Download FH62C14 - User Manual
Transcript
Troubleshooting Troubleshooting Guides Table 6-1. Troubleshooting Guide, continued Malfunction Possible Cause Action Digital electronics Check that all boards and connectors are seated properly defective Replace with spare boards to isolate the problem Cannot change parameters Keypad is not enabled Enable keypad through menu No automatic filter change Reel nuts not tight Tighten reel nuts No more filter tape Replace with new filter tape Filter not properly adhered Adhere filter tape to take-up reel and wind over itself once Optical Tape Counter defective Replace optical tape transport sensor Filter tape transport drive motor defective Replace motor Pressure transducer does not hold calibration or is noisy Pressure transducer Replace pressure transducer Run output noisy Recorder noise Replace or repair recorder and/or leads. Sample concentration varying Run FH62C14 on a zero filter - if quiet, there is no malfunction Foreign material in Clean optical bench defective optical bench Digital electronics Analyzer does not calibrate properly Analog test ramp 6-2 FH62C14 Continuous Ambient Particulate Monitor Instruction Manual defective Replace board with a spare board System leak Find and repair leak Pressure or temperature transducer(s) out of calibration Recalibrate pressure or temperature transducer(s) Digital electronics defective Replace one board at a time with a spare board to isolate the defective board Faulty recorder Replace recorder Zero output on Turn Zero off Span output on Turn span off Thermo Fisher Scientific