Download DIGITAL PRESSURE CONTROLLER

Transcript
MENU | TEST - MENU
4.6.4.1 Menu | Test – Sweep Test
The sweep function can be used to automatically exercise the elastic sensing element of
the device under test prior to performing a calibration. For the DPC to perform the
sweep test, input the high and low set-point pressures, the pressure control tolerance, the
dwell time at set-point, and the number of cycles to perform.
1. Ensure that the units, limits, and control parameters are set to their desired values.
2. The sweep test is entered from the Sweep Test Menu. From the Main Menu, (press
[PREVIOUS] until the Main Menu appears), press Menu [F6], then Test [F5], and
then Sweep [F2].
3. Enter the high and low pressure points, the control tolerance, dwell time, and the
number of cycles.
MENU | TEST | SWEEP – MENU
4. Press Run [F2] to initiate the test.
5. To pause the program, press Pause [F3]. Pause will now be highlighted and the
DPC will continue controlling to the current set-point. The DPC will maintain
control of the current set-point until further instructions from the user. Press
Continue [F4] to resume the program.
4-27
LOCAL OPERATION