Performing Manual Dispenses
Use the Manual Dispense function to perform manual dispensations without
creating and saving a new dispense method, or without running an existing dispense
method. Manual dispenses cannot be stored in memory.
1.
From the Main menu, select the “Manual dispense” function. The “Instrument
configuration” screen appears:
Instrument configuration
One syringe instrument
Two syringe instrument
SELECT menu item
SELECT
ESCAPE
2.
Use the arrow keys to highlight either a one- or a two-syringe instrument
configuration, then press SELECT.
3.
The default value(s) for the selected configuration appear:
Instrument configuration
One syringe instrument
Two syringe instrument
Left Syringe size (uL):
Right Syringe size (uL):
2500.0
250.0
SELECT item to edit-ACCEPT if complete
SELECT
ESCAPE
ACCEPT
3 –3 4
MICROLAB 510B/511C, 530B/531C, and 540B/541C User’s Manual