Download Pegasys to Horizon
Transcript
If images are not being received by the printer but the ping command comes back successfully, restart the Codonics Print daemon: right-click on the desktop to display the Pegasys Main Menu, and then right-click PATIENT UTILITIES > CODONICS UTILITIES > RESTART CODONICS PRINT DAEMON. Setting the Horizon as the Default Printer See also the section Setting the Codonics Printer as the Default in the Codonics Color Dye Sublimation Printers Ver. 7.0 Release Document. To set the Horizon as the default for all sessions, use the Set Default Devices application. Click on SYSTEM MAINTENANCE > SET DEFAULT DEVICES > CHANGE DEFAULT SETTINGS > CHANGE DEFAULT FORMATTER. To change the default printer for a single session: 1. Display the Formatting Control Panel by right-clicking on the Format icon and selecting Control Panel from the pop-up menu. 2. Right-click on the Formatter button and selecting a Codonics printer from the pop-up menu. Click Proceed. 3. When the application is closed, the default printer will revert to the printer set through Set Default Devices or through the Print Preview. Adjusting Image Quality The Horizon provides a number of adjustments that can be used to optimize printed image quality. These adjustments are invoked using the Configuration Info field in the Configuration screen. The most important setting for color printing is Codonics’ exclusive Medical Color Matching (MCM) setting. The setting devised for Horizon is MCM 103. ♦ ♦ ♦ Gamma: Affects the overall lightness or darkness of the printed image while preserving full black and full white. An increase in gamma value will lighten the printed image; a decrease in gamma value will darken the printed image. Contrast: Enhances or reduces the contrast of the printed image. Positive settings will enhance contrast while negative settings will reduce contrast in the printed image. Saturation: Color prints only. As saturation increases colors get brighter and more vibrant. Negative saturation (desaturation) removes color and makes them more muted. An example of the Configuration Info line using three settings is: mcm=103\gamma=1.2\contrast=5 Codonics Driver printing only: If the print is too dark and changes to settings do not seem to have any effect, create a special Custom Job Setting for the printer: 1. Go to the printer and press the left arrow and the Menu key ( + Menu ). 2. Press the down arrow to Custom Job Settings, then press Enter. 3. Go to New… and press Enter. 4. Go to Name and press Enter. 5. Press the left and right arrows ( )simultaneously. Enter the name RAW_JS by using the up and down-arrows. Press Enter when this is complete. 6. Next, go to the LUT setting and press Enter. Change the LUT to Visual. 7. Press Enter, then Menu, then send a print from the Pegasys to verify the change. 8. If further adjustment is required, return to the Custom Job Settings menu (steps 1 and 2), go to Edit, and change the Gamma value to 1.3. Exit using Menu. Test this new configuration, and increase the gamma by .3 until the desired lightness is reached. Gamma should not go above 2.0; if prints are still too dark at that gamma then there is another issue. Erica Sommer Codonics, Inc. Page 5 Philips Medical Systems Pegasys to Codonics Horizon Imager Setup 4/10/2006