Maintenance
Regular and preventive maintenance
10. Prime with distilled and filtered laboratory-grade water.
11. Reconnect the valve tube ends to the dispensing valves
(Figure 5–26).
Figure 5–26. Reconnecting a dispensing valve tube end
12. Finally prime once again with distilled and filtered laboratory-grade
water.
Disinfecting the liquid
path
Routine
maintenance of the
dispensing valves
(tips)
The Multidrop Combi nL liquid path components can be disinfected
with isopropanol, 10% bleach (for example, sodium hypochlorite), or
70% ethanol by running the disinfectant through the liquid path.
To obtain optimal performance and a maximum useful life for the
dispensing valves, it is important that the recommended cleaning
maintenance instructions are followed.
If the dispensing valve is leaking or clogged, refer to “How to clean a
leaking dispensing valve” on page 77 and/or “How to clean a clogged
dispensing valve” on page 79.
Routine maintenance of the dispensing valves covers the sections
described below.
Daily maintenance
The basic maintenance procedure should be performed regularly and on
a daily basis to ensure proper dispensing valve operation:
1. Filter all reagents, detergents, solutions and water before use to
remove solid particles.
76
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific