Electromagnetic Immunity
The SureSight is intended for use in the electromagnetic environment specified below. The
customer or the user of the SureSight should assure that it is used in such an environment.
Immunity test
IEC 60601
Compliance level Electromagnetic
test level
environment – guidance
Electrostatic
±6 kV contact
±6 kV contact
Floors should be wood,
discharge (ESD)
concrete or ceramic tile. If
floors are covered with
synthetic material, the
IEC 61000-4-2
±8 kV air
±8 kV air
relative humidity should be
at least 30%.
Electrical fast
±2 kV for power
NA*
Mains power quality
transient/burst
supply lines
should be that of a typical
commercial or hospital
environment.
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential
NA*
Mains power quality
mode
should be that of a typical
commercial or hospital
IEC 61000-4-5
environment.
±2 kV common mode
Voltage dips, short
>95% dip in 0.5 cycle NA*
Mains power quality
interruptions and
should be that of a typical
voltage variations
commercial or hospital
60% dip in 5 cycles
on power supply
environment. If the user of
input lines
the SureSight requires
30% dip for 25 cycles
continued operation during
power mains interruptions,
IEC 61000-4-11
>95% dip in 5
it is recommended that the
seconds
SureSight be powered
from an uninterruptible
power supply or a battery.
Power frequency
3 A/m
3 A/m
Power frequency magnetic
(50/60 Hz) magnetic
fields should be at levels
field
characteristic of a typical
location in a typical
IEC 61000-4-8
commercial or hospital
environment.
29