3 Technical specification Number of modules Sensor Sensor thickness Pixel size Module size Format Area Intermodule gap Dynamic range Counting rate per pixel Energy range Energy resolution Adjustable threshold range Threshold dispersion Readout time Framing rate Point-spread function Cooling Power consumption Operating temperature Dimensions (W x H x D) Weight 1x3=3 Reverse-biased silicon diode array 320 µm 172 x 172 µm2 83.8 x 33.5 mm2 487 x 619 = 301’453 pixels 83.8 x 106.5 mm2 17 pixels = 2.924 mm 20 Bits = 1’048’576 > 2x106 X-ray/sec 3–30 keV 500 eV 2 – 20 keV 50 eV 2.6 ms 200 Hz 1 pixel Air-cooled 50 W 20 - 30°C 160 x 194 x 289 mm3 11 kg Technical Specification-PILATUS 300K-V1_1.doc 6/25