Download User Manual Defibrillator MonitorE-HEART
Transcript
Defibrillator Monitor E-HEART 16.1.12 OTHER FUNCTIONS Besides the accessible items on the Configuration Menu, there are other accessible functions on the E-HEART BIPHASIC DEFIBRILLATOR MONITOR panel: 16.1.12.1 PACEMAKER (OPTIONAL) A) Turns on / off – Enables or disables the pacemaker; B) MODE (Sync, or Assync.) – Changes between the VOO and VVI modes; C) Inhibits beep – Enables or disables the synchronized beep with the pacemaker pulses; D) Inhibits pulse– Enables or disables the pacemaker pulses deflagration; E) EMERGENCY – Changes the pacemaker configuration for the ones from the Emergency Mode (VOO, 70 rpm, 150 mA, 40 ms). 16.1.12.2 PRINTER (OPTIONAL) When the key is pressed for the first time the ECG printing gets started, when it is pressed again, the procedure is interrupted. 16.1.12.3 NIBP (OPTIONAL) When the key is pressed for the first time the Arterial Pressure scouting gets started, when it is pressed again, the procedure is interrupted. 16.1.12.4 SYNC. ON Enables or disables the electric discharge with the peak of the R wave. When the equipment is turned on, the synchronism is off, when activated, a “Sync, ON” message will appear on the display (right below the heart frequency indication) and its led will appear on. 16.1.12.5 CANCEL LOAD This key should be activated when the deflagration of the electric discharge is no longer desired, in order to discharge the capacitor. 16.1.12.6 2 MIN It inhibits the sound alarms for 2 minutes. 16.1.12.7 SELECTION / MENU It allows adjusting the load level that is going to be used in the next trigger. 15