Limit Inject RF Current Need analyser connected to the second device. In addition, a current probe (or similar) is required. Process is currently under development, but it allows Vasona to limit the amount of current injected into a given line using a current probe. Need an analyser connected to the second device. In addition, a current probe (or similar) is required. There are options in the setup window to allow these to be configured. Cal: Conducted Cal: Radiated Fixed Voltage Allows conducted calibration to be performed. For additional information. See conducted calibration page. Allows radiated calibration to be performed, establishing a fixed voltage within the chamber. See radiated calibration page. Option needs to have an Isotopic field probe attached, noting that Vasona currently only supports serial probes. If forward power is to be recorded then the directional coupler, power meter and power sensor needs to be included in the template. See selection options. Cal: Radiated Fixed Power Check linearity [5dB] Current method is not fully supported. Allows the checking of the linearity of the test level obtained