Download Certus Standard
Transcript
Basic information 1.4. Operating conditions SPM Certus Standard is a complex, high-precision research instrument. To ensure normal operation of the device are advised to observe the following conditions of use and storage: 1 Temperature 1.1 Range 10÷50°С 1.3 Optimum 22±3°С *When delivered, SPM Certus Standard scanner adjusted to the optimum operating temperature. When the room temperature is different from the optimum, but within the permissible operating temperature range, neces sary to calibrate the scanner at current temperature in the room. 2 Humidity 2.1 Maximum 80 % 2.2 Optimum 65±15 % 3 Atmospheric pressure, mm. Hg 760±30 4 Power requirements 4.1 Voltage, V 4.2 Frequency, Hz 4.3 Ground 5 110/220 (+10%/-15%) 50/60 required Noise and vibration 5.1 Vibration sensitivity range, Hz 1÷100 5.2 Maximum vibration magnitude in range 0.5 μm 5.3 Maximum sound pressure within 10000 Hz 35 dB 6 Shield from: 6.1 Direct sun light required 6.2 Heating required 6.3 Electromagnetic radiation required 6.4 Air/gas flow required 7 Extra recommendation 7.1 Frequent and abrupt changes in temperature and humidity, affect to performance and the quality of the results. 7.2 Running compressors, traction control, laminar boxes, cooling devices, and similar sources of periodic or random noise can create background noise. 7.3 Recommended to place Certus Standard with maximum possible distance from the computers, monitors and peripheral devices. 7.4 Avoid strong mechanical influences on Certus Standard, as this may lead to failure of the mechanics and control electronics unit. «Nano Scan Technologies Ltd.» User manual 15