In constant power mode, the DC load will cause a constant power to be dissipated in the load. To set up the DC Load to operate in constant power mode and in the steady state condition, use the following keystrokes: Key Display P-set If you press this key after powering up and there is no currentlystored constant power value, you'll be prompted for the desired power level. If there was a power value already stored, it will be momentarily displayed and the instrument will be in CW mode. If you wish to change the power setting, press P-set again. The OFF annunciator will be displayed. On/Off Turns the constant power load on. The CW annunciator will be on. △ or ▽ Momentarily display the power level and the set power value. On/Off Turns the load off. To operate the DC Load in CW mode in a transient condition, see the Transient operation section. To operate the DC Load in CW mode in a dynamic condition, see the Lists section. Constant resistance mode In constant resistance mode, the DC load will behave as a fixed resistance value. To set up the DC Load to operate in constant resistance mode and in the steady state condition, use the following keystrokes: Key Display R-set If you press this key after powering up and there is no currentlystored constant resistance value, you'll be prompted for the desired resistance. If there was a resistance value already stored, it will be momentarily displayed and the instrument will be in CR mode. If you wish to change the resistance setting, press R-set again. The OFF annunciator will be displayed. On/Off Turns the constant resistance load on. The CR annunciator will be on. △ or ▽ Momentarily display the power level and the set resistance value. On/Off Turns the load off. To operate the DC Load in CR mode in a transient condition, see the Transient operation section. To operate the DC Load in CR mode in a dynamic condition, see the Lists section. Timed operation The DC Load can be set to allow timed operation. When you turn the load on, it stays on for the specified time, then turns itself off. To set up timed operation, use the following keystrokes: 8500 DC Load Series Version: 031015 Page 33 of 77