Download Microlab 500 B & C series user manual
Transcript
– If you are using a ML510B or a ML530B, you can wash the probe directly from the left syringe. – If you are using a ML540B, you cannot wash the right syringe or probe. 13. When you are satisfied with your entries, press ACCEPT. The system displays the “Save method in method memory” screen. Press CONFIRM to save the method, then press any key to continue. Creating a Pipette Method Use this function to create a pipette method. A pipette method aspirates the same volume as it dispenses. The total actual volume dispensed, however, may include an air gap volume. A pipette method performs the following tasks: • With or without a time delay (of up to 60 seconds), aspirate an air gap volume and then aspirate an amount of sample fluid into the left syringe (if using a single syringe instrument) or into the right syringe (if using a dual syringe instrument). • Dispense both the sample fluid and the air gap volumes from the syringe. • Wash the syringe. 1. Follow steps 1–8 under “Creating a New Method.” Select the “Pipette method” function from the “Method type to create” menu. The “Pipette parameters” screen appears: Pipette parameters Right Air gap volume Right Dispense volume Right Syringe aspirate speed: Right Syringe dispense speed: Air gap mode: Air gap delay: 0.0 250.0 2 2 N/A N/A SELECT item to edit-ACCEPT if complete SELECT ESCAPE ACCEPT Chapter 3 Using the MICROLAB 500 System 3 –2 3