Frequency Sweep Table 4-2 No. 1 1 1 Repeat all 1 1 1 Example of Indication of Running Results (Frequency Sweep) Seq Item UT Flatt Harm. Level Angle 3 3 3 Sweep frequency Sweep frequency Sweep frequency 230.0 230.0 230.0 ⎯ ⎯ ⎯ 2 6 8 1.0 1.0 1.0 ⎯ ⎯ ⎯ Duration 1.000 1.000 1.000 Interval ⎯ ⎯ ⎯ ■ Description of Table 4-2 [Number] in the following items corresponds to the numbered steps in Fig. 4-3. No. 1 (1st row): Frequency sweep • • • • • [1] Seq tab: The Seq 3 tab is used. [2] Test type: Sweep in frequency [3] Harm. (harmonic order): 2nd [4] Level: 1.0 % [5] Duration: 1.000 second No. 1 (2nd row): Frequency sweep • • • • • [1] Seq tab: The Seq 3 tab is used. [2] Test type: Sweep in frequency [3] Harm. (harmonic order): 6th [4] Level: 1.0 % [5] Duration: 1.000 second No. 1 (3rd row): Frequency sweep • • • • • [1] Seq tab: The Seq 3 tab is used. [2] Test type: Sweep in frequency [3] Harm. (harmonic order): 8th [4] Level: 1.0 % [5] Duration: 1.000 second Items common to No. 1 (1st to 3rd rows) • • • • • • • • 4-16 IEC 61000-4-13 Repeat (repetition count): The repetition count in a seq tab is not indicated. Trigger: Not indicated Sequences to run: Sequences are run in the Seq 3 tab. Seq end-Output off: Not indicated Bank: Not indicated UT: 230 V Freq: Not indicated Repeat all: The number of repetitions for the sequences to be run in Seq 3 is 1. SD003-PCR-LA