ULTRASONIC THICKNESS GAUGE - A1207 (A1207C)
Change the ultrasound velocity in the device. Put
the device on the sample block and compare the
real thickness with the readings of the device.
If it is necessary you will need to repeat these actions until the device shows the same
thickness of the sample block as it’s real thickness is. Each time you change the meaning
of ultrasound velocity you have to save it in the memory of the device (see art.3.1 of this
manual).