Download NOELLE S554.100 User Guide

Transcript
NOELLE S554.100 User Guide
6.
A green filled oval indicates that the chest
compression was successfully recorded....
7.
Continue the same process to finish the calibration.
At the end of the calibrating session, the wizard shows
the average peak, depth, and duration values for the
procedure. If you feel you performed the procedures
correctly, click the “Save” button. Otherwise, press the
“Back” button to repeat the calibration.
Notice that you can go back, abort or cancel at any time
during the procedure.
Factory Settings
Use the factory settings menu to restore sensors to the
factory calibration. A factory reset will over-ride any
calibrations performed by the facilitator.
First remove the baby from the birthing arm and allow
the motor arm to remain free for 30 seconds. While the
birthing arm is free, click reset the force sensor button to
record the current state as zero force.
Reset Uterine Pressure – Disconnect the uterus from
the pressure port and click reset to restore the
calibration to factory default.
Reset Labor Motor – Recalibrate the position of the
birthing mechanism on the track. Always remove the
birthing baby prior to recalibrating the birthing
mechanism position.
Options
The GaumardUI has several options that can be preset
by the instructor.
NOELLE specific options include: NOELLE Features.
Newborn specific options include: Newborn Add-Ons
and Neonate Features.
Environment
If the FIXED button is selected, the simulator's serial
number must be entered in the text box.
Make sure that when you are restoring the sensors to
the factory settings that no one is practicing chest
compressions or ventilations. Any of these actions may
interfere with the reset. Each time that one of these
sensors is clicked a message will appear at the bottom
left of the screen notifying the user of the status of the
reset (OK, or TRY AGAIN). Should the sensor not
respond, please refer to the troubleshooting guide or
contact Customer Support.
At the end of calibrating a function, the Calibration
Wizard resets the simulator for the changes to take
effect and displays the message "Done". If the wizard
displays the message "Can't reset", it simply means that
the new calibration values will take effect next time you
start the software.
If the changes need to take immediate effect, simply
close the GaumardUI software, wait one minute (for the
simulator to turn off), and then start the GaumardUI
software again.
NOELLE specific censors
The options below are specific to the NOELLE
simulator. Remove the birthing baby from the birthing
mechanism before continuing.
Reset Force sensor – Reset the force sensor if the
labor activity is displaying incorrect force readings.
84
If you cannot find the serial number, first connect to the
simulator using GENERIC and then press on the button
that says “Can't find Serial Number” and follow the
instructions.