Download SAM12 User Manual
Transcript
Calibration Procedure Issue: 1.0 Since the SAM has a very linear response to activity, any source of activity between 100 nCi (3.7 kBq) to 10 μCi (370 kBq) range may be used. When CCM is enabled, the 60Co source used for calibration should have an activity in excess of 5 μCi (185 kBq) in order to minimise the calibration time. Long count times will be required for small sources to maintain statistical accuracy against adverse background influences. 2. A source holder, to securely retain the source for positioning within the cubicle. 3. A generic USB printer, (e.g. Epson LX300) Preparation for Calibration First the user must enter the Calibration Menu (page 5-19). Select HV Scan (page 5-25) and check the detector HV’s are set correctly for the Calibrating isotope. If these are not known, they must be determined as described in Detector HV Selection (page 9-3) before proceeding with the calibration. The SAM12 is then ready for calibration. Calibration Ensure the background is stable, the door is closed and any calibration sources are well away from the SAM12 (at least 8 feet). 1. Ensure the door is shut and the unit is on and ready. 2. Login to the Administration mode, using the Health Physicist role (see Gaining Access to the Administration Mode (page 5-3)). See Cal Check (page 5-19) for more information. To check the calibration, the Cal Check button is selected. This does not allow the calibration factor to be changed, but a comparison is made between the old and new factor. To recalibrate the instrument, i.e. adjust the calibration factor for a particular nuclide, the Calibrate button is selected. The calibration process should be repeated for a number of nuclides that would be found in typical contamination. As a minimum, a calibration to both 137Cs and 60Co should be undertaken. The 137Cs calibration is essential when the NBR facility is used; the 60Co calibration is essential when CCM is required. Notes: Ensure that each source is at least 8 feet away from the SAM during the background measurement. 10-2 SAM12 User Manual Thermo Fisher Scientific