Download A.2 Environmental Specification
Transcript
WARNING z The pressure calibration interface has no occlusion alarm. Do not carry out the infusion operation in order to prevent any severe harm to patients. 5.6 Changing Infusion Tube Brand The infusion tube IS-G-V3(1) of ‘Dragon-heart’ brand is the standard setting for this infusion pump. If infusion tube of other brand be used by the user, it is necessary to reset infusion tube parameters as the following steps: Step 1: prepare a new brand of infusion tube Step 2: start the infusion pump and enter into the interface of setting menu Step 3: fix the infusion tube well according to the correct installation steps Step 4: Press the key and the key simultaneously on the mode setting interface to enter into the interface of Advanced Setting so as to set parameter for the infusion tube. 5.6.1 Selection of IV Set Figure 5-9 There are 3 brands of infusion tube parameters that can be saved as the tube.A, B,or C displayed on the top right corner of LCD stands for each band of the infusion tube being used. Select infusion tube by pressing the key or in the Advanced Setting menu as shown in Figure 4-5, and make it adjustable by pressing the key cyclically choose A, B, or C by pressing the key key again to save the setting. 5-8 or , and press the and