Download NOELLE S554.100 User Guide
Transcript
NOELLE S554.100 User Guide 6. A green filled oval indicates that the chest compression was successfully recorded.... 7. Continue the same process to finish the calibration. At the end of the calibrating session, the wizard shows the average peak, depth, and duration values for the procedure. If you feel you performed the procedures correctly, click the “Save” button. Otherwise, press the “Back” button to repeat the calibration. Notice that you can go back, abort or cancel at any time during the procedure. Factory Settings Use the factory settings menu to restore sensors to the factory calibration. A factory reset will over-ride any calibrations performed by the facilitator. First remove the baby from the birthing arm and allow the motor arm to remain free for 30 seconds. While the birthing arm is free, click reset the force sensor button to record the current state as zero force. Reset Uterine Pressure – Disconnect the uterus from the pressure port and click reset to restore the calibration to factory default. Reset Labor Motor – Recalibrate the position of the birthing mechanism on the track. Always remove the birthing baby prior to recalibrating the birthing mechanism position. Options The GaumardUI has several options that can be preset by the instructor. NOELLE specific options include: NOELLE Features. Newborn specific options include: Newborn Add-Ons and Neonate Features. Environment If the FIXED button is selected, the simulator's serial number must be entered in the text box. Make sure that when you are restoring the sensors to the factory settings that no one is practicing chest compressions or ventilations. Any of these actions may interfere with the reset. Each time that one of these sensors is clicked a message will appear at the bottom left of the screen notifying the user of the status of the reset (OK, or TRY AGAIN). Should the sensor not respond, please refer to the troubleshooting guide or contact Customer Support. At the end of calibrating a function, the Calibration Wizard resets the simulator for the changes to take effect and displays the message "Done". If the wizard displays the message "Can't reset", it simply means that the new calibration values will take effect next time you start the software. If the changes need to take immediate effect, simply close the GaumardUI software, wait one minute (for the simulator to turn off), and then start the GaumardUI software again. NOELLE specific censors The options below are specific to the NOELLE simulator. Remove the birthing baby from the birthing mechanism before continuing. Reset Force sensor – Reset the force sensor if the labor activity is displaying incorrect force readings. 84 If you cannot find the serial number, first connect to the simulator using GENERIC and then press on the button that says “Can't find Serial Number” and follow the instructions.