Download Anritsu 373xxA 40GHz Vector Network Analyzer
Transcript
PERFORMANCE VERIFICATION q Measurement Technique KEY SYSTEM PARAMETERS perature, during warm-up, calibration, and when making measurements on your DUT. The use of proper microwave connector mating techniques. The measurements of the key system performance parameters are performed using the ANRITSU 37XXX Test Software, PN: 2300-178. NOTE The Standard Conditions and Special Precautions described in paragraph 4-2 should be observed when performing these measurements. The software will use the following standards to determine system performance values: q q q q Assurance Air Line Set Offset Termination Frequency Counter Power Meter Equipment required: 37XXX Test Software (ANRITSU PN: 2300-178) External computer/controller; refer to Table 1-1 located on page 1-9. GPIB cable (ANRITSU PN: 2100-2), or equivalent Measurement Procedure: Step 1. Refer to file “Readme.txt” of the distribution disk for in- formation regarding setup of the computer/controller. Step 2. Using the GPIB cable, connect the external computer/controller to the the 373XXA IEEE 488.2 GPIB rear panel connector. Step 3. Program may be run from floppy disk or copied to any folder in the computer and run from there. Step 4. Follow the directions displayed on the computer screen to perform all tests. NOTE Some tests require that the first test selected must be the 12-Term Calibration. The software will guide you through this calibration. If any tests fail, refer to the Troubleshooting and Help selections of the Test Software menu for assistance. 373XXA MM 4-7/4-8