RADOSYS User Manual File Ref. RS_Man82 Page 4-16 Revised at 7/8/2013 4.3.2.7 Neutralization bath Prepare a diluted synthetic vinegar solution by adding 200ml of 15% or 20% vinegar into 4000 ml of distillated water. Pour the solution into the bath tank. After a few minutes drain the neutralized fluid into a laboratory plastic container. 4.3.2.8 Final washing bath Pour 4000ml of washing distillated water into the bath tank. Drain the washing water into a laboratory plastic container. 4.3.2.9 Rinsing and drying Open the cover door of the etching unit and remove the etching drum. Place it to a laboratory tray for rinsing and drying. After an hour of drying the detector holders with the radon test plastic pieces can be taken for evaluation at the Radometer track evaluation unit. 4.3.2.10 Storage of the chemicals The hot sodium-hydroxide solution should be stored in heat-proof laboratory glass. The neutralization and final washing water bath can be stored in any plastic container. 4.3.2.11 Waste chemical treatment Concerning the disposal of the used chemicals, the sodium-hydroxide solution should be treated with extreme attention. It can not be disposed directly because of environmental safety reasons. Two alternative methods can recommended for the disposal. This can be disposed either by neutralization or by proper dilution. One liter of sodium-hydroxide solution can be neutralized by adding four liter 15% vinegar solution. The proper dilution rate in tap water is an other method. Concerning the permitted dilution rate, please consider the local environmental standard or consult a environmental safety engineer. Chapter 4. Routine operation © 2013 by Radosys Kft. All rights reserved