Download USB-2416 User`s Guide - Measurement Computing
Transcript
USB-2416 User's Guide Functional Details Figure 9 below shows a typical thermocouple connection. Figure 9. Differential thermocouple connection example Noise filtering, data rate, and throughput rate Although the USB-2416 A/D converter has a maximum data rate of 3,750 S/s, the actual throughput rate you observe for voltage and temperature data is determined by these formulas. Maximum single-channel throughput: 1 1 + 640 µs data rate Maximum multiple-channel throughput: 1 1 + 640 μs ∑ n data rate where n is the number of channels See Table 18 and Table 19 in the Specifications chapter for details. This drop-off in throughput rate is due to the USB-2416's noise filtering feature. You can control the amount of the noise filtering by adjusting the data rate setting. By reducing the data rate, the averaging of samples increases, and noise drops correspondingly. 15