factory (Factory Settings) Description Resets the TCM2.5/2.6 to the factory settings. The internal modes are set to the following factory defaults b=5 er=e mpcal=d cc et=d sdo=t damping=d timeconst=8 sn=m ec=e halt=d sp=8 em=d ma=u uc=d ep=e mag_dec=0.0 ui=d Syntax factory<cr> Output H0V0m0.00<cr><lf> Example factory<cr> go ut=c (Enter Continuous Sampling Mode) Description Enters the Continuous Sampling Mode (go). The TCM2.5/2.6 immediately begins outputting sensors data at rate specified by the SP parameter, default is 8 Hz. Data is automatically transmitted according to the Output Word Format (sdo=) currently specified. To exit the TCM2.5/2.6 and return to Standby mode, issue the Halt Continuous Sampling (h) command. Syntax go<cr> Output Selected data output word. Example go<cr> Related Commands s? (Single Update of Output Word) h (Halt Continuous Sampling) PNI Sensor Corporation TCM2.5 & TCM2.6 User Manual DOC#1009269 r11 Page 28 of 49