7.12.3 Voltage Subsystem
VOLTage
[:LEVel]
[:IMMediate]
[:AMPLitude] <NRf+>
:TRIGger <NRf+>
Set output voltage
Set voltage value for trigger pending
:PROTection
[:OVER] <NRf+>
Set OVP value
: LOW
:STATe <CRD>
Set UVP|UVL mode
:[LEVel] <NRF+>
Set UVP|UVL value
:CLEar
:MODE <CRD>
Select arbitrary trigger control mode
VOLTage
Sets the output voltage value in Volts. The range of voltage values are described in Table 7-3. The
maximum number of characters is 12.
SCPI Command Syntax
[SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude] <NRf+>
GEN Command Syntax
PV < NRf+> PV?
Default Suffix
V
*RST Value
0
Examples
:VOLT 500 MV
Query Syntax
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
VOLTage? MAX VOLTage? MIN
Returned Parameters
<NR3> VOLT? returns the present programmed voltage level.
VOLT? MAX and VOLT? MIN return the maximum and minimum programmable
current levels.
VOLT:LEV 234.56789
VOLT:MODE
This command selects FIX, LIST, WAVE subsystems control over the power supply output voltage.
SCPI Command Syntax
[SOURce]:VOLTage:MODE <CRD>
GEN Command Syntax
None
Parameters
NONE | FIXed | LIST | WAVE
*RST Value
NONE
Examples
VOLT:MODE LIST VOLT:MODE FIX
Query Syntax
[SOURce]:VOLTage:MODE?
Returned Parameters
NONE | FIX | LIST | WAVE
92