENT FILE ..... :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 91 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Verdict
b) T HERMAL CUT-OUTS with a safety function to
be reset by a soldering operation affecting
operating value not fitted in ME EQUIPMENT as
verified by examination of design and RISK
MANAGEMENT FILE ................................................ :
N/A
c) An independent non-SELF-RESETTING THERMAL
CUT-OUT is, additionally, provided where a
failure of a THERMOSTAT could constitute a
HAZARD as verified by examination of design
and RISK MANAGEMENT FILE ................................. :
N/A
d) Based on design and RISK MANAGEMENT FILE
review, loss of function of ME EQUIPMENT due to
operation of THERMAL CUT-OUT or OVER CURRENT
RELEASE doesn’t result in a HAZARDOUS SITUATION
N/A
e) Capacitors or other spark-suppression
devices not connected between contacts of
THERMAL CUT -OUTS
N/A
f) Use of THERMAL CUT-OUTS or OVER-CURRENT
RELEASES do not affect safety of ME EQUIPMENT
as verified by following tests:
N/A
Positive temperature coefficient devices (PTC’s)
complied with IEC 60730-1: 1999, clauses 15,
17, J.15, and J.17 as applicable
N/A
ME EQUIPMENT containing THERMAL CUT-OUTS and
OVER-CURRENT RELEASES operated under the
conditions of Clause 13 .................................... :
N/A
SELF-RESETTING THERMAL CUT-OUTS and OVERCURRENT RELEASES including circuits performing
equivalent functions (other than PTC's)
Certified according to appropriate standards ..
N/A
In the absence of Certification in accordance
with IEC standards, SELF-RESETTING THERMAL
CUT-OUTS and OVER-CURRENT RELEASES including
circuits performing equivalent functions (other
than PTC's) operated 200 times
N/A
Manual reset THERMAL CUT-OUTS and OVERCURRENT RELEASES Certified in accordance with
appropriate IEC standards
N/A
When certification based on IEC standards, or
data from MANUFACTURER demonstrating
reliability of component to perform its safetyrelated function is not available, manual reset
THERMAL CUT-OUTS and OVER-CURRENT RELEASES
operated 10 times
N/A
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 92 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Thermal protective devices tested separately
from ME EQUIPMENT when engineering judgment
indicated test results would not be impacted
N/A
g) Protective device, provided on ME EQUIPMENT
incorporating a fluid filled container with
heating means, operated when heater switched
on with container empty and prevented an
unacceptable RISK due to overheating
N/A
h) ME EQUIPMENT with tubular heating elements
provided with protection against overheating in
both leads where a conductive connection to
earth could result in overheating as verified by
review of design and RISK MANAGEMENT FILE ..... :
N/A
15.4.2.2
Temperature settings clearly indicated when
means provided to vary setting of THERMOSTATS
N/A
15.4.3
Batteries
Pass
15.4.3.1
Battery housings from which gases can
escape during charging or discharging likely to
result in a HAZARD ventilated to minimize RISK of
accumulation and ignition as verified by review
of design and RISK MANAGEMENT FILE ................. :
N/A
Battery compartments prevent accidental short
circuiting of battery when this could result in a
HAZARDOUS SITUATION as verified by examination
of design and RISK MANAGEMENT FILE
N/A
15.4.3.2
Means provided to prevent incorrect
See appended RM Results
connection of polarity when a HAZARDOUS
Table 15.4.3.2
SITUATION may develop by incorrect connection
After revered the connection
or replacement of a battery .............................. :
of the BT2, the battery was
not hazardous situation
came out.
Pass
15.4.3.3
Overcharging of battery prevented by virtue of See Appended RM Results
design when it could result in an unacceptable Table 15.4.3.3
RISK as verified by review of design ................ :
N/A
15.4.3.4
Lithium batteries that could become a HAZARD See Appended RM Results
complied with appropriate tests of IEC 60086-4 Table 15.4.3.4
N/A
Tests of IEC 60086-4 waived on the lithium
battery based on examination of design
N/A
A properly RATED protective device provided
within INTERNAL ELECTRICAL POWER SOURCE to
protect against fire caused by excessive
currents when (in case of a short circuit) layout
of internal wiring, cross-sectional area, rating
of connected components can result in a fire :
N/A
15.4.3.5
Test Report No.: 1307033001
60601-1TRF_A
Page 93 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
15.4.4
Requirement + Test
Result - Remark
Verdict
Protective device has adequate breaking
capacity to interrupt the maximum fault current
N/A
Justification for OVER-CURRENT RELEASES or FUSE
exclusion is included in RISK MANAGEMENT FILE
N/A
Indicator lights provided to indicate ME
EQUIPMENT is ready for NORMAL USE, except
when apparent to OPERATOR from normal
operating position, and marking of 7.4.1 are
insufficient for this purpose ............................. :
N/A
An additional indicator light provided on ME
EQUIPMENT with a stand-by state or a warm-up
state exceeding 15 s, except when apparent to
OPERATOR from normal operating position
No stand-by or warm-up
state
N/A
Indicator lights provided on ME EQUIPMENT
incorporating non-luminous heaters to indicate
heaters are operational when a HAZARDOUS
SITUATION could exist, except when apparent to
OPERATOR from normal operating position
N/A
Requirement not applied to heated stylus-pens
for recording purposes
N/A
Indicator lights provided on ME EQUIPMENT to
indicate an output exists where an accidental
or prolonged operation of output circuit could
constitute a HAZARDOUS SITUATION
N/A
Colours of indicator lights complied with 7.8.1
N/A
Charging mode visibly indicated in ME
EQUIPMENT incorporating a means for charging
an INTERNAL ELECTRICAL POWER SOURCE
N/A
15.4.5
RISKS associated with pre-set controls
No pre-set controls function
addressed in RISK MANAGEMENT PROCESS when
applicable as verified by review of RISK
MANAGEMENT FILE ................................................ :
N/A
15.4.6
Actuating parts of controls of ME EQUIPMENT
N/A
15.4.6.1
a) Actuating parts cannot be pulled off or
loosened up during NORMAL USE
N/A
b) Indication of scales (e.g., “on” “off”
positions, etc.) always corresponds to position
of controls with adjustment that can result in a
HAZARDOUS SITUATION for PATIENT or OPERATOR
while ME EQUIPMENT is in use
N/A
c) Incorrect connection of indicating device to
relevant component prevented by adequate
construction when it could be separated
without use of a TOOL
N/A
Test Report No.: 1307033001
60601-1TRF_A
No Actuating parts
Page 94 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
When torque values per Table 30 applied
between control knob and shaft of rotating
controls for not less than 2 s, 10 times in each
direction, knobs did not rotate ........................ :
N/A
Tests conducted by applying an axial force of
60 N for electrical components and 100 N for
other components for 1 min when an axial pull
was required in NORMAL USE with no
unacceptable RISK ............................................ :
N/A
Stops of adequate mechanical strength
provided on rotating/ movable parts of controls
of ME EQUIPMENT where necessary to prevent an
unexpected change from max to min, or viceversa, of the controlled parameter when this
could cause a HAZARDOUS SITUATION ................. :
N/A
Torque values in Table 30 applied 10 times in
each direction to rotating controls for 2 sec ... :
N/A
Application of an axial force of 60 N for
electrical components and 100 N for other
components to rotating or movable parts of
controls for 1 min when an axial pull was
required in NORMAL USE .................................... :
N/A
15.4.7
Cord-connected HAND-HELD and foot-operated control devices
N/A
15.4.7.1
a) HAND-HELD control devices of ME EQUIPMENT
complied with 15.3.4.1
N/A
15.4.6.2
15.4.7.2
15.4.7.3
Not Cord-connected handheld and foot-operated
control devices
b) Foot-operated control device supported an
actuating force of 1350 N for 1 min applied over
an area of 30 mm diameter in its position of
NORMAL USE with no damage to device causing
an unacceptable RISK ........................................ :
N/A
Control device of HAND-HELD and foot-operated
control devices turned in all possible abnormal
positions and placed on a flat surface ............ :
N/A
No unacceptable RISK caused by changing
control setting when accidentally placed in an
abnormal position
N/A
a) Foot-operated control device is at least IPX1
& complies with tests of IEC 60529 (IP Code)... :
N/A
b) ENCLOSURE of foot operated control devices
containing electrical circuits is at least IPX6
and complies with IEC 60529 if in NORMAL USE
liquids are likely to be found (IP Code) ............. :
N/A
Probability of occurrence estimated as part of
N/A
RISK MANAGEMENT PROCESS
Test Report No.: 1307033001
60601-1TRF_A
Page 95 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
15.4.8
Aluminum wires less than 16 mm2 in crosssectional area are not used
No aluminum wires
N/A
15.4.9
a) Oil container in PORTABLE ME EQUIPMENT
allows for expansion of oil and is adequately
sealed to prevent loss of oil in any position
No oil container
N/A
b) Oil containers in MOBILE ME EQUIPMENT sealed
to prevent loss of oil during transport
N/A
A pressure-release device operating during
NORMAL USE is, optionally, provided
N/A
c) Partially sealed oil-filled ME EQUIPMENT and its
parts provided with means for checking the oil
level to detect leakage
N/A
ME EQUIPMENT and technical description
examined, and manual tests conducted to
confirm compliance with above requirements
N/A
15.5
MAINS SUPPLY TRANSFORMERS OF ME EQUIPMENT and transformers providing
separation in accordance with 8.5
N/A
15.5.1
Overheating
N/A
15.5.1.1
Transformers of ME EQUIPMENT are protected
Evaluated in part of power
against overheating in the event of short
supply
circuit or overload of output windings and
comply with this Clause and tests of 15.5.1.2 –
3 ......................................................................... :
N/A
During tests, windings did not open, no
HAZARDOUS SITUATION occurred, and maximum
temperatures of windings did not exceed
values in Table 31
N/A
Dielectric strength test of 8.8.3 conducted on
transformer after short circuit and overload
tests .................................................................. :
N/A
Transformer output winding short circuited,
Evaluated in part of power
and test continued until protective device
supply
operated or THERMAL STABILITY achieved .......... :
N/A
Short circuit applied directly across output
windings for transformers not tested according
to 5X frequency and 5X voltage test of 15.5.2
N/A
15.5.1.3
Multiple overload tests conducted on windings Evaluated in part of power
with more than one protective device to
supply
evaluate worst-case NORMAL USE loading and
protection .......................................................... :
N/A
15.5.2
Transformer windings provided with adequate Evaluated in part of power
insulation to prevent internal short-circuits that supply
could cause overheating which could result in
a HAZARDOUS SITUATION
N/A
15.5.1.2
Test Report No.: 1307033001
60601-1TRF_A
Page 96 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Dielectric strength tests were conducted in
accordance with requirements of this clause
with no breakdown of insulation system and no
detectable deterioration of transformer .......... :
N/A
15.5.3
Transformers forming MEANS OF PROTECTION as
required by 8.5 comply with IEC 61558-1:1997,
Clause 5.12 ........................................................ :
N/A
16
ME SYSTEMS
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
16.1
After installation or subsequent modification,
Product is a Slim Medical
ME SYSTEM didn’t result in an unacceptable RISK Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
Only HAZARDS arising from combining various
equipment to form a ME SYSTEM considered
N/A
– ME SYSTEM provides the level of safety within
the PATIENT ENVIRONMENT equivalent to ME
EQUIPMENT complying with this standard
N/A
– ME SYSTEM provides the level of safety outside
PATIENT ENVIRONMENT equivalent to equipment
complying with their respective IEC or ISO safety
standards
N/A
– tests performed in
as specified
except
N/A
– tests performed under operating conditions
specified by MANUFACTURER of ME SYSTEM
N/A
Safety tests previously conducted on
individual equipment of ME SYSTEM according to
relevant standards not repeated
N/A
RISK MANAGEMENT methods, optionally, used by
MANUFACTURER of an ME SYSTEM reconfigurable
by RESPONSIBLE ORGANIZATION or OPERATOR to
determine configurations with highest RISKS
and measures to ensure any configuration of
ME SYSTEM will not present unacceptable RISKS
N/A
Non-ME EQUIPMENT used in ME SYSTEM complied
with applicable IEC or ISO safety standards
N/A
Test Report No.: 1307033001
60601-1TRF_A
NORMAL CONDITION,
Page 97 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Equipment relying only on BASIC INSULATION for
protection against electric shock not used in
N/A
ME SYSTEM
16.2
ACCOMPANYING DOCUMENTS of an ME SYSTEM
Documents containing all data necessary for
ME SYSTEM to be used as intended by
MANUFACTURER including a contact address
accompany ME SYSTEM or modified ME SYSTEM
ACCOMPANYING DOCUMENTS regarded as a part of
N/A
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
N/A
ME SYSTEM
ACCOMPANYING DOCUMENTS are, optionally,
provided in electronic format (e.g. electronic
file format or CD ROM ) and ME SYSTEM is capable
of displaying or printing these documents
N/A
a) ACCOMPANYING DOCUMENTS provided for each
item of ME EQUIPMENT supplied by MANUFACTURER
N/A
b) ACCOMPANYING DOCUMENTS provided for each
item of non-ME EQUIPMENT supplied by
N/A
MANUFACTURER
c) the required information is provided:
N/A
– specifications, instructions for use as
intended by MANUFACTURER, and a list of all
items forming the ME SYSTEM
N/A
– instructions for installation, assembly, and
modification of ME SYSTEM to ensure continued
compliance with this standard
N/A
– instructions for cleaning and, when
applicable, disinfecting and sterilizing each
item of equipment or equipment part forming
part of the ME SYSTEM
N/A
– additional safety measures to be applied
during installation of ME SYSTEM
N/A
– identification of parts of ME SYSTEM suitable
for use within the PATIENT ENVIRONMENT
N/A
– additional measures to be applied during
preventive maintenance
N/A
– a warning forbidding placement of MULTIPLE
SOCKET-OUTLET, when provided and it is a
separate item, on the floor
N/A
– a warning indicating an additional MULTIPLE
SOCKET-OUTLET or extension cord not to be
connected to ME SYSTEM
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 98 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– a warning to connect only items that have
been specified as part of ME SYSTEM or specified
as being compatible with ME SYSTEM
N/A
– maximum permissible load for any MULTIPLE
SOCKET-OUTLET(S) used with ME SYSTEM
N/A
– instructions indicating MULTIPLE SOCKETOUTLETS provided with the ME SYSTEM to be used
only for supplying power to equipment
intended to form part of ME SYSTEM
N/A
– an explanation indicating RISKS of connecting
non-ME EQUIPMENT supplied as a part of ME
SYSTEM directly to wall outlet when non-ME
EQUIPMENT is intended to be supplied via a
MULTIPLE SOCKET - OUTLET with a separating
transformer
N/A
– an explanation indicating RISKS of connecting
any equipment supplied as a part of ME SYSTEM
to MULTIPLE SOCKET-OUTLET
N/A
– permissible environmental conditions of use
for ME SYSTEM including conditions for transport
and storage
N/A
– instructions to OPERATOR not to,
simultaneously, touch parts referred to in 16.4
and PATIENT
N/A
d) the following instructions provided for use
by RESPONSIBLE ORGANIZATION:
N/A
– adjustment, cleaning, sterilization, and
disinfection PROCEDURES
N/A
– assembly of ME SYSTEMS and modifications
during actual service life shall be evaluated
based on the requirements of this standard
N/A
16.3
Instructions for use of ME EQUIPMENT intended to
receive its power from other equipment in an
ME SYSTEM , describe the other equipment to
ensure compliance with these requirements
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
16.4
Parts of non-ME EQUIPMENT in PATIENT
ENVIRONMENT subject to contact by OPERATOR
during maintenance, calibration, after removal
of covers, connectors, etc., without use of a
TOOL operated at a voltage  voltage in 8.4.2 c)
supplied from a source separated from SUPPLY
MAINS by two MEANS OF OPERATOR PROTECTION
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 99 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
16.5
Safety measures incorporating a SEPARATION
DEVICE applied when FUNCTIONAL CONNECTION
between ME EQUIPMENT and other items of an ME
SYSTEM or other systems can cause allowable
values of LEAKAGE CURRENT to exceed
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
SEPARATION DEVICE has dielectric strength,
CREEPAGE and CLEARANCES required for one
MEANS OF OPERATOR PROTECTION appropriate for
highest voltage occurring across SEPARATION
DEVICE during a fault condition
N/A
WORKING VOLTAGE was highest voltage across
SEPARATION DEVICE during a fault condition, but
not less than MAXIMUM MAINS VOLTAGE (V)......... :
N/A
16.6
LEAKAGE CURRENTS
N/A
16.6.1
T OUCH CURRENT in NORMAL CONDITION, from or
between parts of ME SYSTEM within the PATIENT
ENVIRONMENT , did not exceed 100 μA ................ :
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
T OUCH CURRENT did not exceed 500 μA in event
of interruption of any non-PERMANENTLY
INSTALLED PROTECTIVE EARTH CONDUCTOR, from or
between parts of ME SYSTEM within PATIENT
ENVIRONMENT ..................................................... :
N/A
16.6.2
Current in PROTECTIVE EARTH CONDUCTOR of
MULTIPLE SOCKET - OUTLET did not exceed 5 mA .. :
N/A
16.6.3
PATIENT LEAKAGE CURRENT and total PATIENT
LEAKAGE CURRENT of ME SYSTEM in NORMAL
CONDITION did not exceed values specified for
ME EQUIPMENT in Tables 3 and 4 ........................ :
N/A
Measurements made using a device as in
clause 8.7.4.4
N/A
16.7
ME SYSTEM complied with applicable
requirements of Clause 9 when a MECHANICAL
HAZARD existed .................................................. :
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
16.8
Interruption and restoration of relevant power
connections of ME SYSTEM one at a time and all
connections simultaneously did not result in a
HAZARDOUS SITUATION other than interruption of
its intended function
Product is a Slim Medical
Panel PC, application and
assessment for ME system
is not considered in current
status.
N/A
16.9
ME SYSTEM connections and wiring
Test Report No.: 1307033001
60601-1TRF_A
Page 100 of 203
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
16.9.1
Incorrect connection of accessible connectors, Product is a Slim Medical
removable without a TOOL, prevented where a
Panel PC, application and
HAZARDOUS SITUATION could otherwise exist ..... : assessment for ME system
is not considered in current
status.
N/A
– Connectors complied with Clause 15.4.1
N/A
– Plugs for connection of PATIENT leads could
not be connected to other outlets of the same
ME SYSTEM likely to be located in PATIENT
ENVIRONMENT , except when examination of
connectors and interchanging them proved no
HAZARDOUS SITUATION could result
N/A
16.9.2
MAINS PARTS, components and layout
N/A
16.9.2.1
a) – MULTIPLE SOCKET-OUTLET only allows
connection using a TOOL, or
N/A
– MULTIPLE SOCKET-OUTLET is of a type that
cannot accept MAINS PLUGS of any of the kinds
specified in IEC/TR 60083, or
N/A
– MULTIPLE SOCKET-OUTLET is supplied via a
separating transformer
N/A
b) – MULTIPLE SOCKET-OUTLET marked with safety
sign 2 of Table D.2 (i.e., safety sign ISO 7010W001) visible in NORMAL USE, and
N/A
– marked either individually or in combinations,
with the maximum allowed continuous output
in amperes or volt-amperes, or
N/A
– marked to indicate the equipment or
equipment parts it may safely be attached to
N/A
– MULTIPLE SOCKET-OUTLET is a separate item or
an integral part of ME EQUIPMENT or non-ME
N/A
EQUIPMENT
c) MULTIPLE SOCKET-OUTLET complied with IEC
60884-1 and the following requirements:
N/A
– CREEPAGE and CLEARANCES complied with 8.9
N/A
– It is CLASS I, and PROTECTIVE EARTH CONDUCTOR
is connected to earthing contacts in socketoutlets
N/A
– PROTECTIVE EARTH TERMINALS and PROTECTIVE
EARTH CONNECTIONS comply with 8.6, except total
impedance for ME SYSTEM was up to 400 m, or
higher when conditions of 8.6.4 b) met (m).. :
N/A
– ENCLOSURE complied with 8.4.2 d)
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 101 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– MAINS TERMINAL DEVICES and wiring complied
with 8.11.4, when applicable
N/A
– RATINGS of components are not in conflict
with conditions of use ..................................... :
N/A
– Electrical terminals and connectors of
MULTIPLE SOCKET - OUTLETS prevent incorrect
connection of accessible connectors
removable without a TOOL
N/A
– POWER SUPPLY CORD complied with 8.11.3
N/A
d) Additional requirements applied when
MULTIPLE SOCKET - OUTLET combined with a
separating transformer:
N/A
– Separating transformer complied with IEC
61558-2-1, except requirements of maximum
RATED output power of 1 kVA and degree of
protection IPX4 were not applied ..................... :
N/A
– Separating transformer is CLASS I
N/A
– Degree of protection against ingress of water
specified as in IEC 60529
N/A
– Separating transformer assembly marked
according to 7.2 and 7.3
N/A
– MULTIPLE SOCKET-OUTLET permanently
connected to separating transformer, or
socket-outlet of separating transformer
assembly cannot accept MAINS PLUGS as
identified in IEC/TR 60083
N/A
Removal of any single item of equipment in ME
SYSTEM will not interrupt the protective earthing
of any other part without simultaneous
disconnection of electrical supply to that part
N/A
Additional PROTECTIVE EARTH CONDUCTORS can be
detachable only by use of a TOOL
N/A
16.9.2.3
Conductors connecting different items within
an ME SYSTEM protected against mechanical
damage
N/A
17
ELECTROMAGNETIC COMPATIBILITY OF ME EQUIPMENT AND ME
SYSTEMS
N/A
RISKS associated with items addressed in RISK
MANAGEMENT PROCESS as confirmed by review . :
N/A
16.9.2.2
– electromagnetic phenomena at locations
where ME EQUIPMENT or ME SYSTEM is to be used
as stated in ACCOMPANYING DOCUMENTS ............. :
Test Report No.: 1307033001
60601-1TRF_A
Page 102 of 203
Not evaluated by Victronic
Technology Corporation.
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– introduction of electromagnetic phenomena
into environment by ME EQUIPMENT or ME SYSTEM
that might degrade performance of other
devices, electrical equipment, and systems
N/A
ANNEX G PROTECTION AGAINST HAZARDS OF IGNITION OF FLAMMABLE
ANESTHETIC MIXTURES
N/A
G.2
Locations and basic requirements
N/A
G.2.1
Parts of CATEGORY APG ME EQUIPMENT in which a
FLAMMABLE ANESTHETIC MIXTURE WITH AIR occurs
are CATEGORY AP or APG ME EQUIPMENT and
complied with G.3, G.4, and G.5
G.2.2
FLAMMABLE AESTHETIC MIXTURE WITH AIR occurring
due to a leakage or discharge of a FLAMMABLE
AESTHETIC MIXTURE WITH OXYGEN or NITROUS OXIDE
from an ENCLOSURE considered 5 to 25 cm from
point of occurrence
N/A
G.2.3
A FLAMMABLE AESTHETIC MIXTURE WITH OXYGEN or
NITROUS OXIDE contained in a completely / partly
enclosed ME EQUIPMENT part and in PATIENT’S
respiratory tract 5 cm from an ENCLOSURE part
where leakage or discharge occurs
N/A
G.2.4
ME EQUIPMENT or parts thereof specified for use
with FLAMMABLE AESTHETIC MIXTURE WITH AIR (in a
location as in G.2.2) are CATEGORY AP or APG ME
EQUIPMENT and complied with G.4 and G.5
N/A
G.2.5
ME EQUIPMENT or parts thereof for use with
N/A
Equipment not suitable for
use in the presence of
FLAMMABLE ANESTHETIC
MIXTURES
N/A
FLAMMABLE AESTHETIC MIXTURE WITH OXYGEN OR
NITROUS OXIDE (location per G.2.2) are CATEGORY
APG ME EQUIPMENT and comply with G.4 and G.6
ME EQUIPMENT in G.2.3 to G.2.5 met appropriate
tests of G.3-G.5 conducted after tests of 11.6.6
and 11.6.7
N/A
G.3
Marking, ACCOMPANYING DOCUMENTS
N/A
G.3.1
CATEGORY APG ME EQUIPMENT prominently
marked. with a green-coloured band  2 cm
wide with letters “APG” according to symbol 23
in Table D.1 ....................................................... :
N/A
Length of green-coloured band is  4 cm, and
size of marking is as large as possible for
particular case
N/A
When above marking not possible, relevant
information included in instructions for use ... :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Equipment not suitable for
use in the presence of
FLAMMABLE ANESTHETIC
MIXTURES
Page 103 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Marking complied with tests and criteria of
7.1.2 and 7.1.3
N/A
CATEGORY AP ME EQUIPMENT prominently marked,
with a green-coloured circle  2 cm in diameter,
with characters “AP” according to symbol 22 in
Table D.1 ........................................................... :
N/A
Marking is as large as possible for the
particular case
N/A
When above marking not possible, the relevant
information included in instructions for use ... :
N/A
Marking complied with tests and criteria of
7.1.2 and 7.1.3
N/A
G.3.3
The marking according to G.3.2 and G.3.3
placed on major part of ME EQUIPMENT for
CATEGORY AP or APG parts, and not repeated on
detachable parts that can only be used with the
marked EQUIPMENT
N/A
G.3.4
ACCOMPANYING DOCUMENTS contain an indication
enabling the RESPONSIBLE ORGANIZATION to
distinguish between CATEGORY AP and APG parts
N/A
G.3.5
Marking clearly indicates which parts are
CATEGORY AP or APG when only certain ME
EQUIPMENT parts are CATEGORY AP or APG
N/A
G.4
Common requirements for CATEGORY AP and CATEGORY APG ME EQUIPMENT
N/A
G.4.1
a) CREEPAGE and CLEARANCES between points of Equipment not suitable for
POWER SUPPLY CORD connection are according to use in the presence of
Table 12 for one MEANS OF PATIENT PROTECTION
FLAMMABLE ANESTHETIC
MIXTURES
N/A
b) Connections, except those in circuits
described in G.5.3 and G.6.3, protected against
accidental disconnection in NORMAL USE or
connection and disconnection can be
performed only with a TOOL
N/A
c) CATEGORY AP and APG not provided with a
DETACHABLE POWER SUPPLY CORD, except when
circuit complied with G.5.3 and G.6.3
N/A
Construction details
N/A
a) Opening of an ENCLOSURE providing
protection against penetration of gases or
vapours into ME EQUIPMENT or its parts possible
only with a TOOL
N/A
b) ENCLOSURE complies with requirements to
minimize arcing and sparking due to
penetration of foreign objects .......................... :
N/A
G.3.2
G.4.2
Test Report No.: 1307033001
60601-1TRF_A
Page 104 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– no openings on top covers of ENCLOSURE,
except for openings for controls covered by
control knobs
N/A
– openings in side-covers prevented
penetration of a solid cylindrical test rod of 4
mm in diameter applied in all possible
directions without appreciable force
N/A
– openings in base plates prevented
penetration of a solid cylindrical test rod of 12
mm in diameter applied in all directions without
appreciable force
N/A
c) Short circuiting conductor(s) to a conductive
part without presence of explosive gasses
where insulation may contact a part containing
a FLAMMABLE AESTHETIC MIXTURE WITH OXYGEN or
NITROUS OXIDE, ignitable gases alone, or oxygen,
did not result in loss of integrity of the part, an
unacceptable temperature, or other HAZARD
N/A
a) Electrostatic charges prevented on CATEGORY
AP and APG ME EQUIPMENT by a combination of
appropriate measures
N/A
– Use of antistatic materials with a limited
electrical resistance as specified in G.4.3 b) ... :
N/A
– Provision of electrically conductive paths
from ME EQUIPMENT or its parts to a conductive
floor, protective earth or potential equalization
system, or via wheels to an antistatic floor of
medical room
N/A
b) Electrical resistance limits of aesthetic
tubing, mattresses and pads, castor tires, and
other antistatic material complied with ISO
2882 based on measurements according to ISO
1853, ISO 2878 and ISO 23529 .......................... :
N/A
G.4.4
Corona cannot be produced by components or
parts of ME EQUIPMENT operating at more than
2000 V a.c. or 2400 V d.c. and not included in
ENCLOSURES complying with G.5.4 or G.5.5
N/A
G.5
Requirements and tests for CATEGORY AP ME EQUIPMENT, parts and components
N/A
G.5.1
ME EQUIPMENT, its parts or components do not
ignite FLAMMABLE AESTHETIC MIXTURES WITH AIR
under NORMAL USE and CONDITIONS based on
compliance with G.5.2 to G.5.5 (inclusive)
N/A
G.4.3
Test Report No.: 1307033001
60601-1TRF_A
Page 105 of 203
Equipment not suitable for
use in the presence of
FLAMMABLE ANESTHETIC
MIXTURES
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Alternatively, ME EQUIPMENT, its parts, and
components complied with requirements of IEC
60079-0 for pressurized ENCLOSURES (IEC 600792); for sand-filled ENCLOSURES, IEC 60079-5; or
for oil immersed equipment, IEC 60079-6; and
with this standard excluding G.5.2 to G.5.5 ..... :
N/A
G.5.2
ME EQUIPMENT, its parts, and components in
contact with gas mixtures in NORMAL USE and
CONDITIONS not producing sparks and not
resulting in surface temperatures above 150 °C
in case of restricted or 200 °C in case of
unrestricted vertical air circulation measured at
25 °C comply with G.5.1.................................... :
N/A
G.5.3
ME EQUIPMENT, its parts, and components
producing sparks in NORMAL USE and CONDITION
complied with temperature requirements of
G.5.2, and Umax and Imax occurring in their
circuits, and complied as follows:
N/A
Measured Umax ≤ UzR with IzR as in Fig. G.1 ..... :
N/A
Measured Umax ≤ Uc with Cmax as in Fig. G.2 ... :
N/A
Measured Imax ≤ IzR with UzR as in Fig G.1 ....... :
N/A
Measured Imax ≤ IzL with Lmax and a Umax ≤ 24 V
as in Fig G.3 ..................................................... :
N/A
– Combinations of currents and corresponding
voltages within the limitations IzR.UzR ≤ 50 W
extrapolated from Fig G.1
N/A
No extrapolation made for voltages above 42 V
N/A
– Combinations of capacitances and
corresponding voltages within limitations of
C/2U2  1.2 mJ extrapolated from Fig G.2
N/A
No extrapolation made for voltages above 242V
N/A
Umax, additionally, determined using actual
resistance R when the equivalent resistance R
was less than 8000 
N/A
– Combinations of currents and corresponding
inductances within limitations L/2I2  0.3 mJ
extrapolated from Fig G.3
N/A
No extrapolation made for inductances larger
than 900 mH
N/A
– Umax was the highest supply voltage
occurring in circuit under investigation with
sparking contact open, taking into
consideration MAINS VOLTAGE variations in 4.10
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 106 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
G.5.4
Requirement + Test
Verdict
– Imax was the highest current flowing in circuit
under investigation with sparking contact
closed, taking into consideration MAINS VOLTAGE
variations required in 4.10
N/A
– Cmax and Lmax taken as values occurring at the
component under investigation producing
sparks
N/A
– Peak value considered when a.c. supplied
N/A
– An equivalent circuit calculated to determine
equivalent max capacitance, inductance, and
equivalent Umax and Imax, either as d.c. or a.c.
peak values in case of a complicated circuit ... :
N/A
Temperature measurements made according to
11.1, and Umax, Imax, R, Lmax , and Cmax
determined with application of Figs G.1-G.3 .. :
N/A
Alternatively, compliance was verified by
examination of design data ..............................:
N/A
External ventilation with internal overpressure
N/A
ME EQUIPMENT, its parts, and components
enclosed in an ENCLOSURE with external
ventilation by means of internal overpressure
complied with the following requirements:
N/A
a) FLAMMABLE AESTHETIC MIXTURES WITH AIR that
might have penetrated into ENCLOSURE of ME
EQUIPMENT or part removed by ventilation before
EQUIPMENT energized, and penetration of such
mixtures during operation was prevented by
maintenance of overpressure by means of air
without flammable gases, or by physiologically
acceptable inert gas (e.g., nitrogen)
N/A
b) Overpressure inside ENCLOSURE was 75 Pa,
min., in NORMAL CONDITION (Pa) ......................... :
N/A
Overpressure maintained at the site of potential
ignition even when air or inert gas could
escape through openings in ENCLOSURE
necessary for normal operation of ME EQUIPMENT
or its parts
N/A
ME EQUIPMENT could be energized only after the
required minimum overpressure was present
long enough to ventilate the ENCLOSURE so that
the displaced volume of air or inert gas was at
least five times the volume of ENCLOSURE
N/A
ME EQUIPMENT energized at will or repeatedly
when overpressure was continuously present
N/A
Test Report No.: 1307033001
60601-1TRF_A
Result - Remark
Page 107 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
c) Ignition sources de-energized automatically
by means used where G.4 does not apply, or
complied with G.5 when during operation
overpressure dropped below 50 Pa (Pa) ......... :
N/A
d) External surface of ENCLOSURE in which
internal overpressure was maintained did not
exceed 150 °C in 25 °C ambient under NORMAL
USE and CONDITION (°C) ...................................... :
N/A
ENCLOSURES with restricted breathing
N/A
ME EQUIPMENT, its parts, and components
enclosed in an ENCLOSURE with restricted
breathing complied with the following:
N/A
a) A FLAMMABLE AESTHETIC MIXTURE WITH AIR did
not form inside ENCLOSURE with restricted
breathing when it was surrounded by a
FLAMMABLE AESTHETIC MIXTURE WITH AIR of a high
concentration for at least 30 min without any
pressure difference inside ENCLOSURE
N/A
b) Gasket or sealing material used to maintain
tightness complied with aging test B-b of IEC
60068-2-2, Clause 15, at 70 °C ± 2 °C and 96 h. :
N/A
c) Gas-tightness of ENCLOSURE containing inlets
for flexible cords maintained when the cords
were stressed by bending or pulling
N/A
Cords are fitted with adequate anchorages to
limit stresses
N/A
After the test in G.5.4 b), an internal
overpressure of 400 Pa was created and 30
pulls of the value in Table G.1 applied to each
flexible cord in axial direction of cord inlet and
in the least favourable direction for 1 s
N/A
Overpressure not reduced below 200 Pa
N/A
Tests waived when examination of ENCLOSURE
indicated it is completely sealed or gas-tight
without a doubt (100 % degree of certainty)
N/A
Operating temperature of external surface of
ENCLOSURE was  150 °C in 25 °C (°C) .............. :
N/A
Steady state operating temperature of
ENCLOSURE also measured (°C) ........................ :
N/A
G.6
CATEGORY APG ME EQUIPMENT,
N/A
G.6.1
ME EQUIPMENT, its parts, and components did
not ignite FLAMMABLE AESTHETIC MIXTURE WITH
OXYGEN OR NITROUS OXIDE under NORMAL USE and
G.5.5
parts and components thereof
SINGLE FAULT CONDITION
Test Report No.: 1307033001
60601-1TRF_A
Page 108 of 203
Equipment not suitable for
use in the presence of
FLAMMABLE ANESTHETIC
MIXTURES
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
ME EQUIPMENT, its parts, and components not
complying with G.6.3 subjected to a CONTINUOUS
OPERATION test after attaining thermal steady
state (max. 3 h) over a period of 10 min in a
12.2 % ± 0.4 ether by volume/oxygen mixture
N/A
G.6.2
Parts and components of CATEGORY APG ME
EQUIPMENT operating in a FLAMMABLE AESTHETIC
MIXTURE WITH OXYGEN OR NITROUS OXIDE supplied
from a source isolated from earth by insulation
equal to one MEANS OF PATIENT PROTECTION and
from electrical parts by insulation twice the
MEANS OF PATIENT PROTECTION ............................. :
N/A
G.6.3
Test of G.6.1 waived when the following
requirements were met in NORMAL USE and
under NORMAL and SINGLE FAULT CONDITIONS ...... :
N/A
a) no sparks produced and temperatures did
not exceed 90 °C, or
N/A
b) a temperature limit of 90 °C not exceeded,
sparks produced in NORMAL USE, and SINGLE
FAULT CONDITIONS, except Umax and Imax occurring
in their circuits complied with requirements,
taking Cmax and Lmax into consideration:
N/A
Measured Umax ≤ UzR with IzR as in Fig. G.4 .... :
N/A
Measured Umax ≤ UzC with Cmax as in Fig. G.5 ... :
N/A
Measured Imax ≤ IzR with UzR as in Fig G.4 ....... :
N/A
Measured Imax ≤ IzL with Lmax and a Umax ≤ 24 V
as in Fig G.6 .................................................... :
N/A
– Extrapolation from Figs G.4, G.5, and G.6 was
limited to areas indicated
N/A
– Umax was the highest no-load voltage
occurring in the circuit under investigation,
taking into consideration mains voltage
variations as in 4.10
N/A
– Imax was the highest current flowing in the
circuit under investigation, taking into account
MAINS VOLTAGE variations as in 4.10
N/A
– Cmax and Lmax are values occurring in relevant
circuit
N/A
– Umax additionally determined with actual
resistance R when equivalent resistance R in
Fig G.5 was less than 8000 
N/A
– Peak value taken into consideration when a.c.
supplied
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 109 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
G.6.4
Requirement + Test
Result - Remark
Verdict
– An equivalent circuit calculated to determine
max capacitance, inductance, and Umax and Imax,
either as d.c. or a.c. peak values in case of a
complicated circuit .......................................... :
N/A
– When energy produced in an inductance or
capacitance in a circuit is limited by voltage or
current-limiting devices, two independent
components applied, to obtain the required
limitation even when a first fault (short or open
circuit) in one of these components
N/A
Above requirement not applied to transformers
complying with this standard
N/A
Above requirement not applied to wire-wound
current-limiting resistors provided with a
protection against unwinding of the wire in
case of rupture
N/A
Compliance verified by examination of
CATEGORY APG ME EQUIPMENT, parts, and
components , or
N/A
Temperature measurements made in
accordance with 11.1, or
N/A
Umax, Imax, R, Lmax and Cmax determined together
with application of Figs G.4-G.6 ...................... :
N/A
Alternatively, compliance verified by
comparison with design data ...........................:
N/A
ME EQUIPMENT, its parts, and components
heating a FLAMMABLE AESTHETIC MIXTURE WITH
OXYGEN OR NITROUS OXIDE provided with a nonSELF-RESETTING THERMAL CUT- OUT and complied
with 15.4.2.1 ....................................................... :
N/A
Current-carrying part of heating element is not
in direct contact with FLAMMABLE AESTHETIC
N/A
MIXTURE WITH OXYGEN OR NITROUS OXIDE
G.7
Test apparatus for flammable mixtures
Test apparatus used was in accordance with
this Clause and Fig G.7
ANNEX L
Equipment not suitable for
use in the presence of
FLAMMABLE ANESTHETIC
MIXTURES
INSULATED WINDING WIRES FOR USE WITHOUT INTERLEAVED
INSULATION
Test Report No.: 1307033001
60601-1TRF_A
N/A
Page 110 of 203
N/A
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
L.1
BASIC, SUPPLEMENTARY , DOUBLE, and REINFORCED Evaluated in part of power
INSULATION in wound components without
supply
interleaved insulation complied with this Annex
covering round winding wires between 0.05
mm and 5.00 mm diameters
N/A
L.2
Wire construction
N/A
Overlap of layers when wire is insulated with
two or more spirally wrapped layers of tape is
adequate to ensure continued overlap during
manufacture of wound component
N/A
Layers of spirally wrapped wire insulation are
sufficiently secured to maintain the overlap
N/A
Type Test
N/A
The wire subjected to tests of L.3.1 to L.3.4 at a Evaluated in part of power
temperature and a relative humidity specified
supply
N/A
L.3
L.3.1
L.3.2
Result - Remark
Temperature (°C).............................................. :
—
Humidity (%) ..................................................... :
—
Dielectric strength
N/A
Dielectric strength test of Clause 8.8.3 for the
Evaluated in part of power
appropriate type and number of MOP(s)
supply
conducted by preparing the sample according
to IEC 60851-5:1996, Clause 4.4.1 for a twisted
pair with test voltages at least twice Tables 6 &
7, but not less than below with no breakdown:
N/A
– 3000 V for BASIC and SUPPLEMENTARY
INSULATION (V).................................................... :
N/A
– 6000 V for REINFORCED INSULATION (V) ............ :
N/A
Flexibility and adherence
N/A
Sample subjected to flexibility and adherence
test 8 of IEC 60851-3:1996, clause 5.1.1, using
mandrel diameters of Table L.1
Evaluated in part of power
supply
N/A
Sample examined according to IEC 60851-3:
1997, clause 5.1.1.4, followed by dielectric test
of clause 8.8.3, except test voltage applied
between wire and mandrel with no breakdown
N/A
Test voltage was at least the voltage in Tables
6 and 7but not less than the following:
N/A
– 1500 V for BASIC and SUPPLEMENTARY
INSULATION (V).................................................... :
N/A
– 3000 V for REINFORCED INSULATION (V) ............ :
N/A
Test Report No.: 1307033001
60601-1TRF_A
Verdict
Page 111 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
L.3.3
Requirement + Test
Result - Remark
Tension applied to wire during winding on
mandrel calculated from the wire diameter
equivalent to 118 MPa ± 11.8 MPa .................. :
N/A
Heat Shock
N/A
Sample subjected to heat shock test 9 of IEC
60851-6:1996, followed by dielectric strength
test of clause 8.8.3, except test voltage applied
between the wire and mandrel
L.3.4
Verdict
Evaluated in part of power
supply
N/A
Test voltage was at least the voltage in Tables
6 and 7, but not less than the following:
N/A
– 1500 V for BASIC and SUPPLEMENTARY
INSULATION (V).................................................... :
N/A
– 3000 V for REINFORCED INSULATION (V) ............ :
N/A
Oven temperature based on Table L.2 (C) ..... :
—
Mandrel diameter and tension applied as in
clause L.3.2, (MPa; N/mm2) .............................. :
N/A
Dielectric strength test conducted at room
temperature after removal from the oven
N/A
Retention of electric strength after bending
N/A
Five samples prepared as in L.3.2 subjected to
dielectric strength and bending tests
Evaluated in part of power
supply
N/A
Test voltage was at least the voltage in Tables
6 and 7, but not less than the following:
N/A
– 1500 V for BASIC and SUPPLEMENTARY
INSULATION (V).................................................... :
N/A
– 3000 V for REINFORCED INSULATION (V) ............ :
N/A
Test voltage applied between the shot and
conductor.
N/A
Mandrel diameter and tension applied as in
L.3.2, (MPa; N/mm2).......................................... :
N/A
L.4
Tests during manufacture
N/A
L.4.1
Production line dielectric strength tests
conducted by the manufacture according to
L.4.2 and L.4.3 ...................................................:
L.4.2
Test voltage for routine testing (100 % testing)
is at least the voltage in Tables 6 and 7 but not
less than the following:
N/A
– 1500 V r.m.s. or 2100 V peak for BASIC and
SUPPLEMENTARY INSULATION (V) ...........................:
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 112 of 203
Not required since all
related insulation will be
confirmed during power
supply manufacturing.
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
– 3000 V r.m.s. or 4200 V peak for REINFORCED
INSULATION (V) ....................................................:
L.4.3
Sampling tests conducted using twisted pair
samples (IEC 60851-5:1996, clause 4.4.1)
N/A
Not required since all
related insulation will be
confirmed during power
supply manufacturing.
N/A
Minimum breakdown test voltage at least twice
the voltage in Tables 6 and 7 but not less than:
N/A
– 3000 V r.m.s. or 4200 V peak for BASIC and
N/A
SUPPLEMENTARY INSULATION
– 6000 V r.m.s. or 8400 V peak for REINFORCED
N/A
INSULATION
Test Report No.: 1307033001
60601-1TRF_A
Page 113 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
4.2
RM RESULTS TABLE: Risk Management Process for ME Equipment or ME
Systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
3.3a

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Additional
documents
The Risk Management Procedure define the
policy for determining acceptable risk,
taking into account relevant International
Standards, and national or regional
regulations
Pass
3.5e

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Definition of
Risk Graph
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Criteria
The Risk Management Plan defined the
criteria for risk acceptability.
Pass

Pass
Verdict
4.1

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19),
Section - Release
Criteria
Risk Analysis procedure
Pass
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
Intended use and identification of
characteristics are well defined
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis
All hazards associated with the device have
been identified. The associated risk for some
of them would need to be re-assessed based
on the end system assembly and its
intended use.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures
The estimates of the risk(s) were recorded in
the risk management file
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 114 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
4.2
RM RESULTS TABLE: Risk Management Process for ME Equipment or ME
Systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file
Pass

Pass
Verdict
6.1

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Decision about
measures
The manufacturer follow the process
specified in 6.2 to 6.7 of ISO 14971 to control
the risk(s) so that the residual risk(s)
associated with each hazard is judged
acceptable
Pass
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures
The risk control measures selected were
recorded in the risk management file
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Evaluation
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan.
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 115 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
4.2
RM RESULTS TABLE: Risk Management Process for ME Equipment or ME
Systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.6

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Evaluation
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Evaluation of
Residual Risk
The risk control measures have been
reviewed to identify if other hazards are
introduced.
Pass

Pass
Verdict
6.7

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
The manufacturer assures that the risk(s)
from all identified hazards have been
evaluated.
Pass
7

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Evaluation of
Residual Risk
After all risk control measures have been
implemented and verified, the manufacturer
decide the overall residual risk posed by the
monitor is acceptable using the criteria
defined in the risk management plan.
Pass
4.3
TABLE: ESSENTIAL PERFORMANCE
List of ESSENTIAL
PERFORMANCE functions
N/A
MANUFACTURER’S document number
reference or reference from this standard
or collateral or particular standard(s)
Remarks
Supplementary Information:
ESSENTIAL PERFORMANCE is performance, the absence or degradation of which, would result in an
unacceptable risk.
4.3
RM RESULTS TABLE: Essential Performance
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
Test Report No.: 1307033001
60601-1TRF_A
Page 116 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
4.3
RM RESULTS TABLE: Essential Performance
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
5
4.5
RM RESULTS TABLE: Equivalent Safety for ME Equipment of ME System
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
4.6
RM RESULTS TABLE: ME Equipment or system parts contacting the patient
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
4.7
RM RESULTS TABLE: Single Fault Condition for ME Equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the unit in both normal and fault
conditions.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 117 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
4.7
RM RESULTS TABLE: Single Fault Condition for ME Equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.4

The estimates of the risk(s) were recorded in
the risk management file.
4.8
RM RESULTS TABLE: Components of ME Equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the use of the components under both
normal and single-fault conditions have
been evaluated.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures, F-01, H-01, C07, C-08
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures, F-01, H-01, C07, C-08
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures, F-01 (H-01, H02); F-02 (H-01, H-02); F03 (H-01); F-04 (H-01); F05 (H-01); F-06 (H-01), F07 (H-01, H-02, H-03); F08 (H-01); F-09 (H-01); F10 (H-01)
Test Report No.: 1307033001
60601-1TRF_A
Page 118 of 203
Pass
Verdict
Pass
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
4.8
RM RESULTS TABLE: Components of ME Equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures,
RC-001, RC-002, RC031, RC-033
The risk control measures selected were
recorded in the risk management file
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures,
RC-001, RC-002, RC031, RC-033
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures, F-01, H-01, C07, C-08
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan.
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 119 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
4.9
RM RESULTS TABLE: Use of components with high-integrity characteristics
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
4.11
TABLE: Power Input
Pass
Voltage
(V)
Frequency
(Hz)
Current
(A)
Power
(W or VA)
Power factor
(cos φ)
Maximum normal load / --
90
50
1.38
64
--
Maximum normal load / --
90
60
1.40
66
--
Maximum normal load / 1.8A
100
50
1.27
64
--
Maximum normal load / 1.8A
100
60
1.25
64
--
Maximum normal load / 0.8A
240
50
0.64
63
--
Maximum normal load / 0.8A
240
60
0.61
62
--
Maximum normal load / --
264
50
0.59
63
--
Maximum normal load / --
264
60
0.58
63
--
24Vdc
--
2.63
63
--
Operating Conditions / Ratings
Maximum normal load / 2.91A
Supplementary Information: Tests were conducted at TFT LCD with 1280 x 1024 resolution, support 2.5”
Hard Disk Drive storage function and one mini card slot for WLAN expansion, feature flexible I/O ports,
such as six USB2.0 each USB load 0.5A (6 USB ports total load 15W), one RS-232, one RS-232/422/485
and two Gigabit LANs.
5.1
RM RESULTS TABLE: Type Tests
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

The RMF mention the “Intended
use/intended purpose”
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
Test Report No.: 1307033001
60601-1TRF_A
Page 120 of 203
Pass
Verdict
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
5.1
RM RESULTS TABLE: Type Tests
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the risk analysis of unit type.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5.4 a)
RM RESULTS TABLE: Other Conditions
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis
The manufacturer had compiled a list of
known or foreseeable hazards situation have
been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures
The estimates of the risk(s) were recorded in
the risk management file
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 121 of 203
Pass
Verdict
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
5.7
RM RESULTS TABLE: Humidity preconditioning treatment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis,
F-01, H-01, C-09
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the humidity preconditioning treatment.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures, F-01, H-01, C09
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C09
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures,
RC-024
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures,
RC-024
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 122 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
5.7
RM RESULTS TABLE: Humidity preconditioning treatment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.4

All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
5.9.2
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures, F-01, H-01, C09
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
N/A
Pass
Verdict
TABLE: Determination of ACCESSIBLE parts
Pass
Location
Determination method (NOTE1)
Comments
Output connector
rigid test finger
The plastic / metal enclosure serves as
mechanical / electrical / fire enclosure.
Enclosure
rigid test finger
The plastic / metal enclosure serves as
mechanical / electrical / fire enclosure
Supplementary information:
NOTE 1 - The determination methods are: visual; rigid test finger; jointed test finger; test hook.
5.9.2.3
RM RESULTS TABLE: Actuating mechanisms
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
Test Report No.: 1307033001
60601-1TRF_A
Page 123 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
7.1.2
TABLE: Legibility of Marking
Result - Remark
Verdict
Pass
Markings tested
Ambient illuminance
(lx)
Outside Markings (Clause 7.2) ............... :
1500 (lx) and 100 (lx)
Remarks
Clearly readable
Inside Markings (Clause 7.3) .................. :
Controls & Instruments (Clause 7.4) ...... :
Safety Signs (Clause 7.5) ........................ :
Symbols (Clause 7.6) .............................. :
Supplementary information:
Observer, with a visual acuity of 0 on the log Minimum Angle of Resolution (log MAR) scale or 6/6 (20/20), reads marking
at ambient illuminance least favourable level in the range of 100 lx to 1,500 lx. The ME EQUIPMENT or its part was positioned
so that the viewpoint was the intended position of the OPERATOR at any point within the base of a cone subtended by an
to the axis normal to the centre of the plane of the marking and at a distance of 1 m.
7.1.3
TABLE: Durability of marking test
Pass
Characteristics of the Marking Label tested:
Remarks
Material of Marking Label .................................... :
Car Tong Co type CT-M002
--
Ink/other printing material or process.................. :
Ink
--
Material (composition) of Warning Label ............ :
--
--
Ink/other printing material or process.................. :
--
--
Other..................................................................... :
--
--
Supplementary information:
Marking rubbed by hand, first for 15 s with a cloth rag soaked with distilled water, then for 15 s with a cloth
rag soaked with methylated spirit, and then for 15 s with a cloth rag soaked with isopropyl alcohol.
7.2.2
RM RESULTS TABLE: Identification
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-07, H-01, H-02, H-03
The manufacturer had compiled a list of
known or foreseeable hazards about
identification of equipment.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 124 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.2.2
RM RESULTS TABLE: Identification
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-07, H-01, H02, H-03
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-07, H-01, H02, H-03
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-07, H-01, H02, H-03
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan.
Pass

6.4



Pass
Verdict
7.2.5
RM RESULTS TABLE: ME EQUIPMENT powered from other equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
Test Report No.: 1307033001
60601-1TRF_A
Page 125 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.2.5
RM RESULTS TABLE: ME EQUIPMENT powered from other equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
6.4
7.2.13
RM RESULTS TABLE: Physiological effects (safety signs and warning)
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.3
7.2.17
RM RESULTS TABLE: Protective packaging
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Risk Analysis:
F-01, H-01, C-20
The manufacturer had compiled a list of
known or foreseeable hazards associated
with packing conditions.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Risk Evaluation
Before Measures: F-01,
H-01, C-20
The estimates of the risk(s) were recorded
in the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Risk Evaluation
Before Measures: F-01,
H-01, C-20
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Test Report No.: 1307033001
60601-1TRF_A
Page 126 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.2.17
RM RESULTS TABLE: Protective packaging
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Release
Protocols of Measures:
RC-029
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -New evaluation
after measures: F-01, H01, C-20
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of the
Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan.
Pass


Pass
Verdict
7.3.3
RM RESULTS TABLE: Batteries
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Risk Analysis
The manufacturer had compiled a list of
known or foreseeable hazards associated
with information of battery of the device
have been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Risk Evaluation
Before Measures, F-10,
H-01
The estimates of the risk(s) were recorded
in the risk management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 127 of 203
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
7.3.3
RM RESULTS TABLE: Batteries
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Risk Evaluation
Before Measures: F-10,
H-01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Release
Protocols of Measures:
RC-031, RC-033
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass

Pass
6.3

7.3.7
RM RESULTS TABLE: Supply terminals
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
Verdict
N/A
Verdict
4.3
7.4.2
RM RESULTS TABLE: Control devices
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
7.5
RM RESULTS TABLE: Safety signs
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
Test Report No.: 1307033001
60601-1TRF_A
Page 128 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
7.5
RM RESULTS TABLE: Safety signs
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
6.3
7.9.1
RM RESULTS TABLE: General accompanying documents (See Table C.4)
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-07, H-01, H-02, H-03
The manufacturer had compiled a list of
known and foreseeable hazards, associated
with the content and format of the
accompanying documents have been
evaluated.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-07, H-01, H02, H-03
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-07, H-01, H02, H-03
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-011
The results of this risk evaluation were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-011
The results of this risk evaluation were
recorded in the risk management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 129 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
7.9.2.4
RM RESULTS TABLE: Electrical power source
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.3
7.9.3.2
RM RESULTS TABLE: Replacement of fuses, power supply cords, other
parts
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
8.1 b(1)
RM RESULTS TABLE: Fundamental rule of protection against electric shock
- interruption of any one power-carrying conductor
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01
The manufacturer had compiled a list of
known and foreseeable hazards associated
with interruption of any one supply
conductor have been considered.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01
The estimates of the risk(s) were recorded in
the risk management file.
Pass
8.1 b(2)
RM RESULTS TABLE: Fundamental rule of protection against electric shock
- unintended movement of a component
Pass
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Test Report No.: 1307033001
60601-1TRF_A
Result - Remarks
Page 130 of 203
Pass
Verdict
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.1 b(2)
RM RESULTS TABLE: Fundamental rule of protection against electric shock
- unintended movement of a component
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01
The manufacturer had compiled a list of
known and foreseeable hazards associated
with unintended movement of components
have been considered.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures,
RC-002
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures,
RC-002
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 131 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.1 b(2)
RM RESULTS TABLE: Fundamental rule of protection against electric shock
- unintended movement of a component
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.4

All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures, F-01, H-01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
Pass
Verdict
6.5
N/A
8.1 b(3)
RM RESULTS TABLE: Fundamental rule of protection against electric shock
- accidental detachment of conductors and connectors
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

The manufacturer had compiled a list of
known and foreseeable hazards associated
with accidental detachment of conductors
and connectors have been considered.
8.2.2
RM RESULTS TABLE: Connection to an external d.c. power sources
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-08, H-01
The manufacturer had compiled a list of
known and foreseeable hazards associated
with connection to an external d.c. power
sources have been considered.
Pass
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01
Test Report No.: 1307033001
60601-1TRF_A
Page 132 of 203
Pass
Verdict
Pass
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.2.2
RM RESULTS TABLE: Connection to an external d.c. power sources
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-08, H-01
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-08, H-01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Verdict
8.3 d
RM RESULTS TABLE: Requirements of Type BF or CF Applied Parts
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
6.2
Pass
Result - Remarks
N/A
Verdict
See also RM Table for 4.6
8.4.2
TABLE: TABLE: Working Voltage / Power Measurement
N/A
1
Test supply voltage/frequency (V/Hz) ........................................................... :
Location
From/To
Vrms
Vpk or
Vdc
Measured values
Peak-toPower
peak ripple2
W/VA
Energy
(J)
Remarks
Supplementary Information:
1. The input supply voltage to the ME EQUIPMENT shall be the RATED voltage or the voltage within the RATED voltage
range which results in the highest measured value. See clause 8.5.4.
2. If the d.c peak-to-peak ripple >10%, waveform considered as a.c. See clause 8.4.2.2
The adapter Eljintek Inc. type GMPU70A-6 was certified by IEC 60601-1: 2005 and EN 60601-1:
2006
8.4.2 c
RM RESULTS TABLE: Accessible parts including applied parts
Test Report No.: 1307033001
60601-1TRF_A
Page 133 of 203
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-04, H-01
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the accessible parts have been
identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-04, H-01
The estimates of the risk(s) were recorded in
the risk management file.
Pass
8.4.3
Verdict
TABLE: ME EQUIPMENT for connection to a power source by a plug
- measurement of voltage or calculation of stored charge 1 s after
disconnection of plug from mains supply
N/A
Maximum allowable voltage (V) .....................................................................................: 60
Voltage measured (V)
Voltage Measured Between:
1
2
3
4
5
6
7
8
9
10
9
10
Plug pins 1 and 2
Plug pin 1 and plug earth pin
Plug pin 2 and plug earth pin
Plug pin 1 and enclosure
Plug pin 2 and enclosure
Maximum allowable stored charge when measured voltage exceeded 60 v (c) .......: 45
Calculated stored charge (c)
Voltage Measured Between:
1
2
3
4
5
6
7
8
Plug pins 1 and 2
Plug pin 1 and plug earth pin
Plug pin 2 and plug earth pin
Plug pin 1 and enclosure
Plug pin 2 and enclosure
Test Report No.: 1307033001
60601-1TRF_A
Page 134 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Supplementary information:
The adapter Eljintek Inc. type GMPU70A-6 was certified by IEC 60601-1: 2005 and EN 60601-1: 2006
8.4.4
TABLE: Internal capacitive circuits – measurement of residual voltage or
calculation of the stored charge in capacitive circuits (i.e., accessible
capacitors or circuit parts) after de-energizing ME EQUIPMENT
Maximum allowable residual voltage (V):
60 V
Maximum allowable stored charge when residual voltage exceeded 60 V :
45 C
Description of the capacitive circuit Measured residual
(i.e., accessible capacitor or circuit
voltage (V)
parts)
Calculated stored
charge (C)
N/A
Remarks
Supplementary information:
8.5.2.2
RM RESULTS TABLE: Type B applied parts
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
8.5.2.3
RM RESULTS TABLE: PATIENT Leads
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
8.5.5.1a
Test
Condition:
Figs. 9 & 10
TABLE: defibrillation-proof applied parts – measurement of hazardous
electrical energies
Measurement
made on
accessible part
Test Report No.: 1307033001
60601-1TRF_A
Applied part with
test voltage
Test voltage
Measured
voltage between
polarity
Y1 and Y2 (mV)
Page 135 of 203
N/A
Remarks
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Supplementary information:
8.5.5.1b
TABLE: defibrillation-proof applied parts – verification of recovery time
Applied part with test
voltage
Test voltage Recovery time from
Measured
polarity
documents (s)
recovery time (s)
N/A
Remarks
Supplementary information:
8.5.5.2
TABLE: DEFIBRILLATION-PROOF APPLIED PARTS or PATIENT CONNECTIONS of
DEFIBRILLATION- PROOF APPLIED PARTS - Energy reduction test –measurement of
Energy delivered to a 100  load
Measured Energy Measured Energy
E1 (mJ)
E2 (mJ)
Test Voltage applied to
N/A
Energy E1
as % of E2 (%)
PATIENT CONNECTION 1 or APPLIED PART with
PATIENT CONNECTIONS 2, 3, and 4 of the same
APPLIED PART connected to earth
PATIENT CONNECTION 2 or APPLIED PART with
PATIENT CONNECTIONS 1, 3, and 4 of the same
APPLIED PART connected to earth
PATIENT CONNECTION 3 or APPLIED PART with
PATIENT CONNECTIONS 1, 2, and 4 of the same
APPLIED PART connected to earth
PATIENT CONNECTION 4 or APPLIED PART with
PATIENT CONNECTIONS 1, 2, and 3 of the same
APPLIED PART connected to earth
Supplementary information: For compliance: E1 must at least 90% of E2
E1= Measured energy delivered to 100  with ME Equipment connected;
E2= Measured energy delivered to 100  without ME equipment connected.
8.6.3
RM RESULTS TABLE: Protective earthing of moving parts
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
Test Report No.: 1307033001
60601-1TRF_A
Page 136 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
8.6.3
RM RESULTS TABLE: Protective earthing of moving parts
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
6.3
6.4
6.5
8.6.4
TABLE: Impedance and current-carrying capability of PROTECTIVE EARTH
N/A
CONNECTIONS
Type of ME EQUIPMENT & impedance
measured between parts
Test current Voltage drop
(A)
measured
between
/Duration (s)
parts (V)
Maximum
calculated
impedance
(m)
Maximum
allowable
impedance
(m)
PERMANENTLY INSTALLED ME EQUIPMENT,
impedance between PROTECTIVE EARTH
TERMINAL and a PROTECTIVELY EARTHED part
100
ME EQUIPMENT with an APPLIANCE INLET,
impedance between earth pin in the
APPLIANCE INLET and a PROTECTIVELY EARTHED
part
100
ME EQUIPMENT with a non-DETACHABLE POWER
SUPPLY CORD, impedance between the
protective earth pin in the MAINS PLUG and a
PROTECTIVELY EARTHED part
200
Supplementary information:
The adapter Eljintek Inc. type GMPU70A-6 was certified by IEC 60601-1: 2005 and EN 60601-: 2006
Test Report No.: 1307033001
60601-1TRF_A
Page 137 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.7
Requirement + Test
Result - Remark
Verdict
TABLE: leakage current
Type of leakage current and test
condition (including single faults)
Pass
Supply
voltage
(V)
Supply
Measured
frequency max. value
(Hz)
(µA)
Remarks
Fig. 13 - Earth Leakage (ER)
—
—
—
Maximum allowed values:
5 mA NC; 10 mA SFC
Before Humidity Condition:
—
—
—
—
ER, NC
264
60
229
S1 = 1, S5 = 1
ER, NC
264
60
220
S1 = 1, S5 = 0
ER, SFC (Neutral Open)
264
60
304
S1 = 0, S5 = 1
ER, SFC (Neutral Open)
264
60
305
S1 = 0, S5 = 0
—
—
—
ER, NC
264
60
231
S1 = 1, S5 = 1
ER, NC
264
60
222
S1 = 1, S5 = 0
ER, SFC (Neutral Open)
264
60
306
S1 = 0, S5 = 1
ER, SFC (Neutral Open)
264
60
306
S1 = 0, S5 = 0
Measured with a non-frequencyweighted device
—
—
—
Maximum allowed values:
10 mA NC; 10 mA SFC
Before Humidity Condition:
—
—
—
—
ER, NC
264
60
242
S1 = 1, S5 = 1
ER, NC
264
60
240
S1 = 1, S5 = 0
ER, SFC (Neutral Open)
264
60
347
S1 = 0, S5 = 1
ER, SFC (Neutral Open)
264
60
348
S1 = 0, S5 = 0
—
—
—
ER, NC
264
60
244
S1 = 1, S5 = 1
ER, NC
264
60
242
S1 = 1, S5 = 0
ER, SFC (Neutral Open)
264
60
349
S1 = 0, S5 = 1
ER, SFC (Neutral Open)
264
60
350
S1 = 0, S5 = 0
Fig. 14 - Touch Current (TC)
—
—
—
Maximum allowed values:
100 uA NC; 500 uA SFC
Before Humidity Condition:
—
—
—
TC, NC
264
60
46
MD1: between earth and
plastic enclosure
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
46
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
68
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
68
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
76
S1 = 1, S5 = 1, S7 = 0
After Humidity Condition:
After Humidity Condition:
Test Report No.: 1307033001
60601-1TRF_A
Page 138 of 203
—
—
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
TC, SFC (Ground Open)
Result - Remark
Verdict
264
60
75
—
—
—
TC, NC
264
60
48
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
48
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
70
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
70
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
77
S1 = 1, S5 = 1, S7 = 0
TC, SFC (Ground Open)
264
60
76
—
—
—
S1 = 1, S5 = 0, S7 = 0
MD2: between plastic
enclosure with foil and metal
enclosure
TC, NC
264
60
4
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
4
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
6
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
6
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
7
S1 = 1, S5 = 1, S7 = 0
TC, SFC (Ground Open)
264
60
7
—
—
—
S1 = 1, S5 = 0, S7 = 0
MD2: between plastic
enclosure with foil and metal
enclosure
TC, NC
264
60
6
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
6
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
8
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
8
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
9
S1 = 1, S5 = 1, S7 = 0
TC, SFC (Ground Open)
264
60
9
S1 = 1, S5 = 0, S7 = 0
Measured with a non-frequencyweighted device
—
—
—
Maximum allowed values:
10 mA NC; 10 mA SFC
Before Humidity Condition:
—
—
—
TC, NC
264
60
175
MD1: between earth and
plastic enclosure
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
170
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
210
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
211
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
235
S1 = 1, S5 = 1, S7 = 0
TC, SFC (Ground Open)
264
60
230
S1 = 1, S5 = 0, S7 = 0
After Humidity Condition:
Before Humidity Condition:
After Humidity Condition:
Test Report No.: 1307033001
60601-1TRF_A
Page 139 of 203
S1 = 1, S5 = 0, S7 = 0
MD1: between earth and
plastic enclosure
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
MD1: between earth and
plastic enclosure
—
—
—
TC, NC
264
60
177
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
171
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
211
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
213
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
236
S1 = 1, S5 = 1, S7 = 0
TC, SFC (Ground Open)
264
60
232
—
—
—
S1 = 1, S5 = 0, S7 = 0
MD2: between plastic
enclosure with foil and metal
enclosure
TC, NC
264
60
12
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
12
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
17
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
17
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
18
S1 = 1, S5 = 1, S7 = 0
TC, SFC (Ground Open)
264
60
18
—
—
—
S1 = 1, S5 = 0, S7 = 0
MD2: between plastic
enclosure with foil and metal
enclosure
TC, NC
264
60
14
S1 = 1, S5 = 1, S7 = 1
TC, NC
264
60
14
S1 = 1, S5 = 0, S7 = 1
TC, SFC (Neutral Open)
264
60
19
S1 = 0, S5 = 1, S7 = 1
TC, SFC (Neutral Open)
264
60
19
S1 = 0, S5 = 0, S7 = 1
TC, SFC (Ground Open)
264
60
20
S1 = 1, S5 = 1, S7 = 0
TC, SFC (Ground Open)
264
60
20
S1 = 1, S5 = 0, S7 = 0
Fig. 15 - Patient Leakage Current (P)
—
—
—
Maximum allowed values:
Type B AP:
100 uA NC; 500 uA SFC (a.c.)
Before Humidity Condition:
—
—
—
—
P, NC
264
60
3
S1 = 1, S5 = 1, S7 = 1
P, NC
264
60
3
S1 = 1, S5 = 0, S7 = 1
P, SFC (Neutral Open)
264
60
4
S1 = 0, S5 = 1, S7 = 1
P, SFC (Neutral Open)
264
60
4
S1 = 0, S5 = 0, S7 = 1
P, SFC (Ground Open)
264
60
5
S1 = 1, S5 = 1, S7 = 0
P, SFC (Ground Open)
264
60
5
S1 = 1, S5 = 0, S7 = 0
—
—
—
—
264
60
5
S1 = 1, S5 = 1, S7 = 1
After Humidity Condition:
Before Humidity Condition:
After Humidity Condition:
After Humidity Condition:
P, NC
Test Report No.: 1307033001
60601-1TRF_A
Page 140 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
P, NC
264
60
5
S1 = 1, S5 = 0, S7 = 1
P, SFC (Neutral Open)
264
60
6
S1 = 0, S5 = 1, S7 = 1
P, SFC (Neutral Open)
264
60
6
S1 = 0, S5 = 0, S7 = 1
P, SFC (Ground Open)
264
60
7
S1 = 1, S5 = 1, S7 = 0
P, SFC (Ground Open)
264
60
7
S1 = 1, S5 = 0, S7 = 0
Measured with a non-frequencyweighted device
—
—
—
Maximum allowed values:
10 mA NC; 10 mA SFC
Before Humidity Condition:
—
—
—
—
P, NC
264
60
10
S1 = 1, S5 = 1, S7 = 1
P, NC
264
60
10
S1 = 1, S5 = 0, S7 = 1
P, SFC (Neutral Open)
264
60
13
S1 = 0, S5 = 1, S7 = 1
P, SFC (Neutral Open)
264
60
13
S1 = 0, S5 = 0, S7 = 1
P, SFC (Ground Open)
264
60
18
S1 = 1, S5 = 1, S7 = 0
P, SFC (Ground Open)
264
60
18
S1 = 1, S5 = 0, S7 = 0
—
—
—
—
P, NC
264
60
12
S1 = 1, S5 = 1, S7 = 1
P, NC
264
60
12
S1 = 1, S5 = 0, S7 = 1
P, SFC (Neutral Open)
264
60
14
S1 = 0, S5 = 1, S7 = 1
P, SFC (Neutral Open)
264
60
15
S1 = 0, S5 = 0, S7 = 1
P, SFC (Ground Open)
264
60
19
S1 = 1, S5 = 1, S7 = 0
P, SFC (Ground Open)
264
60
19
S1 = 1, S5 = 0, S7 = 0
—
Maximum allowed values:
Type B: N/A
Type BF AP: 5000 uA
Type CF AP: 50 uA
After Humidity Condition:
Fig. 16 - Patient leakage current with
mains on the F-type applied parts (PM)
—
—
Fig. 17 - Patient leakage current with
external voltage on Signal Input/Output
part (SIP/SOP)
—
—
—
Maximum allowed values:
Type B or BF AP: 10 uA NC; 50 uA
SFC(d.c. current);
100 uA NC; 500 uA SFC (a.c.) ;
Type CF AP: 10 uA NC; 50 uA SFC
(d.c. or a.c. current)
Fig. 18 - Patient leakage current with
external voltage on metal Accessible
Part that is not Protectively Earthed
—
—
—
Maximum allowed values:
Type B or BF AP: 500 uA
Type CF: N/A
Test Report No.: 1307033001
60601-1TRF_A
Page 141 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Fig. 19 – Patient Auxiliary Current
Fig. 15 and 20 – Total Patient Leakage
Current with all AP of same type
connected together
Fig. 17 and 20 – Total Patient Leakage
Current with all AP of same type
connected together with external
voltage on SIP/SOP
Result - Remark
—
—
—
—
—
—
Verdict
—
Maximum allowed values:
Type B or BF AP: 10 uA NC; 50 uA
SFC (d.c. current);
100 uA NC; 500 uA SFC (a.c.) ;
Type CF AP: 10 uA NC;50 uA SFC
(d.c. or a.c. current)
—
Maximum allowed values:
Type B or BF AP: 50 uA NC;
100uA SFC (d.c. current);
500 uA NC; 1000 uA SFC (a.c.);
Type CF AP: 50 uA NC; 100 uA
SFC (d.c. or a.c. current)
—
Maximum allowed values:
Type B or BF AP: 50 uA NC;
100uA SFC (d.c. current);
500 uA NC;1000 uA SFC (a.c.);
Type CF AP: 50 uA NC; 100 uA
SFC (d.c. or a.c. current)
Fig. 16 and 20 – Total Patient Leakage
Current with all AP of same type
connected together with external
voltage on F-type AP
—
—
—
Maximum allowed values:
Type B: NA
Type BF: 5000uA
Type CF: 100 uA
Fig. 18 and 20 – Total Patient Leakage
Current with all AP of same type connected
together with external voltage on metal
Accessible Part not Protectively Earthed
—
—
—
Maximum allowed values:
Type B & BF: 1000 uA
Type CF: N/A
Supplementary information:
Note 1: For EARTH LEAKAGE CURRENT see 8.7.3 d) and 8.7.4.5;
Note 2: For TOUCH CURRENT see 8.7.3 c) and 8.7.4.6;
Note 3: For PATIENT LEAKAGE CURRENT SEE 8.7.3.b) and 8.7.4.7
Note 4: Total PATIENT LEAKAGE CURRENT values are only relative to equipment with multiple APPLIED PARTS of the same type.
See 8.7.4.7 h). The individual APPLIED PARTS complied with the PATIENT LEAKAGE CURRENT values.
Note 5: In addition to conditions indicated in the Table, tests conducted at operating temperature and after humidity
preconditioning of 5.7, EQUIPMENT energized in stand-by condition and fully operating, max rated supply frequency, at 110 %
of the max RATED MAINS VOLTAGE, and after relevant tests of Clause 11.6 (i.e., overflow, spillage, leakage, ingress of water
and particulate matter, cleaning & disinfection, & sterilization).
Test Report No.: 1307033001
60601-1TRF_A
Page 142 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
ER - Earth leakage current
TC – Touch current
P - Patient leakage current
PA – Patient auxiliary current
TP – Total Patient current
PM - Patient leakage current with mains on the applied parts
MD - Measuring device
8.8.3
Verdict
A - After humidity conditioning
B - Before humidity conditioning
1 - Switch closed or set to normal polarity
0 - Switch open or set to reversed polarity
NC - Normal condition
SFC - Single fault condition
TABLE: Dielectric strength test of solid insulating materials with safety
function – MEANS OF OPERATOR PROTECTION (MOOP) / MEANS OF PATIENT PROTECTION
(MOPP)
Insulation under test
(area from insulation
diagram)
Insulation Type
(1 or 2
MOOP/MOPP)
Pass
Reference Voltage
PEAK WORKING
VOLTAGE (U)
V peak
Dielectric
A.C. test
breakdown
PEAK WORKING voltages in V
after 1 minute
VOLTAGE (U)
r.m.s1
Yes/No2
V d.c.
Adapter primary to
Adapter earth (4)
1MOPP
354
--
1500Vac
No
Adapter primary to
monitor plastic
enclosure with foil (5)
2MOPP
536
--
4517Vac
No
Adapter primary to
Touch Panel with foil
(5)
2MOPP
536
--
4517Vac
No
Adapter primary to
monitor SIP/SOP (6)
2MOPP
536
--
4517Vac
No
SIP/SOP to Patient
connection (7)
1 MOPP
354
--
1500Vac
No
Adapter primary to
Patient connection (8)
2 MOPP
536
--
4517Vac
No
Supplementary information:
Test reference voltage determined by power adapter type:
Eljintek type GMPU70A-x, Maximum working voltage 536Vpk
1
Alternatively, per the Table (i.e., __dc), a d.c. test voltage equal to the peak value of the a.c. test voltage used.
A) Immediately after humidity treatment of 5.7, ME EQUIPMENT de-energized, B) after required sterilization PROCEDURE,
ME EQUIPMENT de-energized, C) after reaching steady state operating temperature as during heating test of 11.1.1, and D)
after relevant tests of 11.6 (i.e., overflow, spillage, leakage, ingress of water, cleaning, disinfection, and sterilization).
2
8.8.4.1
TABLE: Resistance to heat - Ball pressure test of thermoplastic parts
Allowed impression diameter (mm) ............... :  2 mm

Force (N) .......................................................... : 20

Part/material
Plastic Enclosure, Sabic type C2800
Test Report No.: 1307033001
60601-1TRF_A
Pass
Page 143 of 203
Test
temperature (C)
Impression
diameter (mm)
76
0.8
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Supplementary information: -8.8.4.1
RM RESULTS TABLE: Mechanical strength and resistance to heat
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01, C-17
The manufacturer had compiled a list of
known and foreseeable hazards under both
normal and single-fault conditions,
associated with the insulating materials,
have been evaluated.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C17
The estimates of the risk(s) were recorded in
the risk management file RMF.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C17
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-004
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-004
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 144 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.8.4.1
RM RESULTS TABLE: Mechanical strength and resistance to heat
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.4

All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
8.9.2
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-01, C17
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
N/A
Verdict
TABLE: Short circuiting of each single one of the CREEPAGE DISTANCES and AIR
CLEARANCES for insulation in the MAINS PART between parts of opposite
polarity in lieu of complying with the required measurements in 8.9.4
Specific areas of circuits shortcircuited and test conditions
Supplementary information:
Note 1:
AC - AIR CLEARANCE
Test Report No.: 1307033001
60601-1TRF_A
Pass
Test in lieu of
HAZARDOUS SITUATION
CREEPAGE
DISTANCE or AIR
1
CLEARANCE
observed (i.e., fire hazard,
shock hazard, explosion,
discharge of parts, etc.)?
Yes/No
N/A
Remarks
CD - CREEPAGE DISTANCE
Page 145 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.9.3.2
Test
Sequence
No.
Requirement + Test
Verdict
Table: Thermal cycling tests on one sample of insulating compound forming
solid insulation between conductive parts
Each test duration and
temperature
68 h at T1 ± 2 °C = ___°C
1
Result - Remark
Dielectric test voltage
(V = Test voltage in 8.8.3
times 1.6)
N/A
Dielectric strength test
after humidity
preconditioning per cl. 5.7
except for 48 h only,
Breakdown: Yes/No
1
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
68 h at T1 ± 2 °C = ___°C 1
2
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
68 h at T1 ± 2 °C = ___°C 1
3
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
68 h at T1 ± 2 °C = ___°C 1
4
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
Supplementary information:
1
T1 = 10 °C above the maximum temperature of relevant part determined per 11.1.1, or 85 °C, the higher of
the two. 10 °C not added to T1 when temperature measured by an embedded thermocouple. Used gradual
transition from one temperature to another.
Test Report No.: 1307033001
60601-1TRF_A
Page 146 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.9.3.4
Test
Sequence
No.
Requirement + Test
Result - Remark
Verdict
Table: Thermal cycling tests on one sample of cemented joint (see 8.9.3.3)
Each test duration and
temperature
Dielectric test voltage
(V = Test voltage in 8.8.3
times 1.6)
N/A
Dielectric strength test
after humidity
preconditioning per cl. 5.7
except for 48 h only,
Breakdown: Yes/No
68 h at T1 ± 2 °C = ___°C 1
1
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
68 h at T1 ± 2 °C = ___°C
2
1
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
68 h at T1 ± 2 °C = ___°C 1
3
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
68 h at T1 ± 2 °C = ___°C 1
4
1 h at 25 °C ± 2 °C
2 h at 0 °C ± 2 °C
1 or more h at 25 °C ± 2 °C
Supplementary information:
1
T1 = 10 °C above the maximum temperature of relevant part determined per 11.1.1, or 85 °C, the higher of
the two. 10 °C not added to T1 when temperature measured by an embedded thermocouple. Used gradual
transition from one temperature to another.
Test Report No.: 1307033001
60601-1TRF_A
Page 147 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
8.10
Requirement + Test
Result - Remark
Verdict
TABLE: List of critical components
Component/
Part No.
Manufacturer/
Trademark
Type No./model
No./
Technical data
Standard No./,
Edition
I/P: 100240Vac,
50/60Hz, 1.80.8A;
O/P: 21-27Vdc,
3.34-2.60A 70W
max.
ANSI/AAMI
ES60601-1:
2005,
CSA C22.2 NO.
60601-1: 2008,
IEC 60601-1:
2005
Mark(s) &
Certificates of
1
conformity
Power
Adapter
Eljintek Inc.
Plastic
Enclosure
Sabic Innovative C2800 (GG)
Plastics US L L
C
Min. V-0, 70°C, UL 94
3.0mm
thickness
minimum, see
AttachmentTechnical
Documentation
for details.
UL
Bottom side
metal
bracket
--
--
Metal, overall
-see
Attachment Technical
Documentation
for details.
Minimum
1.0mm
thickness.
--
- Bottom side -opening
--
Provided circle -shape opening,
each opening
approx.
measured
diameter
1.9mm.
--
Metal
Enclosure
--
--
Metal, overall
-see
AttachmentTechnical
Documentation
for details.
--
Internal Rear -Bracket
--
Metal, overall
-see
AttachmentTechnical
Documentation
for details.
--
Heatsink for
--
Metal, overall
--
--
Test Report No.: 1307033001
60601-1TRF_A
GMPU70A-6
Pass
Page 148 of 203
--
UL, CSA
TUV/Rh
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
CPU
Verdict
see
AttachmentTechnical
Documentation
for details.
Hard Disk
Drive
(Optional)
Various
Various
5Vdc, 1.1A
max.
UL 60950-1,
EN 60950-1
UL, TUV, VDE
Chock (L35)
ZenithTek Inc.
ZPWM-Series
125°C
--
--
RTC Battery
(BT2)
Mitsubishi
Electric Corp
CR2032
Maximum
UL 1642
abnormal
charge current
10 mA.
UL
- alternate
Hitachi Maxell
Energy Ltd
CR2032
Maximum
UL 1642
abnormal
charge current
10 mA.
UL
- alternate
Sony Energy
Devices Corp
CR2032
Maximum
UL 1642
abnormal
charge current
10 mA.
UL
- alternate
Panasonic
Corporation,
Panasonic
Corporation of
North America
CR2032
Maximum
UL 1642
abnormal
charge current
10 mA.
UL
- alternate
Mitsubishi
Electric Home
Appliance Co
Ltd
CR2032
Maximum
UL 1642
abnormal
charge current
10 mA.
UL
Printed
Various
Wiring Board
Various
Minimum V-1,
105°C
UL
LCD Panel
AU Optronics
Corp
M215HTN01.1
21.5 inch, TFT - UL 60950-1
LCD module
with LED
Backlight
UL
Internal
Plastic Part
Various
Various
Minimum V-2
UL 94
UL
--
Each 3 W, 4
ohm
--
--
Speaker (two -provided)
UL 796
Supplementary information:
1) An asterisk indicates a mark which assures the agreed level of surveillance. See Licenses and
Certificates of Conformity for verification.
8.10.1
RM RESULTS TABLE: Fixing of components
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Test Report No.: 1307033001
60601-1TRF_A
Result - Remarks
Page 149 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.10.1
RM RESULTS TABLE: Fixing of components
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01, C-11
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the forces to which the device would be
subject to unwanted movement of
components used has been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C11
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C11
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-019
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-019
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 150 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.10.1
RM RESULTS TABLE: Fixing of components
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.4

All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-01, C11
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
Pass
Verdict
6.5
N/A
8.10.2
RM RESULTS TABLE: Fixing of wiring
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01, C-18
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the forces to which the device would be
subject to unwanted movement of wiring
has been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C18
The estimates of the risk(s) were recorded in
the risk management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 151 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.10.2
RM RESULTS TABLE: Fixing of wiring
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C18
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-026
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-026
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-01,
C-18
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
N/A
8.10.5
RM RESULTS TABLE: Mechanical protection of wiring
Test Report No.: 1307033001
60601-1TRF_A
Page 152 of 203
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01, C-21, C-22
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the forces to which the device would be
subject to unwanted movement of wiring has
been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C21, C-22
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C21, C-22
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-008, RC-026
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-008, RC-026
The effectiveness of the risk control
measures had been verified and the results
of the verification were recorded in the risk
management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 153 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
8.10.5
RM RESULTS TABLE: Mechanical protection of wiring
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.4

All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
8.11.3.5
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-01, C21, C-22
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
N/A
Pass
Verdict
TABLE: Cord anchorages
Cord under test
N/A
Pull (N)
Mass of
equipment (kg)
Torque Nm)
Remarks
Supplementary information:
8.11.3.6
TABLE: Cord guard
Cord under test
N/A
Test mass
Measured curvature
Remarks
Supplementary information:
8.11.5
RM RESULTS TABLE: Mains fuses and over-current releases
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.3
4.4
5
Test Report No.: 1307033001
60601-1TRF_A
Page 154 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
8.11.5
RM RESULTS TABLE: Mains fuses and over-current releases
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
6.2
6.3
6.4
6.5
9.2.1
RM RESULTS TABLE: HAZARDS associated with moving parts - General
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.2.2.2
TABLE: Measurement of gap “a” according to Table 20 (ISO 13852: 1996)
Part of body
Allowable adult
gap1, mm
Measured adult
gap, mm
Allowable children
gap1, mm
Body
> 500
> 500
Head
> 300 or < 120
> 300 or < 60
Leg
> 180
> 180
Foot
> 120 or < 35
> 120 or < 25
Toes
> 50
> 50
Arm
> 120
> 120
Hand, wrist, fist
> 100
> 100
Finger
> 25 or < 8
N/A
Measured children
gap, mm
> 25 or < 4
1
Supplementary information: In general, gaps for adults used, except when the device is specifically
designed for use with children, values for children applied.
9.2.2.4.3
RM RESULTS TABLE: Movable guards
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
Test Report No.: 1307033001
60601-1TRF_A
Page 155 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9.2.2.4.3
RM RESULTS TABLE: Movable guards
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
Verdict
N/A
Verdict
5
6.2
6.3
6.4
6.5
9.2.2.4.4
RM RESULTS TABLE: Protective measures
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.2.2.5 c)
RM RESULTS TABLE: Continuous activation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
Test Report No.: 1307033001
60601-1TRF_A
Page 156 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9.2.2.6
RM RESULTS TABLE: Speed of movement(s)
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
Verdict
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.2.3.2
RM RESULTS TABLE: Over travel
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.2.4
RM RESULTS TABLE: Emergency stopping devices
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
6.6
Test Report No.: 1307033001
60601-1TRF_A
Page 157 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9.2.5
RM RESULTS TABLE: Release of patient
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.3
RM RESULTS TABLE: Hazards associated with surfaces, corners and edges
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-02, H-01, C-03
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the enclosure shape have been
identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Evaluation of
risk (before measures:
F-02, H-01, C-03
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-02, H-01, C03
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-018
The risk control measures selected were
recorded in the risk management file.
Pass

6.2

Test Report No.: 1307033001
60601-1TRF_A
Page 158 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
9.3
RM RESULTS TABLE: Hazards associated with surfaces, corners and edges
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.3

xxxx-ONYX2121DTyxxxxxxx
Section - Release
Protocols of Measures:
RC-018
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-02, H-01, C03
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
N/A
9.4.2.1
Verdict
TABLE: Instability—overbalance in transport position
ME EQUIPMENT
preparation
Test Condition (transport position)
Pass
N/A
Remarks
Supplementary information:
Test Report No.: 1307033001
60601-1TRF_A
Page 159 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
9.4.2.2
Requirement + Test
Result - Remark
Verdict
TABLE: Instability—overbalance excluding transport position
ME EQUIPMENT
preparation
N/A
Test Condition (excluding transport
position) Test either 5 ° incline and verify
Warning marking or 10 ° incline)
Remarks
Supplementary information:
9.4.2.3
TABLE: Instability—overbalance from horizontal and vertical forces
ME EQUIPMENT
preparation
Test Condition (force used, direction of
force, weight of equipment, location of
force
N/A
Remarks
Supplementary information:
9.4.2.4.2 TABLE: Castors and wheels – Force for propulsion
ME EQUIPMENT
preparation
Test Condition (force location and height)
N/A
Remarks
Supplementary information:
9.4.2.4.3 TABLE: Castors and wheels – Movement over a threshold
ME EQUIPMENT
preparation
Test Condition (speed of movement)
N/A
Remarks
Supplementary information:
9.4.2.4.3
RM RESULTS TABLE: Movement over a threshold
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Test Report No.: 1307033001
60601-1TRF_A
Result - Remarks
Page 160 of 203
N/A
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9.4.2.4.3
RM RESULTS TABLE: Movement over a threshold
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.4.3.1
TABLE: Instability from unwanted lateral movement (including sliding) in
transport position
ME EQUIPMENT
Preparation
Test Condition (transport position,
working load, locking device(s), caster
position)
N/A
Remarks
Supplementary information:
9.4.3.2
TABLE: Instability from unwanted lateral movement (including sliding)
excluding transport position
ME EQUIPMENT
Preparation
Test Condition (working load, locking
device(s), caster position, force, force
location, force direction)
N/A
Remarks
Supplementary information:
Test Report No.: 1307033001
60601-1TRF_A
Page 161 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
9.4.4
Requirement + Test
Result - Remark
Verdict
TABLE: Grips and other handling devices
Clause and Name of Test
Test Condition
N/A
Remarks
Supplementary information:
9.5.1
RM RESULTS TABLE: Protective means
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
9.6.1
RM RESULTS TABLE: Acoustic energy - General
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.6.2.2
RM RESULTS TABLE: Infrasound and ultrasound energy
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
Test Report No.: 1307033001
60601-1TRF_A
Page 162 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9.6.2.2
RM RESULTS TABLE: Infrasound and ultrasound energy
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
6.4
6.5
9.7.2
RM RESULTS TABLE: Pneumatic and hydraulic parts
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
9.7.4
RM RESULTS TABLE: Pressure rating of ME equipment parts
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
9.7.5
TABLE: Pressure vessels
Hydraulic,
Pneumatic or
Suitable Media
and Test
Pressure
Vessel Burst
N/A
Permanent
Deformation
Leaks
Vessel fluid
substance
Remarks
Supplementary Information:
Test Report No.: 1307033001
60601-1TRF_A
Page 163 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
9.7.6
RM RESULTS TABLE: Pressure-control device
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
9.7.7
RM RESULTS TABLE: Pressure-relief device
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
9.8.1
RM RESULTS TABLE: Hazards associated with support systems - General
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”.
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-02, H-01, C-02
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the support systems have been
considered.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-02, H-01, C02
The estimates of the risk(s) were recorded in
the risk management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 164 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
9.8.1
RM RESULTS TABLE: Hazards associated with support systems - General
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-02, H-01, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-010
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-010
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-02, H-01, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
N/A
9.8.2
RM RESULTS TABLE: Tensile safety factor
Test Report No.: 1307033001
60601-1TRF_A
Page 165 of 203
Pass
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-02, H-01, C-02
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the support systems have been
considered.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-02, H-01, C02
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-02, H-01, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-010
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-010
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 166 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
9.8.2
RM RESULTS TABLE: Tensile safety factor
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.4

All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-02, H-01, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
Pass
Verdict
6.5
N/A
9.8.3.1
RM RESULTS TABLE: Strength of patient or operator support or suspension
systems - General
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
9.8.3.2
TABLE: PATIENT support/suspension system - Static forces
ME EQUIPMENT
or area
part
Position
Test Report No.: 1307033001
60601-1TRF_A
Load
Page 167 of 203
Area
N/A
Remarks
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
9.8.3.2
Requirement + Test
Result - Remark
Verdict
TABLE: PATIENT support/suspension system - Static forces
N/A
Supplementary Information:
9.8.3.2a, b
RM RESULTS TABLE: Static forces due to loading from persons
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
9.8.3.3
TABLE: Support/Suspension System – Dynamic forces due to loading from
persons
ME EQUIPMENT
part or area
Position
Safe Working
Load
Area
N/A
Remarks
Supplementary Information:
9.8.4.1
RM RESULTS TABLE: Systems with mechanical protective devices - General
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
9.8.4.3
RM RESULTS TABLE: Mechanical protective device for single activation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.3
Test Report No.: 1307033001
60601-1TRF_A
Page 168 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
9.8.4.3
RM RESULTS TABLE: Mechanical protective device for single activation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.4
5
6.2
6.3
6.4
6.5
9.8.5
RM RESULTS TABLE: Systems without mechanical protective devices
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the support systems have been
considered.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-02, H-01, C02
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-02, H-01, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-010
The risk control measures selected were
recorded in the risk management file.
Pass

6.2

Test Report No.: 1307033001
60601-1TRF_A
Page 169 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
9.8.5
RM RESULTS TABLE: Systems without mechanical protective devices
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-010
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-02, H-01, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
N/A
10.1.1
TABLE: Measurement of X - radiation
Pass
Verdict
N/A
Maximum allowable radiation pA/kg ( μSv/h) (mR/h) 36 (5 Sv/h) (0.5 mR/h)
Surface area under test
Surface no./ Description1
1/
/
2/
/
3/
/
4/
/
5/
/
6/
/
7/
/
8/
/
9/
/
10/
Measured Radiation,
Remarks
pA/kg (μSv/h) (mR/h)
/
Test Report No.: 1307033001
60601-1TRF_A
Page 170 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
1
Supplementary information: Measurements made at a distance of 5 cm from any surface to which
OPERATOR (other than SERVICE PERSONNEL) can gain access without a TOOL, is deliberately provided with
means of access, or is instructed to enter regardless of whether or not a TOOL is needed to gain access
10.1.2
RM RESULTS TABLE: ME equipment intended to produce diagnostic or
therapeutic X-radiation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
10.2
RM RESULTS TABLE: Alpha, beta, gamma, neutron & other particle radiation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
10.3
RM RESULTS TABLE: Microwave radiation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
Test Report No.: 1307033001
60601-1TRF_A
Page 171 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
10.5
RM RESULTS TABLE: Other visible electromagnetic radiation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
10.6
RM RESULTS TABLE: RISK associated with infrared radiation other than
emitted by lasers and LEDs
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
10.7
RM RESULTS TABLE: RISK associated with ultraviolet radiation other than
emitted by lasers and LEDs
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
Test Report No.: 1307033001
60601-1TRF_A
Page 172 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
TABLE: Excessive temperatures in ME EQUIPMENT
xxxxModel No................................................ :
ONYX-2121DTyxxxxxxx
xxxxTest ambient (C) ................................. :
-ONYX2121DTyxxxxxxx
4
Test supply voltage/frequency (V/Hz) .. : A.90/60
-B.264/60
Verdict
11.1.1
Model No.
Thermocouple
No.
--
See
below
--
1.
--
Thermocouple location3
--
--
--
--
--
--
--
--
--
Max allowable
Max
temperature1 from
measured
Table 22, 23 or 24
temperature2,
or RM file for AP5
(C)
(C)
Remarks
See below
A.
B.
BT2 body (Main board)
100
81.1
82.6
2.
L35 coil (Main board)
125
109.3
110.0
--
3.
PWB near CPU (Main board)
130
81.0
82.2
--
4.
PWB near U29 (Main board)
130
89.4
91.0
--
5.
PWB near USB port
130
75.6
77.9
--
6.
L1 coil (LCD driver board)
105
74.3
76.1
--
7.
C7 Body (LCD driver board)
130
65.7
67.7
--
8.
HDD body
--
58.1
59.4
--
9.
Panel body
60
45.0
47.3
--
10.
Switch body
60
41.9
44.9
11.
Plastic enclosure inside near
CPU
70
62.6
63.7
12.
Plastic enclosure outside near
CPU
60
49.2
51.6
(10 ‘s <t<
1min)
13.
Plastic enclosure outside near
USB
60
46.5
49.2
(10 ‘s <t<
1min)
14.
Metal enclosure outside near
CPU
74
59.9
61.1
15.
Test corner near rear enclosure
90
42.2
43.6
16.
Test corner near output
connector
90
40.6
43.4
17.
T1 coil (power supply)
120
93.0
91.0
--------
See below
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 173 of 203
--
(10 ‘s <t<
1min)
(10 ‘s <t<
1min)
(t< 1 ‘s)
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
----
Requirement + Test
Result - Remark
Verdict
18.
IC2 body (power supply)
100
84.6
80.3
19.
Enclosure outside near T1
(power supply)
86
64.0
59.5
20.
Ambient
--
(t< 1 ‘s)
40.0
40.0
(39.6) (39.5)
Supplementary information:
Maximum allowable temperature on surfaces of test corner is 90 C
Max temperature determined in accordance with 11.1.3e)
3
When thermocouples used to determine temperature of windings, limits of Table 22 reduced by 10 °C.
4
Supply voltage:
- ME EQUIPMENT with heating elements - 110 % of the maximum RATED voltage;
- Motor operated ME EQUIPMENT - least favourable voltage between 90 % of the minimum RATED and 110 % of
the maximum RATED voltage. ME EQUIPMENT operated under normal load and normal DUTY CYCLE.
- Combined heating and motor operated and other ME EQUIPMENT - tested both at 110 % of the maximum
RATED voltage and at 90 % of the minimum RATED voltage.
5
APPLIED PARTS intended to supply heat to a PATIENT - See RISK MANAGEMENT FILE containing temperatures and clinical
effects. Also, see instructions for use.
1
2
Test Report No.: 1307033001
60601-1TRF_A
Page 174 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
11.1.1
RM RESULTS TABLE: Maximum temperature during normal use
(Table 23 or 24)
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-02, C-01
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the heat generated by the device have
been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C01
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-032
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-032
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 175 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
11.1.1
RM RESULTS TABLE: Maximum temperature during normal use
(Table 23 or 24)
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.4

All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-02, C01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
Pass
Verdict
6.5
N/A
11.1.2.1
RM RESULTS TABLE: Applied parts intended to supply heat to patient
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
11.1.2.2
RM RESULTS TABLE: Applied parts not intended to supply heat to patient
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
Test Report No.: 1307033001
60601-1TRF_A
Page 176 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
11.1.2.2
RM RESULTS TABLE: Applied parts not intended to supply heat to patient
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
6.3
6.4
6.5
11.1.3
TABLE: Temperature of windings by change-of-resistance method
Temperature T of winding:
t1 (°C)
R1 ()
t2 (°C)
R2 ()
N/A
T (C)
Allowed Insulatio
Tmax (C) n class
Supplementary information:
11.1.3
RM RESULTS TABLE: Measurements
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-02, C-01
The manufacturer had compiled a list of
known or foreseeable hazards associated
with the heat generated by the device have
been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C01
The estimates of the risk(s) were recorded in
the risk management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 177 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
11.1.3
RM RESULTS TABLE: Measurements
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
6.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-032
The risk control measures selected were
recorded in the risk management file.
Pass
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-032
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-02, C01
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


6.5
N/A
11.2.2.1
RM RESULTS TABLE: Risk of fire in an oxygen rich environment
Test Report No.: 1307033001
60601-1TRF_A
Page 178 of 203
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Clause of
ISO 14971
Requirement + Test
Result - Remark
Document Ref. in RMF
(Document No. & paragraph)
Verdict
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
11.2.2.1
TABLE: Alternative method to 11.2.2.1 a) 5) to determine existence of an
ignition source
Areas where sparking might cause ignition:
N/A
Remarks
1.
2.
3.
5.
6.
Materials of the parts between which sparks could occur (Composition, Grade
Designation, Manufacturer):
Remarks
1.
2.
3.
4.
5.
6.
Test parameters selected representing worst case conditions for ME EQUIPMENT:
Remarks
Oxygen concentration (%) ............... :
Fuel .................................................. :
Current (A) ....................................... :
Voltage (V) ....................................... :
Capacitance (F).............................. :
Test Report No.: 1307033001
60601-1TRF_A
Page 179 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Inductance or resistance (h or ).... :
No. of trials (300 Min)....................... :
Sparks resulted in ignition (Yes/No) :
Supplementary information:
Test procedure of 11.2.2.1 a) 5) & Figs 35-37 used for tests. For circuits not in Figs 35-37, test voltage or
current set at 3 times the worst case values with other parameters set at worst case values to determine if
ignition can occur.
11.3
RM RESULTS TABLE: Constructional requirements for fire enclosures of ME
equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-02, C-02
The manufacturer had compiled a list of
known or foreseeable hazards associated
with potential ignition sources and spread of
fire has been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section -Release
Protocols of Measures:
RC-006
The risk control measures selected were
recorded in the risk management file.
Pass

6.2

Test Report No.: 1307033001
60601-1TRF_A
Page 180 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
11.3
RM RESULTS TABLE: Constructional requirements for fire enclosures of ME
equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-006
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-02, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


Verdict
6.5
N/A
11.5
RM RESULTS TABLE: ME equipment and ME systems intended for use in
conjunction with flammable agents
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
Pass
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
Test Report No.: 1307033001
60601-1TRF_A
Page 181 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
11.6.1
Requirement + Test
Result - Remark
Verdict
TABLE: overflow, spillage, leakage, ingress of water, cleaning, disinfection,
sterilization, compatibility with substances
Clause / Test Name
Test Condition
Part under test
Pass
Remarks
Cleaning (11.6.6)
Cleaning (Gently wipe it with a piece
of soft dry cloth)
Unit
No dielectric breakdown
and leakage current test
compliances.
Cleaning (11.6.6)
Cleaning (Gently wipe it with a piece
of cloth moistened with water)
Unit
No dielectric breakdown
and leakage current test
compliances.
Cleaning (11.6.6)
Cleaning (Gently wipe it with a piece
of cloth moistened with rubbing
alcohol)
Unit
No dielectric breakdown
and leakage current test
compliances.
Cleaning (11.6.6)
Cleaning (with Vacuum cleaner)
Unit
No dielectric breakdown
and leakage current test
compliances.
11.6.2
RM RESULTS TABLE: Overflow in ME equipment
N/A
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Supplementary information: --
Result - Remarks
Verdict
4.2
4.3
4.4
5
11.6.3
RM RESULTS TABLE: Spillage on ME equipment and ME system
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
11.6.5
RM RESULTS TABLE: Ingress of water or particulate matter into ME
EQUIPMENT and ME SYSTEMS
Test Report No.: 1307033001
60601-1TRF_A
Page 182 of 203
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Clause of
ISO 14971
Requirement + Test
Document Ref. in RMF
(Document No. & paragraph)
Result - Remark
Verdict
Result - Remarks
Verdict
4.2
4.3
4.4
5
11.6.6
RM RESULTS TABLE: Cleaning and disinfection of ME equipment and ME
systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RMF mention the “Intended
use/intended purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-01, C-23
The manufacturer had compiled a list of
known or foreseeable hazards associated
with use of the device in environment and
the required cleaning and disinfection of the
device and its accessories have been
identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C23
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-01, C23
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-022
The risk control measures selected were
recorded in the risk management file.
Pass

6.2

Test Report No.: 1307033001
60601-1TRF_A
Page 183 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
11.6.6
RM RESULTS TABLE: Cleaning and disinfection of ME equipment and ME
systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
6.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Release
Protocols of Measures:
RC-022
The effectiveness of the risk control
measures had be verified and the results of
the verification were recorded in the risk
management file.
Pass
6.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - New
evaluation after
measures: F-01, H-01, C23
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
After Measures
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Overview of
the Results of the Risk
All residual risk that remains after the risk
control measure(s) are applied and
evaluated using the criteria that defined in
the risk management plan
Pass
No reduction of risk through Risk/benefit
analysis
N/A


Verdict
6.5
N/A
11.6.7
RM RESULTS TABLE: Sterilization of ME equipment and ME systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
Pass
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
11.6.8
RM RESULTS TABLE: Compatibility with substances used
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Test Report No.: 1307033001
60601-1TRF_A
Result - Remarks
Page 184 of 203
N/A
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
11.6.8
RM RESULTS TABLE: Compatibility with substances used
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
Verdict
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
12.1
RM RESULTS TABLE: Accuracy of controls and equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
12.3
RM RESULTS TABLE: Alarm systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
12.4.1
RM RESULTS TABLE: Intentional exceeding of safety limits
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
Test Report No.: 1307033001
60601-1TRF_A
Page 185 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
12.4.1
RM RESULTS TABLE: Intentional exceeding of safety limits
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.3
4.4
5
6.2
6.3
6.4
6.5
12.4.2
RM RESULTS TABLE: Indication of parameters relevant to safety
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
12.4.3
RM RESULTS TABLE: Accidental selection of excessive output values
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
12.4.4
RM RESULTS TABLE: Incorrect output
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
Test Report No.: 1307033001
60601-1TRF_A
Page 186 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
12.4.4
RM RESULTS TABLE: Incorrect output
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.4
5
6.2
6.3
6.4
6.5
12.4.5.2
RM RESULTS TABLE: Diagnostic X-ray equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
12.4.5.3
RM RESULTS TABLE: Radiotherapy equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
12.4.5.4
RM RESULTS TABLE: Other ME equipment producing diagnostic or
therapeutic radiation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
Test Report No.: 1307033001
60601-1TRF_A
Page 187 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
12.4.5.4
RM RESULTS TABLE: Other ME equipment producing diagnostic or
therapeutic radiation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
5
6.2
6.3
6.4
6.5
12.4.6
RM RESULTS TABLE: Diagnostic or therapeutic acoustic pressure
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
13.1.2
TABLE: measurement of power or energy dissipation in parts & components to
waive SINGLE FAULT CONDITIONS in 4.7, 8.1 b), 8.7.2, and 13.2.2 relative to
emission of flames, molten metal, or ignitable substances
Power dissipated less than (W) ............................. :
15
Energy dissipated less than (J) ............................. :
900
Part or component
tested
Measured
power
dissipated (W)
Calculated energy SINGLE FAULT CONDITIONS
dissipated (J)
waived (Yes/No)
N/A
Remarks
Supplementary information: --
Test Report No.: 1307033001
60601-1TRF_A
Page 188 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
13.2
TABLE: SINGLE FAULT CONDITIONS in accordance with 13.2.2 to 13.2.13, inclusive
Clause
No.
13.2.2
Result - Remark
Verdict
Results observed
D33 short for RTC Battery circuit
HAZARDOUS
SITUATION
Description of SINGLE FAULT CONDITION
Electrical SINGLE FAULT CONDITIONS per Clause 8.1:
Pass
(Yes/No)
—
—
Normal operation. No
component damaged.
NO
Battery current: 3.0mA
Test duration: 15 min
NO
13.2.3
Overheating of transformers per Clause 15.5:
—
—
13.2.4
Failure of THERMOSTATS according to 13.2.13 &
15.4.2, overloading - THERMOSTATS short circuited
or interrupted, the less favourable of the two:
—
—
13.2.5
Failure of temperature limiting devices according
to 13.2.13 & 15.4.2, overloading, THERMOSTATS
short circuited or interrupted, the less favourable
of the two:
—
—
13.2.6
Leakage of liquid - RISK MANAGEMENT FILE
examined to determine the appropriate test
conditions (sealed rechargeable batteries
exempted)
—
—
13.2.7
Impairment of cooling that could result in a
HAZARD using test method of 11.1:
—
—
Test Report No.: 1307033001
60601-1TRF_A
Page 189 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Clause
No.
Requirement + Test
Result - Remark
Verdict
Results observed
SITUATION
Description of SINGLE FAULT CONDITION
Ventilation opening blocked
HAZARDOUS
(Yes/No)
Normal operation. No
hazards. No component
damaged. Measure max.
temp at L35 body: 119.7°C,
Plastic enclosure inside near
heatsink of CPU:71.4°C ,
Plastic enclosure outside
near heatsink of CPU:
63.8°C, Test corner: 43.8°C,
Ambient: 36.2 °C
NO
13.2.8
Locking of moving parts – Only one part locked
at a time – Also see 13.2.10 below:
—
—
13.2.9
Interruption and short circuiting of motor
capacitors – Motor capacitors short & open
circuited 1 – Also see 13.10
—
—
V measured =
V measured =
13.2.10
Additional test criteria for motor operated ME
EQUIPMENT in 13.2.8 &13.2.9:
—
—
—
—
For every test in SINGLE FAULT CONDITION of 13.2.8
and 13.2.9, motor-operated EQUIPMENT stared
from COLD CONDITION at RATED voltage or at the
upper limit of RATED voltage range for specified
time:
Temperatures of windings determined at the end
of specified test periods or at the instant of
operation of fuses, THERMAL CUT-OUTS, motor
protective devices
Temperatures measured as specified in 11.1.3 d)
Temperatures did not exceed limits of Table 26
13.2.11
Failures of components in ME EQUIPMENT used in
conjunction with OXYGEN RICH ENVIRONMENTS:
Test Report No.: 1307033001
60601-1TRF_A
Page 190 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Clause
No.
13.2.12
Result - Remark
Verdict
Results observed
SITUATION
Description of SINGLE FAULT CONDITION
Failure of parts that might result in a MECHANICAL
HAZARD (See 9 & 15.3):
HAZARDOUS
(Yes/No)
—
—
Supplementary information:
1
Test with short-circuited capacitor not performed when motor provided with a capacitor complying with IEC 60252-1 and the ME
intended for unattended use including automatic or remote control. See Attachment # and appended Table 8.10.
EQUIPMENT not
13.2.6
RM RESULTS TABLE: Leakage of liquid
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
14.1
RM RESULTS TABLE: Programmable electrical medical systems - General
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
14.6.1
RM RESULTS TABLE: Identification of known and foreseeable hazards
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.3
14.6.2
RM RESULTS TABLE: Risk control
Test Report No.: 1307033001
60601-1TRF_A
Page 191 of 203
N/A
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Clause of
ISO 14971
Requirement + Test
Document Ref. in RMF
(Document No. & paragraph)
Result - Remark
Verdict
Result - Remarks
Verdict
14.7
RM RESULTS TABLE: Requirement specification
N/A
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
6.1
Result - Remarks
Verdict
6.3
14.8
RM RESULTS TABLE: Architecture
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
6.3
14.9
RM RESULTS TABLE: Design and Implementation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
6.2
6.3
14.10
RM RESULTS TABLE: Verification
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
6.3
14.11
RM RESULTS TABLE: PEMS validation
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
6.3
Test Report No.: 1307033001
60601-1TRF_A
Page 192 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
14.13
RM RESULTS TABLE: Connection of PEMS by NETWORK/DATA COUPLING
to other equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
15.1
RM RESULTS TABLE: Construction of ME equipment – Arrangements of
controls and indicators of ME equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
Test Report No.: 1307033001
60601-1TRF_A
Page 193 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
15.3
Clause
15.3.2
Requirement + Test
Result - Remark
TABLE: Mechanical Strength tests 1)
Name of Test
Force = 250 N ± 10 N for 5 s
Part under test :
Unit (Top),
Unit (Side),
Unit (Bottom)
15.3.3
Pass
Test conditions
Push Test
Impact Test
Steel ball (50 mm in dia., 500 g ± 25
g) falling from a 1.3 m
Part under test :
Unit (Top),
Unit (Side),
Unit (Bottom)
Verdict
Observed results/Remarks
There was no cracking of the
enclosure and were no live parts
that became accessible
There was no cracking of the
enclosure and were no live parts
that became accessible
15.3.4.1 Drop Test (handheld)
Free fall height (m) =
--
15.3.4.2 Drop Test (portable)
Drop height (cm) = 5 cm
There was no damage to the
interior or exterior of the tested
sample which caused exposure to
live parts or increased risk of
electric shock.
15.3.5
Rough handling test
Travel speed (m/s) =
--
15.3.6
Mould Stress Relief
7 h in oven at temperature (°C) =
74°C
There was no warping, cracking,
breaking and exposure of live
parts.
1)
Supplementary information: As applicable, Push, Impact, Drop, Mould Stress Relief and Rough
Handling Tests (delete not applicable rows).
Test Report No.: 1307033001
60601-1TRF_A
Page 194 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
15.3.2
RM RESULTS TABLE: Push test
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RM mention the “Intended use/intended
purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-02, C-02
The manufacturer had compiled a list of
known or foreseeable hazards associated
with failure of the protective enclosure have
been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
15.3.3
RM RESULTS TABLE: Impact test
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RM mention the “Intended use/intended
purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-02, C-02
The manufacturer had compiled a list of
known or foreseeable hazards associated
with failure of the protective enclosure have
been identified.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 195 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
15.3.3
RM RESULTS TABLE: Impact test
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
The estimates of the risk(s) were recorded in
the risk management file.
Pass
5

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
The results of this risk evaluation were
recorded in the risk management file.
Pass

Pass
Verdict
15.3.4.2
RM RESULTS TABLE: Portable ME equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
4.2

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Version Data
The RM mention the “Intended use/intended
purpose”
Pass
4.3

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Analysis:
F-01, H-02, C-02
The manufacturer had compiled a list of
known or foreseeable hazards associated
with failure of the protective enclosure have
been identified.
Pass
4.4

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
The estimates of the risk(s) were recorded in
the risk management file.
Pass
Test Report No.: 1307033001
60601-1TRF_A
Page 196 of 203
Pass
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
15.3.4.2
RM RESULTS TABLE: Portable ME equipment
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
5

The results of this risk evaluation were
recorded in the risk management file.

xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk
Evaluation Before
Measures: F-01, H-02, C02
xxxx-ONYX-2121DTyxxxxxxx
_RMF_2013_07_19,
Section - Risk Graph
Before Measures
15.3.5
RM RESULTS TABLE: Rough handling test
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Pass
Verdict
Pass
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
15.4.1
RM RESULTS TABLE: Construction of connectors
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
15.4.2.1 a
RM RESULTS TABLE: THERMAL CUT-OUTS and OVER-CURRENT RELEASES
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
Test Report No.: 1307033001
60601-1TRF_A
Page 197 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
15.4.2.1 b
RM RESULTS TABLE: T HERMAL CUT-OUTS with a safety function
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
15.4.2.1 c
RM RESULTS TABLE: Independent non-SELF-RESETTING THERMAL CUT-OUT
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
15.4.2.1 d
RM RESULTS TABLE: Loss of function of ME EQUIPMENT
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
15.4.2.1 h
RM RESULTS TABLE: ME EQUIPMENT with tubular heating elements
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
15.4.3.1
RM RESULTS TABLE: Housing
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
15.4.3.2
RM RESULTS TABLE: Connection
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
Test Report No.: 1307033001
60601-1TRF_A
Page 198 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
15.4.3.2
RM RESULTS TABLE: Connection
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
Verdict
N/A
Verdict
4.4
15.4.3.3
RM RESULTS TABLE: Protection against overcharging
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
15.4.3.4
RM RESULTS TABLE: Lithium batteries
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
15.4.3.5
RM RESULTS TABLE: Excessive current and voltage protection
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
15.4.4
RM RESULTS TABLE: Indicators
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
15.4.5
RM RESULTS TABLE: Pre-set controls
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Test Report No.: 1307033001
60601-1TRF_A
Result - Remarks
Page 199 of 203
N/A
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
15.4.5
RM RESULTS TABLE: Pre-set controls
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Verdict
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
15.4.6
TABLE: actuating parts of controls of ME EQUIPMENT – torque & axial pull tests
Rotating control
under test
Gripping
diameter “d” of
control knob
(mm) 1
Torque from
Table 30 (Nm)
Axial force
applied
(N)
Unacceptable
RISK occurred
Yes/No
N/A
Remarks
Supplementary information: 1 Gripping diameter (d) is the maximum width of a control knob regardless of its
shape (e.g. control knob with pointer)
15.4.7.3 b
RM RESULTS TABLE: Entry of liquids
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
15.5.1.2
TABLE: transformer short circuit test short-circuit applied at end of windings
or at the first point that could be short circuited under SINGLE FAULT CONDITION
N/A
Primary voltage (most adverse value from 90 % to 110 % of RATED voltage)(V)1 .... :
—
RATED input frequency (Hz) ............................................................................ :
—
Winding
tested
Class of
insulation
(A, B, E, F,
or H)
Type of
protective device
(fuse, circuit
breaker) /Ratings
Test Report No.: 1307033001
60601-1TRF_A
Protective
device
operated
Yes/No
Time to THERMAL
STABILITY (when
protective device
did not
operate)(Min)
Page 200 of 203
Maximum
allowed
temp from
Table 31
(ºC)
Maximum
winding
temp
measured
(ºC)
Ambient
(ºC)
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Supplementary information:
1
Loads on other windings between no load and their NORMAL USE load. Short-circuit applied at end of
windings or at the first point that could be short circuited under SINGLE FAULT CONDITION.
EVALUATED AT CERTIFICED POWER SUPPLY.
15.5.1.3
TABLE: transformer overload test – conducted only when protective device
under short-circuit test operated
N/A
1
Primary voltage, most adverse value between 90 % to 110 % of RATED voltage (V) .. :
RATED input frequency (Hz) ......................................................................................... :
Test current just below minimum current that would activate protective device &
achieve THERMAL STABILITY under method a) (A) .......................................................... :
Test current based on Table 32 when protective device that operated under method a)
is external to transformer, and it was shunted (A)...................................................... :
Winding tested
Class of
insulation
(A, B, E, F, H)
Type of protective device
used (fuse, circuit
breaker)/Ratings
Maximum allowed
temp from
Table 31 (ºC)
Maximum
winding temp
measured (ºC)
Ambient
(ºC)
Supplementary information:
1
Loads on other windings between no load and their NORMAL USE load.
Time durations: - IEC 60127-1 fuse: 30 min at current from Table 32.
Non IEC 60127-1 fuse: 30 min at the current based on characteristics supplied by fuse manufacturer, specifically, 30 min
clearing-time current. When no 30 min clearing-time current data available, test current from Table 32 used until THERMAL
STABILITY achieved.
- Other types of protective devices: until THERMAL STABILITY achieved at a current just below minimum current operating the
protective device in a). This portion concluded at specified time or when a second protective device opened.
15.5.2
TABLE: Transformer dielectric strength after humidity preconditioning of 5.7
Transformer
Model/Type/ Part
No
Test voltage applied between
Test
voltage,
(V)
Test
frequency
(Hz)
Breakdown
Yes/No
N/A
Deterioration
Yes/No
Primary & secondary windings
Primary winding & frame
Secondary winding & frame
Test Report No.: 1307033001
60601-1TRF_A
Page 201 of 203
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
Supplementary information: Tests conducted under the conditions of 11.1, in ME EQUIPMENT or under
simulated conditions on the bench. See Clause 15.5.2 for test parameters & other details
16.1
RM RESULTS TABLE: General requirements for ME Systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
16.6.1
TABLE: LEAKAGE CURRENTS in ME SYSTEM _ TOUCH CURRENT MEASUREMENTS
Specific area where TOUCH
measured (i.e., from
or between parts of ME SYSTEM
within PATIENT ENVIRONMENT)
CURRENT
Allowable
Measured
TOUCH CURRENT
in NORMAL
TOUCH CURRENT
in NORMAL
CONDITION (A)
CONDITION (A)
Allowable TOUCH CURRENT
in event of interruption of
N/A
PROTECTIVE EARTH
Measured TOUCH
in event of
interruption of
CONDUCTOR,
PROTECTIVE EARTH
(A)
CURRENT
CONDUCTOR,
100
500
100
500
100
500
100
500
100
500
(A)
Supplementary information:
16.9.1
RM RESULTS TABLE: Connection terminals and connectors
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
N/A
Result - Remarks
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
17
RM RESULTS TABLE: Electromagnetic compatibility of ME equipment and
ME systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Test Report No.: 1307033001
60601-1TRF_A
Result - Remarks
Page 202 of 203
N/A
Verdict
Issued: 22 August. 2013
VICTRONIC TECHNOLOGY CORPORATION
IEC 60601-1
Clause
Requirement + Test
Result - Remark
Verdict
17
RM RESULTS TABLE: Electromagnetic compatibility of ME equipment and
ME systems
Clause of
ISO 14971
Document Ref. in RMF
(Document No. & paragraph)
Result - Remarks
N/A
Verdict
4.2
4.3
4.4
5
6.2
6.3
6.4
6.5
SP
TABLE: Additional or special tests conducted
Clause and Name of Test
Test type and condition
N/A
Observed results
Supplementary information:
Test Report No.: 1307033001
60601-1TRF_A
Page 203 of 203
Issued: 22 August. 2013