Download Horiba ABX Micros 60 OS-OT - User manual
Transcript
Micros 60 RDW Calibration form a Quality Control Standard continued: 4 - Take the Minotrol Assay sheet that comes with the control kit and note “Only”, the Normal Control Mean Assay value for RDW. 5 - Calculate the New RDW coefficient as followed: • Take the Normal Control Mean Assay value for RDW, • Divide it by the Mean value of the Normal control ran (6) times as a sample, • Times the current RDW calibration coefficient, This will equal the New RDW Calibration coefficient. 6 - To Enter into the calibration coefficient menu, from the Main Menu, select 3 - CALIBRATION, 2 - COEFFICIENTS, 1 - CALIB. COEFF., enter the password, and use the “Down” arrow key to 7 - RDW. 7 - Press the “ENTER” key to enter the New coefficient for RDW. Press the “ENTER” key again to accept the New coefficient after it has been entered. 8 - Press the Escape “ESC” key until you return to the Main Menu. 9 - Now take all (3) Levels of controls, LOW, NORMAL, HIGH, and run them (1) time each as a regular patient analysis. When the cycles are complete, note only the RDW results and compare them to the Minotrol control Assay values for RDW, on the Assay sheet. Verify that the RDW results are somewhat close to the Mean Assay values and within the Ranges as specified for all (3) Levels. 10 - The RDW calibration is now complete. Be sure to monitor your RDW results and verify that they fall within your patient population. Verify that Normal RDW results will be within the established ranges set in the set-up menu. See Section 5 - Instrument Configuration, 2 - Change Laboratory Limits, (2.1./2.2. Result Low Limits - Result High Limits). 10 MICROS 60 OS/OT ! Note: Certain Pathological conditions will affect the RDW results. See Section 1 - Specifications, 4Limitations, (4.3. Known Interfering substances) 1.5. Calibration Coefficients Calibration may also be achieved by changing the “Calibration Coefficients” directly. From the Mani Menu, select 3 - CALIBRATION, 2 COEFFICIENTS. The following menu will be displayed on the LCD screen as indicated: COEFFICIENTS 10:23 1 - CALIB. COEFF. 2 - PRINT COEFF. In this Menu, the operator has the option of either “Editing” the current calibration coefficients or “Printing” the current calibration coefficients. MICROS 60 OS/OT ! Important: The ABX MICROS 60 is an accurate and reliable instrument when properly maintained. Should any variation of Quality Control results outside the Assayed ranges occur after calibration, it is Highly suggested that you contact your local ABX Technical Support Representative before Manually editing the calibration coefficients!!! Changing Calibration Coefficients When in the Coefficients menu, manually editing the calibration coefficients is performed by selecting 1 - CALIB. COEFF. The LCD screen on the ABX MICROS 60 will then ask for the password which allows the operator to enter into and edit the coefficients. The display will state as indicated: PASSWORD ? : 10:24 RAB042DA /section 4: Calibration and Quality Control