Download TH1941 Multimeter
Transcript
TH1942 MED 50000 Count Multimeter DC I AC I DC V AC V Ω Period Freq dB dBm AUTO CATⅠ(1000V) VΩ ! CATⅡ(300V) 500mA COM 1000V MAX AC+DC % Comp Rel OFF ON Auto mA 2nd Hold Rate Trig Menu Local Esc Shift T1AL 250V 20A ALL INPUT 1KV MAX - + DC POWER CURRENT SOURCE DC Current measurement at Range: 5A, 20A TH1942 MED 50000 Count Multimeter DC I AC I DC V AC V Ω Period Freq dB dBm AUTO CATⅠ(1000V) ! CATⅡ(300V) 500mA VΩ COM 1000V MAX AC+DC % Comp Rel OFF ON Auto mA 2nd Hold Rate Trig Menu Local Esc Shift T1AL 250V 20A ALL INPUT 1KV MAX AC POWER CURRENT SOURCE AC Current measurement at Range: 5A, 20A CAUTION: Maximum Input = 20A DC or RMS Maximum test times : < 20s Figure 3-2 DC and AC Current Measurements 3.3.2 Front Panel Fuse Replacement WARNING: Make sure the instrument is disconnected from the power line and other equipment before replacing the AMPS fuse. 1. Turn off the power and disconnect the power line and test leads. 2. From the front panel, use a screwdriver to rotate the fuse carrier several turns counter-clockwise. Take the fuse carrier out of the socket. 3. Remove the fuse and replace it with the same type (T1AL, 250V, 5×20mm). CAUTION: Do not use a fuse with a higher current rating than specified or instrument damage may occur. If the instrument repeatedly blows fuses, try to find out the reason before replacing the fuse. 4. Install the new fuse by reversing the procedure above. 21