Download user manual - TDK
Transcript
7.12.3 Voltage Subsystem VOLTage [:LEVel] [:IMMediate] [:AMPLitude] <NRf+> :TRIGger <NRf+> Set output voltage Set voltage value for trigger pending :PROTection [:OVER] <NRf+> Set OVP value : LOW :STATe <CRD> Set UVP|UVL mode :[LEVel] <NRF+> Set UVP|UVL value :CLEar :MODE <CRD> Select arbitrary trigger control mode VOLTage Sets the output voltage value in Volts. The range of voltage values are described in Table 7-3. The maximum number of characters is 12. SCPI Command Syntax [SOURce]:VOLTage[:LEVel] [:IMMediate][:AMPLitude] <NRf+> GEN Command Syntax PV < NRf+> PV? Default Suffix V *RST Value 0 Examples :VOLT 500 MV Query Syntax [SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? VOLTage? MAX VOLTage? MIN Returned Parameters <NR3> VOLT? returns the present programmed voltage level. VOLT? MAX and VOLT? MIN return the maximum and minimum programmable current levels. VOLT:LEV 234.56789 VOLT:MODE This command selects FIX, LIST, WAVE subsystems control over the power supply output voltage. SCPI Command Syntax [SOURce]:VOLTage:MODE <CRD> GEN Command Syntax None Parameters NONE | FIXed | LIST | WAVE *RST Value NONE Examples VOLT:MODE LIST VOLT:MODE FIX Query Syntax [SOURce]:VOLTage:MODE? Returned Parameters NONE | FIX | LIST | WAVE 92