Download USB-2416 User`s Guide - Measurement Computing

Transcript
USB-2416 User's Guide
Specifications
Table 10. E type thermocouple accuracy specifications, including CJC measurement error.
All specifications are (±).
A/D data
rate
Sensor temperature
range
Accuracy error
maximum
Accuracy error typical
Tempco
(°C/°C)
3750 S/s
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
-200 °C
0 °C
1000 °C
2.413 °C
1.069 °C
1.575 °C
2.413 °C
1.069 °C
1.575 °C
2.413 °C
1.069 °C
1.575 °C
2.306 °C
1.017 °C
1.539 °C
2.306 °C
1.017 °C
1.539 °C
2.306 °C
1.017 °C
1.539 °C
2.306 °C
1.017 °C
1.539 °C
2.306 °C
1.017 °C
1.539 °C
2.306 °C
1.017 °C
1.539 °C
2.306 °C
1.017 °C
1.539 °C
2.306 °C
1.017 °C
1.539 °C
1.352 °C
0.551 °C
1.211 °C
1.352 °C
0.551 °C
1.211 °C
1.352 °C
0.551 °C
1.211 °C
1.244 °C
0.499 °C
1.175 °C
1.244 °C
0.499 °C
1.175 °C
1.244 °C
0.499 °C
1.175 °C
1.244 °C
0.499 °C
1.175 °C
1.244 °C
0.499 °C
1.175 °C
1.244 °C
0.499 °C
1.175 °C
1.244 °C
0.499 °C
1.175 °C
1.244 °C
0.499 °C
1.175 °C
0.017
2000 S/s
1000 S/s
500 S/s
100 S/s
60 S/s
50 S/s
25 S/s
10 S/s
5 S/s
2.5 S/s
0.017
0.017
0.017
0.017
0.017
0.017
0.017
0.017
0.017
0.017
To achieve the thermocouple accuracies listed above, the USB-2416 should be warmed up for 45 minutes after
the initial power on. The accuracies listed above are only guaranteed if the USB-2416 is housed in its plastic
enclosure.
30