Download DIGITAL PRESSURE CONTROLLER
Transcript
CAUTION: The DPC should only be opened by qualified electrical/mechanical service technicians. Lethal voltages are present and exposed in the power supply and display. 1. Identify which dual range sensor requires mechanical zeroing. 2. Remove the DPC’s top cover. 3. Locate and remove the plastic zeroing plug located in the side of the sensor oven assembly. 4. Enter the Zeroing screen by selecting MENU | CALIBRATE. 5. Select the Primary sensor by pressing the Sensor [F6] key until the words Primary is displayed above the calibration coefficients. 6. Select Zero [F1]. Do NOT press the Calibrate button. 7. Select the Mechanical Zero [F1] key. 8. Referring to Figure 6-5, slightly loosen the Allen screw on the back of the sensor bracket, making sure to maintain a snug fit on the screw. Using the T-shaped Photocell Centering Tool (RIC # 7250sys-111), adjust the photocells by placing the tool in the hole in the back of the sensor and turning the tool slightly. The screen will show a number of horizontal lines. The top yellow bar represents the current zero setting. The bottom portion of the bar has a green bar in the center and two red bars on either end. If the top yellow bar is over the red bar, then mechanical zeroing is required. If the yellow bar is over the green bar, then no adjustment is required. MAINTENANCE 6-14