C Alarm Information C.1 Alarm Information Note: Column A means it can be totally eliminated; Column B means it can be eliminated to acousto-optic; Column L means the alarm level. Alarm A B L Cause Solutions Press Air bubble in the infusion tube. Air bubble Over Yes Yes Yes No remove the air bubble in the set, then press to restart infusion. High High to stop alarm, Infusion tube installation error. Reinstall the infusion tube. Malfunction with the sensor. Contact manufacturer. Preset infusion volume have been finished. Press key to stop infusion and stop alarm, then press key to eliminate the accumulated volume, press to restart infusion. key key to stop Press infusion and silence alarm , Infusion tube occlusion. Occlusion Yes No High The occlusion value is much too sensitive. Infusion set Yes No Low then press key to restart infusion after moving the occlusion in the loop. Increase the cclusion level. Malfunction with the sensor. Contact manufacturer. Infusion set installation error. Restall the infusion tube. C-1