Download Stat Profile pHOx Service Manual
Transcript
Service Manual Preface pHOx Service Manual Stat Profile® pHOx® Service Manual Part Number and Ordering Information The Stat Profile® pHOx® Service Manual (PN 24301) can be ordered from Nova Biomedical Order Services. Write or call: Nova Biomedical 200 Prospect Street Waltham, MA 02454-9141 U.S.A. Tel: 1-800-458-5813 FAX: (781) 893-6998 (in the U.S.A.) or (781) 899-0417 (outside the U.S.A.) Technical Assistance (USA) For technical assistance, call Nova Biomedical Technical Services at: Tel: 1-800-545-NOVA or (781) 894-0800 FAX: (781) 894-0585 Trademarks Stat Profile® is a registered trademark of Nova Biomedical. pHOx® is a registered trademark of Nova Biomedical. Copyright Printed in the U.S.A. Copyright 2005, Nova Biomedical, Waltham, MA 02454. i Preface ii pHOx Service Manual Stat Profile pHOx Service Manual Contents 1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 1.1 1.2 1.3 The Stat Profile pHOx Analyzer ............................................................................1-1 1.1.1 Environmental ............................................................................................1-1 1.1.2 Energy consumption ..................................................................................1-1 1.1.3 Physical Dimensions ..................................................................................1-1 1.1.4 Regulatory Compliance .............................................................................1-1 1.1.5 Location - Stable stationary surface. .........................................................1-2 1.1.6 Operational.................................................................................................1-2 1.1.7 General .......................................................................................................1-2 Cautions and Hazards ............................................................................................1-3 Required tools .......................................................................................................1-4 2.0 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 2.1 2.2 Mechanical Assemblies..........................................................................................2-1 2.1.1 Pump ..........................................................................................................2-2 2.1.2 Sampler Assembly ....................................................................................2-2 2.1.3 Rotary Valve Assembly .............................................................................2-2 2.1.4 Sensor Module Assembly .........................................................................2-3 Electronic Assemblies. ...........................................................................................2-3 2.2.1 Power Entry Module .................................................................................2-3 2.2.2 Power Supply ............................................................................................2-3 2.2.3 Digital Control Board ...............................................................................2-4 2.2.4 Analog Board .............................................................................................2-6 2.2.5 SO2 Control Board.....................................................................................2-6 2.2.6 Electrode Interface Boards .........................................................................2-7 2.2.7 Door Assembly Electronics........................................................................2-7 PN 24301 Rev. B 2005/11 TOC-1 Table of Contents 3.0 Replacement Procedures . . . . . . . . . . . . . . . . . . . . 3-1 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 3.10 3.11 3.12 3.13 3.14 3.15 3.16 3.17 3.18 3.19 Cover Removal.......................................................................................................3-1 Tubing Harness .....................................................................................................3-2 Power Supply Removal ..........................................................................................3-4 Sampler Assembly .................................................................................................3-5 Rotary Valve Assembly ........................................................................................3-10 Valve Assembly ....................................................................................................3-11 Rotary Valve Motor ..............................................................................................3-12 Pump Assembly ...................................................................................................3-13 Digital Control Board ..........................................................................................3-14 Analog Board ......................................................................................................3-15 Printer Assembly .................................................................................................3-16 Sensor Module Light Source ...............................................................................3-17 Door Removal /Reinstallation Procedure ...........................................................3-18 Display Assembly ................................................................................................3-20 Distribution board ...............................................................................................3-21 Keypad Assembly ...............................................................................................3-21 Sensor Module Assembly ....................................................................................3-22 Electrode Interface Boards ...................................................................................3-23 Pinch Valve Assemblies .......................................................................................3-24 4.0 Troubleshooting Information . . . . . . . . . . . . . . . . . . 1-1 4.1 4.2 4.3 TOC-2 Service Screens ......................................................................................................1-3 4.1.1 System Test Screen ....................................................................................1-4 4.1.1.1 Operator Flow Test .........................................................................1-4 4.1.1.2 Service Flow test ............................................................................1-5 4.1.2 Analog Input Screen ..................................................................................1-6 4.1.3 Sensor Subsystem .....................................................................................1-7 4.1.4 Printer Menu ..............................................................................................1-8 4.1.5 Communication Tests .................................................................................1-9 Flow Problems .....................................................................................................1-10 4.2.1 Isolating a Plug in the W/R to Waste Tubing ...........................................1-11 4.2.2 Isolating a flow problem in the Sampler/Sensor module .........................1-11 4.2.3 Isolating a flow problem between the Fluid Pack and the Fluid Fountain1-12 Status Codes .......................................................................................................4-13 4.3.1 Analyte Troubleshooting ( Series code 1-6 and Digit codes 1-4) ............4-18 4.3.2 Single Channel Failures ......................................................................4-20 4.3.3 Temperature Problems .............................................................................4-21 4.3.4 Air Detector (ADT ) Problems ................................................................4-21 Operator Flow Test ...............................................................................................4-26 Sensor Module Back Flush Procedure .................................................................4-27 PN 24301 Rev. B 2005/11 Stat Profile pHOx Service Manual A Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1 A.1 A.2 A.3 A.4 A.5 A.6 S Quick Reference Guide .........................................................................................A-1 General Maintenance/QC......................................................................................A-8 A.2.1 Other Analyzer Maintenance ..................................................................A-12 pHOx Electrode/Standards..................................................................................A-13 Electronics and Mechanicals Overview ..............................................................A-13 Shutdown Procedure ...........................................................................................A-14 Start Up Procedure ..............................................................................................A-14 Installation Check List ........................................................................................A-15 Preventive Maintenance Check List....................................................................A-17 Explosion Parts List ............................................................................................A-19 Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . S-1 #10-1087A S-2 #10-1071B #10-1072B #10-1073B #10-1074B #10-1075A #10-1076B #10-1077B #10-1078B #10-1079B #10-1080B #10-1081B #10-1082B #10-1083B #10-1084B #10-1085B #10-1086A Wiring Doucument ...................................................................................... S-1 Wireing Charts ............................................................................................. S-2 24292-01 pHOx Digital Board Processor and Memory ............................. S-3 24292-02 pHOx Digital Board Comm. Printer, Display Touch Memory ... S-4 24292-03 pHOx Digital Board Motor & Valve Drivers .............................. S-5 24292 pHOx Digital Board Assembly Drawing .................................... S-6 21642 pHOx SO2 Board ........................................................................ S-7 24727 pHOx Distribition Board ............................................................ S-8 24291 Analog Board - Cortex Assembly Drawing .............................. S-9 24292-01 New Gen. Sensor Board ........................................................... S-10 24292-02 New Gen. Sensor Board ........................................................... S-11 24292-03 New Gen. Sensor Board ........................................................... S-12 24292-04 New Gen. Sensor Board ........................................................... S-13 23897 Pump Assembly ........................................................................ S-14 23899 Sampler Assembly .................................................................... S-15 23898 Rotary Valve Drive Assembly ................................................... S-16 24614 Printer Interface IF3001 Seiko ................................................. S-17 24720 Power Supply Universal input .................................................. S-18 PN 24301 Rev. B 2005/11 TOC-3 Table of Contents TOC-4 PN 24301 Rev. B 2005/11 1 Introduction This manual contains information and procedures used in servicing the Stat Profile pHOx Analyzer. The intent is to supplement the information provided in the Stat Profile pHOx Instructions for Use Manual (PN 37865) and the Stat Profile pHOx Advanced User Applications Manual (PN 35235). This manual does not supersede any current performance specification claims, expendable maintenance procedures, or warranty criteria as outlined in the reference manual. 1.1 The Stat Profile pHOx Analyzer 1.1.1 Environmental Ambient Temperature: Ambient Humidity: 15° - 32° C ( 59° - 86° F) 0% - 95% Non-condensing 1.1.2 Energy consumption Air Conditioning Load: Peak 44 BTU/Hour Power Consumption: ~130 watts peak Power requirements: 100-120; 230-240 VAC, 50/60 Hz • 2 Amp Time Delay (SB 2A or T2A) at 100-120VAC line • 1 Amp Time Delay (T1A) at 220-240 VAC line 1.1.3 Physical Dimensions Weight: Dimensions: 8.1 Kg (18 lbs) 38 cm high (15 in) x 31 cm (12 in) wide x 38 cm (15 in) depth 1.1.4 Regulatory Compliance The analyzer is tested and acceptable to attach the CSA, GS (Europe), JIS (Japan), and CE (self declaration) safety marks, and compliant with EN1010-1, EN55011, and IEC 801-2. 1-1 1. Intro. 1.0 Introduction pHOx Service Manual 1.1.5 Location - Stable stationary surface. • • • 8 cm ( 3 in) clearance to a wall or adjacent instrumentation Not intended for use on vibrating (helicopter, etc.) surface Placement on a cart is acceptable if the cart is stationary during operation. 1.1.6 Operational The analyzer is designed to be on at all times. It may be turned off without special procedure or consideration for up to 1 hour with no detrimental effect on the sensors or fluidics. The tubing should be flushed and the pump tubing relaxed prior to extended shutdown periods. Any shutdown period will require a 2-point calibration and verification by control material upon restarting operation. 1.1.7 General The analyzer utilizes an Intel 386EX processor, PCMCIA ROM programming. There are three RS-232 connectors, utilizing ASTM protocol, and one Bar Code reader connection. The display is a backlit liquid crystal (LCD) with type that is CCFT backlit with 320 x 240 pixel resolution. The internal optional printer is a 20 column thermal international font /166 dot graphics printer with ~5 cm (2 in) wide paper. NOTE: 1-2 The circuit boards use fine pitch surface mount technology. Field repair of these circuit boards is not possible. DO NOT attempt to repair any components on the board. 1 Introduction 1. Intro. 1.2 Cautions and Hazards There are NO user serviceable assemblies inside the analyzer. Only a trained, authorized service representative should remove the cover of this analyzer. WARNING: Removal of the top cover allows access to power supply voltages. Care should be taken to avoid electrical shock. Remove the power cord prior to accessing any internal assemblies. WARNING: Blood samples and blood products are potential sources of hepatitis and other infectious agents. Handle all blood products and flow path components (waste-line, capillary adapter, probe, sensor module, etc.) with care. Gloves and protective clothing are recommended. NOTE: This International Caution Label appears on the rear of the pHOx Analyzer and means refer to the manual. It is the responsibility of the service representative to decontaminate any assembly or analyzer being returned to Nova for repair or warranty claim. Decontamination of external surface of the flow path may be accomplished by a wipe down with cleaning agent or 10% bleach solution. The Internal surfaces of the flow path and tubing should always be considered contaminated. Prior to removal from the analyzer, ensure that the flow path and tubing are emptied. WARNING: Internal surfaces may have sharp edges. Care should be taken to avoid cuts and scrapes when accessing internal assemblies. Used tubing, biosensors, electrodes, reagents, controls, etc. may be disposed by normal laboratory waste procedures. 1-3 pHOx Service Manual 1.3 Required tools The analyzer is designed to require no unique or special tools. Flow path cleaning wires/kits are provided in the accessory kit. Avoid metal wires or other material that may scratch the internal surfaces of the flow path. • Digital Volt Meter (2 decimal accuracy) • Phillips® screwdriver (crosshead) preferably one with 15 cm (6 in) shank • Spring clip expander wrench or comparable means of releasing tension of retention clips • Pliers - needle nose or small snub nose 1-4 2 Product Description 2.0 Product Description 2.1 Mechanical Assemblies The analyzer is designed to calibrate using calibration standards and reference solution contained in a reagent pack. The tubing automatically connects with the pHOx when the pack is placed into the analyzer. Likewise, a space is provided for a plug-in control pack. The content and serial number of the pack is coded in a “Dallas Semiconductor Add-Only Memory,” which is read by a One-Wire MicroLAN® system. The analyzer has 4 major mechanical assemblies to move calibrator or sample in front of the biosensors. Figure 2-1 Flow Diagram 2-1 2. Descrip. The pHOx analyzer accepts syringe or capillary samples through a sample probe. The operator selects the sample mode via the keypad. The probe moves from its home position in the analyzer to an extended 45° position for syringe sampling. If the capillary sample is selected, the probe moves to a horizontal position that is back inside the capillary adapter. pHOx Service Manual 2.1.1 Pump The analyzer uses one pump (PN 21492) for 2 functions. The pump aspirates either sample or calibration standards from the sample probe, through the sensor module/reference electrode, through a solenoid actuated pinch valve, and delivers to the waste bottle of the reagent pack. The second function of the pump is to aspirate the reference solution from the reagent pack and deliver it to the reference electrode through a solenoid actuated pinch valve. 2.1.2 Sampler Assembly The sampler assembly (PN 21504) utilizes a probe that rotates to 1 of 3 possible positions. The HOME position is vertical with the tip of the probe seated against a "Fluid Fountain” assembly that is the common port of the rotary valve. The syringe position is the probe tip extended out in front of the analyzer at approximately a 45° angle. The capillary position is horizontal, but the tip is totally inside the capillary adapter. This allows the capillary to be used without a separate adapter. (This position is also used to change the capillary adapter.) The capillary adapter is pushed onto the end of the sampler. Extending the probe outward from the capillary position ensures that the probe will be seated inside of the adapter. The sampler will exert enough force to push the adapter off of the sampler if it is not aligned correctly. 2.1.3 Rotary Valve Assembly The rotary valve assembly (PN 22215) has one common port connected to the sampler’s fluid fountain. Each calibration standard and quality control standard is connected to a port on the rotary valve. The rotor positions a central port connecting one of the fluids to the outlet port. The outlet port is connected to the Fluid Fountain by the S-Line. Starting the pump aspirates the selected fluid into the flow path. The rotary valve's manifold also has positions that allow air to enter the S-line. Refer to the replacement procedures for actual positions. 2-2 2 Product Description 2.1.4 Sensor Module Assembly 2.2 Electronic Assemblies. 2.2.1 Power Entry Module The power entry module (PN 22818) provides the interconnect to wall power. The fuses are found in the accessory kit for the operating voltage (120 VAC PN 22814, 236 VAC PN 22817). The power entry module connects to the power supply through TB1. 2.2.2 Power Supply The power supply (PN 21602) provides all of the power requirements for the analyzer. The analog and logic voltage used in the pHOx are + 5VDC. The stepper motors are driven by +24 VDC. This module has no repairable parts inside. The power supply connects to the digital control board by TB2. 2-3 2. Descrip. The Sensor Module Assembly contains the measuring sensors, flow cell, and preheater. The measuring sensors clip into place upon insertion. The sensors have contacts instead of cabled connectors. When mounted, the sensor module forces the sensor contacts to align with the matching contacts on the left/right interconnect boards. The module also contains the light sources and detector for the SO2% channel and 2 air detectors. Please refer to the drawing section and Appendix A for individual part numbers. pHOx Service Manual 2.2.3 Digital Control Board There are 2 Digital Control Boards: all units with monochromatic display use PN 24292, and all units with a color display use PN 40797. The main controlling board of the analyzer uses an Intel 386X processor with the following integrated peripheral functions: • Clock and Power Management unit • Watchdog Timer Circuit • Asynchronous Serial I/O unit • Synchronous Serial I/O unit • Parallel I/O unit • Refresh control unit • JTAG- Complaint Test Logic Unit The digital control board provides one 72 pin socket for a 4 MB SIMM module of RAM. This volatile memory stores the calibration, slope, and millivolt data. Major circuit identification: U25 is the main clock generator; it is crystal controlled 14.318 MHz and a phaselocked-loop frequency generator. This generates a 50.11 MHz CPU clock, 14.318 MHz for VGA video timing, and 1.84 MHz for Baud rate generation. U35 is the real time clock containing a battery which will retain the date/time, setup options, daily QC (up to 144 points), Error log (last 96 errors FIFO), and sample counter data. The battery has an expected 10 year life. U37 PEROM stores the monthly (256 data points) and YTD QC data, QC setup information, and recognition of the QC fluid pack presence. PCMCIA contains the operational program and boot program. This card must be in place for the unit to function. U43 is the video control chip. U26 controls the printer output and the I/O ports Comm 1 and Comm 2. U36 is the DUART that controls the Comm 3 I/O port. U16 provides the serial interface for the (RMS) interface and fluid pack memories. The mechanical devices are controlled by U18. The Keypad and “home” sensors for the rotary valve and sampler are ported through U18. The stepper motors are 4 phase stepper motors. U18 controls the speed and direction each phase is powered. U3 and U4 provide the power for each winding of the pump motor. U5 and U6 power each phase winding of the sampler motor. U7 and U8 power each phase of the rotary valve motor. U18 also provides the logic signal to operate the other mechanical devices. Typically, the devices are connected to either 12 VDC or 24 VDC and the logic provides a ground path through a transistor. The waste valve solenoid (12 VDC) grounds through Q3, the reference solenoid (12 VDC) through Q4, the sensor module lamp (24 VDC) through Q6, and the preheater (24 VDC) through Q2. The 24 VDC valve boost circuit is on the board but not used on the pHOx analyzer. 2-4 2 Product Description Power is received from the power supply through the J7 connector on the digital control board. Power Supply Connector TB2 5 Ground 6 Ground 7 +5 VDC 8 +5 VDC Digital Board Connector J7 Pin 1 2 3 Ground +24 VDC. +5 VDC 4 +5 VDC 5 6 7 8 Ground +12 VDC. Ground -12 VDC Sensing Wire Color Black Violet White Red Black Orange Green Yellow Individual circuits are protected from an over current condition by the use of Positive Temperature Coefficient Resistors (PTC). These resistors act as a direct short if the current is below a set current. If the current increases, the resistors heat up and open, similar to a fuse. When the device cools off, the connection is remade, allowing the circuit to function. The following lists these devices: R22 - Solenoid common boost circuit (not used on pHOx) R24 - Heater power circuit R29 - Printer motor power R43 - Print Head heater power R44 - +24 VDC R45 - +5 VDC R71 - Bar Code Reader (+5 VDC out) R249 - Distribution board power (This board is mounted in the door.) The LED indicators on this board are mostly unused. All will light at power up and then go off. The exception is D8 - 100 Hz Clock. D9 - 386 Strobe will appear to be always on. D7 is your program fail indicator and will come on with a software reset. Connectors: J1 Bar Code Reader J2 COMM 1 J3 COMM 2 J4 COMM 3 J5 JTAG J6 PCMCIA J7 Power Supply J8 Mechanical Assemblies J9 Printer Connector J10 Analog Board J11 - Display/Keypad (Motors/heater/lamp/solenoids and door) 2-5 2. Descrip. 1 2 3 4 +12 VDC. -12 VDC +24 VDC. Ground pHOx Service Manual 2.2.4 Analog Board There are 2 Analog Boards: pHOx, pHOx Basic, BiopHOx use PN 24291 and pHOx Plus, pHOx Plus L, and pHOx Plus C use PN 33390. U34 is the microprocessor that controls the signal acquisition and digital conversion. The signals are connected through several connectors. • Left Sensor Interface board J6 - PCO2, pH, Reference, Air Detector (ADT) 4 • Right Sensor Interface board J7 - PO2, Na+ • Sensor module J4 -SO2% detector, Preheater temp, ADT2, ADT3/Hct • Mechanical assemblies J3 - ADT1 Barometric pressure sensor U9 is on the analog board. The Analog to Digital Conversion (ADC) is performed by U31. The signals are multiplexed through U26 and U32. Preheater temperature control circuits are on the analog board. +24 VDC are ported through J3 to the SO2%/sensor module board (J4). The heater power wiring is protected on the digital control board by a PTC Resistor; the Heater has an in line thermostat (50°C) in the sensor module. The thermistor has 18.25 Kohms resistance at 37°C. The signal will change at a rate of 22.5 millivolts/0.1°C. There are 4 ADTs in the system. ADT1 holds the sample probe (40 uL setting). ADT2 at the inlet to the sensor module (70 uL). ADT3 is at the input to the sensors and is also used for the Hct. ADT4 is in the reference electrode housing. Logic and analog power is supplied to the analog board from the digital control board. Logic voltage (+5 VDC) enters J1 pin 14. The analog +5 VDC enter J1 pin 3 and -5 VDC J1 pin 4. The analog board provides power to the SO2 board through +5 VDC J4 pin 5 and -5 VDC J4 pin 6. 2.2.5 SO2 Control Board SO2 Control Board is an integral part of the Sensor module. It controls the 2 LED light sources that glow at 880 nM and 660 nM, respectively. 2-6 2 Product Description 2.2.6 Electrode Interface Boards The electrode interface boards are found only on pHOx and pHOx Basic. The Electrode Interface Boards have 2 part numbers: PN 20641 (Left) and PN 20674 (Right). The electrodes are wireless. The electrodes have point contacts which engage the Interface board contacts when inserted into the sensor module. The left interface board, connecting to the analog board through J6, has 3 sets of contacts: one set for the reference electrode/ADT4, one set for Na+, and one set for PCO2 sensors. The reference electrode/ADT4 has 3 individual point contacts while the Na+ and PCO2 have 2 individual point contacts. The right interface board has 2 sets of contacts: one for the pH and one for the PO2 sensors; each consisting of 2 individual point contacts. This board is connected through J7 to the analog board. 2.2.7 Door Assembly Electronics The door assembly electronics include the Display Assembly, the Distribution Board, and the Keypad Assembly. The digital control board sends/receives signals to the door via the flat cable J104 on the distribution board. NOTE: See Section 3.13 for part numbers that are model specific. The keypad, which is glued onto the door, connects to the distribution board via J102 and has a ground cable clamped to the conductive surface on the inside of the door. The Analyze key also connects to the distribution board through this connector. The distribution board also provides the signal path to the display assembly and power to the display back lighting lamp. The display is a liquid crystal, graphics display. It has a 320 x 240 dot matrix (1/4 VGA) yielding a 122 x 92 mm (4.99 x 1.33 inch) viewing area. This translates to a 40 x 20 character display using a standard character set. The display is backlit by cold cathode fluorescent tube (CCFT). An 8 ohm voice coil loudspeaker is mounted on the distribution board, producing all associated sounds. 2-7 2. Descrip. NOTE: pHOx Service Manual 2-8 3 Replacement Procedures 3.0 Replacement Procedures CAUTION: Assemblies may have loose hardware mounting. Ensure no loose hardware is left in the unit before restoring power to the analyzer. Cover Removal 1. 2. 3. 4. Turn the unit OFF, disconnect the line cord, and have the back of the unit facing you. R e m o ve t h e 4 Screws (2) rear-panel-cornermounting screws. Lift the printer cover and remove the 2 cover mounting screws. Slightly spread the bottom sides and slide the cover toward the back. This Mount will disengage the Screws (4) internal clips at the front of the cover. Lift the cover off the unit.Take care not to damage any cables or wiring during removal. Figure 3-1 Cover Removal 3-1 DWG #10-1029A 3.1 3. Replace. WARNING: Removal of the analyzer cover will allow access to line voltage wiring. Power should be turned off and the power cord disconnected before any assembly replacement procedure is performed. pHOx Service Manual Tubing Harness 1. 2. Remove the analyzer cover per Section 3.1. Remove the digital control shield. There are 4 mounting screws: One at the top front One at the bottom front Two at the bottom rear (above and below the PCMCIA card) NOTE: Take care not to lose the spacers on the screws. NOTE: It is optional to remove the digital control board and mount plate when accessing the reference and waste line. To remove the digital control board and mount plate, follow these steps: a. Disconnect the electrical connectors (J7, J8, J9, and J10) from the top of the board. b. Disconnect J11 connector at the bottom front of the board. c. The digital control board and mount plate come off as one assembly. There are 3 mounting screws: two on the rear panel and one at the top front on the printer box. (Take care not to damage the board.) d. Slightly lift the board and mount plate upward and out from the side of the analyzer. e. Remove the ground strap on the back of the board. Digital Control Board Mount Plate Mount Screws (3) (4) Digital Control Board DWG #10-1030A 3.2 Figure 3-2 Removing the Digital Control Board/Mount Plate 3. 3-2 Remove the R and W-lines from their front panel fitting. (Looking at the front panel, R is on the left.) If you have not removed the digital control board, it may be easier to remove the panel fitting nut on the front panel and move the fitting to the new tubing. 3 Replacement Procedures 4. Carefully remove the tubing from the rotary valve including the center port to fluid fountain. Take care not to damage the valve ports. Push the tubing end off the port to reduce the risk of damaging the valve. 3. Replace. G DW 0 0-1 #1 A 40 NOTE: Figure 3-3 Removing Tubing from Rotary Valve 6. 7. 8. 9. NOTE: The tubing bracket is held in place on each end by a screw. Two screws hold the bracket in alignment; access them from the bottom of the analyzer. Remove both screws. Lift the bracket and tubing set, removing it from the sampler side of the analyzer. Slide the new assembly into the analyzer and reconnect the W and Front Panel W and R Fittings R-tubing to the front panel ports. WR Ensure that they go to the correct ports and brackets. Reinstall the 2 mount screws at the bottom of the analyzer and the 2 end mount screws. Place the reagent tubing and S-line W onto their ports on the rotary valve. R Take care that the metal ports are not damaged. DWG #10-1039A 5. The A line connects at the lower left (7 O’clock position). Figure 3-4 Reconnect W and R-lines 10. If the control board and mount plate were removed, replace them. 11. Reinstall the cover and restore the unit to operation. 3-3 pHOx Service Manual 3.3 Power Supply Removal The Power Supply (PN 24720) is removed as follows: WARNING: Unplug the Analyzer before proceeding. 1. 2. 3. NOTE: 4. Remove the analyzer cover per Section 3.1. Disconnect the power harness connectors: P1 and P3 (top of the power supply assembly). Remove the 4 power supply mounting screws from the rear analyzer panel. Support the power supply before removing the mounting screw. Carefully lower the supply. Rotate it to access the line input module wiring (TB1) on the bottom of the power supply. P3 P1 Use Center Holes to Mount Rear Panel Power Supply Mount Screws (4) DWG #10-1038B Power Supply Power Plug On/Off Switch Figure 3-5 Power Supply Removal 5. 6. 3-4 Disconnect TB1 and remove the power supply. Position the new power supply with the lower side toward the analyzer. Plug in TB1. 3 Replacement Procedures NOTE: NOTE: NOTE: 9. The holes closest to the middle of the supply bracket are the ones used to mount the supply. Brown Connect P1 and P3. The wires leave the top of P1. The white wire of P3 goes to the digital display board P7. Black Green & Yellow TB1 Guide Power Supply Reinstall the unit cover and power cord. Figure 3-6 Power Supply Position 3.4 Sampler Assembly The Sampler Assembly (PN 23899) is removed as follows: 1. Remove the probe and the air detector. a. From the Menu screen, select Flowpath/Probe Maintenance and press Enter. b. Press Move Probe (soft key). Open the door. c. Remove the capillary adapter from the front of the probe by gently pulling it off. d. Disconnect the air detector's sample line from the sensor module. e. Disconnect the 2-prong cable. f. Push the air detector down and pull the air detector with the probe out of the sampler assembly. WARNING: Unplug the Analyzer before proceeding. 2. 3. NOTE: Remove the analyzer cover per Section 3.1 Cut the tie wraps on the rear of this assembly. Care should be taken to avoid cutting wires. 3-5 3. Replace. 8. Position the new power supply to align with the mounting holes in the back panel. DWG #10-1037A 7. TB1 male end has 3 wires. The female end has 5 sockets. The Green/Yellow wire is closest to you. pHOx Service Manual 4. 5. STOP: Disconnect the following connectors: J5 Motor cable connector J10 Home detector P6 door switch cable Sensor module light cable Ground wire Remove the S-Line from the fluid fountain assembly. There are washers and nuts that are NOT captive. Do not lose any hardware into the unit. Flow Cell Light Source Mount Screw Mount Screws Ground Power Supply S-Line J5, J10 DWG #10-1032B CN3 Figure 3-7 Sampler Assembly 3-6 3 Replacement Procedures 6. DWG #10-1034B NOTE: Remove sampler subassemblies. a. Motor assembly 1. From the Gear set side of the assembly, remove the flywheel (solid metal disk) by loosening the set screws on its side. 2. On the motor side remove the 2 mount screws. 3. Remove the motor (some Mount Screws tie wraps may need cutting). Install the new motor by reversing these steps. Use new tie wraps as required. b. Home detector (PN 21504) 1. Remove the sensor module light and bracket. Touching the light glass will shorten use life. DWG #10-1035A 8. Figure 3-9 Removal of Motor Assembly 3-7 3. Replace. 7. Remove the 4 mounting screws. Mount • One at the sensor modScrew ule light bracket to front panel • One at the lower front of bracket • Two at the rear of the sampler bracket, they Mount Screws also mount the rotary valve. If NOT replacing the entire assembly, skip to Step 8, otherwise: a. Install the sensor module light bracket onto the new sampler asMount Screw sembly. b. Position the new assembly onto the unit. Install the mounting screws and tie wraps. Figure 3-8 Removal of Sampler Assembly pHOx Service Manual 2. 3. 4. NOTE: Remove the 2 mounting screws from the home detector. Install the new detector by reversing these steps. Use new tie wraps as required to prevent fouling of the wires. After restoring power, ensure the capillary adapter seals against the Fluid fountain in the Home position. If not, adjust the Home flag on the top gear. Ensure the home flag enters the detector without obstruction. DWG #10-1045A C. Gear Set (PN 22634) 1. The gears are held in position by compression springs. These can be removed by spreading the open end and lifting. 2. Remove the old gears noting the position and orientation of Gear each gear. The noise Compression Spring damper is behind the middle gear. Spread Compression 3. After installing the new Spring to loosen gears, ensure that the Post gears run freely. Figure 3-10 Removal of the Compression Spring 3-8 3 Replacement Procedures DWG #10-1041A Figure 3-11 Fluid Fountain Assembly 3-9 3. Replace. D. Fluid Fountain Assembly (PN 24620) 1. Rotate the gears so the capillary adapter lifts off the fluid fountain assembly. 2 . Remove the S-Line from the bottom of the spike assembly. 3. Loosen the mounting nut (under the Capillary sampler bracket). Adapter 4. Slide the fluid fountain assembly off the sampler. 5. Move the washer and mount nut to the new assembly. Fluid 6. Slide the new assemFountain bly onto the sampler. 7. Rotate the sampler Sampler gears until the capillary presses down on the fluid fountain Washer assembly. Nut NOTE: The fluid fountain should be centered and should move S-Line downward between 0.06 to 0.08 inches (1.5 to 2.0 cm). pHOx Service Manual Rotary Valve Assembly The Rotary Valve Assembly (PN 23898) is removed as follows. 1. Remove the cover per Section 3.1. 2. Optional: Perform this step to enhance accessibility. Remove the sampler assembly per Section 3.4. 3. Remove the 7 tubes and SLine from the rotary valve. CAUTION: Care should be taken not to scratch the metal tubes of the valve. 4. 5. 6. NOTE: Mount Screws (2) Disconnect the motor connector J2 and the optical detector connector J9. Loosen the 2 mounting screws at the bottom of the assembly. Slide the assembly toward the front and lift. The front of the valve bracket is hook shaped. Slide the assembly toward the back of the unit to clear the sampler bracket. G DW #1 0-1 0 04 A 7. Flow Cell Light Source DWG #10-1036B 3.5 Figure 3-12 Removing Rotary Valve Assembly 8. NOTE: 3-10 If replacing the entire assembly, install the new rotary valve assembly by following Steps 1 thru 7 in reverse order. Ensure the red, nonmetallic washers for the sampler mount screws stay in place during the remounting process. If only repairing the rotary valve assembly, refer to Section 3.6. If repairing the rotary valve motor, refer to Section 3.7. 3 Replacement Procedures Valve Assembly The Valve Assembly (Valve Repair Kit PN 24626) is removed as follows: NOTE: The rotary valve assembly must be removed first before proceeding. (If not already removed, see Section 3.5.) 1. Remove “C” clamp on the valve shaft and remove the drive gear. CAUTION: Do not lose the shaft pin when removing the drive gear. With the drive gear removed, access and remove the 2 optical-detector mount screws. Remove the optical detector. Remove the 4 screws from the compression spring washer. Remove the washer and compression spring. Slide the rotor and tube manifold out of the valve body. 3. 4. 5. 6. NOTE: Before proceeding, note the alignment pin in the valve body that aligns with the slot in the manifold. NOTE: Before installing the new parts, lubricate with a small drop of sealing oil on the mating surfaces. 7. Slide the new rotor into the valve body. The large flange side goes in first. Align with the drive pins. B C A D S J9 DWG #10-1033A E G F Figure 3-13 Valve Assembly Removal 3-11 3. Replace. 2. DWG #10-1019A 3.6 pHOx Service Manual 8. 9. NOTE: Align the manifold with the pin and slide it into place. Reinstall the compression spring and washer Ensure the drive gear goes all the way over the shaft pin. 10. Install the “C” clamp on the drive gear shaft. 11. Install the valve by reversing Steps 1 - 10 above. 12. Install the rotary valve assembly. See Section 3.5 and reverse Steps 1 thru 7. 3.7 Rotary Valve Motor WARNING: Unplug the power from the analyzer. The Rotary Valve Motor (PN 24627) is removed as follows: NOTE: 1. 2. 3. 4. 5. The rotary valve assembly must be removed first before proceeding. (If not already removed, see Section 3.5.) Remove the drive gear/opto detector as in Steps 1 and 2 in Section 3.6. Remove the 4 mounting screws from the motor. Remove the fly wheel from the motor shaft byFlow loosening the Source Allen (hex) Cell Light screw. Mount the new motor and opto detector. Reinstall the rotary valve assembly. See Section 3.5 and reverse Steps 1 thru 7. Access Holes in Gear Optical Detector DWG #10-1026B DWG #10-1027A Optical Detector on Reverse Side Mounting Screws Figure 3-14 Optical Detector 3-12 3 Replacement Procedures Pump Assembly The Pump Assembly (PN 23897) is removed as follows: 1. Remove the analyzer cover per Section 3.1. 2. Remove the digital control board shield. There are 4 mounting screws: One at the top front One at the bottom front Two at the bottom rear (above and below the PCMCIA card) 3. Take care not to lose the spacers on the screws. Remove the digital control board and mount plate. a. Disconnect the electric connectors (J7, J8, J9, and J10) from the top of the board. b. Disconnect J11 connector at the bottom front of the board. c. The digital control board and mount plate come off as one assembly.There are 4 mounting plate screws: two WR on the rear panel and two under the printer box. NOTE: Take care not to damage the board. d. Slightly lift the mount plate and the board upward and out the side of the analyzer. Pump Assembly W R Figure 3-15 Pump Assembly 4. 5. 6. 7. Disconnect the motor connector J1. Remove the pump cap (on the front side of the pHOx) by pulling it off. Remove the 4 corner mounting screws then remove the pump. Install the new pump by reversing the above steps. 3-13 3. Replace. NOTE: DWG #10-1048A 3.8 pHOx Service Manual 3.9 Digital Control Board WARNING: Unplug the power from the analyzer. The Digital Control Board is removed as follows: 1. 2. Units manufactured after June 1, 2005, use PN 40797; Units manufactured through June 1, 2005, use PN 24292. Remove the analyzer cover per Section 3.1. Remove the digital control board shield. There are 4 mounting screws: -One at the top front, -One at the bottom front -Two at the bottom rear (above and below the PCMCIA card) NOTE: The shield is not present on units manufactured after June 1, 2005. Mount Screws (4) Shield DWG #10-1047A NOTE: Figure 3-16 Digital Board Removal 3. 4. Take care not to lose the spacers on the screws. At the top of the board, disconnect the electrical connectors: • J 7 - Power Supply • J 8 - Mechanical devices harness • J 9 - Printer (Pull side tabs outward to release the cable.) • J10 - Analog Board At the front lower corner disconnect J11 – Display/ PCMCIA keyboard (pull side tabs Release outward to release cable). Button Mount Screws Digital Control Board Figure 3-17 Install PCPMIC Card 3-14 Shield Stand Offs DWG #10-1046B NOTE: 3 Replacement Procedures 5. 6. 7. 3.10 Analog Board The Analog Board is removed as follows: 1. 2. 3. Basic/BiopHOx, pHOx/pHOx use PN 24291; pHOx Plus, Plus L,Plus C use PN 33390; all jumpers will be in the 3 to 4 position. To access the top/rear mounting screw, use a long shank (15 cm or 6 inch) Phillips® head (cross head) screwdriver. Remove the analyzer cover per Section 3.1. Disconnect the sampler ground strap. Disconnect the cable connectors from the Analog board Sensor Interface board connectors: J6 - 3 sensor board J7 - 2 sensor board J4 - Sensor module/SO2 cable J3 - RMS/Fluid Deck (Pull side tabs outward to release cable.) J1 - Digital control board J4 J6 J7 J3 Analog Board Mount Screws J1 DWG #10-1049B NOTE: Figure 3-18 Analog Board Removal 3-15 3. Replace. 8. 9. 10. 11. 12. 13. 14. 15. Remove the PCMCIA card (PN 22805) Push the button below the PCMCIA card (bottom rear corner) to release the card. Remove the 2 mounting screws and 2 shield standoffs: One screw is at the top rear corner One screw is between J8 and J9 connectors One shield standoff is at the top front and the other is at the bottom front Remove the 8 COM connectors screw standoffs at the rear of the pHOx by unscrewing 2 standoffs from each COM port. Remove the board. Position the new board on the mount plate. Install the 2 mounting screws and the 2 shield standoffs. Install the 8 COM connector screw standoffs. Reinstall the PCMCIA card into its slot. Reconnect all the cable connectors. Reinstall the digital board shield (4 screws). Reinstall the cover and restore the analyzer to operation. pHOx Service Manual 4. 5. Remove the 3 mounting screws: one in each corner of the board. You may need to place a screw driver through the holes in the power supply to access the screw in the upper back corner of the board. Install the analog board by reversing the above steps. 3.11 Printer Assembly WARNING: Unplug the power from the analyzer. DWG #10-1050A The Printer Assembly (PN 24615) is removed as follows: 1. Remove the analyzer cover per Section 3.1. 2. Remove the digital control board per Section 3.7. 3. Open the front door; remove the 2 screws on top of the printer cover plate; remove the printer cover plate by pushing the plate from the inside outward; then remove the paper and roller bar. 4. Remove the 3 print Printer Cover head mounting screws: two at the top of the print head 2 Screws and one at the lower right. 5. The flat ribbon cable from the print head sits in a connector. Slide the top of the connector upward to release the cable. The connector stays on the board. Tilt the print head from the assembly. Remove the paper advance motor cable that is connected into the printer control board. Figure 3-19 Printer Cover Removal 6. NOTE: 3-16 The Printer Control Board (PN 24614) is mounted to the paper box by 2 screws that fit into the slots. Loosen the screws and slide out the board. Save cable that goes from the printer control board to J9 of the digital control board. 3 Replacement Procedures 7. 8. NOTE: When installing the new assembly, install the printer control board first. Do not tighten the 2 slotted screws. This allows access when connecting the print head cable. Attach the paper advance motor cable to the printer control board. Install the print head. Connect the flat ribbon cable. The easiest way to plug in the print head cable is using one finger on each side. Tighten the 3 print head screws. The print head cable is clamped in place. Lift up on the white connector to release the flat cable for easy removal or replacement. 9. Use 2 Fingers to Connect Printer Cable to Printer Board DWG #1 0-1051A 3. Replace. After connecting the print head, tighten the printer control board mounting screws inside the paper box. 10. Attach the cable from the printer control board (key faces up, blue or black wire faces towards the front of unit) to the digital control board. 11. Reinstall the digital control board, shield, printer cover, analyzer cover, and paper. Restore unit to operation. Figure 3-20 Attaching the Printer Cable 3.12 Sensor Module Light Source Flowcell Light Source Mount Screws DWG #10-1052A The Sensor Module Light Source (PN 24717) is removed as follows: 1. Remove the analyzer cover per Section 3.1. 2. Remove the 2 screws of the light source mounting bracket. 3. The light bulb is removed by pushing it into the socket, rotating it counterclockwise a 1/4 turn, then releasing it. Figure 3-21 Removing Mounting Screws CAUTION: Do not touch the glass surface of the new light. 4. 5. 6. 7. Place the glass end of the new light bulb into the tubing provided. Insert the light bulb into the socket. Push it in and rotate it clockwise a 1/4 turn. Then pull off the tubing. Reinstall the bracket. Reinstall the covers and restore the unit to operation. 3-17 pHOx Service Manual Push and rotate 1/4 turn counterclockwise to unscrew light. DWG #10-1053B Push plastic tube over lamp . Do not touch glass of new light bulb. Push and rotate 1/4 turn to clockwise to install new bulb into socket Figure 3-22 Tubing for Light 3.13 Door Removal /Reinstallation Procedure CAUTION: Until free of the hinge and cables, support the door to prevent stress to other cables/ assemblies. 105 4B Use a pencil to 3 Mounting trace the position Screws of the hinge onto the door mounting plate. 10- STOP: Turn the unit OFF and disconnect the line cord. DW G# 1. Figure 3-23 Door Removal 3-18 3 Replacement Procedures NOTE: 4. Remove the 3 hinge screws from the door plate. Remove the 5 plate-mounting screws: one in each corner, plus one at right side. Care should be taken not to strip the threads. Keypad Display Shield 5 Plate Mounting Screws Door Distribution Board Figure 3-24 Door Removal Continued 5. Reinstall the Door by reversing the above steps. Use the pencil marks made on the door plate to realign the door to the hinge. Ensure the door opens and closes easily and aligns with the chassis. Analyzer pHOx pHOx Basic/ BiopHOx pHOx Plus pHOx Plus L pHOx Plus C PN of Door 21493 33102 30817 34082 35350 3-19 3. Replace. Remove the plate and disconnect the flat cable (J101). This frees the door. Place it face down on a protective surface to prevent scratching. DWG #101055B 2. 3. pHOx Service Manual 3.14 Display Assembly The Display Assembly is removed as follows: Prior to June 1, 2005 Display Assy. 24716 Door Distribution Board 24727 Digital Control Board 24292 Door Interface Cable 21983 1. 2. 3. 4. June 1, 2005 forward 40910 40795 40797 40495 Remove the analyzer door assembly per Section 3.13. Disconnect the cable from the display board (CN2). Disconnect the light source, J103, on the distribution board. Remove the 4 corner mounting screws. CAUTION: When reinstalling the mount screws, DO NOT overtighten. Assure the door plastic is not stripped. The screws will tighten in about one turn clockwise. 5. 6. Position the new display and install the 4 mounting screws. Reconnect the cables CN2 and J103. Reinstall the door per Section 3.13. CN2 Keypad J104 DWG #10-1065A J101 J102 J102 Light Source J103 Keypad Clip Ground Strip Figure 3-25 Display Removal 3-20 3 Replacement Procedures 3.15 Distribution board CAUTION: When reinstalling the mount screws, DO NOT overtighten. Assure the door plastic is not stripped. The screws will tighten in about one turn clockwise. 5. 6. Reconnect cables J104, J103, J102, and J101. Reinstall the door per Section 3.13. The Keypad Assembly (PN 24625) is removed as follows: 1. Remove the analyzer door per Section 3.13. 2. Disconnect the cable connector J102 from the distribution board. 3. Remove the clip holding the keypad ground strip to the door. Save this for reuse. 4. The Keypad is held by adhesive to the front panel. Peel the old assembly from the front of the door. 5. Clean the surface of any debris. 6. Remove the protective backing from the new keypad. 7. Slide the cable/ground strap through the hole in the door. Align and press new keypad in place. 8. Fold the ground strap over the ledge and reinstall the clip. Keypad Clip 9. Plug in J102 and reinstall the door per Ground Section 3.13. Strip DWG #10-1056A 3.16 Keypad Assembly Figure 3-26 Keypad Removal 3-21 3. Replace. The Distribution Board (PN 24727) is removed as follows: 1. Remove the analyzer door per section 3.13. 2. Disconnect the cable connectors J104 and J103 coming from the display board, J101 coming from the digital control board (should already be disconnected), and J102 coming from the Keypad. (Refer to Figure 3-25.) 3. Remove the 4 corner mounting screws. 4. Position the new distribution board (J101 will be on the hinge side of the door) and remount. pHOx Service Manual 3.17 Sensor Module Assembly WARNING: Unplug the power from the analyzer. NOTE: The preheater and SO2 component are all part of this assembly. R DWG #10-1059A The Sensor Module Assembly is removed as follows: 1. Remove the sensors (electrodes) and sample probe/ADT1 assemblies. 2. Remove the reference electrode assembly. 3. Rotate the retaining clamp on the top and the bottom of the assembly. 4. Pull the assembly toward you. 5. Remount by reversing these steps. W Figure 3-27 Removing the Sensor Module Sensor Module Part Numbers Analyzer pHOx pHOx Basic/BiopHOx pHOx Plus pHOx Plus L pHOx Plus C 3-22 Part Number 21494 34653 27807 34738 37343 3 Replacement Procedures 3.18 Electrode Interface Boards This section pertains to the pHOx, pHOx Basic, and BiopHOx Analyzers only. WARNING: Unplug the power from the analyzer. NOTE: 3 Mount Screws to Remove Electrode Interface Board Bracket DW G# 10-1 057 A The top left corner screw was removed with sampler removal. Figure 3-28 Removing the 3 Mount Screws 6. Pull the electrode interface board out thru the inside of the pHOx. 3-23 3. Replace. The Electrode Interface Boards (PN 24719 {left} and PN 24718 {right}) are removed as follows: 1. Remove the analyzer's cover per Section 3.13. 2. Remove the sampler assembly per Section 3.4. 3. Remove the sensor module assembly per Section 3.17. 4. Remove the spring loaded sensor module retaining clips (front of pHOx). 5. Remove the 3 remaining mount screws. pHOx Service Manual 7. Each interface board is held to the plate by 2 screws. Ensure that the 3 connector board is on left, and the 2 connector board is on the right when looking at it from the front. 3 Connector Board DWG #10-1058A 2 Connector Board Figure 3-29 Correct position of Interface Board 8. 9. Disconnect the affected board from the analog board: 3 connector board to J6, 2 connector board to J7. Install new assembly by reversing the above steps. 3.19 Pinch Valve Assemblies 0-1059A #10 Screw DWG #1 The Pinch Valve Assemblies (PN 16928) are interchangeable. The assembly plugs into the front panel and is held in by one screw. #10 Screw Figure 3-30 Pinch Valve Assemblies 3-24 4 Troubleshooting 4.0 Troubleshooting Information This section contains suggestions for individual error codes, cycle information to help identify the cause of an error condition, and hints on basic troubleshooting. 1. Troubleshooting is recognizing the variance from normal operation. The calibration cycle can be used as the sequence of events where the cycle response and results are known. Fluid Function all 4 ADTs, Hct, PCO2, Na+, K+, iCa, Cl-, Gluc, Lact PCO2 pH, Na+, K+, Cl-, Gluc, Lact, (iCa on pHOx +L) all ADTs, pH, Hct, (iCa on pHOx+C) 4. Trouble. Start cycle Std. A Std. B Std. C Std. D PO2 uses Room Air (A-air position of the rotary valve). SO2/Hb utilizes external calibrators. (These are model specific.) 2. Sensor Methodology ISE sensors - Na+, K+, iCa, Cl-, pH, and PCO2 ISE requiring a reference signal - Na+, K+, iCa, Cl-, pH ISE with its own internal reference - PCO2 Reflected Impedance - Air Detectors 1,2,4, 3/Hct Amperometric - PO2, Glucose, Lactate Optical Reflectance - SO2% 3. Calibrator Functions other than calibration. PO2 Std. B is used to flush the system. PO2 is monitored on all fluids in the Calibrator pack. If the PO2 millivolt reading varies from that of room air, a 72 Delta mV error is generated indicating a contaminated calibrator pack. Std. D has surfactant and is used to clean the flowpath between pump cycles. 4-1 pHOx Service Manual 4. Dependencies (Refer to the quick reference guide in the appendix.) Hct Calibrated Na+ required NOTE: The pHOx Na+ is a 30 day expendable electrode. SO2% Hb PCO2 Calibrated Hct and PO2 required Calibrated Hct and SO2% required Calibrated pH required Calculated values will not be reported if a needed test is suppressed. P50 is reported only if PO2 is between 30 mmHg (4 kPa) and 75 mmHg (10 kPa). 5. Password protection - One password for Set Up, QC, and Service menus. Should the password be lost, enter the following number when asked for the password. Date Code “Back-door” Password = (((month x 31) + day) x 137) + (year x 16)) + 5987 2 digits 2 digits last two digits Example: If today is 02 January 2003 the “back-door” password would be: Password = (((01 x 31) + 02) x 137) + (03 x 16)) + 5987 = (33 x 137) + 48 + 5987 = 4521 + 48 + 5987 = 10556 Change the password to a known value and write it down. 4-2 4 Troubleshooting 4.1 Service Screens Pressing the “MENU” soft key from the “Ready Screen” displays an “Operation Menu.” Note that the soft key labels change. The “QC” label changes to “Service” (second key from the left). Pressing this key will display the Service Menu. If the home key is pressed, the system returns to the “Ready Screen.” Manual Device Control Service Menu Use to select. Press to change / enter. System T est Active analog input signal (1 - 25) Analog Input - Real Time mV Calibration data Sensor Subsystem Printer enable, Error log, System configuration print Printer Menu Error log (Last 100 FIFO) I/O Signals / Loop Back Test Error Log Comm unications Tests Home Exit 4-3 4. Trouble. Manual Device Control + Flow test pHOx Service Manual 4.1.1 System Test Screen System Test Rev: 123.456 Measured Parameters/ Current signal (millivolt) Use to select. Press to select / enter. pH: +047.43 LED1: +314.3 Na: +147.4 AD3: +314.3 PCO2: -082.12 LED2: +314.3 AD4: +314.3 PO 2: -051.20 Device Names (column 1&3) / Current state of the device (column 2&4) 11-23-96 10:54 am AD1: +314.3 HCT: +007.43 AD2: +314.3 Rot. Valve: STD A Sampler: Pump: CTRL 1 SO 2% LEDs: Waste valve: Ref. valve: Home Open Closed Exit Temp: Home On Air Osc: On X level: Hct Operator Exit Flow Test 37.0 Service Flow Test Test Flow Manual Control of a mechanical device: • Move the cursor over the device /Press “ENTER.” • The Option selections will appear. (If there are only 2 options, the device will toggle its state when the “ENTER” key is pressed, i.e., solenoid valves.) • Move the cursor over the option wanted /Press “ENTER.” NOTE: To successfully move fluids, a knowledge of the complete flow path is required. 4.1.1.1 Operator Flow Test Pressing the soft key under the Operator Flow Test turns on the pump and raises the sample probe. Use the guide in this Section titled Flow Problems in Conjunction with the Operator Flow Test. 4-4 4 Troubleshooting 4.1.1.2 Service Flow test 4-5 4. Trouble. Pressing the soft key under the Service Flow Test brings up a screen that displays the results of this test. On this screen, pressing the soft key under Run Test causes the unit to pump each fluid twice: once with the Reference valve shut and once with the Reference valve Open. The expectation is that the flow time will increase with the decrease in vacuum caused by pumping reference solution into the waste line junction in the reference electrode. Ideally the 2 pump windings are equal and the flow time will double with the reference solenoid valve open. The pumping times will yield 2 pieces of information: 1. The individual flow times appear at the end of pumping that fluid. 2. All the fluids will pump at the same rate. Usually Standard D is the one standard that varies in flow rate from the other fluids. A good rule of thumb for the flow rates is pHOx Basic/BiopHOx and pHOx is in the 1000 – 1200 range. pHOx Plus/C is in the 1200 – 1300 range. pHOx Plus L pumps in the 1300 – 1400 range. pHOx Service Manual At the end of the Test, the Average flow time for Reference Valve Closed and Reference Valve Open is displayed. A ratio of the average times is presented. The ratio should fall between 0.5 – 0.65. • Ratios approaching 1 indicate the loss of reference flow. • Ratios approaching 0.0 indicate an obstruction or air leak in the system before the flowpath reaches the Reference electrode. FLOW TEST RESULTS CAL A 1180 CAL B 2010 CAL C CAL D 0.58 QC 1 QC 2 1190 AVG FLOW 1 1195 AVG FLOW 2 1190 1170 RATIO Average Flow time Reference Valve Shut Average Flow time Reference Valve Open Flow 1/Flow 2 (nominal 0.5 – 0.65) 1195 1193 QC 3 1195 Print EXIT Run Test 4.1.2 Analog Input Screen Selecting this option, displays the 25 possible input channels. The display is active. Chnl Signal Unit Chnl Signal Unit PO2 pHOx+, pHOx+ C 1 Analog GND ALL 15 + 2 K pHOx+,+L,+C 15 Lac pHOx+L pHOx+,+L 16 Glu pHOx+,+L,+C 3 Cl-/Ca++ 3 ClpHOx+C 17 PO2 pHOx+L ++ 4 Ca pHOx+C 18 PO2 pHOx+,+C 5 pH pHOx 18 Lac pHOX+ L pHOx+,+L,+C 20 Temp 0.00=37C 5 Na+ pHOx 22 LED1 ALL 6 Na+ 6 pH pHOx+,+L,+C 23 LED2 ALL 7 PCO2 All 24 ISE Ref ALL PCO2,Ref ALL 8 Barr.Press. All 25 9 +5V Ref All 13 HCT pHOx, pHOx+C 14 PO2 pHOx, pHOx+C 10 Voltage Monitor Sums +24 and +5 4-6 4 Troubleshooting 4.1.3 Sensor Subsystem This set of screens shows the calibration data for each analyte measured by the pHOx. Use the “NEXT” soft key to page to the analyte of interest. Last 2-point calibration data Millivolts for each standard Slope calculated Last 1 point data (Millivolts should be similar to the same Standard in 2 point cal) pH Sensor Callibration & Analysis Data mV mV STDA + 43.1 STDA STDB + 43.1 SAMPLE STDC + 43.1 CONC STDD + 43.1 SLOPE + 43.1 + 43.1 97.4 - 7.1 Next Page Exit Print Sample millivolts and concentration - Unknown and therefore not relevant to troubleshooting Useful to the clinician for any possible dilution SO2 LED 1 SO2 sensors - Each LED has its own screen. Displays similar data to the other Measured parameters, plus The “dark” (off) and “Light” (full) Signals are shown Callibration & Analysis Data STD1 mV + 379.1 SAMPLE STD2 + 379.1 CONC DARK + 379.1 LIGHT + 379.1 SLOPE +11.4 Home Print Exit AIR + 117.0 FLUID THRESHOLD AD1 + 54.0 + 99.0 mV + 379.1 Next Page Air Detector Subsystems Air Detectors Displays the millivolts for Air and Liquid. (Air column is actually Standard D.) Shows the AIR/Liquid threshold (set by the analyzer) millivolt level NOTE: Numbers shown on this screen are not real values. Callibration Data AD2 + 117.0 + 54.0 + 99.0 AD3 + 117.0 + 54.0 + 99.0 AD4 + 117.0 + 54.0 + 99.0 Exit Next Page Home Print 4-7 4. Trouble. Home pHOx Service Manual 4.1.4 Printer Menu This menu is straight forward. Move the cursor to the desired action and press “ENTER” Printer Menu Use to select. Press to change / enter. Enable/Disable internal printer Printer: Enabled Prints last errors Print System Error Log Print System Dump Prints setup configuration and millivolt data Character Set Test Test Print Home 4-8 Cancel Printout Exit 4 Troubleshooting 4.1.5 Communication Tests This screen allows you to transmit out a particular port for testing. Likewise using a test plug you can perform a Loopback test for each I/O function. Communications Tests Use to select. Press to change / enter. Loop Back Tests Send test characters. 4. Trouble. Home Exit Loopback Tests Use to select. Press to change / enter. COOX 1. Select Port and install loopback plug. ASTM 2. Press to begin testing. Remote Tests Completed: 0 Bar Wand RXD DTR RTS TXD DSR DCD CTS RI Errors Home 0 0 0 0 0 Exit Loopback Tests Use to select. Press to change / enter. COOX 1. Select Port and install loopback plug. ASTM 2. Press to begin testing. Remote Tests Completed: 0 Bar Wand RXD DTR RTS TXD DSR DCD CTS RI Errors Home 0 0 0 0 0 Exit Pressing the “HOME” key returns to the READY screen, or EXIT to go back one screen. 4-9 pHOx Service Manual 4.2 Flow Problems NOTE: Each of the following troubleshooting procedures has an associated flow diagram that may be found at the end of this chapter. General procedures when isolating flow problems: NOTE: Refer to the FLOW TEST decision tree. Do not be confused with the Flow Test Cycle that only shows the flow times. Unit calibrates, but the sample will not aspirate. This usually means you have an air leak into the system. The sample, being more viscous than the aqueous standards, causes more vacuum to be created by the roller pump. If a membrane is leaking, this increase in vacuum will pull the aqueous fill solution from the electrode, easier than pulling sample from the probe. Likewise, a poor connection at the interconnect tubing or between the electrode sealing washer and the sensor module will allow air to enter easier than pulling sample up the flow path. • Check the connections between probe and sensor module, sensor module, and reference electrode. • Check the electrodes for the loss of internal fill solution. • Check the flow cell window during a pump cycle, looking for bubbles entering the flow path or pump and waste tubing. A No flow or obstructed flow cause can be isolated by performing the FLOW TEST. The fluidic system can be isolated into three large sections. This will limit the area of your search. The 3 sections are Fluid Pack to Fluid Fountain, Sampler/Sensor module, and W/R to Waste tubing. Use the SYSTEM TEST screens to control the mechanical devices. 1. 2. 3. 4. 5. NOTE: 4-10 Extend the sampler to the syringe position. Open the waste line solenoid. Turn on the pump and attempt to aspirate water then air from a sample cup. If water is aspirated from the cup, the sample probe, sensor module, external tubing, and waste line are functioning. The problem exists between the Fluid Fountain and Fluid Pack. If water is NOT aspirated from the cup, remove the tubing from the outlet port of the Reference electrode. Place it in a cup of water. This assumes the waste solenoid has functioned and is not pinching the tubing. 4 Troubleshooting 6. 7. If step 5 fails to aspirate from the cup, the problem lies between the W/R tubing and the waste bottle of the Fluid Pack. If Step 5 does aspirate water, the problem is in the Sampler/ Sensor module. 4.2.1 Isolating a Plug in the W/R to Waste Tubing NOTE: 1. 3. 4. 5. Disconnect the waste tubing from the reference electrode and place in a cup of water. If water is NOT aspirated, there is a plug. Lift the waste tubing out of the waste solenoid. If flow starts, the valve is dirty. Clean the dirt from the waste solenoid or replace it. If flow does NOT start, remove the waste line from the front panel connection. If water flows from this tube there is a plug in the internal waste tubing or waste bag of the fluid pack. Turn off the pump. Connect a syringe of water to the waste port on the front panel. Attempt to inject water through the line to clear the plug. If there is no flow, remove the Fluid Pack to ensure the problem is not a poorly seated or damaged waste connection. Should the problem continue, replace the internal tubing set. 4.2.2 Isolating a flow problem in the Sampler/Sensor module 1. 2. 3. 4. NOTE: You noted flow when aspirating from the waste tubing in step 2 of Section 4.2.3. Raise the Sampler to the Syringe position. The Pump should be OFF for this test. Connect a syringe of water to the reference electrode outlet port. Attempt to inject the water. If water flows freely, air is entering the system preventing aspiration. a. Inspect each electrode to assure the internal fill solution level is not dropping. b. If yes, replace the electrode or membrane as required. c. Inspect the seal between the reference electrode and sensor module (interconnect tubing) . If the flow path starts to empty while the syringe is still attached, only the water below the leak will leave the flow path out of the tip of the probe. 4-11 4. Trouble. 2. If the reference flow is obstructed, the sample may not stop in the sensor module due to increased vacuum. Flow time decreases due to increased flow rate. pHOx Service Manual d. 5. Inspect the sensor sealing washers for distortion or misalignment. Replace as necessary. e. The seal between the air detector and the sample probe may be tested by attaching the syringe to the sample probe, blocking the air detector outlet tube. Apply pressure to the syringe and look for water leaking out. If water does not flow freely, there is a blockage. a. Loosen the reference electrode. Water should pass through or the electrode is plugged. Flushing with cleaning agent or hot water may free the blockage. b. If the reference is clear, reinstall it and remove the bottom electrode. c. Apply a moderate pressure with the syringe. d. If water does not enter the sensor module of the bottom electrode the blockage is above it. e. Removing one electrode at a time. Look for water entering the sensor module. If no water enters the module as the electrode is removed, the plug is above it. f. If water did enter the bottom sensor module in step C above, the plug is in the Sample Probe/Air detector. Remove these assemblies and flush through them or replace them. 4.2.3 Isolating a flow problem between the Fluid Pack and the Fluid Fountain If you have successfully aspirated a sample, but the analyzer will not calibrate or aspirate on-board controls, proceed as follows. 1. Position the sampler in the syringe position and inspect the capillary adapter for signs of wear. Replace if necessary. 2. Lower the sampler to the Home position; connect a syringe to the reference outlet port. 3. Remove the analyzer cover. 4. Disconnect the S-Line (center tube) from the rotary valve. 5. Inject the water and look for flow from the S-Line. 6. Look for a leak between the fluid fountain and the capillary adapter. Realign if necessary. 7. If there is no flow from the S-Line, remove the fluid fountain/S-Line and flush with hot water or cleaning agent, or replace the assembly. 8. If there is free flow through the S-Line, the problem is in the rotary valve, tubing harness, or connection to the fluid pack. 9. Reconnect the S-Line and remove the fluid pack. 10. Move the rotary valve to all the possible positions, observing which ports do not allow flow. 11. If there is no flow in ALL positions, remove the valve and clean or replace it. 4-12 4 Troubleshooting 12. Remove the tubing for each position that does not flow freely. 13. If water can be flushed through the individual lines, remove and clean or replace the valve. 14. If water cannot be flushed through the affected lines, replace the tubing harness. 4.3 Status Codes Error codes come in 2-digit groups. The tens (first) number, called the “series code,” identifies the affected component. Series 1 through 6 are sensors, 0 = pH 1 = PCO2 2 = PO2 3 = SO2 % 4 = Hb 5 = Hct 6 = Na+ Series Digit code -1 -2 -3 -4 -5 -6 -7 1-6 7 8 Slope No Std.A No QC1 Instability No Std.B No QC2 Overload No Std.C No QC3 Drift No Std.D Dependency Sample flow Air flow pH Status Codes Code Status Screen Text 01 pH Slope 02 pH Instability 03 pH Overload 04 pH Drift 96 Hardware 97 Software 9 10 Barometer Not assigned Printer Flow Cell Temp Communication Digital Board Analog Board (ADX) Software Error Log Text pH Slope pH Instability pH Overload pH Drift pH ADC pH Math 4-13 4. Trouble. Series 7 = flow, 8 = Q.C. 9 = Electronic/Mechanical/Hardware 10 = CO-Oximeter codes. The ones (second) number, called the “digit code,” provides more information on the failure. pHOx Service Manual PCO2 Status Codes Code 11 12 13 14 15 96 97 Status Screen Text PCO2 Slope PCO2 Instability PCO2 Overload PCO2 Drift PCO2 Dependency Hardware Software PO2 Status Codes Code 21 22 23 24 96 97 Status Screen Text PO2 Slope PO2 Instability PO2 Overload PO2 Drift Hardware Software SO2 LED 1 Status Codes Code 31 32 33 34 35 96 97 Status Screen Text SO2 % Slope SO2 % Instability SO2 % Overload SO2 % Drift SO2 % Dependency Hardware Software SO2 LED 2 Status Codes Code 31 32 33 34 96 97 Status Screen Text SO2 % Slope SO2 % Instability SO2 % Overload SO2 % Drift Hardware Software Hb Status Codes Code Status Screen Text 41 Hb Slope 45 Hb Dependency 97 Software 4-14 Error Log Text PCO2 Slope PCO2 Instability PCO2 Overload PCO2 Drift PCO2 Dependency PCO2 ADC PCO2 Math Error Log Text PO2 Slope PO2 Instability PO2 Overload PO2 Drift PO2 ADC PO2 Math Error Log Text SO2 % LED1 SO2 % LED1 SO2 % LED1 SO2 % LED1 SO2 % Dependency SO2 % LED1 ADC SO2 % LED1 Math Error Log Text SO2 % LED2 SO2 % LED2 SO2 % LED2 SO2 % LED2 SO2 % LED2 ADC SO2 % LED2 Math Error Log Text Hb Slope Hb Dependency Hb Math 4 Troubleshooting Error Log Text Hct Hct Hct Hct Hct Hct ADC Hct Math Na+ Status Codes Code Status Screen Text 61 Na Slope 62 Na Instability 63 Na Overload 64 Na Drift 96 Hardware 97 Software Error Log Text Na Slope Na Instability Na Overload Na Drift Na ADC Na Math AD1 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software Error Log Text AD1 AD1 AD1 AD1 ADC AD1 Math AD2 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software Error Log Text AD2 AD2 AD2 AD2 ADC AD2 Math AD3 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software Error Log Text AD3 AD3 AD3 AD3 ADC AD3 Math 4. Trouble. Hct Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 54 Hct/Air Det Drift 55 Hct/Air Det Dependency 96 Hardware 97 Software 4-15 pHOx Service Manual 4-16 AD4 Status Codes Code Status Screen Text 51 Hct/Air Det Slope 52 Hct/Air Det Instability 53 Hct/Air Det Overload 96 Hardware 97 Software Error Log Text AD4 AD4 AD4 AD4 ADC AD4 Math Flow Status Codes Code Status Screen Text 71 StdA Flow 72 StdB Flow 73 StdC Flow 74 StdD Flow 75 Flowtime 76 Sample Flow 77 Air Flow 78 Back Flow Error Log Text StdA Flow StdB Flow StdC Flow StdD Flow Flowtime Sample Flow Air Flow Back Flow Control Status Codes Code Status Screen Text 81 Ctrl 1 Flow 82 Ctrl 2 Flow 83 Ctrl 3 Flow Error Log Text Ctrl 1 Flow Ctrl 2 Flow Ctrl 3 Flow System Status Codes Code Status Screen Text 91 Barometer 92 Printer 93 Temperature 94 Communication 95 Hardware 95 Hardware 95 Hardware 96 Hardware 97 Software 99 Scheduled QC Not Run Error Log Text Barometer Printer Temperature Communication PEROM Write PEROM Read PEROM CRC Hardware Software Scheduled QC Not Run RMS Status Codes Code Status Screen Text 688 N/A 690 N/A 691 N/A 693 N/A 694 N/A Error Log Text Software (Pack Not Present) Software (Pack Empty) Software (Pack Lot Date Expired) Software (Pack Use Life Expired) Communication Error 4 Troubleshooting CMS Status Codes Code Status Screen Text 696 N/A 698 N/A 699 N/A 701 N/A 702 N/A Error Log Text Software (Pack Not Present) Software (Pack Empty) Software (Pack Lot Date Expired) Software (Pack Use Life Expired) Communication Error Standards Information: Sensor Hct& ADET’s A&D 4. Trouble. Standards Used Delta mV Calibration To calibrate (between standards) _______________________________________________________ pH C&D 33 + Na (pHOx) A&D 17 (pHOx+/C/L) A&C 16 PO2 Electronic 0.00 / Room Air, Std A used for 1 pt Ca++ PCO2 A&B 17 Insert sheet shows HCO3- value. HCO3 x 1.5=CO2 in mmHg (all ADET’s) > 500 Hct ideal > 800 (Hb) Based on SO2 & Hct (mV not shown) SO2 External Stds 1 & 2 n/a + K A&C 24 iCa A&D 12 Cl A&C 15 Gluc A&C 38-55 Lact A&C 35-50 _______________________________________________________ 4-17 pHOx Service Manual 4.3.1 Analyte Troubleshooting ( Series code 1-6 and Digit codes 1-4) Multiple Channel failures: Slope • Ensure the Reagent Pack is fully plugged in. • The pack box edge closest to you should lie inside the small lip of the chassis. • Check the flow times displayed on the service screens. • Inspect the Reference tubing to assure flow. • Inspect the waste line for bubbles. • Prime the system, check each fluid for air. • pH and Na+ require a working reference electrode. Figure 4-1 Reagent Pack Placement • • • • • Note if the channels that fail use the same standard from the fluid pack. Loss of the +5 Vdc on the analog board Loss of the Analog board A/D/ converter Loss of +5 Vdc to the Analog Board Analog Board over heated (PTC resister open) - If unit is allowed to cool off by shutting it off for 5 minutes. Investigate why the unit or Analog board is overheating. Dependency Rules • Hct requires calibrated Na+ for calibration and results. • PCO2 requires calibrated pH for results (pH checked on standards to assure bicarbonate is correct). • SO2 requires calibrated Hct and PO2 for results. Calculated SO2% is reported if measured SO2 is not available. • Hb requires Hct and SO2 for reporting. Priority: 1. If linked with COOX, the COOX HGB is reported. 2. If no COOX link, the measured result is reported, if available. 3. If no COOX or measured result, the calculated result from the Hct reading (Hct/3) is reported. 4. If no COOX, no measured, and no Hct, the default value (Hb) is reported. 4-18 4 Troubleshooting • • • • • • • • Qsp/Qt requires 2 separate blood samples for determination: mixed venous and arterial. RI utilizes input or default FIO2% for calculation. Calculated Results require calibrated and proper results by measured tests used in the calculation. Suppression rules - If measured test is suppressed in the Setup, the calculated results based on test also will not be shown or printed. Password (One) is used for Setup, Service Menu, and QC. (No user passwords are available.) Interfaces are NOVA COOX, PDM, and ASTM. Standby Mode: There is no time dependency. Analyzer will prime and calibrate after coming out of standby. Idle pumping is active, but no fluidic or gas calibrations are performed. Features Auto-Wake-Up, Auto-Cal, and Auto-QC Micro Sample Analysis: Micro-Mode Tests: pH, PCO2, and PO2 Stepped Analysis: Stage 1 = Tip of probe to ADT1 (40 µL) Stage 2 = Up to ADT3 Stage 3 = Between ADT3 (up to ADT4): Other Tests Reading (including contact with reference electrode) Instability • Inspect sensor module for bubbles during the measurement cycle • Ensure that the waste valve stops the fluid flow during analysis. Fill the flow path with any standard, raise the probe to the syringe position Look for droplets forming at the sampler tip. • Ensure that the waste and Reference solenoid valve are functioning. • Fill the flow path with any standard and view the system test screen. The signals should be steady and there is no flow of fluid there are three Possible conditions, a loose sensor module, an analog board failure, or a poor connection between a set of electrodes and an interconnect board. • Inspect the electrode washers to assure proper alignment. Nonalignment may cause a flow problem. • In rare cases the tubing harness may have become contaminated. Using the cleaning adapter, flush the system with cleaning agent followed by distilled water. 4-19 4. Trouble. • P50 is measured and is reported only when PO2 is between 30-75 mmHg. P50c is reported when PO2 is outside those limits. P50 also requires calibrated SO2% and Hb/Hct for reporting. pHOx Service Manual 4.3.2 Single Channel Failures Slope View the Sensor Subsystem screens. If the change in millivolts between standards is 0 or less than that mentioned in Section 4.3.1, there are several possibilities: • Membrane or sensor has reached the end of its use life. • Bubble of Air is covering the sensing material at the electrode tip. • The electrode is not making good contact with the Interconnect board contacts. • Loss of the Analog board multiplexer will cause a zero input to the A/D converter. • Ensure calibration on required channels: Sodium is required for Hct, Hct is required for SO2%. • Viewing the Sensor subsystem screens you note an overload condition (999.9 mv). • Check for NO connection to the Interconnect board. • Check the PO2 /PCO2 for loss of the Internal Fill solution (membrane rupture). • Loss of the multiplexer or individual channel circuit on the Analog board. • The pH electrode may need conditioning. • Na+ sensor over 30 days of use. The Na+ sensor will be destroyed by conditioning. Instability Single channel instability is seldom a flow problem unless a standard is unique to a particular channel. • De-bubble the electrodes. • Ensure a good connection between the electrode and the Interconnect Board. • Ensure a good connection between the sensor module and Chassis. • Ensure the analyzer's door is closed and the ground strap is making good connection. 4-20 4 Troubleshooting 4.3.3 Temperature Problems • • • • • Ensure that the room ambient temperature is within specification ( 18o - 32o C). Ensure that the sensor module is plugged in correctly. View the Analog Input screen channel 20. This channel will be a 0 if the module is at 37o C and change at a rate of 22.5 mv/1oC, + if above and - if below 37o C. The power for the preheater comes from the digital control board. The heater control circuitry is on the analog board. 4.3.4 Air Detector (ADT) Problems • • • NOTE: • • Air Detectors are only calibrated during a 2-point calibration. Typically the Fluid /Air millivolt difference will be about 700. Check the sample probe for air being introduced below the sensor module. Check the sensor module for air being introduced at the electrode/sensor module seal. Ensure that the electrode sealing washers are centered and seated correctly. The Air detector signals are processed by the analog board. They are only displayed on the System Test screen. Alternately aspirate air and then fluid to see the change in the signal. Remember to turn on the air oscillator to obtain an ADT reading. Flow Test During a flow test, the rotary valve rotates to each fluid position. The pump turns on to fill the lines from the rotary valve S-line port to the ADT1. The rotary valve rotates to an air position. The pump turns on and the time in milliseconds until the ADT senses air is recorded. The time is normally 1200 msec. The acceptable range is 100 to 1700 milliseconds. Run a Flow Test; press the soft key on the Service screen. Typical result for Standard A, B, C, D is 1300. If all are high (3000-5000), there is a flow path obstruction. If only one is high, 1. Check for correct fluid pack installation. 2. Check that the valve is in the correct position. 3. Connect a syringe to the S-line port of the valve. Remove the reagent pack and move the valve to each position. Verify that the link is clear by purging air through the S-port. 4-21 4. Trouble. • • • pHOx Service Manual 4-22 4 Troubleshooting 4. Trouble. 4-23 pHOx Service Manual Inspect for air leak. Loss of IFS, Replace or use plasClear the plug by flushbad washer, ruptured membrane, Replace or or interconnect tubing seal. backflush the tic cleaning line to ing plastic line or reclear the plug. place the module. 4-24 air detector. 4 Troubleshooting 4. Trouble. 4-25 pHOx Service Manual ������������������ ������������������������������������������������� � ��� �������������������������������������������� � � ��������������������������������������������������������������� � ������������������������������������������� ������������������������������������������������ � ��� ���������������������������������������������� � ���� �������������������� � ��� �������������������������������������������������������� ���� ������������������������������������������������������ �������������������������������������������� � ��� ������������������������������������������������ ���� � �������������������������������������������������������� �� � ������������������� � ���� ��������������������� ��� ������������������������������������������������������������� � � �������������������������������������������������������������� � ����������������������������������������������������������� ����������������������������������������������������������������� � ��� ��������������������� � ���� �������������������������������������������������� � ��� ���������������������������������������������� ��� ������������������������������������������������ ��� ����������������������������� �� ����������������������������������������������������� ��� ����������������������������������������������� ��� ��������������������������������������������������� ������� ����������������������������������������� ��� �������������������������������������������������� � ������������������������������������������������������������ � ����������������������� ������������������������������������������������������������� �� ����������������������������������������������������������������� � ������������������������������������������������������������� � ��������������������������������������������������������� � � � 4-26 4 Troubleshooting ���������������������������������� ��� � ������������������������������������������������ � ����������������������������������������������� � ��� � ������������������������������������������� � ���������������������������������������������������������������� � ��������������������������������� � ������������������������������������������������������� ��� ��� ������������������������ � ������������������������������������������������ ���� ��������������������������������������������������������� ��������������������������������� � ������������������������������������������������������� ���� ��������������������������������������������������������� �������������������������������������������������������� ����������������������������� � �� 4. Trouble. �� �� ���� ��������������������������������������� � ������������������������������������������������������� � �������������������������������� ������������������������������������������������������������ ���������������������������������������������������������� �� ��� �������������������������������������������������������� �� �������� �� ���� �������������������������������������������������������� � � � ������������������������������������������ ��� ����������������������� ���������������������������� ���� ���������������������������������������������� �� � ��������������������������������� ������������������������������������������������ �������������������������������������������������������� �������������������������������������������������������� ����������������������������������������������� �� �� �������������������������������������������������������������� �� �������������������������������������������������������� ������������������������������������������������������������ �������������������������������������������������������� �������������������������������������������������� �������������������������������������������������������� �� ���������������������������������������������������������� ���������������������������������������������������� ������������������������������������������������������������� ����������������������������������������������������������� �������������������������������������������������������� ��������� �������������������������������������������� ������������������������������������������������ �� 4-27 pHOx Service Manual 4-28 Appendix A Appendix A.1 Quick Reference Guide Measured Tests: pH, PO2, PCO2, SO2%, Hb/Hct, (Na+ is not reported) Alveolar Oxygen Concentration Arterial Alveolar Oxygen Tension Ratio Arterial Alveolar Oxygen Base Excess, Extracellular Fluid Base Excess, Blood Actual Bicarbonate Concentration Oxygen Content P50 Calculated (when PO2 is <30 or >75) Physiologic Shunt Respiratory Index Standard Bicarbonate Oxygen Saturation % (Calculated) Total Carbon Dioxide pH/PCO2/PO2 (Temperature Patient Corrected, if selected) Slope Limits: pH 9.1 to 11.6 (1,7) PO2 -15.0 to -1.6 (1,2,4,8) Na+ 8.8 to 11.5 (1,7,8,9) Electrode Notes: 1 = Long Life Sensor 2 = Int’l Ref Electrode 3 = Membrane Cap; Push-On 4 = Membrane Cap; Screw-On 5 = Reflectance PCO2 7.9 to 12.6 SO2% 6.9 to 18.2 Hct: 12.0 to 50. Appendix Calculated Tests: • A • a/A • A-aDO2 • BEecf • BEblood • HCO3• O2Ctc • P50c • Qsp/Qt • RI • SBC • SO2%c • TCO2 • Temp (1,2,4,8) (1,5,8) (1,6,7) 6 = Impedance 7 = No Washer 8 = Internal Washer 9 = Non-Reported Test 10= Obtained from SO2% and Hct A-1 pHOx Service Manual Flow Limits: Flow Test (Service Menu): Typical/Ideal Value = 1300 (milliseconds) ** High Values (ie: > 3000) or “ERROR” note on Std Channel: Indicates major flow blockage. Offset Limits (Slope/Intercept): Test Y-Intercept Range pH +/- 100 mpH H+ NONE PCO2 +/- 10 mm Hg PO2 +/- 20 mm Hg SO2 % NONE Hct +/- 10.0 % Hb +/- 2.0 g/dl Slope Range NONE 0.8 to 1.2 0.8 to 1.2 0.9 to 1.1 0.9 to 1.1 0.8 to 1.2 0.9 to 1.1 Flow Cell Positions: Top Adet 1 Adet 2 Adet 3 Adet 4 Reference Biosensor Located above Sample Probe Na+ Used to detect 40 uL MicroSample In Sensor Mod/Flow Cell pCO2 Used to detect 70 uL sample Reads Hct, then blood pulled up and air is seen telling stop for electrode sample sO2 reading; sample position Positions top of sample for electrode readings Reference ADET 4 pH pO2 ADET 3 ADET 2 Calibration Details (Analysis Time/Throughput): Two Point Calibration: User Selects (not auto control), 2,4,6 hour intervals (turned off in Standby Mode). One Point Cal: User selects for 30 or 45 minutes (same as STP/UL), One point cal for ALL (gases plus pH/Hct), One point SO2 done with each full 2-point calibration. Standard One Point Cal: Done with blood samples: Single Mode (each sample); NOT available with product release; will be for low volume users only. A-2 Appendix Throughput specs subject to change without notice due to software sequence changes. Throughput Details: Mode A = 44 sec.; Mode B = 64 sec (time to view results) Mode A = 62 sec; Mode B = 74 sec (full cycle) Approx 50/Hour (varies by mode) Analysis (view) = 43 sec; Run Time = 65 sec. pHOx+C: Analysis Cycle: Run Time: Throughput: Micro Mode: Mode A = 47 sec.; Mode B = 100 sec (time to view results) Mode A = 65 sec; Mode B= 109 sec (full cycle) Approx 33/Hour (varies by mode) Analysis (view) = 45 sec; Run Time = 65 sec. pHOx+-L: Analysis Cycle: Run Time: Throughput: Micro Mode: Mode A = 60 sec.; Mode B = 126 sec (time to view results) Mode A = 89 sec; Mode B= 134 sec (full cycle) Goal: 25-30/Hour Analysis (view) = 71 sec; Run Time = 95 sec. Default Values: Hb Temperature FIO2 % 14.5 g/dL 37 ° C 20.9 % (room air) Sample Volume: Regular Mode Micro Mode 70 uL 115 uL 120 uL 40 uL 55 uL (Range 10.0 to 20.0 g/dL) (Range 10.0 to 40.0°C) (Range 20 to 100.0%) pHOx/B pHOx-Plus pHOx-Plus L pHOx/B pHOx-Plus- Plus L (all tests) (all tests) (all tests) (gases) (gases) Software Version: 4.xx Sample Type: Arterial, venous, capillary, mixed venous, Quality Controls, and Proficiency A-3 Appendix pHOx/B: Analysis Cycle: Run Time: Throughput: Micro Mode: pHOx Service Manual One-Point Cal Modes: Mode A One point fluid cal performed with each Gas Cal I (30-45 min intervals). (Best for medium-high volume customers) Mode B One point fluid and Gas Cal performed with EACH analysis. (Best for LOW volume customers only - No separate Gas Cal I) Rotary Valve Positions: Closed ("Air Std D") AIR ("Air Cntrl 1") Std D Std C Cntrl 1 AIR ("Air Std C") Closed ("Air Cntrl 2") Std B Cntrl 2 Sample AIR ("Air Cntrl 3") Closed ("Air Std B") Cntrl 3 Std A AIR ("Air Std A") Home: Closed "Dummy"/Closed (not used) Figure A-2 Valve Manifold Assembly (Tubing Connections and Rotation Positions) Control/Position Valve Manually and View Positions: Service Menu, System Test A-4 AIR Positions: Closed Positions: Air Std A, Air Std C, Air Cntrl 1, Air Cntrl 3 HOME, Air Std B, Air Std D, Air Cntrl 2, Dummy Fluid Positions: Connections on Valve: Std A (A) Control (Fluid) Positions: Connections on Valve: Cntrl 1 (E) Not Used Position: Dummy Std B (B) Cntrl 2 (F) Std C (C) Std D (D) Cntrl 3 (G) Appendix Linear (Measurement) Ranges: pH 6.50 to 8.00 pH units PCO2 3.0 to 200 mmHg (0.4 to 26.7 kPa) PO2 0 to 800 mmHg (0.0 to 106.7 kPa) SO2% 0.0 to 100% Hct 12% to 70% Hb 4 - 23.3 g/dL - based on Hct Na+ not reported - 1.0 to 200.0 mmol/L) Memory Storage Capability: Patient Storage 18 Samples (FIFO) QC Storage 6 QC Files (FIFO) Stores up to 10 daily points/level - condensed to 1 monthly point Monthly Data - 32 points for each level Error Codes 96 total codes stored (FIFO) 6 shown per screen (16 potential screens) Appendix QC Files Details: 6 QC Files plus “File 0” (proficiency) 3 On-Board QC files (combined gases/Hct/SO2 controls; 3 levels) 3 External QC files (3 levels for BG; 2 levels for Hct/SO2) Features Auto QC (on-board QC) w/ preset times Control reminder/lockout; Setting daily QC analysis times. Tracking Daily, Monthly, Cumulative, Levey-Jennings. Fluids Details: Std A - It is used to calibrate PCO2, Na+, Hct, and PO2 (1-point check) and sits in flow cell during idle. Std B - It is used to calibrate PCO2, contains surfactant, and is used to clean flow path during cycles. Std C - It is used to calibrate pH. Std D - It is used to calibrate pH, Na+, and Hct. External SO2 Calibrators - There are 2 levels used to calibrate SO2 (approx 50 and 95 %; final values TBD). Calibrate monthly (or as needed). A-5 pHOx Service Manual Dependency Rule and Setup Information: HCT Requires Na+ calibrated for calibration/results. PCO2 Requires pH calibrated for results. (pH checked to assure bicarbonate is correct.) SO2% Requires Hct calibrated, uses SO2%c if measured SO2 is not available. Hb Requires Hct and SO2 for reporting. Hb method priority: (1) If linked with COOx, then COOx Hb will be reported. (2) If no COOx link, measured results are reported (if available). (3) If no COOx or measured result, calculated result from Hct reading (Hct/3) is reported. (4) If no Coox, no measured, and no Hct: Reports ‘Default’ Value (Hb). P50 Measured Reported only with PO2 between 30 and 75 mmHg. Also requires calibrated SO2% and Hb/Hct for reporting. Qsp/Qt Requires 2 separate blood samples for determination: Mixed Venous and Arterial. RI Utilizes input or default FIO2 % for calculation. nCa pHOx-Plus Only. Requires properly installed/calibrated iCa and pH. Anion Gap pHOx-Plus Only. Requires properly calibrated/installed Na+,K+,Cl-, PCO2. Calculated Results require calibrated and ‘proper results’ by measured tests used in calculation. Suppression (Hb): Only Hb can be suppressed: Option to turn Hb from measured to calculated (one or the other will always appear). Passwords: • One password (System Password) for Set Up • 200 user unique passwords available - 3 levels of priviledge: 1. Privilege Level 1: Operators have access to all areas of the analyzer except those protected by the existing System Password. Level 1 operators do not require PDM review and may override this feature. Level 1 operators can override QC lockout. 2. Privilege Level 2: Operators have access to all areas of the analyzer except those protected by the existing System Password. Level 2 operators DO require PDM review, but may override this feature. Level 2 operators cannot override QC lockout. 3. Privilege Level 3: Operators have access to analysis only. Level 3 operators require PDM review and cannot override this. Level 3 operators cannot override. Test Suppression: If measured test is suppressed, any calculated/associated tests will also be suppressed. Tests still will be calibrated w/errors displayed on suppressed channels. A-6 Appendix Interfaces: NOVA COOX, PDM, ASTM interfaces. Standby Mode: No time dependency will prime and calibrate after coming out of Standby, In Standby only does idle pumping, no fluidic/gas calibration, Auto-Wake-Up, Auto-Cal, and Auto-QC Micro-Sample Analysis: Micro-Mode Tests: pH, PCO2, PO2 Stepped Analysis: Stage 1 = Tip of Probe to Adet 1 (45/55 uL Stage 2 = Up to Adet 3. Stage 3 = Between Adet 3 (up to Adet 4): Other Tests Reading (including contact with Reference Electrode) A-7 Appendix QC Lockout: Mode A Lockout: If lockout enabled, and tests fail, unit can be ‘unlocked’ by running any control level and passing all tests. Mode B Lockout: If lockout enabled, and tests fail, unit can only be ‘unlocked’ by running specific control level(s) that failed and passing all tests. QC Storage Options: Mode A QC Storage: If all QC results ‘pass’ range limits, QC results are automatically stored in memory. If any results ‘fail’ range limits, all results will remain on unit display until either user accepts/rejects results or another QC or sample result is run/displayed (at which time past [non-released] QC results) will be deleted. Mode B QC Storage: All QC results, whether passing or failing range limits are automatically stored in memory at end of control analysis. Glucose (Overrange Testing) pHOx-Plus/C/L only: Gluc Overrange Solution included in Reagent Pack: Done Automatically Daily; Elimination of manual - ampule running of Performance Check Solution Automatic Glucose Performance done at operator-programmed time daily. (Nova recommends to set time at 5:00 AM for optimum performance.) PASS - FAIL is reporting. Change Gluc Membrane if test lists as FAILED and repeat. Glucose Auto Polishing Cycle: Automatic polishing of Gluc sensor occurs each day at midnight, followed by 2-point cal, and additional 2-pt calibrations at 2 and 4 AM. pHOx Service Manual Error Code Overview: Code Groups Summary Group: 1 - 6 Group: 7 - 13 Group: 14 - 19 Group: 20 - 32 Group: 33 - 35 Group: 36 - 42 Group: 43 - 48 Group: 49 - 53 Group: 54 - 58 Group: 59 - 63 Group: 64 - 68 Group: 69 - 75 Group: 76 - 78 Group: 79 - 104 Group: 602 - 684 Group: 686 - 703 pH PCO2 PO2 SO2% Hb Hct Na+ ADET 1 ADET 2 ADET 3 ADET 4 Flow-Related QC-Related Hardware-Software Related QC Related RMS Reagent Pack Related (Refer to the reference manual for specific code details.) A.2 General Maintenance/QC 1. Flowpath Conditioning (AS NEEDED) (Use serum/plasma/whole blood - Automated Cycle.) (Sales/Customer Typical Information: Customers as below; Sales- Do once at start of demo/evaluation.) a. Done daily for low-mid volume users (as needed). b. Not required for higher volume users. 2. Control(s) Analysis (DAILY) (Sales Information: SetUp QC Files fully before demo/evaluation; Customers: Run QC as they wish.) NOTE: Internal & External Controls are unique for all pHOx models. Controls for pHOx cannot be used on pHOx-Plus or “L”, and visa-versa. a. b. A-8 Internal Controls - 3 levels (All Tests including gases and SO2) External Controls - 3 levels in Multi-Pack Only (pHOx/B =All Tests inc. SO2 & Gases; pHOx+/L = All Tests, inc. electrolytes/Glu/Lac) Appendix Glucose Overrange Solution (Analysis) pHOx+/L/C Only (DAILY) (Sales Information: Verify Performance Chk passes- AutoRun at start of demo/evaluation & at each followup) Software Versions 2.0.x or Higher (Elimination of manual-ampule running of Performance Check Solution.) Automatic Glucose Performance done at operator-programmed time daily. PASS - FAIL reporting: Change Gluc Membrane if test lists as FAILED then repeat test. For Failed Test: Gluc is BACKLIT on Ready screen indicating failure; tests are unavailable for use until resolved. 4. Flowpath Cleaning (MONTHLY) a. For product release: Do monthly. b. Automated aspiration cycle, use Deproteinizing Solution. c. For pHOx only, remove Na+ electrode (non-glass sensor) and use a blank electrode. 5. Calibrate SO2 Channel pHOx/+/L/C (MONTHLY) a. Calibrate using 2 external Calibrators (latex suspension). b. Do monthly or more frequently as needed. 6. Clean SO2 Sensor (Optics) Surface: (pHOx/+/L/C) (QUARTERLY) a. Wipe optics surface of SO2 Sensor with lint-free tissue soaked with Cleaning Solution or 10% Bleach. Rinse with H2O. 7. Replace Na+ Electrode pHOx (AS NEEDED) a. Required to be done based on slope or results performance. b. Correct Hct QC is NOT reliable indicator of Na+ channel performance. Condition Na+ Electrode pHOx+/L/C (AS NEEDED) a. Required to be done based on slope or results performance. b. Use Na+ Conditioning Solution (PN 36856). Remove Na+ and soak in Conditioning Solution for 2 minutes. 8. Electrode Maintenance Details pH Electrode • Condition (manually) troubleshooting w/ pH Cond Soln (PN 06856); 10 min soak. • NO need to condition at start of use (Cond Soln in electrode pouch: pre-soaked) • Long-Life Sensor (similar to Stat Profile M) • Expected Lifetime: 1-2 Years; Warranty = 6 months from installation A-9 Appendix 3. pHOx Service Manual PO2 Electrode • • • • • • • Screw-On Membrane Cap (w/ Washer attached) Polish electrode (manually) (10 sec. circular on paper) w/ each membrane change. Electrode = Long Life; Membrane = Short Life Electrode estimated life = 1-3 Years; Membrane estimated life = 1-5 months Warranty: Electrode: 6 months; Membrane: Free of defects at installation Blood Condition after new electrode is installed (membrane changed as needed). Prefilled Membrane Caps (no IFS addition required or to be added) PCO2 Electrode • • • • • • • • Screw-on Membrane Cap (NEW Version; Different than STP/UL); washer attached. Membrane Life (new membranes/sensors = 2-12 weeks NO Conditioning (PCO2 Condition Solution) at StartUp (Caps presoaked in package) Electrode = Long Life; Membrane = Short Life Electrode estimated life = 1-3 Years; Membrane estimated life = 1-5 months Warranty: Electrode: 6 months; Membrane: Free of defects at installation Blood Condition after new electrode install, membrane change and as needed. Pre-filled Membrane Caps (no IFS addition required or to be added) Na+ Electrode pHOx: • Short- Life Disposable Electrode. NO maintenance. • Expected Lifetime: 2-4 months. Warranty = 1 month (from install). • Do NOT condition with pH Cond Solution (will damage electrode). • Must remove Na+ when doing Flowpath Cleaning (use ‘blank’). pHOx-Plus/L/C: • Long-Life Sensor (similar to SP-M) • Expected Lifetime: 1-3 Years. Warranty = 6 months (from install). • Condition (as needed): Requires manual soak in Na+ Conditioning Solution. • NO need to remove Na+ (from flowpath) when doing Flowpath Cleaning. A-10 Appendix SO2 Channel • • • pHOx/+/L/C Only Clean flow-channel with cotton-tipped swab. Wipe optics surface of SO2 Sensor with lint-free tissue soaked w/ Cleaning Solution or 10% Bleach. Rinse with H2O. (Quarterly) Expected Lifetime: 6-12 Months Warranty = 3 months from installation K+ Electrode pHOx-Plus/L/C Only • Mid-life (age) Sensor (similar to SP-M); Max Use Life = 5 months • Blood/protein conditioning for low-volume customers (as needed) iCa Electrode pHOx-Plus/L/C Only • Short-life (age) Sensor (similar to STP-M) • Expected Lifetime: 1-3 Months; Warranty = 7 days or 300 samples from installation • Blood conditioning (optional) can improve sensor performance. Glucose Electrode pHOx+/L/C Only • Electrode Short-life (age) Sensor (similar to STP-M) • Electrode = Long Life; Membrane = Short Life • Electrode estimated life = 1-3 Years; Membrane estimated life = 1-4 weeks (replace membrane if Glucose Performance Check fails or slope drops out) • Warranty: Electrode: 6 months; Membrane: 3 days from installation • Polish using new YELLOW (1 u) Polishing Paper only. • NO CONDITIONING (soaking) of Sensor required routinely. • Polish sensor ONLY FOR TROUBLESHOOTING (not with each membrane change). (Polish AS NEEDED.) • Polish 10 seconds. (Polish DRY on paper: no water added.) • Rinse with water; blot dry. • Replace new membrane directly. • Condition sensor in Soaking Solution (PN 13409) for 5 minutes with polishing. A-11 Appendix Cl- Electrode pHOx-Plus/L/C Only • Short-life (age) Sensor (similar to STP-M) • Expected Lifetime: 1-3 Months; Warranty = 7 days or 300 samples from installation • Blood conditioning (optional) can improve sensor performance • No membrane cap; pre-filled/sealed sensor (similar to STP-M) pHOx Service Manual Lactate pHOx-Plus-L Only • Electrode = Long Life; Membrane = Short Life • Electrode estimated life = 1-3 Years; Membrane estimated life = 1-3 weeks (Replace membrane if Lact Std D Hi or Lo errors occur or slope drops out.) • Warranty: Electrode: 6 months; Membrane: 3 days from installation • Polish using new YELLOW (1 u) Polishing Paper only. • NO CONDITIONING (soaking) of Sensor required routinely. • Polish sensor ONLY FOR TROUBLESHOOTING (not with each membrane change). (Polish AS NEEDED.) • Polish 10 seconds. (Polish DRY on paper: no water added.) • Rinse with water; blot dry. • Condition sensor in Soaking Solution (PN 13409) for 5 minutes with polishing. A.2.1 Other Analyzer Maintenance 1. a. b. c. d. 2. Analyzer Shutdown a. Short-Term (<2 Hours) • Power-off the analyzer • No need to rinse/clean unit for this period of time b. A-12 Tubings Details Pump Tubings QUARTERLY May need more frequent changes due to any flow-related problems. External Tubings: Waste/Reference Lines (through Solenoids) YEARLY Other External Tubings YEARLY Internal Tubings (Manifold) YEARLY (done only at time of PM by field engineer) AS NEEDED Long Term (> 48 Hours) • Requires water and air purge of flowpath and all reagent lines • Utilize special Purge Manifold to attach to reagent needle ports and place • Put all lines (except waste) in distilled water. Process 2-3 fluid primes with water. • Then remove lines from water and leave open to air; process 2 fluid primes with air. • After air prime, unit may be shut-off and remove electrodes as needed. • Take pump tubing off pump roller to relieve tubing tension. Appendix A.3 pHOx Electrode/Standards Electrode Standards Used in Approximate Delta mV Calibration/Slope (between standards) _________________________________________________________________ Stds C & D 33 Na+ Stds A & D 17 PO2 Stds Room Air, Std A and 0 (Ratio Air to Fluid (A) for 2 pt Cal; Std A used for 1 pt Cal) Not Shown (Air to 0) 1 (typical ratio Air/A) (should be almost same) PCO2 Stds A and B (HCO3- for dissolved gas>>monitor pH) 17 Hct and ADET’s Stds A and D (all ADET’s) > 500 (Adet 3 = Hct) (ideal > 800) Hb Based on SO2 and Hct TBD Appendix pH SO2% External Stds 1 and 2 n/a _________________________________________________________________ A.4 Electronics and Mechanicals Overview Electronics Circuit Boards and Main Assemblies NOTE: • • • Almost all circuit boards on the pHOx utilize surfacemount technology, therefore board component repair is not recommended except for socketed chips. Analog Board/Sensor Board Digital/Controller Board Display Assembly A-13 pHOx Service Manual • • • • • • • A.5 Distribution Board Electrode (Left and Right) Interconnect Boards Keypad Power Entry Module Power Supply Assembly Printer Printhead and Printer Interface Board SO2 Sensor Board Shutdown Procedure If the analyzer is to be turned off for more than 24 hours, flush the tubing harness first with distilled water then with air. Use the Flush Fixture (PN 24327) to perform this procedure. 1. Remove the Reagent Pack from the analyzer. 2. Install the Flush Fixture into the analyzer in the same way as the Reagent Pack. 3. Place the W-line of the fixture into an empty container. 4. Place the other tubing ends into a beaker of distilled water. 5. From the Ready/Not Ready screen, press Menu (soft key). 6. Select Flowpath Maintenance and press Enter. 7. Press Purge (soft key). 8. When the cycle ends, take all the tubings out of the distilled water. Leave the W-line in its container. 9. Repeat the purge with air. 10. When this cycle is completed, remove the flush fixture. 11. The pHOx Analyzer is now ready to be powered off for extended time. A.6 Start Up Procedure 1. 2. 3. 4. 5. 6. A-14 Insert the Reagent Pack and the Quality Control Pack. Reinstall the pump tubing. Power up the analyzer. Perform normal fluid priming and biosensor maintenance. Calibrate the analyzer. Run QC per your normal procedures. Stat Profile pHOx Installation Check List DONE DONE Model _____________________ Serial # ____________________ Instrument Check Out Set Up Check Out W/R Pump Tubing Harness Reagent Cartridge Control Cartridge (when available) Tubing Harness Electrode / Membranes SO2 Sensor Module Software Card and Version Probe/Microsample Adapter W and R Lines Inspect Unit for Shipping Damage Remove Cover, Check all Cable Connections Fuses Correct Line Voltage N/A YES SMR(s) ____________________ Barometric Pressure N/A YES Sensor Module Temperature Troubleshooting Training Subjects Solenoid Valve Tubing Incorrectly Seated Water Test Flow Blockage Other __________________________ Other __________________________ Other __________________________ Operator Training Subjects System Identificaion Running an Analysis Operational Status Calibrating the Analyzer Replacing Reagents and Paper Troubleshooting Flow Test and Flushing DONE Using and Handling Controls Fill out and review PACE forms Review customer warranty or service financing coverage Maintenance Set Up The above checks have been performed on your NOVA analyzer as part of your INSTALLATION. We at NOVA BIOMEDICAL believe that you, our valued customer, are entitled to the highest quality support available. As part of this service, we are leaving a copy of this check list for your records. If any questions arise, feel free to contact us TOLL FREE at 1-800-545-NOVA. Nova Support Representative Date Customer Signature Date A-15 A-16 Nova Biomedical pHOx Preventive Maintenance Check List SMR _____________________ CLEAN ADJUST REPLACE Mechanical Checkout INSPECT CLEAN ADJUST REPLACE INSPECT Model _____________________ Serial # _____________________ Electronic Checkout LCD Screen Electronics Bay Electrical Grounds RS232 Connections Software Version ______ Pump Assy Sampler Assy W & R Valve Rotary Valve CLEAN ADJUST REPLACE INSPECT Printer Assy Fluidic Checkout Flowcell SO2 Sensor Sensors (replace membrane, if applicable) DONE External Tubing Internal Tubing Fluids Bay Fluid Fountain Sample Probe Capillary Adapter The above checks have been performed on your NOVA analyzer as part of your scheduled PREVENTIVE MAINTENANCE service call. We at NOVA BIOMEDICAL believe that you, our valued customer, are entitled to the highest quality support available. As part of this service, we are leaving a copy of this check list for your records. If any questions arise, feel free to contact us TOLL FREE at 1-800-545-NOVA. The PM sticker placed on the upper, front corner of the left side cover indicates today’s date, the month your next scheduled PM is due, the initials of the Nova Support Representative and the service report number. Nova Support Representative Date Customer Signature Date A-17 A-18 A-19 � � �� � � � � ����� ���������������� ����� �������������� ����� ��������������� ����� �������������������������� ����� ����������������������������� � ����� ���������������� ��� � ���������������������������� ��� � ������������������������������ ��� � ��������������������������������� ��� � ���������������� ��� � �������������� ���� ��������������������������������������������� ������������������������������������������������� ������������������������������������������������������ ������������������������������������������������� ������������������������������������������������������ �������������������������������������������������� � � ������������������������������ ��������������������������������������������������� � � ������������������������������ ������������������������������������ ��������������������������������������� ���� �������������������������� ���� ���������������������� � � �� ��������� �� �� �� �� �� �� �� �� �� �� �� �� �� �� � � �� �� �� �� �� ���� ��������������������������������� ���� ������������������������������������������ ���� ����������������������������������������������� ���� ����������������������������� ���������������������������������������������� ������������������������������������ ������������������������������������ ���������������������������� �������������������������������������������������������� ������������������������������� ������������������������������������������������� ������������������������������� �������������������������������� ����������������������������� ���� ��������������������������������� ���� ������������������������ ���� ������������������������������������ ���� ������������������������������� ������������������������������������������������ ���� �������������������������� ������� ��������� �� �� �� ������������� � � � � A-20 PCMCIA COM3 COM2 COM1 BARCODE POWER SUPPLY J7 TB2 AC INPUT TB1 J8 DIGITAL CONTROL BOARD J9 CN3 P1 P5 PUMP J11 SAMPLER J10 SEIKO PRINTER J102 J104 J103 J6 J3 FLUID PACK FP P4 FRONT PANEL LEFT ELECTRODE CLIP TEST J7 CN1 CCFL KEYPAD J2 ANALOG BOARD CN2 OPTREX LCD PANEL P2 J1 DISTRIBUTION BOARD SENSOR MODULE RIGHT ELECTRODE CLIP J101 ROTARY VALVE CP Title: Page S-1 Size CONTROL PACK WIRING DOCUMENT FOR PRODUCT: NOVA PHOX Drawing Number Rev D Do not scale drawing Sheet 1 of 1 DWG #10-1087A AC Mains Entry Module To Power Supply TB1 AC Line Brown 1 AC Neutral Earth Ground 3 5 Blue Green/Yellow Analog Board J3 to Sensor Module CN1 ADT1-E Yellow 1 ADT1-M 2 THERMISTOR FPACK AGND Blue Red 3 White 4 Black 5 --- 6 HEATER+ Red 7 HEATER- Brown 8 N.C. N.C. --- 9 N.C. --- 10 Digital Control Board J9 to Seiko Printer CN3 PV1 --- 1 GND PV2 PV2 ------- 2 3 4 PGND PGND ----- 5 6 P STB# P DR1 P DR2 ------- 11 12 13 P DR3 P DR4 P DR5 ------- 14 15 16 P DR6 P DR7 P DR8 ------- 17 18 19 P BUSY P ACK# P ERR# ------- 20 21 22 P PE GND P RESET# ------- 23 24 25 P ONLINO P ONLINI P FEED# ------- 26 27 28 P BFEED# P NMI# ----- 29 30 PV2 PV2 GND ------- 31 32 33 PV1 --- Digital Control Board J11 to Distribution Board J101 GND 1 --2 D LP --- GND DM GND D FLM GND D CP GND D DD0 GND D DD1 GND D DD2 GND D DD3 GND D SPARE GND D ENAVDD D VEEN D ENABKL D YELLOW# D GREEN# D VCC D VCC D VCC D VCC KBY5# KBY4# KBY3# KBY2# KBY1# KBY0# KBX3# KBX1# KBX2# KBX0# SPCLK DACLK DADAT SPKRSPKR+ Digital Control Board J8 to Main Wire Harness PGND P9 Black ANA V+ +5VANA -5VANA ANAGND ANAGND ANAGND ANARXD# ANATXD# VCCPROT GND GND INT7 HTRCTL# FAIL# $1N3313 PGND 1 --- 2 PGND P6 Black 3 ----- 3 4 PMD1 PMD3 P1 P1 Red Yellow ----- 5 6 PMD0 PMD2 P1 P1 Black Grey --- 10 ----- 12 13 ------- ----- 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 ------- 30 31 32 ----- 33 34 ------- 35 36 37 ----- 38 39 Main Wire Harness J8 to Rot Val P9 N.C. --- 1 ------- 41 42 43 N.C. ROVHOME# ROVLED --White Red 2 3 4 --- 44 GND GND Black Black 5 6 --------------------------------------------------- 34 Main Wire Harness J8 to Control Pack CP TOUCH Orange 1 Main Wire Harness J8 to Lamp P7 LAMPRed 1 LAMP+ 2 N.C. N.C. GND 2 3 4 Brown 1 Digital Control Board J10 to Analog Board J1 PGND --- ----Black 5 6 ------- 2 3 4 7 8 KBY1# KBY0# ----- 5 6 9 10 11 KBX3# KBX2# KBX1# ------- 7 8 9 12 13 14 KBX0# G AN G KA ------- 10 11 12 Y AN Y KA ----- 13 14 VALVECOM TOUCH P4 FP Black Orange P5 Red 14 15 16 DOORSW# P6 White SMD2 PGND P5 P10 Grey Black ----- 18 19 SMD1 P5 Yellow 15 --- 20 GND SMD10 P10 P5 Black Black 16 17 ----- 22 24 PGND RVD3 SAMPLED FP P2 P10 Black Red Red 18 19 20 RVD2 ROVLED HEATER+ P2 P9 P3 Grey Red Red 21 22 23 HOME# DOORLED SAMHOME# P9 --P10 White N.C. White 24 25 26 HEATERLAMP- P3 P7 Brown Brown 27 28 GND --- N.C. 29 LAMP+ VALVE1# P7 P4 Red White 30 31 P2 P4 P2 Black Violet Yellow 32 33 34 P9 --- Black N.C. 2 4 D CP DM VADJ VCCS GND VEE D DD3 D DD2 D DD1 D DD0 DISPOFF# Power Supply TB2 to Digital Control Board J7 PGND Black 1 Main Wire Harness J8 to Rotary Valve P2 RVD1 Yellow 1 +24VDC POWER-FAIL +5VDC Violet White Red 2 3 4 RVD3 RVD0 RVD2 Red Black Grey 2 3 4 GND +12VDC Black Orange 5 6 N.C. N.C. ----- 5 6 GND -12VDC Green Yellow 7 8 Main Wire Harness J8 to Sampler P10 N.C. --- 1 Main Wire Harness J8 to Fluid Pack FP TOUCH Orange 1 N.C. N.C. GND 2 3 4 N.C. SAMHOME# SAMPLED --White Red 2 3 4 GND GND Black Black 5 6 1 VALVE0# VALVECOM N.C. Violet Black --- 2 3 4 ADT1-E ADT1-M Yellow Blue 5 6 Main Wire Harness J8 to Pump P1 PDM1 Red 1 PDM3 PDM0 PDM2 Yellow Black Grey 2 3 4 Main Wire Harness J8 to Sampler P5 SMD1 Yellow 1 RVD0 VALVE0# RVD1 2 3 PGND PGND N.C. N.C. 1 KBY4# KBY3# KBY2# SMD3 Main Wire Harness J8 to Fluid Deck P4 VALVE1# White SMD3 SMD0 SMD2 Distribution Board J102 to Membrane Keypad KBY5# --- Red Black Grey 4 ----- 5 6 Main Wire Harness J8 to Door Switch P6 DOORSW# White N.C. --- 1 2 PGND 3 Black ----Black Distribution Board J103 to Optrex Backlight CN1 COM White 1 N.C. N.C. HOT ----Black 2 3 4 Distribution Board J104 to Optrex LCD Panel CN2 D FLM --1 D LP --2 ----- 3 4 ----- 5 6 --- 7 8 9 10 11 12 13 ------------- 7 6 5 4 3 H P2 DB0 1 16 DB1 2 15 DB2 3 14 DB3 4 13 GND DB8 DB5 6 11 DB6 7 DB7 8 ADB16 15 1 3 14 ADB3 4 13 DB12 ADB4 5 12 DB13 ADB5 6 11 DB14 ADB6 7 10 DB15 ADB7 8 9 ADB17 2 15 ADB18 3 14 ADB10 ADB11 ADB19 4 13 ADB20 5 12 VCC ADB25 BADS# 2 ADB21 6 11 ADB22 7 10 ADB23 8 RAMADD0 1 16 RAMADD8 RAMADD1 2 15 RAMADD9 RAMADD2 3 14 RAMADD10 RAMADD3 4 13 GND RAMADD4 5 12 RAS0# RAMADD5 6 11 RAS1# RAMADD6 7 10 CASH# RAMADD7 8 9 CASL# CLKOUT 3 14 BRSTSYNC 4 13 LATERD# 5 12 CHOKEWR# 6 11 BD/C# READY# 7 10 DREQ0 BM/IO# GND 8 9 DACK0# BW/RD# ADB15 15 BWR# BBLE# ADB14 P6 16 BCLK2 BRD# BBHE# ADB13 9 ADB24 1 GND ADB12 10 TP2 P4 16 ADB9 ADB2 DB11 12 2 ADB1 DB10 5 P3 16 ADB8 DB9 DB4 1 PMPCLK 9 1.8MCLK BUSY D8 R137 221 R138 221 R123 221 GREEN GREEN GREEN D9 LOGIC ANALYZER TEST POINTS TP1 1 DIAGNOSTIC LEDS VCC P1 2 H D7 U21 DB15 1 1 3 4 OC CLK100 DB14 7406D CLOCK GENERATOR 1 OHM 1W C75 C64 47 16V G C63 0.1 0.1 9 R136 10 X1 14 6 11 7 1 X2 24MHZ F0 14.3MHZ F1 12MHZ F2 1.8MHZ 50.113 MHz 11 CLK2 R150 33 R133 33 MR# 1.8MCLK RTC / NVRAM SYSTEM BUS BUFFERS D10 8 7 5 RST RST# C96 4 PFI PFO# CLK2 RST# RST# RSTSYNC RSTSYNC U32 W/RD# POWER ON RESET R114 4.75K D/C# M/IO# RD# WR# R134 A25 R145 4.75K PWRFAIL# 221 C65 BARCODE A24 A23 VCC A22 100PF A21 A20 CLK2 A19 install BARCODE pullup OR pulldown only KU386EXTC so software sees the right levels when nothing is connected J1 DSR RXD# RTS TXD# CTS DTR RI E SGND U24 5 6 4 1 BCRXD R170 4.75K 4.75K 74HCT14 2 8 BCTXD VCC 9 U23 BCVCC TMRGATE0 LAMPON# 74HCT14 C7 100PF NMI R135 U23 BCODE C28 R175 4.75K 74HC02D 221 R46 1 6 2 7 3 8 4 9 5 BS8# 221 R49 DCD VCC 4.75K 4.75K AUDIOENB DOOR 100PF R148 110 33 41 90 73 120 A18 RESET CLK2 BS8# ADS# NA# W/R# NMI D/C# SMI# M/IO# SMIACT# READY# LBA# 101 105 129 102 131 132 104 106 P1.0/DCD0# RD# P1.3/DSR0# WR# DRIVENB# SHELL MECHENB# 4.75K NMIENB# 12 OD13 CS3 R71 CS3# ACT05 SMD050 CS2# CS1# U14 111 108 107 126 125 124 123 122 CS0# P1.1/RTS0# A25 P2.6/TXD0 A24 P2.7/CTS0# A23 P1.2/DTR0# A22 P1.4/RI0# A21 P1.7/HLDA A19 P1.6/HOLD A18 P1.5/LOCK# A17 P2.4/CS4# A16 P2.3/CS3# A15 P2.2/CS2# A14 P2.1/CS1# A13 P2.0/CS0# A12 A11 VCC R47 1.8MCLK 4.75K U23 R51 DOORSW# 3 4 INT2 INT1 1.00K C8 INT3 74HCT14 CLK100 22 40V C29 100PF TMROUT1 TMROUT0 R164 93 94 95 96 INT4 INT6 5 INT7 6 3 74LS14D U19 4 OD 92 89 91 PMPCLK 10 RXD3 TXD3 IPCMRX 74HCT14 IPCM PORT DREQ0 12 13 ANATXD# P3.6/PWRDWN A9 P3.5/INT3 A8 P3.4/INT2 A7 P3.3/INT1 A6 P3.2/INT0 A5 P3.1/TMROUT1 A4 P3.0/TMROUT0 A3 TMRCLK0/INT4 A1 TMRGATE0/INT5 BHE# TMRCLK1/INT6 BLE# DACK0# IPCMTX XCDONE 4.75K XCRLD# R183 4.75K BUSY#/TMRGATE2 D14 PEREQ/TMRCLK2 D13 ERROR#/TMROUT2 D12 D11 118 112 117 128 113 DRQ1/RXD1 D10 DACK1#/TXD1 D9 DRQ0/DCD1# D8 DACK0#/CS5# D7 EOP#/CTS1# D6 R147 74HCT14 U23 XCDATA XINIT D5 98 77 79 78 STXCLK/DSR1# D4 SRXCLK/DTR1# D3 SSIOTX/RTS1# D2 SSIORX/RI1# D1 D0 C 76 25 24 26 119 J5 BRST# VCC TDI TDO TMS 4.75K R182 10 8 6 4 2 UCS# CS6#/REFRESH# WDTOUT ADS# R176 W/RD# D/C# 221 M/IO# R189 72 70 68 67 66 65 63 62 61 59 58 57 56 55 54 53 52 51 50 49 48 45 44 43 42 39 37 2 3 5 6 8 9 10 12 13 14 15 16 17 19 20 21 23 24 1B2 1A2 1B3 1A3 1B4 1A4 1A5 1A6 1B7 1A7 1B8 1A8 1B9 1A9 2B1 2A1 2B2 2A2 2B3 2A3 2B4 2A4 2B5 2A5 2A6 2B6 2B7 2A7 2B8 WR# 2A8 2B9 2A9 1DIR TDI 2DIR TDO 1OE# TCK 2OE# TMS 55 54 52 51 49 48 47 45 44 43 42 41 40 38 37 36 34 33 1 2 114 A16 A25 A15 A24 A14 A23 A13 A22 A12 A21 A11 A20 A10 A19 A9 A18 A8 A17 A7 A16 A6 A15 A5 A14 A4 A13 A3 A12 A2 A11 A1 A10 BHE# A9 BLE# 2 3 5 6 8 9 10 12 13 14 15 16 17 19 20 21 23 24 A8 R174 A6 A16 A13 PFO A12 VCC A11 WE 31 BCLK2 ADB12 30 BRST# ADB11 29 BRSTSYNC ADB10 28 BADS# ADB9 27 A9 ADB8 BW/RD# A10 BD/C# ADB7 25 A7 BM/IO# ADB6 24 A6 BRD# ADB5 23 A5 BWR# ADB25 ADB4 22 A4 ADB3 21 A3 ADB24 ADB2 20 A2 ADB23 ADB1 19 A1 ADB22 BBLE# 18 A0 ADB21 ADB15 3 ADB16 ADB15 ADB14 ADB13 4 CE 6 CHOKEWR# 7 BRD# 8 CS2# DQ6 DQ5 DQ4 DQ3 10 DB6 11 DB5 12 DB4 13 DB3 DQ1 DQ0 ADB10 ADB9 ADB8 ADB7 ADB6 ADB5 ADB4 ADB3 DB2 14 DQ2 ADB11 DB7 9 DQ7 ADB12 VCC 5 OE 26 A8 2 15 DB1 16 DB0 ADB2 ADB1 BBLE# 17 GND DS1644P ADB20 CS1# BRD# ADB19 U35 ADB18 CHOKEWR# 30 27 29 28 1 26 56 31 1B1 1A1 1B2 1A2 1B3 1A3 1B4 1A4 1B5 1A5 1B6 1A6 1B7 1A7 1B8 1A8 1B9 1A9 2B1 2A1 2B2 2A2 2B3 2A3 2B4 2A4 2B5 2A5 2B6 2A6 2B7 2A7 2B8 2A8 2B9 2A9 1DIR TDI 2DIR TDO ADB14 XTDO 6 75 64 17 ADB13 ADB12 ADB5 71 CS3B# ADB4 CS4A# 72 CS4A# CS4B# 69 CS4B# M/IO# 10 IO# CS5A# 66 CS5A# ADS# 11 ADS# CS5B# 65 CS5B# RD# 27 RD# W/RD# 26 W/R# CHWR# 80 D/C# 95 D/C# LTRD# 81 BRST# 91 /RESET SYSTEM DRAM ADB1 U28 ADB9 UCS# BRD# CHOKEWR# ADB7 BWR# LATERD# ADB6 ADB5 33 33 44 34 MECHENB# R196 R198 R195 R197 33 33 33 33 40 43 41 42 R154 33 47 79 RAS0# CLK2 89 CLK2 RAS1# 78 RAS1# CLKOUT 92 CLKOUT CASL# 82 CASL# CASH# 86 CASH# ADB3 ADB2 ADB1 BBHE# WDTOUT 96 WDTOUT BBLE# MECHENB# 30 27 29 28 8 A3 RA2 54 RAMADD2 A4 RA3 67 RAMADD3 BHE# ADB5 22 A5 RA4 74 RAMADD4 BLE# ADB6 33 A6 RA5 73 RAMADD5 ADB7 100 A7 RA6 57 RAMADD6 ADB8 98 A8 RA7 62 RAMADD7 ADB9 99 A9 RA8 60 RAMADD8 A10 RA9 58 RAMADD9 59 RAMADD10 D3 D2 D1 D0 VCC UCS# CS6# CLKOUT 4.75K R166 R165 4.75K 30 27 29 28 1A1 1B1 1A2 1B2 1A3 1B3 1A4 1B4 1A5 1B5 1A6 1B6 1A7 TMS 1A8 1B8 1A9 1B9 2A1 2A2 2B2 2A3 2B3 2A4 2B4 2A5 2B5 2A6 2B6 2A7 2B7 2A8 2B8 2A9 2B9 TDI 1DIR TDO 2DIR TCK 1OE# TMS 2OE# U27 DB15 1MEG = No jumper DB14 4MEG = No jumper DB13 ADB10 DB11 DB10 ADB11 3 ADB12 1 ADB13 23 ADB14 18 DB8 ADB15 37 DB7 ADB16 34 DB6 ADB17 24 DB5 ADB18 25 DB4 R213 DB3 4.75K DB2 DB1 4.75K R162 4.75K DB0 R209 VCC E7 1 BWR# ADB19 7 ADB20 9 ADB21 16 ADB23 30 A13 DRVEN# 77 A14 RSTSYN 56 A15 POEH A17 CAS0# DQ3 CAS1# DQ4 CAS2# DQ5 CAS3# DQ6 DQ7 WE# DQ8 A0 DQ10 A1 DQ11 A2 DQ12 A3 DQ13 A4 DQ14 A5 DQ15 A6 DQ16 A7 DQ17 A8 DQ18 A9 DQ19 A10 DQ20 55 DRIVENB# POEL BUSY A19 DQ23 POEH# A20 A21 ACK 44 PDEN A23 DACK0 MEMSIZ0# 15 SIZ0 MEMSIZ1# 12 SIZ1 46 47 48 DB00 BWR DQ25 4.75K READY# POEL# DQ26 DQ27 DQ28 BUSY 43 DREQ0 DQ24 R200 42 PSTB# A22 19 52 DQ22 RSTSYNC 51 RDYIN A18 21 VCC DQ29 PSTB# ACK# PDEN DQ30 MEMSIZ0# DREQ0 DACK0# 67 68 69 70 DQ31 AT0 35 DB1 PP38 38 DB2 PP49 49 DB3 TDO TMS TCLK TCLK CARD DB1 DB2 AT2 NC0 AT3 NC1 ACCESS 8-BIT 16-BIT DB4 DB5 E6 R211 4.75K DB6 R215 221 DB7 DB8 4.75K R212 DB9 BRSTSYNC DB10 R218 C119 DB11 DB12 DB13 4.75K R144 DB14 4.75K 47 D0 DB15 R156 4.75K R140 4.75K R129 4.75K R125 4.75K R142 4.75K DB0 16V D1 DB1 D2 DB2 DB3 D3 DB4 D4 DB5 DB6 D5 4.75K R127 DB7 D GND4 D6 DB8 R157 4.75K R139 4.75K R124 4.75K R141 4.75K R126 4.75K DB9 DB10 FUJITSU FCN565PO68G0231E4 D7 D8 DB11 DB12 DB13 DB14 VCC DB15 D9 FRAME OF PCMCIA D10 D11 4.75K D12 33 45 R173 R155 4.75K R167 4.75K 4.75K R168 4.75K R169 4.75K CONNECTS TO CHASSIS GND D13 D14 C D15 256K X 32, 1M X 32, OR 4M X 32 SIMM DRAM R17 PP63 63 SUICI J7 JUMPER SELECTS SIZE +24VDCIN 1.0 OHM 1W 1 PGND R18 2 1.0 OHM 1W J10 +12VDC 3 PWRFAIL# R1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 R16 ANA_V+ ALTERA Epm7128SQC100-7 +24VDC +5VANA T -5VANA 4 VCC 5 1 6 C33 22 40V GND 4 7 B 8 -12VDC ANAGND R146 4.75K C2 R15 ANARXD# C41 47 16V +24VDC VCC MC78M05CDT 3 OUT IN U10 +5VANA 1.0 OHM 1W SMD050 ANAGND R98 100 R70 10 OHM 1W 1.0 OHM 1W R44 B E 16 BSY# 33 WP 36 CD1 42 CE2# 43 RFSH 44 RFU1 45 RFU2 57 RFU3 58 RST 59 WT# 60 RFU4 61 REG# 62 BVD1 63 BVD2 67 CD2 18 VPP1 52 VPP2 17 VCC1 51 VCC2 1 GND1 34 GND2 35 GND3 68 2 BS8# DB3 AT1 VCC TMS DB0 R12833 DB0 PP35 386BIST R208 4.75K TDI DQ1 DQ2 12 13 14 15 16 17 18 28 31 32 19 A16 87 BIST# RAS1# F WE# 2 4 6 8 20 22 24 26 49 51 53 55 57 61 63 65 3 5 7 9 21 23 25 27 50 52 54 56 58 60 62 64 DQ21 RD# R143 1.00K DQ0 A12 14 ADB22 RA10 A11 VCC DB9 R163 4.75K 26 56 31 4 16MEG = Yes jumper DB12 10 12 13 14 15 16 17 19 20 21 23 24 2B1 DRAM SELECTION RAS0# DQ9 TCLK 2 3 5 6 8 9 1B7 BWR# 11 A9 12 A8 22 A7 23 A6 24 A5 25 A4 26 A3 27 A2 28 A1 29 A0 7 CE1# 9 OE# 15 MT8D132M-6 R187 R199 RAS0# ADB4 U33 55 54 52 51 49 48 47 45 44 43 42 41 40 38 37 36 34 33 BBLE# ADB8 39 TMS ADB2 ADB10 2 2OE# TCK ADB3 ADB11 ADB4 D4 ADB10 CS3B# ADB3 D5 ADB11 TDI TDO TCK TMS RAMADD1 D6 ADB12 CS3A# 83 D7 ADB13 70 RA1 D8 ADB14 CS3A# A2 D9 ADB15 CS0# 29 D10 ADB16 ADB6 94 CS0# ADB2 1OE# A3 D11 ADB17 ADB15 RAMADD0 D12 ADB18 ADB16 85 D13 ADB19 ADB7 55 54 52 51 49 48 47 45 44 43 42 41 40 38 37 36 34 33 RA0 D14 ADB20 U38 A1 D15 ADB21 ADB8 31 R188 221 ADB22 ADB9 ADB1 A1 ADB23 56 A25 55 A24 54 A23 53 A22 50 A21 49 A20 48 A19 47 A18 46 A17 19 A16 20 A15 14 A14 13 A13 21 A12 10 A11 8 TMS BLEN# A2 ADB24 DRAM CONTROLLER TCLK 32 A4 ADB25 U37 BBLE# 221 J6 A10 BHEN# A5 AT29C010A-90JC 2 A16 21 DB7 D7 3 A15 20 DB6 D6 29 A14 19 DB5 D5 28 A13 18 DB4 D4 4 A12 17 DB3 D3 25 A11 15 DB2 D2 23 A10 14 DB1 D1 26 A9 13 DB0 D0 27 A8 5 A7 6 A6 ADB18 1 7 A5 A18 8 A4 30 ADB17 9 A3 A17 10 A2 11 A1 12 A0 128K x 8 22 CE# 24 RD# 31 WR# ADB17 50 103 U29 A14 ADB16 1 BBHE# WDTOUT FLT# A15 ADB13 74ABT18245ADL 23 22 21 20 19 18 16 14 13 12 11 10 8 7 6 5 NC2 32 DB1 G D0 NC1 74ABT18245ADL A7 DB2 U34 RD# 4.75K 9 7 5 3 1 1A1 1B6 1 26 56 31 221 1B1 1B5 READY# TRST# CLKOUT 99 A17 TCK R158 JTAG PORT DB3 DB0 NC3 ADB14 TMRGATE1/INT7 D15 U11 11 A10 BIST# ACT05 U23 ANARXD# P3.7/COMCLK A2 4.75K D 87 86 85 84 82 80 75 74 115 40 30 29 27 32 4 34 35 P2.5/RXD0 A20 T 34 74ABT18245ADL ADS# 100PF DB4 STORAGE 1.8474 MHz MAX707CSA 1 R254 PATIENT DATA 33 SUICIDE R50 DB5 VIDCLK U25 VCC F 10 7406D 14.318 MHz OE R177 4.75K E5 DB8 DB6 OC DRIVENB# VCC MANUAL RESET DB9 DB7 VCC 14.318MHZ DB10 8 U21 MBRS140T Y1 DB12 DB11 OC BADS# VCC2 50MHZ 1 DB13 VCC1 2 16 15 10 9 4 "DRV" 7406D 4 13 3 VCC "ADS" U21 AV9154-27CS16 R149 "RST" 41 D15 40 D14 39 D13 38 D12 37 D11 66 D10 65 D9 64 D8 6 D7 5 D6 4 D5 3 D4 2 D3 32 D2 31 D1 30 PCMCIA CARD CONNECTOR 8 R45 4.75K SUICIDE ANATXD# E501 1000 Ripple I => 1.5AS12 35V 105c / 3000 hr 1 GND C34 47 16V ESR =< 0.1 ohms 3 -5VANA U9 OUT IN MC79M05CDT SUICI 4 C27 22 40V R48 10 OHM 1W VCCPROT T INT7 SMD050 HTRCTL# FAIL# 6 U23 12 13 74LS14D A RIGHT NEXT TO J7 5 74HCT14 POWER SUPPLY CONNECTOR LOCATE THESE PARTS NET MUST CARRY 5A CURRENT DRIVENB# NOVA Biomedical ANARST DRAWING TITLE Schematic, Cortex Digital Board Processor & Memory U44 PGND SIZE Page S-3 ANALOG BOARD CONNECTOR DRAWING NUMBER E DO NOT SCALE DRAWING 8 200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141 7 6 5 4 3 2 REV. 24292-01 LAST MODIFIED ON: 10-5-1998_13:54 SHEET A 1 OF 3 1 DWG #10-1071B 8 7 6 5 4 3 2 1 U15 MAX211ECAI 12 H C1- C55 DCD DSR RXD# RTS TXD# CTS DTR RI SGND 1 6 2 7 3 8 4 9 5 16 0.1 17 R99 25 1.0 OHM 1W 4.75k R3D R3Q R2D R2Q R4D R4Q T3Q T3D T1Q T1D R1D R1Q T2Q T2D R5D R5Q T4Q T4D SHDN 26 5 22 20 7 8 6 19 21 DCD1# DSR1# RXD1 RTS1# TXD1 CTS1# DTR1# RI1# 1.00K 1.00K MAX211ECAI C66 IPCM DEBUG PORT G 2 J3 1 6 2 7 3 8 4 9 5 DSR RXD# RTS TXD# CTS DTR RI SGND 16 R100 17 C70 R3D T4D 26 5 22 20 7 8 6 19 21 EN# 24 R3Q R2D R2Q R4D R4Q T3Q T3D T1Q T1D R1Q R1D T2Q T2D R5D R5Q T4Q SHDN U20 0.1 IPCMRX 1.00K 2 5 6 9 R29 3 4 7 8 T SMD050 EN D0 TRI 2 RXD0 D5 D4 RTS0# TXD0 D3 DCD1# A2 DSR1# A1 RXD1 A0 CS0# CTS1# CS1# DTR1# CSP# PD1 CLK PD2 RESET# PD3 BDO PD4 INTSEL# PD5 INT0 PD6 INT1 PD7 INTP SLCT AFD# BUSY STB# PE SLIN# ACK# BIDEN DB12 DB13 DB14 DB15 T R31 221 221 221 221 R6 R32 R7 R33 R8 R34 R9 2 5 6 9 221 221 221 221 EN R10 221 R36 221 R37 221 R39 221 R11 4 39 44 43 45 60 59 63 65 66 67 68 1 1.8MCLK BRST# R159 1.00K INT2 INT3 PERR_INT BUSY PE_INT DB0 DB1 DB2 DB3 VCC R203 4.75k R153 DB4 DB5 DB6 DB7 4.75k R35 4.75k R38 OE 4.75k 1 4.75k 12 C1+ V+ 4.75k 221 CLKIN MD6 MD5 4.75k 1.00k LDEV# MD4 LRDY# MD3 BCFG2 MD1 BCFG1 MD0 VMD11 VMD10 VMD9 VMD8 VMD7 VMD6 VMD5 VMD4 VMD3 VMD2 VMD1 VMD0 6 ADDHI 108 107 106 105 104 103 102 101 100 MA7 MA6 5 4 3 2 1 153 152 151 150 149 148 147 146 145 144 143 141 140 138 137 136 135 134 133 MA5 A23 A22 MA4 A21 MA3 A20 MA2 A19 MA1 A18 MA0 A16 VMA7 VMA6 VMA5 VMA4 VMA3 VMA2 VMA1 VMA0 A15 HCAS# A14 LCAS# A13 R225 110 112 113 111 RAS# WE# TMROUT0 4.75k VCC CLICK 8 9 10 2 3 2 R223 R222 R224 R30 D_VCC T SMD100 R214 221 R233 4.75K A10 A9 R231 4.75K 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 A7 A6 A5 A4 A3 A2 A1 A0 D_VEEN D_YELLOW# RAS# HCAS# D_VCC LCAS# KBY5# WE# KBY3# KBY1# OE# KBX3# U40 KBX2# SPCLK DACLK A8 A7 27 4 23 1 2 9 3 18 28 DCD3 DSR3 RXD3# RTS3 TXD3# CTS3 DTR3 RI3 R171 R160 1.00K C97 0.1 C86 C2+ R2D R2Q R4D R4Q T3Q T3D T1Q T1D R1D R1Q T2Q T2D R5D R5Q T4Q 25 SHDN DCD3# DSR3# RXD3 RTS3# TXD3 CTS3# DTR3# 29 31 41 24 26 28 25 30 IPCMRX IPCMTX 8 5 62 12 10 13 11 6 PERR_INT PE_INT RXD4 PGMEN# TXD4 PROGPULSE 221 R179 SYSTEM STATUS LEDS D12 D13 D_DD2 D_DD3 D_SPARE D_ENAVDD D_ENABKL D_GREEN# D_VCC KBY4# KBY2# KBY0# KBX1# KBX0# F DADAT SPKR+ A3 A2 N74F244D A1 2 4 6 8 BLE# D15 M D14 FLM LP CP M 53 54 56 55 SHFCLK D12 D11 ENAVDD D10 ENAVEE D9 ENABKL 48 47 76 PSTB# PNL7 D6 PNL6 D5 PNL5 D4 PNL4 U43 D3 66 65 64 63 61 60 58 57 PNL3 D2 PNL2 D1 PNL1 D0 PNL0 R4 VO2 BYPASS GND +IN VDD -IN VO1 BPSTB# 19 R210 E 221 N74F244D DD3 DD2 DD1 DD0 2 4 6 8 18 16 14 12 17 15 13 11 221 R255 SHUTDOWN D_ENAVDD D_VEEN D_ENABKL U41 YELLOW# GREEN# 1 2 3 4 D_FLM D_LP D_CP D_M 3 5 7 9 1 U39 D8 D7 18 16 14 12 17 15 13 11 D13 D_DD3 D_DD2 D_DD1 D_DD0 3 5 7 9 1 D_YELLOW# D_GREEN# D_SPARE 19 R217 221 8 7 6 5 VCC LM4861M 20.5K C5 D +15VDC C142 C141 0.1 20.5K R102 3 C44 R3 49.9 0.1 10K 74HC02D U24 4.75k R401 C400 560 1000PF U16 MIC6211 .1UF DSR0# D6 D5 RTS0# D4 D3 TXD0 CTS0# D2 DTR0# D1 RI0# D0 21 20 19 18 17 16 15 14 DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0 1 4 2 OC 3 5 V+ D7 RXD0 3 U21 PROGPULSE DCD0# DCD1# A2 DSR1# A1 RXD1 A0 1 R104 R121 20K 3 5 V+ 4 C56 11.8VDC 1 D5 10PF Q5 2 E 4.75K MIC6211 +2.5VREF B 1 V- 2 7406D C 22657 C V- 2 R122 C58 100PF 2 R120 4.75K R501 10K +15VDC VCC 24.9K CS0# CTS1# CS1# YELLOW# GREEN# GREEN D14 XCCLK R207 R206 DTR1# CSP# IORD# 57 56 55 58 32 3 38 36 37 PD0 PD1 CLK PD2 RESET# PD3 BDO PD4 INTSEL# PD5 INT0 PD6 INT1 PD7 INTP ERR# 221 221 221 221 ADB2 ADB1 BBLE# D500 R119 MBRS140T 1.00K CS4A# CS4B# CS5B# R502 324K RI1# IOWR# 53 52 51 50 49 48 47 46 33 34 35 R118 3.32K RTS1# TXD1 220 GREEN D_DD1 24 EN# U30 GREEN D_CP D_DD0 SPKR- U500 T4D D_FLM A4 DUART TRI 1.0 OHM 1W R172 0.1 17 26 5 22 20 7 8 6 19 21 D_M A5 TL16C552FN R3Q D_LP VCC C2- R3D 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 A6 +15VDC 13 G J11 A8 14 28 29 13 27 33 A11 R28 CW 1 33 33 33 A12 13 14 15 16 17 18 20 21 23 24 25 26 27 28 30 31 OPTION 1 26 25 24 23 22 19 18 17 16 VMA8 A17 R504 OPTION R249 BCFG0 MA8 .22uF BPSTB# C500 V- 16 0.1 D11 MD7 VMD12 VGA VIDEO CONTROLLER C1- C87 GREEN CPURST VMD13 VCC 1.5uF COM3 J4 B 34 35 40 39 38 C143 MAX211ECAI 14 15 VCC MD9 MT4C16257-70 39 D15 38 D14 37 D13 36 D12 34 D11 33 D10 32 D9 31 D8 10 D7 9 D6 8 D5 7 D4 5 D3 4 D2 3 D1 2 D0 U24 0.1 SHELL MD10 RD# DOORBOARD FUSE 16V VMD14 U45 AUDIOENB C85 1 6 2 7 3 8 4 9 5 MD11 M/IO# 36 37 DB15 DB14 DB13 DB12 DB11 DB10 DB9 DB8 DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0 1 221 221 221 R42 ADS# FLM stb used for b_feed output PRINTER CONNECTOR SGND CHOKEWR# LATERD# R152 1.00K 74HC02D RI 36 37 221 R41 R12 MD12 4.75k ADB23 ADB22 ADB21 ADB20 ADB19 ADB18 ADB17 ADB16 ADB15 ADB14 ADB13 ADB12 ADB11 ADB10 ADB9 ADB8 ADB7 ADB6 ADB5 ADB4 ADB3 ADB2 ADB1 BBLE# R503 R40 BHE# CS3 CS3A# CS3B# CS5A# ACK# 221 R13 DTR 32 3 38 221 R5 MD14 H C90 0.1 VMD15 1.5uF P_RESET# P_ONLINO P_ONLINI P_FEED# P_BFEED# P_NMI# CTS R244 PDEN 25 26 27 28 29 30 RTS R204 POEL# P_BUSY P_ACK# P_ERR# P_PE TXD# R202 R161 20 21 22 23 RXD# ADB2 ADB1 BBLE# TP3 13 14 17 18 11 DSR R205 33 34 35 U26 U31 3 4 7 8 12 15 16 19 P_STB# P_DR1 P_DR2 P_DR3 P_DR4 P_DR5 P_DR6 P_DR7 P_DR8 MD15 14.3MCLK LP N74F373D R43 SMD100 PV2 RESET 131 130 129 128 127 126 125 124 123 122 121 120 119 117 116 114 MD2 POEH# 2 5 6 24 33 DCD BCLK2 1 PV1 7 8 9 10 11 12 13 14 15 16 17 18 19 C R201 READY# PD0 INIT# CVCC2 157 158 MD8 RI1# IORD# 8 9 10 33 BBHE# ADS# M/IO# W/RD# DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0 RTS1# TXD1 BE3# OE 11 D 57 56 55 58 DB8 DB9 DB10 DB11 13 14 17 18 12 15 16 19 VCC E RI0# D6 ERR# P_INIT# U2 N74F373D 3 4 31 32 53 52 51 50 49 48 47 46 FSELA FSELB FSELC FSELD RS232 DRIVER/RECEIVERS 1 34 8 5 62 12 10 13 11 6 DCD2# DSR2# RXD2 RTS2# TXD2 CTS2# DTR2# RI2# R117 25 1.0 OHM 1W J9 D1 DSR0# IOWR# 1.00K F D2 DTR0# D7 0.1 C2- 27 4 23 1 2 9 3 18 28 SHELL CTS0# 21 20 19 18 17 16 15 14 DCD0# VCC C2+ CVCC1 MD13 C1- V- DCD2 DSR2 RXD2# RTS2 TXD2# CTS2 DTR2 RI2 R151 13 V+ C67 14 15 C68 0.1 DCD C1+ 0.1 E8 COM2 12 159 155 TL16C552FN 29 31 41 24 26 28 25 30 STANDBY# C71 0.1 R220 10 C115 47 BRSTSYNC VIDCLK IPCMTX R72 24 EN# DUART/PRINTER PORT C120 0.1 F65535A 11 7 R219 4.75k C116 47 16V C121 0.1 R232 4.75K R230 C2- 27 4 23 1 2 9 3 18 28 DCD1 DSR1 RXD1# RTS1 TXD1# CTS1 DTR1 RI1 R73 SHELL R216 4.75K C54 0.1 C2+ V- J2 VCC 0.1 14 15 COM1 13 V+ C1+ 0.1 10 VCC MBRS140T3 C42 R221 C43 INIT# SLCT AFD# BUSY STB# PE SLIN# ACK# BIDEN R192 4 39 44 43 45 60 59 63 65 66 67 68 1 R400 RXD4 R101 6.81K MBRS140T U22 CHOKEWR# LATERD# 8 9 TXD4 1.8MCLK BRST# U21 10 74HC02D 5 OC D6 1.0 OHM 1W 6 DS2505 R190 U46 option for either package 221 install only one ds2505 TOUCH MEMORY INTERFACE INT4 INT6 VCC R194 R193 R186 R185 R184 221 221 221 221 221 R181 221 B TOUCH 7406D R103 NC 4.75K DATA GND 3 2 U501 1 1 6 GND NC DATA NC NC NC 2 3 5 4 DS2505P U36 R191 NOVA Biomedical 200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141 DRAWING TITLE Schematic, Cortex Digital Board Comm, Printer, Display Touch Memory TP4 A 1 Page S-4 SIZE DRAWING NUMBER E DO NOT SCALE DRAWING 8 7 6 5 4 3 2 A REV. 24292-02 LAST MODIFIED ON: 10-5-1998_16:11 SHEET 2 OF 3 1 DWG #10-1072B 8 7 6 5 4 3 2 1 FUNCTION SELECT H SYSTEM CLOCK DIVIDER VCC KEYBOARD SCANNER MOTOR CONTROLLER R132 4.75K R130 4.75K R131 4.75K R110 R109 R108 4.75K VCC 4.75K 4.75K XC5202-6PQ100C PGND E1 FSELD 4.75K E2 FSELC 1 85 22 13 36 11 93 87 DB7 DB6 DB5 DB4 DB3 DB2 DB1 DB0 M2 DB6 M1 DB5 M0 SP7 DB1 SP8 AB2 PUMP1 AB1 PUMP2 BBLE PUMP3 CE# 38 10 81 82 83 84 BRSTSYNC DB2 PUMP4 PUMPHOLD RD# PMPCMP12 AB4 SAMP2 CCLK SAMP3 DIN SAMP4 PROG# SAMPHOLD SMPCMP12 79 7 8 9 5 7 11 DENNA OUT1 IN1 BOOT1 IN2 ENABLE XXSP7 R253 XXSP8 9 6 C9 0.1 4.75K R252 4.75K BOOT2 GND SENSE ROV1 TCK ROV2 ROV4 INIT# ROVHOLD HDC ROVGOOSE# LDC# ROVCMP12 DOUT 57 56 PGMEN# PROGPULSE +24VDC PUMP1 PUMP2 5 7 11 PUMP3 PUMP4 DENNA PUMPHOLD 1 C14 SAMP2 5 SAMP3 U11 9 6 VREF BOOT2 GND SENSE 1.0 OHM 1W R75 24.9K ACT05 1 SMPCMP12 U12 COMCLK SP2 CLK100 SP3 SMPCMP34 9 ROV2 R74 24.9K ROV3 ACT05 VCC ROVHOLD 2 U12 ROVCMP12 2 ROV2IN 73 CLICK SPCLK +24VDC 5 7 11 DENNA DACLK eepot increment C15 0.1 U19 XOUT1 XOUT2 16 221 ROVHOME YIN0 74LS14D YIN1 3 ROVHOM2# YIN2 4 YIN3 R506 221 U19 +24VDC 1 SMD2 OUT2 VREF BOOT2 GND SENSE 8 10 L6203 2 5 7 11 VALEN0# DENNA XOUT0 4 IN1 0.01 XOUT1 C18 0.1 XOUT2 9 6 XOUT3 C19 BOOT2 0.01 GND SENSE 8 10 74LS14D YIN4 13 46 59 37 48 49 60 YIN0 U11 YIN1 SMD0 YIN2 14 VCC YIN3 YIN4 U13 U11 11 9 R83 24.9K U13 4.75K R93 R228 R229 R238 KBY5# R239 4.75K R240 D R241 4.75K U42 4.75K KBY4# 4.75K KBY3# 4.75K 4.75K KBY2# +24VDC R226 R227 221 R234 221 UNUSED GATES 13 221 R237 C21 9 6 9 +24VDC 14 8 U8 U12 5 7 11 2 DENNA R245 100 KBX2# R247 100 KBX1# R246 100 KBX0# 10 11 6 100 R248 LM339AD C24 13 8 5 9 U12 1 U14 3 9 74LS14D 9 6 0.1 10 74LS14D 4 74LS14D 100PF 100PF 100PF 100PF 100PF 100PF 100PF 100PF 100PF 100PF C U44 IN2 R106 U14 1000 U44 BOOT2 GND SENSE L6203 LAMP- 8 10 OUT1 IN1 BOOT1 4 IN2 C23 BOOT2 SENSE ACT05 OD ACT05 13 RVD0 MBRS140T C124 C125 C129 C130 C131 C132 C135 C137 C136 C138 3 U14 11 ACT05 OD KEYBOARD CONNECTOR VALEN1# 2 3 DRIVENB# +15VDC ROTARY VALVE MOTOR: RUN: 1700 MA 21.5K U13 C49 11 C74 C107 C39 C139 C35 C51 C57 C140 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 D 18.2K R84 21.5K Q3 2 R59 U13 4.75K IN OUT3 R92 10K C3 C1 22 4 0.1 1 C47 220PF Local +15VDC 5 +2.5V REFERENCE R86 10K 1 G 3 S R26 MTD3055EL R88 +2.5VREF C46 0.001 100K 2 8 4 VCC VCC R251 VCC VIN COMP GND VO TR 1000 C45 MAX872CSA 0.1 0.1 0.1 0.1 C69 0.1 C77 0.1 C37 0.1 C91 0.1 C126 0.1 C122 0.1 C89 MBRS140T 0.1 C80 0.1 VP0610T 0.1 C81 Q1 VALVE0# 6 5 3 0.1 DRIVENB# 1 OD 0.1 C88 0.1 C40 0.1 C38 0.1 C62 0.1 C50 0.1 C60 0.1 C79 0.1 C78 0.1 D4 0.1 C103 0.1 C104 0.1 C95 0.1 C93 0.1 C101 0.1 C102 0.1 C83 0.1 C82 0.1 C72 0.1 C76 0.1 C73 0.1 4 0.1 VALEN0# C117 +24VDC C84 C36 DRIVENB# 5 6 D C6 Q4 U22 4 C134 C123 C111 C110 C114 C106 C105 C118 C112 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 D2 MBRS140T 2 1 40V 1 G 74HC02D R27 221 0.1 3 S VALVECOM R21 324K MTD3055EL T OC NOVA Biomedical R22 12 0.1 Schematic, Cortex Digital Board Motor & Valve Drivers Page S-5 VALVE BOOST SIZE DRAWING NUMBER E DO NOT SCALE DRAWING 8 7 6 5 200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141 DRAWING TITLE U21 VALBOOST VALVE DRIVERS SMD050 C4 7406D C108 B U11 22 13 C127 MBRS140T DENNA D 0.1 C128 G 0.1 C53 S 0.1 C59 D1 2 TEMP 3 +24VDC C52 C144 2.49K ACT05 C100 C 220 GND 4.75K U17 C99 R19 R20 B C98 OUT2 R68 LM339AD U22 +12VDC A +24VDC 2 3 6 7 OUT4 INSTALL ONE DIODE ONLY C92 OUT1 1 220PF 221 C109 LM317LM 100K R82 10 VCC 4 74HC02D C113 U1 4 D3 C94 ROV Motor 40V ACT05 +24VDC LAMP IS T3/4 BULB 28V .17A R69 0.01 4 OD VALVE1# 1 LAMP DRIVER C26 1.5uF U14 MTD3055EL 18.2K 4.75K R64 S G R128 4.75K 8 R94 3 Q6 1 74HC02D DRIVENB# 1.0 OHM 1W R96 D 221 1 GND 13 R66 OUT2 VREF 11 12 LAMPON# RVD1 C25 4 U22 R67 0.01 0.01 8 10 D RVD2 3 VS LAMP+ 0.01 1 VREF +24VDC 1.0 OHM 1W 2 OD VCC U44 TP5 C133 MTD3055EL Jumper 2-3: inverted sense (requires new sw) C22 OUT2 R90 C61 221 R243 RVD3 LM339AD VCC G HEATER CONTROL POLARITY 3 10 U44 1 74HC02D R87 20K BOOT1 ENABLE 11 74LS14D 13 +2.5VREF 4 IN1 2 74LS14D KBX3# S E DRIVENB# 1.0 OHM 1W 74LS14D 1 74LS14D 3 Q2 R250 11 12 Jumper 1-2: normal sense (original state) LM339AD 10 12 5 7 11 0.1 74LS14D 11 D HTRCTL# OUT1 ENABLE 6 8 74LS14D 221 R236 KBY0# 9 2 DENNA 1000 5 74LS14D 4 U24 220PF 7 U7 VCC 4 221 R235 KBY1# 3 U14 R85 21.5K L6203 VS R58 221 Sampler Motor C48 221 U42 HEATER DRIVER 6 OD ACT05 4.75K 20K R89 100K U42 5 R105 LM339AD 221 U42 HEATER- E500 6 1 R60 U42 4.75K 220PF R95 21.5K 100K 10 OD VCC U42 4.75K C32 R91 ACT05 ROVLED2 R500 LM339AD YIN5 VCC R505 R65 0.01 8 R63 4.75K HOME DETECTORS R2 221 ROVLED R14 C20 1.0 OHM 1W R97 24.9K ACT05 1000 DOORLED 4.75K 12 OD R25 R61 OUT2 VREF SMD100 VCC 1 ENABLE U18 R23 T SMD1 BOOT1 IN2 HEATER+ R62 OUT1 U6 VALEN1# R24 +24VDC C17 3 VS YIN5 SAMLED 0.01 IN2 2 R115 74LS14D 71 DADAT eepot direction 74LS14D 42 43 47 45 BOOT1 XOUT3 SAMHOME ROVHOME# 9 6 SMD3 C16 1.0 OHM 1W VALBOOST VALVE0# 4 IN1 ENABLE 12 8 99 R113 20.5K F OUT1 U5 9 VBOOST XOUT0 E +2.5VREF 3 VS CLICK 6 10 100K L6203 2 U19 74 21.5K R76 20K 74LS14D VALVE1# 11 R78 LM339AD ROVCMP34 U19 GCK2 2 220PF 5 ROVGOOSE# R54 4.75K C31 4 ROV4 G PUMP and SAMPLER MOTORS: RUN: 1300 MA 4.75K R79 20K 8 OD PMD1 PMD0 VALVECOM SMD3 SMD2 SMD1 SMD0 RVD3 RVD2 HEATER+ DOORLED HEATERGND VALVE1# VALVE0# J8 R56 100K R81 ROVHOM2# 220PF R77 21.5K LM339AD SP4 30 1 7 R55 4.75K Pump Motor C30 6 ROV1 23 24 51 72 SAMLED ROVLED ROVHOME# SAMHOME# LAMPLAMP+ RVD0 RVD1 R57 6 OD VCC SAMPHOLD AB3 SAMHOME# 8 10 SAMP4 1 5 R112 221 PMD0 OUT2 SAMP1 74LS14D U19 GND 0.01 C12 0.1 PMPCMP34 VCC R116 20.5K PMD1 4.75K IN2 R80 SP1 PMD3 PMD2 TOUCH DOORSW# BOOT1 PMPCMP12 90 88 70 89 65 17 69 68 ROVLED2 R53 4 IN1 ENABLE 13 R242 3 C13 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 R52 OUT1 U44 PMPCLK 20.5K PMD2 0.01 U4 PGMEN# 100 2 50 ADB3 PMD3 1 C11 0.01 8 10 L6203 2 VS PROGPLS 1.8MCLK CLK100 PMPCLK 3 C10 0.01 4 1.0 OHM 1W ROVCMP34 F OUT2 VREF U11 TDI 39 31 33 76 4.75K U3 VS SMPCMP34 ROV3 VCC L6203 2 TDO TMS R111 BRSTSYNC 32 19 35 18 12 95 34 SAMP1 DONE XINIT +24VDC KYBDINT 77 75 55 53 XTDO TDI TCLK TMS POWER GROUND PMPCMP34 21 44 XCCLK XCDATA XCRLD# XCDONE E4 D C B A ANARST 63 64 58 62 20 86 61 WR# INT1 ADB4 E3 FSELB FSELA ANARST DB3 96 97 94 98 BRD# CHOKEWR# 29 25 27 DB4 DB0 ADB2 ADB1 BBLE# CS6# G DB7 H R107 4 3 2 A REV. 24292-03 LAST MODIFIED ON: 10-5-1998_13:57 SHEET 3 OF 3 1 DWG #10-1073B Page S-6 PN 24292 DWG #10-1074B P10 Page S-7 PN 21642 DWG #10-1075A 8 7 6 5 4 3 2 1 VCC C9 .1UF R16 8.45K U6 X9C103 DADAT 2 D 8 U/D 7 DACLK D VCC 3 CS VH INC VW VSS VL 1 R18 1K 5 4 2 E 1 Q2 MMBT4403LT1 B 6 C 3 U8 74ACT05 11 OD 10 VADJ D_ENAVDD R17 20.5K R2 1K R4 1K G R21 182K J104 Q3 VP0610T 2 3 1 D S VEE VCC 1 2 3 4 5 6 7 8 D_FLM D_LP D_CP D_M VADJ VCCS VEE C16 47 16V note: C2 0.1 J101 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 C D_VEEN D_YELLOW VCCU KBY5# KBY3# KBY1# KBX3# KBX2# SPCLK DACLK 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40 42 44 D_LP D_M D_FLM D_CP D_DD0 D_DD1 D_DD2 D_DD3 R5 1K R37 1K SPAREPIN D_ENAVDD D_ENABKL D_GREEN VCCU 5 3 4 2 KBY4# KBY2# KBY0# KBX1# KBX0# SPDAT CS SHDN EXT REF OUT FB GND 6 7 3 G 1 C N.C. Q6 MTD2955E S 12 11 10 9 13 14 LCD DISPLAY 1 D2 MBRS140T3 4 D 8 R9 324K VCC VCCU C4 47 16V L1 47UH C1 0.1 C3 0.1 C17 22 35V DADAT TP1 SPKR+ D_DD0 D_DD1 D_DD2 D_DD3 DISPOFF# VCC R10 20.5K L2 .15PH -23.5V @ 10mA U5 MAX775 U8 74ACT05 9 OD 8 D_VEEN Y2 .2W SPKR- R6 1K pixels swizzeled here R11 0.47 VCC B J102 R8 1K R33 1K U8 74ACT05 13 OD 12 VCC SPCLK R3 1K R1 1K R29 U4 E1636C 1K R35 1K 1 4 HOT 5 ADJ COM GND D_ENABKL R19 1K 3 C Q1 MMBT4401LT1 B 1 C15 47 16V 3 R36 1K R15 4.75K R14 4.75K R12 221 HOT VIN U8 74ACT05 1 OD 2 VCC COM 4 3 2 1 U8 74ACT05 5 OD 6 D_GREEN R13 221 2 J103 BACKLIGHT U8 74ACT05 3 OD 4 D_YELLOW E KBY5# 1 KBY4# 2 KBY3# 3 KBY2# 4 KBY1# 5 KBY0# 6 KBX3# 7 KBX2# 8 KBX1# 9 KBX0# 10 G_AN 11 G_KA 12 Y_AN 13 Y_KA 14 B 2 KEYPAD VCC A C10 .1UF C11 .1UF C12 .1UF C7 .1UF NOVA Biomedical C8 .1UF A 200 Prospect Street P.O. Box 9141 Waltham, MA 02254-9141 DRAWING TITLE Page S-8 SCHEMATIC, CORTEX DISTRIBITION BOARD BYPASSING NOTES: Place power supply filter as close to J101 as possible. Bypass each chip with one 0.1uF from Vcc to Gnd unless bypass capacitors are explicitly shown. SIZE DRAWING NUMBER D DO NOT SCALE DRAWING 8 7 6 5 4 3 2 REV. 24727 LAST MODIFIED ON: 7-15-1998_12:16 SHEET 1 OF 2 1 DWG #10-1076B PN 24291 Page S-9 DWG #10-1077B 12.4K U4 3 R65 5V REF. 1 LM324 2 BAROMETER 3 4 NC 5 NC U9 6 NC MPX4115AP 2 24.9K COMMON MODE VOLTAGE CLAMP 2 CLAMP2 3 1 +5V R69 2.5V 1 24.9K 24.9K R500 R197 1 C34 A 24.9K 100PF A 100PF BAR A 3 D3 C26 100PF 2 SSTPAD5 D1 2 3 1 1000PF C3 C27 R11 BAV99LT1 SSTPAD5 D2 2 1 3 C2 R51 4.75K A A 1000PF SSTPAD5 D5 4.75K -5V R10 2 1 CLAMP1 D4 2 J6 3 SEE SHEET 3 ISE 5 = PCO2+ SSTPAD5 12 14 200MEG 6 J6 R3 1K 3 5 J6 J6 LMC6484 U1 R167 4 J6 ISE5 13 LOW RES ANALOG BUS 2.5V C12 A 1000PF 7 J6 LMC6484 R166 9 J6 J6 GUARD 3 ~CLEAN0,~CLEAN1,PEAK0,PEAK1 REF2 CLEAN/PEAK CTRL BUS 1 200MEG R2 11 2 U1 1K J6 10 CLAMP2 J6 REF2 R164 14 LMC6484 U1 10 R1 R49 ISE4 24 J6 1 J6 GNDCHS J6 LMC6484 U1 5 22 J7 7 200MEG 6 REF1 6 U21 LMC6484 10.0K R98 5 J8 A A 2.49K 1 J7 +5V GNDCHS A 32 J8 J7 200MEG 6 J7 ISE 3 = NA R171 4 LMC6484 U2 12 14 R9 36 J8 ISE3 J7 10 200MEG 21 J7 ISE 2 R169 19 J7 LMC6484 U2 9 8 R6 GNDCHS ISE2 9 7 J8 R7 A J8 J8 11 J8 J8 ISE 1 24 R174 26 200MEG 2 10 3 8 RG RG I/V16 AD622 + REF 5 U23 A R97A R47 10.0K R177 24.9MEG R55 20K 25 J8 1 - 10.0K ISE1 6 1K J8 R79 C129 0.01 7 R54 8 LMC6484 U2 5 R97 R59 10.0K 1000PF 23 51.1K 14 13 100K C13 22 J8 A LMC6484 U21 12 R60 9 J8 1K 10.0K U21 4 LT1004CD-1 LMC6484 D19 8 C55 .011.235V LT1004 1K J8 20 cw R180 5 R48 1K J7 6 7 6 U21 LMC6484 18 J7 1.235V R83 8 1 J8 .1 R83A LT1004CD-1 LT1004 4 44.2K 40 J8 8 D21 10 1000PF A C56 1K 6 R82 39 R94 3 BAS21LT1 D DG444DY 24.9K C14 17 2 S 38 J8 5 3 R81 37 J8 3 J7 J8 1 1 1000PF I/V 1 35 J8 13 1K J7 33 J8 34 J8 D20 U16 C54 21.5K 2 J7 R98A 51.1K 21 J6 I/V1+ 3 R50 1K 24 J7 CLAMP1 18 5 8 RG RG CLEAN1 REF1 J6 1 U22 3 J8 23 J7 2- 1 6 REF AD622 2 100K J8 1000PF R62 4 J8 C8 A 3+ REF1 20 A 10.0K 2 J8 15 R165 20K R61 0.01 R80 13 J6 10.0K C130 9 1K J6 23 J6 8 200MEG 16 J6 24.9MEG ISE 4 = PH 12 J6 Not used in CORTEX R178 22 J6 A J6 HIGH RES ANALOG BUS A/D+,A/D-,F+,F-,E+,E-,D+,D-,C-,ISE[0:5],BAR,2.27V,HX,CAIR,HCTD,I/V[0:2]+,I/V[0:2]-,PK[0:2],TMPR,REF[1:2],VMN,HTR-,BTMP,MH 8 A 1K C24 A DG444DY 1000PF 11 S 9 27 29 31 J8 LMC6484 U2 3 1 R56 51.1K ISE0 R89 100K 2 1K 28 J8 12 R57 LMC6484 U17 S 10 D LMC6484 U17 PEAK1 8 9 PK1 13 1K .22 C57 C71 C25 4700pf 1000PF 3 2 1 TP1 +5V U5 2N7002 14 30 J8 D 10 G Q4 R196 10.0K D13 J8 ISE 0 R175 200MEG MMBD914LT1 J8 R90 1 100K A R86 A R84 100K C123 .22UF CTHRU2 R170 DG444DY 24.9MEG 6 S D 7 +5V 8 100K C124 PO2+ G Q5 A 2200PF A U5 2N7002 PEAK0 R17 R19 100K R8 9 J7 11 J7 5 2RG 8 RG1 R16 2 1 R87 U14 6 S 51.1K 51.1K 3 1K U13 LMC6484 3 10.0K R74 1 10.0K 2 D10 10 5 LMC6484 U17 PK0 7 6 CLEAN0 C72 4700PF .22 C58 10.0K 3 100K R88 +5V 1000PF R91 1 2.49K A I/V 0 R92 100K 2 2 3 J7 LMC6484 U17 D 1 BAV99LT1 R18 8 J7 I/V0+ REF AD622 D14 3+ MMBD914LT1 7 J7 R52 U16 C18 A 16 21.5K 14 R34 R42A D 6 D11 DG444DY 8 1K C20 .1 LT1004CD-1 R33 LT1004 4 44.2K 9 1.235V cw R172 6 7 10.0K 8 5 R42 10 U13 LMC6484 C23 .01 C127 .22UF CTHRU2 15 S 24.9K R85 100K A U13 4 LT1004CD-1 LMC6484 D9 6 8 A LT1004 HIGH RES ANALOG BUS 1.235V PO251.1K R5 J7 J7 J7 12 1K R15 14 100K 12 LMC6484 U13 APC CTRL BUS R75 14 13 16 R14 J7 J7 I/V0- 10.0K 13 R20 R4 C118 15 2200PF A 2 100K 10.0K 3 1 - 8 RG RG 6 AD622 + REF 5 R168 U15 NOVA BIOMEDICAL 24.9MEG A NEW GEN. SENSOR BD. SIZE REV. E 24292-01 SHEET 1 OF 4 DWG #10-1078B Page S-10 DG444DY 11 S 9 D 10 +5V G Q6 U16 2N7002 R124 12 S D LMC6484 U30 10 100K LMC6484 U30 R125 8 9 14 PEAK2 MMBD914LT1 13 C95 R96A 10.0K 2 C94 4700pf .22 3 D37 51.1K R127 10.0K R126 1 A 100K A V+ R123 C126 100K -5V .22UF CTHRU2 R146 6.81K PK2 A R158 R162 2 4.99K 10.0K 1 3 100K A- R159 100K 13 14 D40 MMBD914LT1 R163 A+ R160 100K 12 ACH6 U33 SEE SHEET 3 LMC6484 +5V R161 100K VMN A/D+,A/D-,F+,F-,E+,E-,D+,D-,C-,ISE[0:5],BAR,2.27V,HX,CAIR,HCTD,I/V[0:2]+,I/V[0:2]-,PK[0:2],TMPR,REF[1:2],VMN,BTMP,MH 2.5V HI RES ANALOG BUS ~CLEAN2,PEAK2 CLEAN/PEAK CTRL BUS LOW RES ANALOG BUS 2.5V,AIR[1:4],PHTR,ACH6,MHTRTEMP GEN CTRL BUS CAIR MHTRTEMP 500mV/C R130 1K 2 A IND11 R108 C61 JUMPER J3 16 AD2M L2 IND11 PAKPRESENT PHTR 10PF 500mV/C R115 C65 .22 TMPR 10.0 R116 C69 .15UH 2 74HC14 A 39 J2 1K MH R110 AIR2 R111 FLOW CELL TEMP 200 J4 475K R138 11 2.5V D7 2 9 3 R31 J4 18.2K 12 C105 8 1 2.5V 1000PF 10 475K U10 LMC6484 BAV99LT1 A A 10PF 9.53MEG HEATER TEMP HCT/AIR DETECTOR #3 J4 18 1 3 U25 1 12 U10 LMC6484 R143 U33 LMC6484 100 5 2 4.75K R105 4 J3 38 J2 14 A R106 R100 13 AIR DETECTOR #2 R144 475K 200 10.0 J4 1 BAV99LT1 +5V 475K 1K 9.53MEG 3 30.1K T .22 .15UH C63 R107 3 J3 AIR1 R133 FLUID_TEMP. BTMP R101 L7 AD1M 2 J3 1000PF D26 10PF AIR DETECTOR #1 C104 +5V 2.5V J4 C17 R32 HCT-M J4 7 1K R145 475K HCTD 6 HCT/AIR DETECTOR R137 13 5 R102 6 14 R142 4700pf HCT1 U6 18.2K R139 R136 475K 18.2K 18.2K R141 7 5 U33 LMC6484 LMC6484 10.0K 10PF C59 A A R109 C101 R140 THERMISTOR RETURN C66 AIR DETECTOR #4 R112 2 475K 3 1 AD4M 19 J6 L1 C70 R117 1000PF 10.0K 9.53MEG AIR3 U10 LMC6484 IND11 .15UH .22 1K +5V 6 7 D38 5 2 3 U10 LMC6484 AIR4 1 BAV99LT1 R114 1.25V D39 2 10.0K 3 A C67 R113 10.0K 1 BAV99LT1 4700pf A D8 HCT EXCITATION J4 17 HCT-E [PASS THRU SIGNALS ONLY] 2 +5V 3 1 2.5V -5V 7 J3 BAV99LT1 1 J3 AD1E C38 .22 J4 15 AD2E 8 C62 AIR EXCITATION MONITOR 4700pf 9 17 AD4E 2.49K C39 J3 DG444DY AIR EXCITATION 3 S 1 D 2 10 U5 C64A 1.0UF J3 J4 24 J4 25 J4 26 LAMPHI PWM0BUF R103 .22 23 HEATERLO HX C37 .22 J6 J3 PWM0BUF,HSO.3 J4 HEATERHI LAMPLO HSO.3 J3 6 R104 49.9 C64B 1.0UF A C64C 1.0UF C64D 1.0UF A C64E 1.0UF NOVA BIOMEDICAL NEW GEN. SENSOR BD. SIZE E REV. 24292-02 SHEET 2 OF 4 DWG #10-1079B Page S-11 1 TP2 J2 J2 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 J5 29 30 A J2 27 GND J2 R181 40 1 AGND 3 ISE0 5 ISE2 REF1 2 ISE1 4 ISE3 6 ISE5 8 11 PK1 PK0 10 13 I/V0+ PK2 12 15 I/V1+ I/V0- 14 17 I/V2+ I/V1- 16 19 MH I/V2- 18 7 ISE4 9 REF2 J5 J5 TO SHEETS 1 & 2 PEAK[0:2],~CLEAN[0:2] J5 J5 J5 J5 J5 J5A J5 23 TMPR 22 25 BTMP VMN 24 27 AIR1 CAIR 26 29 AIR2 HX 28 31 A+ PHTR 30 33 AGND A- 32 PGND 34 J5 J2 J2 9 PHTR,ACH[0:2],ACH6,2.5V,AIR[1:4],MHTRTEMP 2.5V 100 A/D+ -5V 1 PMUX0 2 PMUX1 A2 10.0K PMUX2 4 PMUX3 C90 5 NMUX0 A+ 6 .01 6 NMUX1 5 U30 LMC6484 7 C85 +V 1 F E NC 2 D -V 27 C NC 3 B GND 12 A NC 13 9 U26 8 7 OUT 28 6 5 4 EN 18 3 A3 14 2 A2 15 1 A1 16 0 A0 17 4 5 6 7 8 9 10 11 26 25 24 23 22 21 20 19 EDCBTMP MH VMN REF2 REF1 I/V2I/V1I/V0HCTD CAIR HX .01 3 U30 LMC6484 3 +5V .1 +5V .1 R118 4 5 6 7 8 9 10 11 26 25 24 23 22 21 20 19 E+ D+ SO2DET2 SO2DET1 TMPR PK2 PK1 PK0 I/V2+ I/V1+ I/V0+ HCTD CAIR HX .1 C84 A NMUXENA0 NMUX3 NMUX2 NMUX1 NMUX0 F E D C B A 9 8 7 6 5 4 3 2 1 0 C106 +V 1 NC 2 -V 27 NC 3 GND 12 NC 13 .1 C103 U32 PMUXENA1 OUT 28 EN A3 A2 A1 A0 A PMUXENA0 PMUX3 PMUX2 PMUX1 PMUX0 18 14 15 16 17 8 14 10 R119 DG406DN GND J2 J2 12 31 32 NMUXENA1 +5V PMUXENA2 PMUXENA3 VDD U31 1 AD7710 24 SCLK DGND 2 23 MCLK IN DVdd 3 22 SDATA MCLK OUT 4 21 A0 /DRDY 5 20 /SYNC /RFS 6 19 /TFS MODE 7 18 AIN1(+) AGND 8 17 Iout AIN1(-) 9 16 AIN2(+) REF OUT 10 15 AIN2(-) REFIN(+) 11 14 Vss REFIN(-) 12 13 AVdd Vbias MCLKOUT A/D+ A/D- D38A 8 R198 2.49k .1 C97 A R131 121K R132 4 A C100 .1 R134 8 10 0.0 U33 LMC6484 A A 9 10 11 12 7 6 5 4 BAR ISE5 ISE4 ISE3 ISE2 ISE1 ISE0 9 7 6 5 4 3 2 1 0 C109 22PF WRL FPX-20MHZ -5V J2 GND E+ VDD SYNC DRDY MODCTRL HSO.5 U34-1 .15UH LEDSEL0 A J2 J2 J2 16 17 PMUX0 PMUX1 PMUX2 PMUX3 R148 R147 R149 R150 4.75K 4.75K 4.75K 4.75K 23 J1 17 J1 OC CLK 19 1D 1Q 18 2D 2Q 17 3D 3Q 16 4D 4Q 15 5D 5Q 14 6D 6Q 13 7D 7Q 12 8D 8Q 74HC574 9 J1 7 J1 PEAK0 PEAK1 PEAK2 CLEAN0 CLEAN1 CLEAN2 A/DSYNC BLKHTR R188 SIGOUT 18 U24 1 2 2 3 D33 1 BAV99LT1 LOGIC SUPPLY L3 D28 2 GND MOD0 MOD1 MOD2 MOD3 PMUXENA1 PMUXENA2 PMUXENA3 NMUXENA1 C134 C138 47UF R128 4.75K 4.75K MOD0 MOD1 R120 .1 R122 4.75K J1 IND11 .15UH 1 MMBZ5V6A R129 14 3 16 J1 15 J1 4.75K MOD2 MOD3 P2.3 PAKPRESENT SPARE1 GND PWM0 SPARE2 L6 ANALOG SUPPLIES +5V VDD IND11 C137 D27 C135 47UF 2 47UF C133 .1 3 J1 4 J1 .15UH L5 HSO.3 C136 IND11 .15UH .1 3 1 5 J1 6 J1 10 J1 MMBZ5V6A A MODON 18 LEDSEL3 J2 J1 100 74HC14 VDD MIX/VAL RFS TFS PMUXENA0 NMUXENA0 GND LEDSEL2 1 11 2 3 4 5 6 7 8 9 PMUX0 PMUX1 PMUX2 PMUX3 NMUX0 NMUX1 NMUX2 NMUX3 PMUX0 PMUX1 PMUX2 PMUX3 NMUX0 NMUX1 NMUX2 NMUX3 U34-14,36,68 LEDSEL1 NMUX3 U27 HSO.1 P1.6 P2.6 HSO.0 P1.5 P1.7 HSO.2 28 29 30 31 32 33 34 C109B .01 C109A .1 15 NMUX2 A PMUXENA1 PMUX2 PMUX1 PMUX0 -5V +5V IND11 J2 15 16 17 18 19 20 21 22 23 24 25 26 27 PREHTR +5VF NMUX1 VDD PWM2 L8 3 13 14 8 2 15 16 1 A 26 J2 67 66 65 64 63 62 XTAL1 XTAL2 BUS INST ADV RD 61 CLKOUT P3.0 60 /EA NMI P3.1 59 ACH3 P3.2 58 P3.3 57 ACH1 ACH0 P3.4 56 P3.5 55 ACH2 ACH6 P3.6 54 ACH7 P3.7 53 P4.0 52 ACH5 P4.1 51 ACH4 ANGND P4.2 50 VREF P4.3 49 U34 P4.4 48 P4.5 47 EXTINT P4.6 46 /RESET RXD P4.7 45 TXD P2.3 44 READY 43 P1.0 P1.1 P2.4 42 P1.2 WRH 41 8XC196KC PWM1 WRL 40 PWM2 PWM0 39 P2.7 38 HSI.0 VPP 37 HSI.1 HSO.4 HSO.5 HSO.3 35 2 3 4 5 6 7 8 9 10 11 12 13 FAIL PHTR AIR3 AIR4 MHTRTEMP ACH6 2.5V AIR2 AIR1 ALE 28 J2 R154 VDD GND X1 D+ 4.75K NMUX0 C108 22PF A 33 4.75K A C99 C- 34 R153 4.75K NOT INSTALLED IN CORTEX .1 J2 VV+ GND OUT EN A2 A1 A0 DG508ADY 2.5V SMDXTAL J2 R152 4.75K .1 U28 2.27V +5V MONITOR R151 C96 +5V 100K C98 37 4.75K -5V 100K GND D- 35 R157 4.75K .1 +5V 5 6 .1 J2 R156 4.75K DG406DN R199 +5V E- 36 R155 4.75K VDD A J2 R121 C93 NMUX3 R135 0.0 J2 PMUXENA3 NMUXENA1 PMUXENA2 C91 NMUX2 J2 13 J5A J5 GND J2 J5A 10 LOW RES ANALOG BUS J5 J2 11 J5A 8 J5 LT1009S8 J2 J5A J5A 10.0K J2 6 J5 7 J2 4 SO2DET1 J5 A J2 2 HCTD J5 1 J2 AIR3 J5A ANALOG MUX BUS -5V J2 HX 9 PMUX[0:3],NMUX[0:3],NMUXENA[0:1],PMUXENA[0:3] J2 AIR4 5 7 SO2DET2 J5A A/D- J2 AGND 3 J5A 20 ACH6 1 J5A HI RES ANALOG BUS J5 MHTRTEMP 21 BAR CLEAN/PEAK CTRL BUS J5 19 LEDSEL6 MOD CTRL BUS J2 J2 J2 /LEDSEL 20 LEDSEL0 LEDSEL1 LEDSEL2 LEDSEL3 LEDSEL4 LEDSEL5 LEDSEL6 LEDSEL7 LEDSEL7 23 HSO.5 24 P2.3 U29 15 Y0 14 1 Y1 A 13 2 Y2 B 12 3 Y3 C 11 Y4 10 6 Y5 G1 9 4 Y6 G2A 7 5 Y7 G2B 74HC138 MOD0 MOD1 MOD2 21 TO SHEETS 1 & 2 D31 3 4.75K 1 GND BAV99LT1 LEDDRV GND GND R38 10.0 R193 2 VDD R64 J2 GEN CTRL BUS MOD2 MODON U24 VDD 10.0 PREHTR 3 J2 C33 .1 R195 475 9 J4 10 R37 475 150 Q2 Q1 Q3 MMBT4403LT1 MMBT4403LT1 MMBT4403LT1 R35 J4 R27 1K LED2-A 7 TXD 74HC14 RXD U24 6 5 2.5V 2 RESET U24 8 9 74HC14 EXTINT 74HC14 U24 12 13 U24 10 11 74HC14 A V+ 2 3 1K 5 FAIL 1 PWM0 VP- 7 1K R36 U5 14 LED1 C22 3 12 J1 13 J1 22 J1 11 J1 20 J1 VDD 3 1 2 3 1 BAV99LT1 D29 2 3 1 R189 R182 R184 R191 4.75K 4.75K 4.75K 4.75K A 8 BAV99LT1 D30 9.09K 22 FAILIN D34 2 2 R67 J2 INTIN R190 100 D DG444DY RESETIN R183 100 BAV99LT1 D32 .01UF 100K 15 SRXD R194 100 1 6 R70 16 200K LED1-A 5 LM324 J1 R185 100 D35 VDD U4 R72 S J4 74HC14 J1 21 STXD R187 100 6 19 SYNCIN R192 100 U25 5 74HC14 PWM0BUF R28 24.9 LED1 74HC14 VP+ 4 U3 MIC6211B D6 J4 SYNC R200 681 1 R21 LED2 U25 4 3 HTRCTL R186 100 4 R63 10.0 25 LEDCOM R22 10.0 R23 R24 10.0 10.0 A 3 C36 C29 R73 470PF 2 1 GND 49.9K BAV99LT1 BAV99LT1 D36 2 3 1 BAV99LT1 GND 4700pf R68 1K NOVA BIOMEDICAL A NEW GEN. SENSOR BD. SIZE REV. 24292-03 E SHEET 3 OF 4 DWG #10-1080B Page S-12 VDD V+ C28 U25 C68 C92 C30 C44 C87 C86 C83 11 .1 .1 .1 .1 .1 .1 .1 U25 10 13 74HC14 .1 12 74HC14 P GND +5V C5A C45 C40 C19 C52 C89 C16 C48 C102 C107 .1 .1 .1 .1 .1 .1 .1 .1 .1 .1 C21 C53 C88 C49 C60 .1 .1 .1 .1 .1 C35 C46 .1 .1 A C41 .1 C73 C5 .1 .1 C74 C10 .1 .1 C77 C81 C31 C42 C75 C79 .1 .1 .1 .1 .1 .1 C78 C82 C32 C43 C76 C80 .1 .1 .1 .1 .1 .1 C7 .1 C1 .1 -5V MAKE CONNECTION TO ALL SPARES WITH ETCH SO THAT ECCH CUTS AND JUMPERS CAN USE THESE DEVICES FOR ECO'S J4 1 J4 4 J4 9 5 U4 10 A J4 LM324 U4 12 8 13 14 LM324 +5V 6 -5V R300 U25 9 MCLKOUT CLKOUT 8 U12 MOD0 MOD2 MOD3 1 A 2 B 3 C MODCTRL 6 G1 4 G2A 5 G2B 74HC138 15 14 13 12 11 10 9 7 Y0 Y1 Y2 Y3 Y4 Y5 Y6 Y7 U11 SAM1 B1 SAM1 SAM2 MEAS 1 IN1 3 74HC08 J1 BKGND_IN 9 8 10 74HC08 U11 SAM2 B2 B1 B2 U11 B1 B2 2 8 100 74HC14 4 IN2 6 5 U11 74HC08 SCOPETRIGGER 12 11 13 74HC08 A DG444DY R41 2 D DG444DY S 14 S 16 3 1 R40 12 LMC6484 U6 14 475K 10.0K U7 SO2-1 D15 13 U7 MEAS SO2DET1 SAM1 C114 DG444DY 3 S 1 D 2 C119 U8 DG444DY SAM2 6 S 8 .033UF CTHRU .22UF CTHRU2 C122 .033UF CTHRU DG444DY D 7 11 S 9 D 10 U8 U8 A A C121 IN1 C120 .033UF CTHRU .033UF CTHRU DG444DY 14 S 16 DG444DY 6 S 8 C115 .22UF CTHRU2 R66 D 15 3 U8 DG444DY 11 S 9 SO2DET2 CTHRU2 CTHRU2 IN2 D 10 U7 A 1 .22UF .22UF BKGND_IN LMC6484 U6 2 C116 C117 U7 SO2-2 475K D 7 MEAS A MEAS CO2INPUT A R30 200K J4 R29 2 9 8 J4 10.0K R26 3 U6 LMC6484 10.0K C113 .22UF CTHRU2 SO2DET-C 10 R25 36.5K NOVA BIOMEDICAL A NEW GEN. SENSOR BD. SIZE REV. 24292-04 E SHEET 4 OF 4 DWG #10-1081B Page S-13 Page S-14 PN 23897 DWG #10-1082B ITEM 65 5, 40, 58, 38 31 8 52 P/N 22745 24628 24622 24621 20861 DESCRIPTION Flowcell Lamp Kit Sampler Gears Optosensor Motor, Sampler Fluid Fountain PN 23899 Page S-15 DWG #10-1083B ITEM 2 3 4 5 P/N 24627 24626 24613 24624 DESCRIPTION Motor Rotary Valve Kit Rotary Valve Gear Rep. Opto Sensor Rotary Valve Rep. Kit Ceramic Rotary Valve Page S-16 PN 23898 DWG #10-1084B SPECIFICATION SHEET TITLE DWG.NO. SIZE PRINTER INTERFACE IF3001 SEIKO REV A P/N 24614 SHT 1 OF 1 To Print Head CN01: Head flexible printed circuit connector CN01 CN02: Motor control connector To Printer Motor CN03: Input/Output control connector CN02 SW01: Function switch SW01 a, b: Position selection notches for mounting to the LTP3245 c: Holes for mounting to the LTP3245 d: Two holes for mounting to the outer case (or frame) d c NOTE: CN01: Lift white plastic part to release print head flat cable CN03 PT300 P01 To Control Board J9 PT300 GA1 d Page S-17 REV ECO# APPROVED DRP/CO-DRP/DATE b EFFECTIVE ECO MGR/DATE REV ECO# APPROVED DRP/CO-DRP/DATE EFFECTIVE ECO MGR/DATE For Doc.Cntrl Use Only Disk No. a DWG #10-1085B TITLE SPECIFICATION SHEET INPUT: SIGNALS: OUTPUT: J1 AMP P.C.B. HEADER P/N 640445-5 PIN 1) AC LINE TB1 PIN 2) N/C PIN 1) PIN 3) AC NEUTRAL PIN 2) PIN 4) N/C PIN 3) PIN 5) AC GROUND J2 AMP P.C.B. HEADER P/N 640456-4 PIN 1) N/C PIN 3) PIN 2) +SENSE PIN 4) J3 AMP P.C.B. HEADER P/N 1-640445-6 PIN 1) OUTPUT #4 PIN 14) PIN 2) OUTPUT #4 PIN 15) PIN 3) OUTPUT #3 PIN 16) PIN 4) OUTPUT #3 PIN 5) OUTPUT #2 TB2 PIN 6) OUTPUT #2 PIN 1) PIN 7) COMMON PIN 2) PIN 8) COMMON PIN 3) PIN 9) COMMON PIN 4) PIN 10) COMMON PIN 5) PIN 11) COMMON PIN 6) PIN 12) COMMON PIN 7) PIN 13) COMMON PIN 8) SIZE POWER SUPPLY UNIVERSAL INPUT 5+/-12 24Vo DWG.NO. A P/N 24720 SHT 1 OF 1 1 AC LINE AC NEUTRAL AC GROUND REV 1 TB1 3 F1 T3.5A/250V~ TB2 -SENSE POWER FAIL OUTPUT #1 OUTPUT #1 OUTPUT #1 V1 ADJ. V2 ADJ. OUTPUT #4 OUTPUT #3 OUTPUT #2 COMMON COMMON COMMON OUTPUT #1 OUTPUT #1 NC +S -S PF J2 8.50" (215.9mm) 8.000" (203.2mm) NOTE: 5A MAXIMUM RECOMMENDED CURRENT PER CONNECTOR PIN Page S-18 Model (Note A) Output Output Voltage Minimum Output Current Maximum (Note C) Maximum (Note D) Peak Noise P-P Total Regulation (Note B) GPx130D 1 2 3 4 +5V +24V -12V +12V 3A 0A 0A 0A 16A 3.5A 1.2A 1.2A 20A 3.5A 1.2A 1.2A 25A 5A 1.5A 1.5A 50mV 240mV 120mV 120mV 2% 2% 3% 3% REV ECO# APPROVED DRP/CO-DRP/DATE EFFECTIVE ECO MGR/DATE REV ECO# APPROVED DRP/CO-DRP/DATE EFFECTIVE ECO MGR/DATE For Doc.Cntrl Use Only Disk No. WEIGHT: 2.3 LBS. MAX. DWG #10-1086A