Download Keithley Series 3400 Generators
Transcript
Series 3400 Pulse/Pattern Generators User’s Manual Section 6: Remote Programming :PULS:DOUB[1|2]:DEL:HOLD [:SOURce]:PULSe:DOUBle[1|2]:DELay:HOLD Parameters TIME|PRATio Default TIME Query :PULS:DOUB[1|2]:DEL:HOLD? Description Use this command to set/read the coupling between the pulse period and the double-pulse delay: TIME - The absolute pulse delay is held fixed when the pulse period is varied (pulse phase varies). PRATio - The pulse phase delay (delay as ratio of period) is held fixed when the pulse period is varied (pulse delay varies). Examples :PULS:DOUB1 ON :PULS:DOUB1:DEL 50 PCT :PULS:DOUB1:DEL:HOLD PRAT Enable double pulses on output 1 Set inter-pulse delay to 50% of pulse period Hold inter-pulse delay as fixed percent age of pulse period :PULS:DOUB[1|2]:DEL:UNIT [:SOURce]:PULSe:DOUBle[1|2]:DELay:UNIT Parameters S | SEC | PCT Default S Query :PULS:DOUB[1|2]DEL:UNIT? Description Use this command to set/read the default units for the double-delay parameter. The default unit is the unit used when the parameter is programmed to a value without a unit suffix. Examples :PULS:DOUB1:DEL:UNIT PCT :PULS:DOUB1:DEL 50 3400S-900-01 Rev. B / January 2007 Return to Section Topics Set output 1 double-delay unit to % Set output 1 inter-pulse delay to 50% of period 6-31