Download Prefix
Transcript
Chapter 4
Tabelle 4-1: Table of Available Measurement Channels
Channel
Description
TOPOGRAPHY
In this channel, the height profile of the sample is
measured. This is the output signal of the Z-Feedback loop control. This signal is averaged and then
scaled with the Z-calibration nm. This measurement channel is always automatically selected.
ERROR SIGNAL
(CONTACT)
The "Error Signal" is the input signal of the Z-FEEDBACK loop control ("setpoint value" minus "actual
value"). The setpoint value is "0" (see figure 4-6 on
page 4-10), and the actual value is the Interferometer signal.
ERROR SIGNAL
(NON CONTACT)
Calculated value from the input signal of the ZFEEDBACK loop control (it is called "Error Signal",
because it is setpoint value - actual value). Actual
value here is the scaled Amplitude, in nm, of the
cantilever oscillation (respectively, the interferometer signal). The setpoint value is the nm scaled "SETPOINT"
Z-SENSOR
Optional channel for Nanoss with a Z-Sensor
(metrology sensor) for linear topography values
(averaged).
EXTERN 1
Input channels for SCANControl C electronic
measuring equipment equipped with additional
optional signal processing equipment.
EXTERN 2
EXTERN 3
For these channels in the SCANControl C electronic
measuring equipment, three BNC inputs are available. They are labeled "IN1", 2, and 3. These channels
are recorded with averaging. For special measurement modes of the optional PM Control a special
low-pass filter averaging is utilized (for PM channels, see PM instructions)
EXTERN 4
The "Z-Sensor" input connector can also be configured as an "Extern 4" input, in cases where it is not
already needed for the metrology sensor. An extern
signal, connected to the "ZS" plug at the scan control can be synchronously recorded with the scan
PHASE
(NON CONTACT)
The phase relationship between the excitation oscillation of the cantilever and the oscillation of the
cantilever; averaged for forward scanning.)
AMPLITUDE
(NON CONTACT)
Amplitude of the cantilever oscillation. This signal
can be scaled in nm with "amplitude calibration".
FURTHER CHANNELS
(FIELDCONTRAST, LOC,
ETC.)
Other channels are available for additional measurement modes. These are described in the included
instructions.
Version 1.2 (May. 2009) - MHBasic Operating Instructions
4 - 13