Download GasAlertMicroClip Technical Reference Guide
Transcript
GasAlertMicroClip Sensor Configuration STEL Interval O2 Auto-Calibration on Startup (Automatic O2 Calibration) STEL Interval provides protection for workers from over exposure to high concentrations of gas, and is based on used-defined 5-15 minute intervals. When the maximum STEL is reached, the detector alarms to notify the worker. When enabled, the O2 sensor is automatically calibrated during the startup self-tests. a Caution Follow all safety procedures as defined by your employer. Enter the interval (5-15 minutes) in the STEL Interval field. The detector is shipped with the factory default setting of 15 minutes. The detector is shipped with the O2 Auto-Calibration on Startup option enabled. LEL By Vol CH4 When enabled, the detector displays the LEL value as CH4%, assuming a methane environment. Auto Zero on Startup When enabled, the sensors automatically zero during the startup self-tests. The Auto-Zero on Start-up option is available for the CO, H2S, and LEL sensors. Not applicable to O2. 1. Click the checkbox of each sensor that will be auto zeroed during startup. The LEL By Vol CH4 option is applicable to the LEL sensor only. The detector is shipped with the LEL by Vol CH4 disabled. The detector is shipped with the Auto-Zero on Start-up option enabled for the CO, H2S, and LEL sensor. 23